Track-ItTM Pressure/Vacuum Temperature Logger
SAFEGUARDS AND PRECAUTIONS:
Read and follow all instructions in this instruction sheet carefully, and retain this sheet for future reference.
Do not use this instrument in any manner inconsistent with these operating instructions or under any
conditions that exceed the environmental specifi cations stated.
This instrument is not user serviceable. For technical assistance, contact the sales organization from which
you purchased the product.
DESCRIPTION:
The Pressure/Vacuum Temperature Loggers are sealed (IP67) Data Loggers used for measuring air or
liquid pressure and ambient temperature. The units are self contained in a small 316L stainless steel
enclosure with a threaded ¼” NPT port to connect to a pressure source. The Loggers can operate
unattended for months or years. Recording parameters can be set in the Track-ItTM PC Software. The
units can be set to record on instantly, at preset times, or on alarm conditions. Up to 64,000 readings can
be saved in non-volatile memory at intervals from 2 seconds to 24 hours. Internally there is a USB port
for direct connection to a laptop or PC, and a button and LED for user interaction as described below.
SPECIFICATIONS:
General
Record: Sample Rates: User confi gured 1 every 2 seconds up to 1 every 24 hours
No. of Samples: 32,000 or 64,000 (depends on setup)
Record Trigger: Two independent triggers. Multiple trigger modes: Instantaneous, button control, on alarm, time and date
(start and stop), day of week
Record Mode: Fill to end of memory, cyclic, number of samples, time duration
Alarms: 2 user programmable alarms. High or Low
Communication: Direct USB connection
Software: Track-It™ Software—Program device, view data (historic or real time), export to Excel™.
Battery: Lithium 1/2AA 1.2Ah
Life: Up to 5 years typical @ 1 minute sample rate
Enclosure: Material: 316L Stainless Steel
Dimensions: 1.95” (5cm) diameter x 2.19” (5.6cm) excluding port thread
Measurement
Ranges: Pressure: 0-35, 0-150, 0-350, 0-550, 0-2000, 0-5800 PSI
Vacuum: 760-0 Torr and 760-380 Torr
Accuracy: ± 0.25%
Resolution: 0.01%
Repeatability: ± 0.05%
Overpressure Rating: 1.5 times max
Port connections: 1/4” NPT Male
Temperature Range: -20 to +85°C /-4 to 185°F
Accuracy: ± 0.5°C 0 to 50°C / 32 to 122°F
± 2°C -20 to 85°C / -4 to 185°F
Resolution: 0.06°C / 0.11°F
INSTALLATION:
The logger screws into a ¼” NPT port. Use suitable tape or thread sealant/dope on the threads before inserting the logger into
the port. Use an adjustable wrench on the nut to tighten the device into the port and test for leaks.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com

ACCESS:
To access the internals of the unit it is necessary to unscrew the lid (counterclockwise).
This exposes the control panel shown right. The function of the parts is as follows:
Push Button/LED functionality:
The push button can be used to check status or to change the state of the data logger.
For instance, it can be used to start and stop recording if desired. This feature can also
be disabled to ensure the data logger state remains unchanged until connected to a PC.
TO CHECK STATUS
Short press and release: LED will fl ash in various patterns to indicate data logger status:
The fl ash sequence is: Battery status > Record status > Alarm status
One long red fl ash at the beginning of the fl ash sequence indicates low battery. (no red fl ash = battery ok)
One long red fl ash at the end of the fl ash sequence indicates an alarm. (no red fl ash = no alarm)
Slow red/green blink: Connect to PC. Unit is NOT set up or memory is full.
TO CHANGE STATE
Press and hold button, release when LED turns steady green: Changes record state on/off
Press and hold button, release when LED turns steady red: Reset alarm indication
Press and hold button, release after LED goes off: No change
The table below summarizes the push-button / LED functionality.
SERIAL COMMUNICATION:
The logger has a mini USB port. Plug the cable into this port and the other end into the USB port on your PC/Laptop. The provided software needs
to be installed in order to work with the logger.
Track-ItTM PC Software:
The Track-ItTM PC Software allows the Data Logger to be programmed, allows the data to be retrieved, displayed, saved to disk or exported to
ExcelTM. Data can also be displayed in real time (on devices using internal sensors). Some of the features are:
• Delayed recordings, fi xed duration recordings by time or number of samples, multiple record times, manual record by button press, record
on alarms
• Sample rates from 2 seconds to 24 hours. Instantaneous, average, maximum or minimum values
• Two Alarms, high or low, latched or momentary and record under these alarm conditions
• Display of data graphically, digitally or tabular formats with alarm indication
• Export user selected data in ExcelTM or CSV formats. Filter data to be exported
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com

REPLACING THE BATTERY:
Replace the battery as needed with type: ER14250 3.6V Lithium - Tadiran TL-5902 or similar. Remove threaded cap and
insert battery as shown in diagram right (observe polarity):
O-RINGS:
Track-ItTM Pressure/Vacuum Temperature Loggers come directly from the factory with high quality O-Rings that have been
properly installed. As a user, it is up to you to maintain a functional O-Ring seal. Make sure to clean them frequently,
lubricate them regularly and inspect the O-Ring regulary for signs of failure (nicks, cuts, gashes, fl attening, cracking,
discoloration, deformation, etc.). Don’t poke, jab, pry the O-Ring with sharp or pointed objects. Don’t expose the O-Ring
to harsh chemicals or pressure/temperature beyond specifi cations.
REPLACING THE O-RING:
Remove threaded cap and use a small pointed tool to pry the old O-Ring out of its groove. Make sure that the O-Ring
groove is free of any dirt or debris. Apply a thin coat of lubricant to the new O-Ring. Carefully install the new O-Ring into
the threaded cap. Confi rm cap screws back onto the case fully without binding.
ACCESSORIES:
Battery Replacement Lithium 1/2AA 1.2 AH Battery (ER14250)
O-Rings Replacement O-Ring Seal, 3-Pack
Garden Hose Adapter Brass Coupling
Anti-Theft Cover Locking Cover
USB 2.0 Cable USB 2.0 to 2.0 mini 3-foot cable
USB On-The-Go Cable USB Cable for use with Android devices
Track-ItTM Software Track-ItTM Software on CD
In order to comply with EU Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE):
This product may contain material which could be hazardous to human health and the environment. DO NOT DISPOSE of this product
as unsorted municipal waste. This product needs to be RECYCLED in accordance with local regulations, contact your local authorities
for more information. This product may be returnable to your distributor for recycling - contact the distributor for details.
CE Compliant. RoHS Compliant. Meets the safety requirements of IEC61010-1. This product is not waterproof.
Monarch Instrument’s Limited Warranty applies.
Warranty Registration and Extended Warranty c
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com