Thisdocument is providedtocustomers who havepurchased MolecularDevices, Inc.(“MolecularDevices”) equipment,software, reagents,and consumables to use in the operation of such Molecular Devices equipment, software, reagents, andconsumables. This document is copyright protected and any reproduction of this document, in whole or any part, is strictly prohibited, except as Molecular Devices may authorize in writing.
Software that may be described in this document is furnished under a license agreement. It is against the law to copy, modify, or distribute the software on any medium, except as specifically allowed in the license agreement. Furthermore, thelicense agreement may prohibit the software from being disassembled, reverseengineered, or decompiled for any purpose.
Portions of this document may make reference to other manufacturers and/or their products, whichmay contain parts whose namesare registeredastrademarks and/or function as trademarks of their respective owners. Any such usage is intended only to designate those manufacturers' products as supplied by Molecular Devices for incorporation into its equipment and does not imply any right and/or license to useor permit others to use such manufacturers' and/or their product names as trademarks.
Molecular Devices makes no warranties or representations as to the fitness of thisequipment for any particular purpose and assumes no responsibility or contingent liability, including indirect or consequential damages, for any use to which the purchaser mayput the equipment described herein, or for any adverse circumstances arising therefrom.
For research use only. Not for use in diagnostic procedures.
The trademarksmentionedherein arethe propertyofMolecular Devices, Inc. or their respective owners. These trademarks may not be used in any type of promotion or advertising without the prior written permission of Molecular Devices, Inc.
Product manufactured by Molecular Devices, Inc.
1311 Orleans Drive, Sunnyvale, California, United States of America 94089.
The SpectraMax®M3, M4, M5, and M5eMicroplate Readers are a series of dual-monochromator, multidetection, multi-mode instruments with a triple-mode cuvette port and 6-well to 384-well microplate reading capability. Detection modalities are shown inTable1-1.
Table1-1SpectraMax® Multi-Mode Microplate Readers and Applicable Modes
1
Modes
AbsorbanceXXXX
Fluorescence intensityXXXX
Fluorescence polarization
Time-resolved fluorescence
LuminescencexxXX
Note:In this user guide, all references to SpectraMax Multi-ModeMicroplate Readers include the M3, M4, M5, and M5e models. When a feature or capabilityapplies to only certain readers, this exception isnoted.
The optical performance is comparableto a top-of-the-line dedicated spectrophotometer or spectrofluorometer with no trade-off between instrumentperformance and the number of read modes.
The built-in cuvette port can be used for absorbance, fluorescence andluminescencereadings. Dual monochromators allow selection of any absorbance wavelengthbetween 200nm and 1000nm, and any excitation wavelengthbetween 250nm and 850nm for readings in fluorescence intensity,time-resolved fluorescence (M4, M5,M5e models only) or wavelength-selectable luminescence modes, and 400–750nm for readings in fluorescence polarization mode(M5, M5e models only).
Assaysrequiring a read in two ormoremodes can be combined and run on the SpectraMax Multi-ModeMicroplateReaders by issuing a single command in SoftMax® Pro Software, Molecular Devices’ leading microplate data acquisition and analysis software platform.
SpectraMaxM3SpectraMax M4SpectraMaxM5SpectraMax
XX
XXX
M5
e
0112-0115 F7
Description
Applications
Endpoint, kinetic,spectrum, and multi-pointwell-scanning applications combining absorbance and fluorescence in 6-wellto384-well microplates, as well as endpoint, kinetic, and spectrum applicationsin absorbance and fluorescence using cuvettes, can be run with little to no optimization.
The extreme flexibility and high sensitivity of the SpectraMax Multi-Mode Microplate Readersmake them appropriate for applications within the fields of biochemistry, cell biology, immunology, molecular biology,and microbiology.
Typical applications include ELISA,nucleic acid,protein,enzymatic typehomogeneous and heterogeneousassays, microbial growth, endotoxin testing, and pipettor calibration.
Certified SpectraMax® M5e-HTRF Readers
The SpectraMax M5e reader has the same performance specificationsas the M5 but is certified for use with Cisbio Bioassays’ HTRF (Homogeneous Time-Resolved Fluorescence) technology. HTRF is a proprietary time-resolved fluorescence technology that overcomes many of the drawbacks of standardFluorescence Resonance EnergyTransfer (FRET)techniques, suchas the requirements to correct for autofluorescence and the fluorescent contributions ofunboundfluorophores.
Optics
The use of two holographic diffraction grating monochromators allowsfor individual optimization of wavelengths for both excitation and emission in fluorescence readings. Mirrored optics focus the light into the sample volume, and cutoff filters are used toreduce stray light and minimize background interference. The light source is a high-powered Xenon flash lamp. Sensitivity or read-speed can beoptimized by varying the number of lampflashes per read.
Dynamic Range
The dynamic range of detection is from 10-6 to 10Variations in measured fluorescence values are virtually eliminated by internal compensation for detector sensitivity, photomultiplier tube voltage and sensitivity,as wellas excitation intensity. The photometric range is 0–4 ODs with a resolution of 0.001 OD.
A SpectraMax Multi-Mode Microplate Reader with PathCheck Pathlength Measurement Technology allows normalization of variable well volumes to 1-cm cuvette readings. PathCheck Pathlength Measurement Technology allows for multichannel pipettor validation and for experiment comparison from different days.
Automix
Usingthe Automix feature of the SoftMax Pro Software, the contentsof the wells in a microplate can be mixed automatically by linear shaking before each read cycle, making it possible to perform kineticanalysis ofsolid-phase,enzyme-mediated reactions (mixing is notcritical for liquid-phase reactions).
Temperature Control
Temperature in the microplate chamber is isothermal, both at ambient and when the incubator is turnedon.When the incubator is on,the temperature may be controlled from 2°C above ambient to 60°C.
Supported Plates
Microplates having 6, 12, 24, 48, 96, and 384 wells can be used in the SpectraMax Multi-Mode MicroplateReaders. Top and bottom reads are available for fluorescence, time-resolved fluorescence and luminescence detection. When reading optical density at wavelengths below 340nm, special UV-transparent, disposableor quartzmicroplates and cuvettes that allow transmission of the far UV spectra must be used.
One plate carrier adapter isprovidedwith the instrument. The adapter is required for optimum performance with standard 96-well and 384-well format microplates for all top-read applications.
0112-0115 F9
Description
Computer Control
An external computer running SoftMax Pro Software, which provides integrated instrument control, data display, and statistical dataanalysis, controls the SpectraMax Multi-Mode Microplate Readers. Cuvette port functionality can also be controlled using SoftMax Pro Software.
SoftMax Pro Software provides the following functionality:
Instrument Control
SoftMax Pro Software allows you to set up and run a complete protocolfor the SpectraMax Multi-Mode Microplate Reader, as well as all other Molecular Devices' microplate readers.Instrument settings can be saved as a protocol file and used repeatedly for reading different microplates or cuvettes. All stand-alone instrument functions can be controlled using the software. In addition, SoftMax Pro Software provides capabilities that are notavailable when using an instrumentinstand-alone mode such as user-defined kinetic run times, read intervals, Automix parameters, etc.
Data Collection and Display
SoftMax Pro Software collects and stores all raw data received from the instrument. Data is displayed in agrid format that correspondsto the wells in a microplate or individual cuvettes.
SoftMax Pro Software can collect data from one or more microplates orcuvettes and store it in a single data file, using the same or differentinstrument settings for different microplates or cuvettes. For example,microplates containingdifferent samples can be read using the same or differentmodes, all within the same experiment.
Data Reduction and Plotting
You can manipulate or “reduce” the raw data using dozens of built-in formulas ordefine yourown analysis structure to quickly and easily summarize the raw data. Morethanone reduction can be shown, andresults from different microplates orcuvettes can be compared within the same experiment.
Immediate Results Reporting and Analysis
Once you have defined instrument settings, and have customized a SoftMax Pro Software data file with assay information, reduction settings, custom columns in Group sections, and summary objects, you can save this information to create an assay protocol. Protocols can be used throughout a department or company for highly repeatabledata collection and analysis that is completedthe second the plate read has completed.
10 0112-0115 F
Reader Components
Control PanelBack PanelCuvette Chamber
Microplate Drawer
The main components of the SpectraMax Multi-Mode Microplate Readers are:
•Control panel:for cuvette chamber control.
•Microplate drawer: used for all fiveread modes and four read
types.
•Cuvette chamber: used for absorbance, fluorescence intensity,
and luminescence read modes for endpoint, kinetic, and spectrum scanning.
The control panel consistsof a 2-x-20-character LCD and eleven pressure-sensitive membrane keys that can be used to control somefunctions of the instrument. When you press a control panel key, the instrument performs the associated action.
Note:Settings made inSoftMax Pro Software override control panel settings.
The left side of the display shows the temperature inside the cuvette chamber, both actual and set point, and whether or notthe temperature is at the set point (the enunciator blinks if it is not at set point). The temperature of the microplate chamber lags slightly behind the temperaturein the cuvette chamber. The temperature in themicroplate chamber is reported inthe SoftMax Pro Software interface display.
The middle of the display shows the wavelengths for absorbance/excitation and emission.
The right sideof the display shows the data received fromthe readingas absorbance, percent transmission, fluorescence emission or excitation, or luminescence, and indicates whether or nota reference measurement was made(enunciator blinks if no reference reading was taken).
The key enablesand disablesthe incubator that controls the temperature within both the microplate chamber and the cuvette port.
•When the incubator is on,the set temperature and actual
temperature (cuvette chamber only) are shown on the frontpanel LCD display.
•When the instrument is performing a kinetic or spectral scan,
the temperature keys on the front panel are disabled.
Temp
The keys allow you to enter a set point at which to regulate the cuvette and microplate chamber temperature. However, rememberthat the cuvette temperature only is reported on the LCD display, while the microplate chamber temperature is reported in the SoftMax Pro Software interface display.
Pressing this key scrolls the temperature up or down, starting at theprevious temperature setting (or the default of 37.0°C,ifno setting had been made):
•Pressing the up () or down () arrow once increments or
decrements the displayed temperature by0.1°C.
•Pressing and holding either arrow increments or decrements the
displayed temperature by1°Cuntil it is released.
You cannot set a temperature beyond the upper (60°C) or lower (15°C) instrument limits.
Wavelengths ()
Selects the wavelengthto be used for reading the cuvette manually. Two sets ofup or down arrow keys are available for setting absorbance/excitation (fluorescence) wavelengths and emission (fluorescence) wavelengths.
The controlpanel does not display the wavelength selected through the SoftMax Pro application.
Pressing the up or down arrowkey scrolls up or downthrough the available wavelengths, starting at the previoussetting:
•Pressing the up () or down () arrow once increments or
decrements the displayed wavelength by 1nm.
•Pressing and holding either arrow increments or decrements the
displayed wavelength by10nmuntil it is released.
0112-0115 F13
Description
DRAWER
Ref
A reading of buffer, water, or air taken in the cuvette that is used as I0to calculate Absorbance or%Transmittance. If noreference reading is taken, the instrument uses the I0 values stored in the NVRAM (non-volatile memory) of the instrument.
This key is disabled during a computer-controlled run.
Read Cuvette
Initiates the sample reading of the cuvette.This key is disabled during a computer-controlled run.
Mode
A toggle switch used to display cuvette data aspercenttransmittance (%T), absorbance (A),relative fluorescence units(RFU),or relativeluminescence units(RLU).
Drawer
The key opens and closes (toggles) the microplate drawer.
The microplate drawer is located on the right side of the instrument and slides in and out of the reading chamber. An internal latch positions the microplate in the drawer as it closes (allowing for better robot integration-no springs or clips are used).
The drawer remains in the readingchamber during readcycles.
Figure 1-3The microplate drawer.
Microplate drawer operation varies, dependingon the incubator setting:
•If the incubator isoff, the drawer remains open.
•If the incubator is on, the drawer closes after approximately 10
seconds to assist in maintaining temperature control withinthemicroplate chamber.
Do not obstruct the movementof the drawer.If you must retrieve a plate after an error condition or power outage andthe drawer doesnot open, it is possibleto open it manually(see
page61
0112-0115 F15
).
Troubleshooting on
Description
Microplates
The SpectraMax Multi-Mode Microplate Reader can accommodate SBS-standard 6-well to 384-well microplates and strip wells. When reading optical density at wavelengths below 340nm, special UV-transparent, disposableor quartzmicroplates allowing transmission of the deep UV spectra must beused.
Not all manufacturers' microplates are the same with regard todesign,materials, or configuration. Temperature uniformity within the microplate may vary depending on the type of microplate used.
Microplates currently supported by the SoftMax Pro Software for use in this instrument are:
The SoftMax Pro Software plate listalso includes half area and low-volume plates. SoftMax Pro can always be used to define a new plate type using the manufacturer's specifications for wellsize, spacing and distance from the plate edge.
Locatedat the right front of the SpectraMax instrument, the cuvette chamber has a lid that lifts up, allowing you to insert or remove a cuvette. The chamber contains springs that automatically position the cuvette in the proper alignmentfor a reading.The cuvette door mustbe closed beforeinitiating areading.
Cuvettes
The SpectraMax Multi-Mode Microplate Reader can accommodate standard-height(45 mm), 1cm cuvettesand12x75 mm test tubes when used with the test tube cover.
Not all manufacturers' cuvettes are the same with regard to design,materials, or configuration. Temperature uniformity within the cuvette may vary depending on the type of cuvette used.
Cuvettesused for absorbance readingsare frosted on two sides.Be sure to handle cuvettes on the frosted sides only. Place the cuvette intothe chamber so that the “reading” (clear) sides face left and right.
Fluorescence cuvettes are clear on all four sides and should be handledcarefully. Place a frosted cuvette into the chamber so that the “reading” (clear) sides face left and right.Semi-Micro and Ultra-Micro cuvettes can also be used withan adapter. See information about supported cuvettes.
0112-0115 F17
Cuvettes on page78 for more
Description
Power Switch
Power Cord
Receptacle
Label
Fuse Box
Cover
Computer
Port
Printer
Port
Figure 1-5The test tube cover.
The Back Panel
18 0112-0115 F
Figure 1-6Schematic of the back panel of a reader.
The following components are located on the back panel of the SpectraMax instrument:
•Power switch:a rocker switch, labeled I/O (for on and off,
respectively).
•Power cord receptacle: plug the power cord in here.
•Fuse boxcover: cannot be opened while the power cord is
plugged in. When opened, it provides access to the fuse box containing twofuses that are required for operation.
•Computer port (double-shielded 8-pin RS-232 serial, for use with an
external computer): plug one end of an 8-pin DIN serial cable into this port; the other end attaches tothe serial (modem) port of the computer.
•Printer port: not used for the SpectraMax instrument
•Label: provides information aboutthe reader, such as line
voltage rating, cautionary information, serial number, etc. Record the serial number shown on this label for use when contacting Molecular Devices Technical Support.
Principles of Operation
Absorbance
Note:In this user guide, references to the SpectraMax® readers
include the M3, M4, M5, and M5e models. When a feature or capability applies to only certain readers, this exception is noted.
Absorbance is the amount of light absorbed by a solution. Tomeasure absorbance accurately, it is necessary to eliminate light scatter. In the absence of turbidity, absorbance = optical density:
where I is transmitted light, andIO is incidentlight.In this manual, we use the terms absorbance and optical density
interchangeably.
Optical Density
Optical density is the amount of light passing through a sample to a detector relative to the total amount of light available.Optical density includes absorbanceof the sample plus light scatter from turbidity.
2
A = –log(I/IO)
Transmittance
Transmittance is the ratio of transmitted light to the incident light.
T = (I/IO)
%T = 100T
where I is transmitted light, andIO is incidentlight.
PathCheck® Pathlength Measurement Technology
The Beer-Lambert law states that absorbanceis proportional to the distance that light travels through the sample:
A =
bc
where A is the absorbance, is the pathlength, and c is the concentration of the sample. In short, the longer the pathlength, thehigher the absorbance.
0112-0115 F19
is the molar absorptivity of the sample,b
Principles of Operation
Horizontallight path
Vertical light path
CuvetteMicroplate wells
Microplate readers use a vertical light pathso the distance of the light through the sample depends on the volume. This variable pathlengthmakes it difficult to perform extinction-based assays and also makes it confusing to compare results between microplate readers and spectrophotometers.
The standard pathlength of a cuvette is the conventional basis forquantifying the unique absorbtivity properties of compounds in solution. Quantitativeanalyses can be performed on the basis of extinction coefficients, without standard curves (forexample,NADH-based enzyme assays). When using a cuvette, the pathlength is known and isindependent of samplevolume, so absorbance is proportional to concentration.
In a microplate, pathlength is dependent on the liquid volume, so absorbance is proportional to both the concentration and the pathlength of the sample. Standard curves are often used to determineanalyte concentrations in vertical-beam photometry of unknowns, yet errors can still arise from pipetting the samples and standards. The PathCheckdetermines the pathlength of aqueous samples in the microplateand normalizes the absorbance in each well to a pathlength of 1cm. Thisnovel approach tocorrecting the microwell absorbance values isaccurate towithin 2.5%of the values obtaineddirectlyina 1cmcuvette.
Reference measurements made by reading the cuvette (Cuvette Reference) or using factory-stored values derived from deionized water (Water Constant) can beused to normalize the optical density data for microplate wells.
Pathlengthcorrection is accomplishedonly when using the PathCheck Pathlength Measurement Technology with SoftMax
®
Pro Software.PathCheck Pathlength Measurement Technology is patentedbyMolecular Devices and can beperformed only on an Molecular Devices plate reader.
The SpectraMax Multi-ModeMicroplate Reader offers both the Cuvette Reference and the Water Constant methods.
The actual pathlength, d, of a solvent is found from the following equation:
When a Cuvette Reference is usedfor pathlengthcorrection, the valueof k is obtainedby taking optical density measurements on the fluid inthe cuvette at two wavelengths, 1000 and 900nm:
When the WaterConstant is used for pathlength correction, the value ofk is obtained fromthe instrument. This constant is saved in the instrument in the factory and may differ slightly from instrument to instrument.
Once the pathlength d is found,the followingequation is used for the pathlength correction:
PathCheck Pathlength Measurement Technology is applicable toalmostallbiological/pharmaceutical molecules in aqueous solution because they have little or no absorbance between 900nmand 1000nm atconcentrations normally used. PathCheck Pathlength Measurement Technology can also beused withsamples containing small amounts oforganics or high buffer concentrations by using the Cuvette Reference. See
Water Constant or Cuvette Reference?on page22.
0112-0115 F21
Principles of Operation
Water Constant or Cuvette Reference?
The PathCheck Pathlength Measurement is based on the absorbance of water in the near infrared region (between 900nm and 1000nm).If the sample is completelyaqueous,has no turbidity and has a low salt concentration (less than 0.5M), the Water Constant isadequate. The Water Constant is determined duringmanufacture and is stored in the instrument.
If the sample contains an organic solvent such as ethanol or methanol, we recommend using the cuvette reference. It is important thatthe solvent does not absorb in the 900nm to 1000nm range (to determine whether or not a given solvent would interfere, see the discussion ofinterfering substances below).When a non-interference solvent isadded to the aqueous sample, the water absorbance decreases proportionally to the percentage of organic solvent present. For example, 5% ethanol decreases the water absorbanceby 5% and results in a 5% underestimation of the pathlength. You can avoid the error by putting the same water/solvent mixture in a cuvette and using the CuvetteReference.
Touse the Cuvette Reference, placeinto the cuvette port a standard1cm cuvette containing the aqueous/solvent mixturethat is used forthe samples in the microplate. The cuvette must bein placewhen you read the microplate. When you click the Read button in the SoftMax Pro program, the instrument firstmakes the 900nm and 1000nm measurements in the cuvette, and then makes the designated measurements in the microplate. The cuvette values are stored temporarily and used inthePathCheck PathlengthMeasurement Technology calculations for the microplate samples.
Use ofCuvette Reference with PathCheck Pathlength Measurement Technology is different from a reference reading of a cuvette in aCuvetteSet section (by clicking the Ref button in the CuvetteSet section tool bar inthe SoftMax Pro program). The cuvette reference used for PathCheck Pathlength Measurement Technology calculations(measurements at900nm and 1000nm) does notproduce data thatcan be viewed in a CuvetteSet section and is used onlywith data in microplates, not cuvettes.
Background Considerations
Raw optical density measurements of microplate samples include bothpathlength-dependent components (sample and solvent) and a pathlength-independent component (OD of microplate material). The latter must beeliminated from the PathCheckPathlength Measurement Technologycalculation inorder to obtain PathCheck Technology-normalized results. Thereare3 ways to accomplish this: plate blanks, plate background constants, and plate pre-reads, all of which are described in the PathCheck Pathlength Measurement Technology section of the SoftMax Pro User Guide.
PathCheck Pathlength Measurement Technology and Interfering Substances
Any material that absorbs inthe 900nm to 1000nm spectral region could interfere with PathCheck Pathlength Measurement Technology measurements. Fortunately, there are few materials that do interfere at the concentrations typicallyused.
Turbidity is the most common interference: if you can detect any turbidity in your sample, you should not use the PathCheck Technologyfeature.Turbidity elevates the 900nmmeasurement more thanthe1000nm measurement and causes an erroneously low estimate of pathlength. Using Cuvette Reference does not reliably correct forturbidity.
Samples that are highly colored in the upper visible spectrum may haveabsorbance extending into the nearinfrared(NIR) and can interfere with the PathCheck Pathlength Measurement Technology.Examples include Lowry assays, molybdate-based assays and samples containing hemoglobins or porphyrins. In general, if the sampleis distinctly red or purple, you should check for interference before using the PathCheck Pathlength Measurement Technology.
To determine possible color interference, do the following:
•Measure the optical density at 900nm and 1000nm (both
measured with air reference).
•Subtract the 900nm value from the 1000nm value.
•Do the same for pure water.
If the delta OD for the sample differs significantly from the delta OD for water,then it is advisable not to use the PathCheck Technology feature.Use of Cuvette Reference does not correct for the interference with thecurrent calculation scheme in the SoftMax Pro program. Currently,CuvetteReference involves a single (automated) read at 900nm and 1000nm and the automated calculations in the SoftMax Pro program do not compensate for color or solvent interference. However, you could correct for suchinterference by taking two cuvette measurements and using a different set of calculations. For further information, contact Molecular Devices Technical Support.
Organic solvents could interfere withthe PathCheck Technology feature if theyhave absorbance intheregionof the NIR water peak.Solventssuch as ethanol and methanol do not absorb in the NIR region,so they do not interfere, except for causing a decrease in the water absorbance to the extent of their presencein the solution. Their passive interference can be avoided by using the Cuvette Reference. If,however, the solvent absorbs between 900 and 1000nm, the interference would be similar to the interference of highly colored samples described above. If you are considering adding an organic solvent other than ethanol ormethanol, you are advised to run aspectral scan between 900nm and 1000nm to determine if the solventwould interfere with the PathCheck Technologyfeature.
0112-0115 F23
Principles of Operation
Normalizing Absorbance Measurements
SoftMax Pro Softwareautomatically reports absorbance values normalized to a 1-cm pathlength.SoftMax Pro Software automatically reports absorbancevalues normalized toa 1-cm pathlength. The table below shows results obtained with 75µL to 300µL yellow reagent.
Table2-1Yellow reagent results.
Well Volume
Optical pathlengths and raw absorbance values were directly proportional to well columns. After normalization to a 1-cmpathlength, all absorbance values, regardless of the volumein the wells, were within 1% of the value obtained by measuringthe same solution in a 1-cm cuvette.
Fluorescence
Fluorescent materials absorb light energy of a characteristic wavelength (excitation), undergo an electronic state change, and instantaneously emit light of a longer wavelength (emission). Most common fluorescentmaterials havewell-characterized excitation and emission spectra. Figure2-2 shows an example of excitation and emission spectra for a fluorophore. The excitation and emission bands are each fairly broad, with half-bandwidths of approximately 40nm, and the wavelength difference between the excitation and emission maxima (the Stokes shift) is typically fairly small, about 30nm. There is considerable overlap between the excitation andemission spectra (gray area) when a small Stokes shift is present.
Because the intensity of the excitation light is usually many tens of thousands of times greater thanthat of the emitted light,some type ofspectral separation is necessary to reduce the interference of the excitation light with detection ofthe emitted light. The SpectraMax Multi-Mode Microplate Reader incorporates many features designed to restrict interferencefrom reflected excitation light. Among these features is a set of long-pass emission cutoff filters that can be set automatically by the instrument or manually bythe user. If the Stokes shift is small, it may be advisable to choose an excitation wavelength that is as far away from the emission maximum as possible while stillbeing capable ofstimulating the fluorophore so thatlessof the excited light overlaps the emission spectrum, allowing better selection and quantitation of the emitted light.
0112-0115 F25
Principles of Operation
Wavelength (nm)
Relative Fluorescence
Excitationreadingwavelength
Emissionreadingwavelength
Fluorophore’sexcitationmaximim
Fluorophore’semissionmaximim
500550600650
0
0.5
1.0
Figure 2-3Optimized excitation and emission reading wavelengths.
Figure2-3 shows that the best results are often obtained when the excitation and emission wavelengths used for reading are not the sameas the wavelengths of the excitation and emission spectra of the fluorophore. When the reading wavelengths for excitation and emission are separated, a smaller amount of excitation light passes through tothe emission monochromator(gray area) and on to thePMT,resulting in a purer emission signal and more accurate data.
The SpectraMax Multi-Mode Microplate Reader allows scanning of bothexcitation and emission wavelengths, using separate tunable monochromators. One benefit of being able to scan emission spectra isthat you can assess more accurately whether the emission is, infact, the expected fluorophore, or multiple fluorophores, and not one generated by a variety of background sources or by contaminants. Another benefit is that you may be able to find excitation and emission wavelengthsthat avoid interferencewhen interferingfluorescentspecies are present.
For this reason, it may be desirable to scan emission for bothan intermediate concentration of labeled sample, as well as the backgroundof unlabeled sample. The optimum setting is where the ratio of the sample emission to background emission is at themaximum.
26 0112-0115 F
Loading...
+ 60 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.