Thisdocument is providedtocustomers who havepurchased MolecularDevices, Inc.(“MolecularDevices”) equipment,software, reagents,and consumables to use in the operation of such Molecular Devices equipment, software, reagents, andconsumables. This document is copyright protected and any reproduction of this document, in whole or any part, is strictly prohibited, except as Molecular Devices may authorize in writing.
Software that may be described in this document is furnished under a license agreement. It is against the law to copy, modify, or distribute the software on any medium, except as specifically allowed in the license agreement. Furthermore, thelicense agreement may prohibit the software from being disassembled, reverseengineered, or decompiled for any purpose.
Portions of this document may make reference to other manufacturers and/or their products, whichmay contain parts whose namesare registeredastrademarks and/or function as trademarks of their respective owners. Any such usage is intended only to designate those manufacturers' products as supplied by Molecular Devices for incorporation into its equipment and does not imply any right and/or license to useor permit others to use such manufacturers' and/or their product names as trademarks.
Molecular Devices makes no warranties or representations as to the fitness of thisequipment for any particular purpose and assumes no responsibility or contingent liability, including indirect or consequential damages, for any use to which the purchaser mayput the equipment described herein, or for any adverse circumstances arising therefrom.
For research use only. Not for use in diagnostic procedures.
The trademarksmentionedherein arethe propertyofMolecular Devices, Inc. or their respective owners. These trademarks may not be used in any type of promotion or advertising without the prior written permission of Molecular Devices, Inc.
Product manufactured by Molecular Devices, Inc.
1311 Orleans Drive, Sunnyvale, California, United States of America 94089.
The SpectraMax®M3, M4, M5, and M5eMicroplate Readers are a series of dual-monochromator, multidetection, multi-mode instruments with a triple-mode cuvette port and 6-well to 384-well microplate reading capability. Detection modalities are shown inTable1-1.
Table1-1SpectraMax® Multi-Mode Microplate Readers and Applicable Modes
1
Modes
AbsorbanceXXXX
Fluorescence intensityXXXX
Fluorescence polarization
Time-resolved fluorescence
LuminescencexxXX
Note:In this user guide, all references to SpectraMax Multi-ModeMicroplate Readers include the M3, M4, M5, and M5e models. When a feature or capabilityapplies to only certain readers, this exception isnoted.
The optical performance is comparableto a top-of-the-line dedicated spectrophotometer or spectrofluorometer with no trade-off between instrumentperformance and the number of read modes.
The built-in cuvette port can be used for absorbance, fluorescence andluminescencereadings. Dual monochromators allow selection of any absorbance wavelengthbetween 200nm and 1000nm, and any excitation wavelengthbetween 250nm and 850nm for readings in fluorescence intensity,time-resolved fluorescence (M4, M5,M5e models only) or wavelength-selectable luminescence modes, and 400–750nm for readings in fluorescence polarization mode(M5, M5e models only).
Assaysrequiring a read in two ormoremodes can be combined and run on the SpectraMax Multi-ModeMicroplateReaders by issuing a single command in SoftMax® Pro Software, Molecular Devices’ leading microplate data acquisition and analysis software platform.
SpectraMaxM3SpectraMax M4SpectraMaxM5SpectraMax
XX
XXX
M5
e
0112-0115 F7
Description
Applications
Endpoint, kinetic,spectrum, and multi-pointwell-scanning applications combining absorbance and fluorescence in 6-wellto384-well microplates, as well as endpoint, kinetic, and spectrum applicationsin absorbance and fluorescence using cuvettes, can be run with little to no optimization.
The extreme flexibility and high sensitivity of the SpectraMax Multi-Mode Microplate Readersmake them appropriate for applications within the fields of biochemistry, cell biology, immunology, molecular biology,and microbiology.
Typical applications include ELISA,nucleic acid,protein,enzymatic typehomogeneous and heterogeneousassays, microbial growth, endotoxin testing, and pipettor calibration.
Certified SpectraMax® M5e-HTRF Readers
The SpectraMax M5e reader has the same performance specificationsas the M5 but is certified for use with Cisbio Bioassays’ HTRF (Homogeneous Time-Resolved Fluorescence) technology. HTRF is a proprietary time-resolved fluorescence technology that overcomes many of the drawbacks of standardFluorescence Resonance EnergyTransfer (FRET)techniques, suchas the requirements to correct for autofluorescence and the fluorescent contributions ofunboundfluorophores.
Optics
The use of two holographic diffraction grating monochromators allowsfor individual optimization of wavelengths for both excitation and emission in fluorescence readings. Mirrored optics focus the light into the sample volume, and cutoff filters are used toreduce stray light and minimize background interference. The light source is a high-powered Xenon flash lamp. Sensitivity or read-speed can beoptimized by varying the number of lampflashes per read.
Dynamic Range
The dynamic range of detection is from 10-6 to 10Variations in measured fluorescence values are virtually eliminated by internal compensation for detector sensitivity, photomultiplier tube voltage and sensitivity,as wellas excitation intensity. The photometric range is 0–4 ODs with a resolution of 0.001 OD.
A SpectraMax Multi-Mode Microplate Reader with PathCheck Pathlength Measurement Technology allows normalization of variable well volumes to 1-cm cuvette readings. PathCheck Pathlength Measurement Technology allows for multichannel pipettor validation and for experiment comparison from different days.
Automix
Usingthe Automix feature of the SoftMax Pro Software, the contentsof the wells in a microplate can be mixed automatically by linear shaking before each read cycle, making it possible to perform kineticanalysis ofsolid-phase,enzyme-mediated reactions (mixing is notcritical for liquid-phase reactions).
Temperature Control
Temperature in the microplate chamber is isothermal, both at ambient and when the incubator is turnedon.When the incubator is on,the temperature may be controlled from 2°C above ambient to 60°C.
Supported Plates
Microplates having 6, 12, 24, 48, 96, and 384 wells can be used in the SpectraMax Multi-Mode MicroplateReaders. Top and bottom reads are available for fluorescence, time-resolved fluorescence and luminescence detection. When reading optical density at wavelengths below 340nm, special UV-transparent, disposableor quartzmicroplates and cuvettes that allow transmission of the far UV spectra must be used.
One plate carrier adapter isprovidedwith the instrument. The adapter is required for optimum performance with standard 96-well and 384-well format microplates for all top-read applications.
0112-0115 F9
Description
Computer Control
An external computer running SoftMax Pro Software, which provides integrated instrument control, data display, and statistical dataanalysis, controls the SpectraMax Multi-Mode Microplate Readers. Cuvette port functionality can also be controlled using SoftMax Pro Software.
SoftMax Pro Software provides the following functionality:
Instrument Control
SoftMax Pro Software allows you to set up and run a complete protocolfor the SpectraMax Multi-Mode Microplate Reader, as well as all other Molecular Devices' microplate readers.Instrument settings can be saved as a protocol file and used repeatedly for reading different microplates or cuvettes. All stand-alone instrument functions can be controlled using the software. In addition, SoftMax Pro Software provides capabilities that are notavailable when using an instrumentinstand-alone mode such as user-defined kinetic run times, read intervals, Automix parameters, etc.
Data Collection and Display
SoftMax Pro Software collects and stores all raw data received from the instrument. Data is displayed in agrid format that correspondsto the wells in a microplate or individual cuvettes.
SoftMax Pro Software can collect data from one or more microplates orcuvettes and store it in a single data file, using the same or differentinstrument settings for different microplates or cuvettes. For example,microplates containingdifferent samples can be read using the same or differentmodes, all within the same experiment.
Data Reduction and Plotting
You can manipulate or “reduce” the raw data using dozens of built-in formulas ordefine yourown analysis structure to quickly and easily summarize the raw data. Morethanone reduction can be shown, andresults from different microplates orcuvettes can be compared within the same experiment.
Immediate Results Reporting and Analysis
Once you have defined instrument settings, and have customized a SoftMax Pro Software data file with assay information, reduction settings, custom columns in Group sections, and summary objects, you can save this information to create an assay protocol. Protocols can be used throughout a department or company for highly repeatabledata collection and analysis that is completedthe second the plate read has completed.
10 0112-0115 F
Reader Components
Control PanelBack PanelCuvette Chamber
Microplate Drawer
The main components of the SpectraMax Multi-Mode Microplate Readers are:
•Control panel:for cuvette chamber control.
•Microplate drawer: used for all fiveread modes and four read
types.
•Cuvette chamber: used for absorbance, fluorescence intensity,
and luminescence read modes for endpoint, kinetic, and spectrum scanning.
The control panel consistsof a 2-x-20-character LCD and eleven pressure-sensitive membrane keys that can be used to control somefunctions of the instrument. When you press a control panel key, the instrument performs the associated action.
Note:Settings made inSoftMax Pro Software override control panel settings.
The left side of the display shows the temperature inside the cuvette chamber, both actual and set point, and whether or notthe temperature is at the set point (the enunciator blinks if it is not at set point). The temperature of the microplate chamber lags slightly behind the temperaturein the cuvette chamber. The temperature in themicroplate chamber is reported inthe SoftMax Pro Software interface display.
The middle of the display shows the wavelengths for absorbance/excitation and emission.
The right sideof the display shows the data received fromthe readingas absorbance, percent transmission, fluorescence emission or excitation, or luminescence, and indicates whether or nota reference measurement was made(enunciator blinks if no reference reading was taken).
The key enablesand disablesthe incubator that controls the temperature within both the microplate chamber and the cuvette port.
•When the incubator is on,the set temperature and actual
temperature (cuvette chamber only) are shown on the frontpanel LCD display.
•When the instrument is performing a kinetic or spectral scan,
the temperature keys on the front panel are disabled.
Temp
The keys allow you to enter a set point at which to regulate the cuvette and microplate chamber temperature. However, rememberthat the cuvette temperature only is reported on the LCD display, while the microplate chamber temperature is reported in the SoftMax Pro Software interface display.
Pressing this key scrolls the temperature up or down, starting at theprevious temperature setting (or the default of 37.0°C,ifno setting had been made):
•Pressing the up () or down () arrow once increments or
decrements the displayed temperature by0.1°C.
•Pressing and holding either arrow increments or decrements the
displayed temperature by1°Cuntil it is released.
You cannot set a temperature beyond the upper (60°C) or lower (15°C) instrument limits.
Wavelengths ()
Selects the wavelengthto be used for reading the cuvette manually. Two sets ofup or down arrow keys are available for setting absorbance/excitation (fluorescence) wavelengths and emission (fluorescence) wavelengths.
The controlpanel does not display the wavelength selected through the SoftMax Pro application.
Pressing the up or down arrowkey scrolls up or downthrough the available wavelengths, starting at the previoussetting:
•Pressing the up () or down () arrow once increments or
decrements the displayed wavelength by 1nm.
•Pressing and holding either arrow increments or decrements the
displayed wavelength by10nmuntil it is released.
0112-0115 F13
Description
DRAWER
Ref
A reading of buffer, water, or air taken in the cuvette that is used as I0to calculate Absorbance or%Transmittance. If noreference reading is taken, the instrument uses the I0 values stored in the NVRAM (non-volatile memory) of the instrument.
This key is disabled during a computer-controlled run.
Read Cuvette
Initiates the sample reading of the cuvette.This key is disabled during a computer-controlled run.
Mode
A toggle switch used to display cuvette data aspercenttransmittance (%T), absorbance (A),relative fluorescence units(RFU),or relativeluminescence units(RLU).
Drawer
The key opens and closes (toggles) the microplate drawer.
The microplate drawer is located on the right side of the instrument and slides in and out of the reading chamber. An internal latch positions the microplate in the drawer as it closes (allowing for better robot integration-no springs or clips are used).
The drawer remains in the readingchamber during readcycles.
Figure 1-3The microplate drawer.
Microplate drawer operation varies, dependingon the incubator setting:
•If the incubator isoff, the drawer remains open.
•If the incubator is on, the drawer closes after approximately 10
seconds to assist in maintaining temperature control withinthemicroplate chamber.
Do not obstruct the movementof the drawer.If you must retrieve a plate after an error condition or power outage andthe drawer doesnot open, it is possibleto open it manually(see
page61
0112-0115 F15
).
Troubleshooting on
Description
Microplates
The SpectraMax Multi-Mode Microplate Reader can accommodate SBS-standard 6-well to 384-well microplates and strip wells. When reading optical density at wavelengths below 340nm, special UV-transparent, disposableor quartzmicroplates allowing transmission of the deep UV spectra must beused.
Not all manufacturers' microplates are the same with regard todesign,materials, or configuration. Temperature uniformity within the microplate may vary depending on the type of microplate used.
Microplates currently supported by the SoftMax Pro Software for use in this instrument are:
The SoftMax Pro Software plate listalso includes half area and low-volume plates. SoftMax Pro can always be used to define a new plate type using the manufacturer's specifications for wellsize, spacing and distance from the plate edge.
Locatedat the right front of the SpectraMax instrument, the cuvette chamber has a lid that lifts up, allowing you to insert or remove a cuvette. The chamber contains springs that automatically position the cuvette in the proper alignmentfor a reading.The cuvette door mustbe closed beforeinitiating areading.
Cuvettes
The SpectraMax Multi-Mode Microplate Reader can accommodate standard-height(45 mm), 1cm cuvettesand12x75 mm test tubes when used with the test tube cover.
Not all manufacturers' cuvettes are the same with regard to design,materials, or configuration. Temperature uniformity within the cuvette may vary depending on the type of cuvette used.
Cuvettesused for absorbance readingsare frosted on two sides.Be sure to handle cuvettes on the frosted sides only. Place the cuvette intothe chamber so that the “reading” (clear) sides face left and right.
Fluorescence cuvettes are clear on all four sides and should be handledcarefully. Place a frosted cuvette into the chamber so that the “reading” (clear) sides face left and right.Semi-Micro and Ultra-Micro cuvettes can also be used withan adapter. See information about supported cuvettes.
0112-0115 F17
Cuvettes on page78 for more
Description
Power Switch
Power Cord
Receptacle
Label
Fuse Box
Cover
Computer
Port
Printer
Port
Figure 1-5The test tube cover.
The Back Panel
18 0112-0115 F
Figure 1-6Schematic of the back panel of a reader.
The following components are located on the back panel of the SpectraMax instrument:
•Power switch:a rocker switch, labeled I/O (for on and off,
respectively).
•Power cord receptacle: plug the power cord in here.
•Fuse boxcover: cannot be opened while the power cord is
plugged in. When opened, it provides access to the fuse box containing twofuses that are required for operation.
•Computer port (double-shielded 8-pin RS-232 serial, for use with an
external computer): plug one end of an 8-pin DIN serial cable into this port; the other end attaches tothe serial (modem) port of the computer.
•Printer port: not used for the SpectraMax instrument
•Label: provides information aboutthe reader, such as line
voltage rating, cautionary information, serial number, etc. Record the serial number shown on this label for use when contacting Molecular Devices Technical Support.
Principles of Operation
Absorbance
Note:In this user guide, references to the SpectraMax® readers
include the M3, M4, M5, and M5e models. When a feature or capability applies to only certain readers, this exception is noted.
Absorbance is the amount of light absorbed by a solution. Tomeasure absorbance accurately, it is necessary to eliminate light scatter. In the absence of turbidity, absorbance = optical density:
where I is transmitted light, andIO is incidentlight.In this manual, we use the terms absorbance and optical density
interchangeably.
Optical Density
Optical density is the amount of light passing through a sample to a detector relative to the total amount of light available.Optical density includes absorbanceof the sample plus light scatter from turbidity.
2
A = –log(I/IO)
Transmittance
Transmittance is the ratio of transmitted light to the incident light.
T = (I/IO)
%T = 100T
where I is transmitted light, andIO is incidentlight.
PathCheck® Pathlength Measurement Technology
The Beer-Lambert law states that absorbanceis proportional to the distance that light travels through the sample:
A =
bc
where A is the absorbance, is the pathlength, and c is the concentration of the sample. In short, the longer the pathlength, thehigher the absorbance.
0112-0115 F19
is the molar absorptivity of the sample,b
Principles of Operation
Horizontallight path
Vertical light path
CuvetteMicroplate wells
Microplate readers use a vertical light pathso the distance of the light through the sample depends on the volume. This variable pathlengthmakes it difficult to perform extinction-based assays and also makes it confusing to compare results between microplate readers and spectrophotometers.
The standard pathlength of a cuvette is the conventional basis forquantifying the unique absorbtivity properties of compounds in solution. Quantitativeanalyses can be performed on the basis of extinction coefficients, without standard curves (forexample,NADH-based enzyme assays). When using a cuvette, the pathlength is known and isindependent of samplevolume, so absorbance is proportional to concentration.
In a microplate, pathlength is dependent on the liquid volume, so absorbance is proportional to both the concentration and the pathlength of the sample. Standard curves are often used to determineanalyte concentrations in vertical-beam photometry of unknowns, yet errors can still arise from pipetting the samples and standards. The PathCheckdetermines the pathlength of aqueous samples in the microplateand normalizes the absorbance in each well to a pathlength of 1cm. Thisnovel approach tocorrecting the microwell absorbance values isaccurate towithin 2.5%of the values obtaineddirectlyina 1cmcuvette.
Reference measurements made by reading the cuvette (Cuvette Reference) or using factory-stored values derived from deionized water (Water Constant) can beused to normalize the optical density data for microplate wells.
Pathlengthcorrection is accomplishedonly when using the PathCheck Pathlength Measurement Technology with SoftMax
®
Pro Software.PathCheck Pathlength Measurement Technology is patentedbyMolecular Devices and can beperformed only on an Molecular Devices plate reader.
The SpectraMax Multi-ModeMicroplate Reader offers both the Cuvette Reference and the Water Constant methods.
The actual pathlength, d, of a solvent is found from the following equation:
When a Cuvette Reference is usedfor pathlengthcorrection, the valueof k is obtainedby taking optical density measurements on the fluid inthe cuvette at two wavelengths, 1000 and 900nm:
When the WaterConstant is used for pathlength correction, the value ofk is obtained fromthe instrument. This constant is saved in the instrument in the factory and may differ slightly from instrument to instrument.
Once the pathlength d is found,the followingequation is used for the pathlength correction:
PathCheck Pathlength Measurement Technology is applicable toalmostallbiological/pharmaceutical molecules in aqueous solution because they have little or no absorbance between 900nmand 1000nm atconcentrations normally used. PathCheck Pathlength Measurement Technology can also beused withsamples containing small amounts oforganics or high buffer concentrations by using the Cuvette Reference. See
Water Constant or Cuvette Reference?on page22.
0112-0115 F21
Principles of Operation
Water Constant or Cuvette Reference?
The PathCheck Pathlength Measurement is based on the absorbance of water in the near infrared region (between 900nm and 1000nm).If the sample is completelyaqueous,has no turbidity and has a low salt concentration (less than 0.5M), the Water Constant isadequate. The Water Constant is determined duringmanufacture and is stored in the instrument.
If the sample contains an organic solvent such as ethanol or methanol, we recommend using the cuvette reference. It is important thatthe solvent does not absorb in the 900nm to 1000nm range (to determine whether or not a given solvent would interfere, see the discussion ofinterfering substances below).When a non-interference solvent isadded to the aqueous sample, the water absorbance decreases proportionally to the percentage of organic solvent present. For example, 5% ethanol decreases the water absorbanceby 5% and results in a 5% underestimation of the pathlength. You can avoid the error by putting the same water/solvent mixture in a cuvette and using the CuvetteReference.
Touse the Cuvette Reference, placeinto the cuvette port a standard1cm cuvette containing the aqueous/solvent mixturethat is used forthe samples in the microplate. The cuvette must bein placewhen you read the microplate. When you click the Read button in the SoftMax Pro program, the instrument firstmakes the 900nm and 1000nm measurements in the cuvette, and then makes the designated measurements in the microplate. The cuvette values are stored temporarily and used inthePathCheck PathlengthMeasurement Technology calculations for the microplate samples.
Use ofCuvette Reference with PathCheck Pathlength Measurement Technology is different from a reference reading of a cuvette in aCuvetteSet section (by clicking the Ref button in the CuvetteSet section tool bar inthe SoftMax Pro program). The cuvette reference used for PathCheck Pathlength Measurement Technology calculations(measurements at900nm and 1000nm) does notproduce data thatcan be viewed in a CuvetteSet section and is used onlywith data in microplates, not cuvettes.
Background Considerations
Raw optical density measurements of microplate samples include bothpathlength-dependent components (sample and solvent) and a pathlength-independent component (OD of microplate material). The latter must beeliminated from the PathCheckPathlength Measurement Technologycalculation inorder to obtain PathCheck Technology-normalized results. Thereare3 ways to accomplish this: plate blanks, plate background constants, and plate pre-reads, all of which are described in the PathCheck Pathlength Measurement Technology section of the SoftMax Pro User Guide.
PathCheck Pathlength Measurement Technology and Interfering Substances
Any material that absorbs inthe 900nm to 1000nm spectral region could interfere with PathCheck Pathlength Measurement Technology measurements. Fortunately, there are few materials that do interfere at the concentrations typicallyused.
Turbidity is the most common interference: if you can detect any turbidity in your sample, you should not use the PathCheck Technologyfeature.Turbidity elevates the 900nmmeasurement more thanthe1000nm measurement and causes an erroneously low estimate of pathlength. Using Cuvette Reference does not reliably correct forturbidity.
Samples that are highly colored in the upper visible spectrum may haveabsorbance extending into the nearinfrared(NIR) and can interfere with the PathCheck Pathlength Measurement Technology.Examples include Lowry assays, molybdate-based assays and samples containing hemoglobins or porphyrins. In general, if the sampleis distinctly red or purple, you should check for interference before using the PathCheck Pathlength Measurement Technology.
To determine possible color interference, do the following:
•Measure the optical density at 900nm and 1000nm (both
measured with air reference).
•Subtract the 900nm value from the 1000nm value.
•Do the same for pure water.
If the delta OD for the sample differs significantly from the delta OD for water,then it is advisable not to use the PathCheck Technology feature.Use of Cuvette Reference does not correct for the interference with thecurrent calculation scheme in the SoftMax Pro program. Currently,CuvetteReference involves a single (automated) read at 900nm and 1000nm and the automated calculations in the SoftMax Pro program do not compensate for color or solvent interference. However, you could correct for suchinterference by taking two cuvette measurements and using a different set of calculations. For further information, contact Molecular Devices Technical Support.
Organic solvents could interfere withthe PathCheck Technology feature if theyhave absorbance intheregionof the NIR water peak.Solventssuch as ethanol and methanol do not absorb in the NIR region,so they do not interfere, except for causing a decrease in the water absorbance to the extent of their presencein the solution. Their passive interference can be avoided by using the Cuvette Reference. If,however, the solvent absorbs between 900 and 1000nm, the interference would be similar to the interference of highly colored samples described above. If you are considering adding an organic solvent other than ethanol ormethanol, you are advised to run aspectral scan between 900nm and 1000nm to determine if the solventwould interfere with the PathCheck Technologyfeature.
0112-0115 F23
Principles of Operation
Normalizing Absorbance Measurements
SoftMax Pro Softwareautomatically reports absorbance values normalized to a 1-cm pathlength.SoftMax Pro Software automatically reports absorbancevalues normalized toa 1-cm pathlength. The table below shows results obtained with 75µL to 300µL yellow reagent.
Table2-1Yellow reagent results.
Well Volume
Optical pathlengths and raw absorbance values were directly proportional to well columns. After normalization to a 1-cmpathlength, all absorbance values, regardless of the volumein the wells, were within 1% of the value obtained by measuringthe same solution in a 1-cm cuvette.
Fluorescence
Fluorescent materials absorb light energy of a characteristic wavelength (excitation), undergo an electronic state change, and instantaneously emit light of a longer wavelength (emission). Most common fluorescentmaterials havewell-characterized excitation and emission spectra. Figure2-2 shows an example of excitation and emission spectra for a fluorophore. The excitation and emission bands are each fairly broad, with half-bandwidths of approximately 40nm, and the wavelength difference between the excitation and emission maxima (the Stokes shift) is typically fairly small, about 30nm. There is considerable overlap between the excitation andemission spectra (gray area) when a small Stokes shift is present.
Because the intensity of the excitation light is usually many tens of thousands of times greater thanthat of the emitted light,some type ofspectral separation is necessary to reduce the interference of the excitation light with detection ofthe emitted light. The SpectraMax Multi-Mode Microplate Reader incorporates many features designed to restrict interferencefrom reflected excitation light. Among these features is a set of long-pass emission cutoff filters that can be set automatically by the instrument or manually bythe user. If the Stokes shift is small, it may be advisable to choose an excitation wavelength that is as far away from the emission maximum as possible while stillbeing capable ofstimulating the fluorophore so thatlessof the excited light overlaps the emission spectrum, allowing better selection and quantitation of the emitted light.
0112-0115 F25
Principles of Operation
Wavelength (nm)
Relative Fluorescence
Excitationreadingwavelength
Emissionreadingwavelength
Fluorophore’sexcitationmaximim
Fluorophore’semissionmaximim
500550600650
0
0.5
1.0
Figure 2-3Optimized excitation and emission reading wavelengths.
Figure2-3 shows that the best results are often obtained when the excitation and emission wavelengths used for reading are not the sameas the wavelengths of the excitation and emission spectra of the fluorophore. When the reading wavelengths for excitation and emission are separated, a smaller amount of excitation light passes through tothe emission monochromator(gray area) and on to thePMT,resulting in a purer emission signal and more accurate data.
The SpectraMax Multi-Mode Microplate Reader allows scanning of bothexcitation and emission wavelengths, using separate tunable monochromators. One benefit of being able to scan emission spectra isthat you can assess more accurately whether the emission is, infact, the expected fluorophore, or multiple fluorophores, and not one generated by a variety of background sources or by contaminants. Another benefit is that you may be able to find excitation and emission wavelengthsthat avoid interferencewhen interferingfluorescentspecies are present.
For this reason, it may be desirable to scan emission for bothan intermediate concentration of labeled sample, as well as the backgroundof unlabeled sample. The optimum setting is where the ratio of the sample emission to background emission is at themaximum.
For more information regarding optimizing excitation and emission wavelengths using the spectral scanning capabilities of the SpectraMax series readers, see
Optimizing Absorbance Assays on page44.
Time-resolved Fluorescence (M4, M5, and M5e only)
In normal fluorescence mode, the SpectraMax readings are taken while the lamp ison. The mostcommon limitation tosensitivity innormalfluorescence is excitation energy or background fluorescence that cannot be eliminated from the emission signal. Since the lamp is the source of excitation energy, turning itoff provides the best means of eliminating background excitation. The elimination of background excitation is the critical difference betweenfluorescence intensity measurements and TRF measurements.
Time-resolved fluorescence is performed by flashing the excitation lamp and, after it is off,collecting the delayed emission for a period of time before the lamp is flashed again. Long-lifetime rare-earth lanthanide dyes are typically used to provide a long-lived fluorescent signal that persists after the lamp is turned off. Backgroundfluorescence usually fades after 50µs, while lanthanide chelates and cryptates have fluorescent lifetimes between 100µs and 2ms.
Tooptimize data collection for a particular assay, the user can select when to start and end data acquisition-the minimum is 50µs after the lamp has been turned off, and the maximum is 1450µs, in 50- or 200-µs steps.
By using a fluorescent dye to label a small molecule, its binding to alarge molecule can be monitored through its speed of rotation.
Fluorescence Polarization mode returnstwo setsof data: one for fluorescence intensity parallel (P) tothe excitation plane, and the other for fluorescence intensity perpendicular (S) to the excitation plane. These S and P values are used tocalculatethepolarization (mP) and anisotropy (r) values inSoftMax Pro Software. Although the Raw S&P value is the true raw data returned from the instrument, the calculated polarization (mP) and anisotropy (r) values are treated as the raw data, and these values become the basis for further reduction calculations in SoftMax Pro Software.
Polarization (mP) is calculated as follows:
Anisotropy (r) is calculated as follows:
Luminescence
Luminescence is the emission of lightby processes that derive energy from essentially non-thermal changes, the motion of subatomic particles, or the excitation of an atomic system by radiation.
When the SpectraMax Multi-Mode Microplate Reader is in luminescence mode, no excitation is necessary as the species being measured emit light naturally.For this reason, the lamp does not flash, so no background interference occurs. A dark estimate is done over a dark reference, and multiple readings are averaged together intoone reading per well.
The default setting for luminescence is the “zero order” position where the grating monochromator acts as a mirror thatreflects all light tothe PMT detector. If wavelengthselection is desired, you can choose the wavelength where peak emission is expected to occur. In addition, multiple wavelength choices allow species with multiple components to be differentiated and measured easily. In luminescence read mode, no emission cutoff filter is used.
The full power of aSpectraMax Multi-Mode Microplate Reader can only be harnessed when the instrumentis controlled using SoftMax® Pro Software running on a computer connected to theinstrument. For a complete description of themodes of operation, howto choose instrumentsettings, etc. refer to the SoftMax® Pro Software User Guide.
However, some functionality is available directly on the instrumentwithout having to use SoftMax Pro Software:
•Temperature control
•Wavelength control
•Fixed-point cuvette readings
Temperature Regulation
The SpectraMax Multi-Mode MicroplateReaders havebeen designed to regulate the temperature of the cuvette and microplatechamber from2°C above ambientto60°C. Upon power up, when the incubator is off,the temperaturein the chambers isambient and isothermal. Turningon the incubator bypressing the key causes the instrument tobegin warming the cuvette and microplate chambers. The temperature set point defaults to 37.0°C atstart-up.
Accuracy of the temperature set point is guaranteed only if the setpoint is at least 2°C above ambient. If the temperature set point is lower than the ambienttemperature, the chamber temperature remainsat ambient.Temperature regulation is controlled by heaters only and, therefore, cannot cool the temperature to a setting lower than ambient. Additionally, the highest setting (60°C) can be achieved only if the ambienttemperature is greater than 20°C.
Typically, the cuvette and microplate chambers reach 37.0°C inless than 30 minutes. The temperature is maintained at the set point untilyou press the incubator key again, turning temperature regulation off.
Ifyou turn the incubator back on aftera momentary shutdown, allow about ten minutes for the control algorithm tofully stabilize the temperature.
Temperature regulation and control is achieved through electric heaters, a fan, efficient insulation, and temperature sensors. The heaters are located in the microplatechamber, which is insulated to maintain the temperature set point. The sensors are mounted inside the chamber and measure the air temperature.
0112-0115 F29
Principles of Operation
The temperature feedback closed-loopcontrol algorithms measure the chamber air temperature, compare itto the temperature set point, and use the difference to calculate the regulation of the heating cycles. Thistechnique results inaccurate, precise control of the chamber temperature with a temperature variation of the air inside the chamber of less than 1.0°C. The temperature uniformitywithin the microplatedepends on its design and composition.
Read Types
The SpectraMax Multi-Mode Microplate Reader can performfour typesofread: endpoint,kinetic, spectrum and well scan. Instrument setupparameters for each read type are discussed inthe SoftMax® Pro Software User Guide.
Endpoint Read
In an endpoint read, a reading of each microplate well is taken at a single or multiple wavelengths.
Depending on the data mode selected in the Reduction window,values can be reported as optical density or %Transmittance.
Kinetic Read
In a kinetic read the data are collected over time with multiple readings taken at regular intervals. Toachieve the shortest possible interval for kinetic readings, choose wavelengths in ascending order.
Kinetic analysis can be performed for upto 99 hours. The kinetic read interval depends upon the instrument setup parameters chosen in SoftMax Pro Software.
Kinetic analysis hasmany advantages when determining the relativeactivityof an enzyme indifferent types of microplate assays, includingELISAs and the purificationandcharacterization ofenzymes and enzyme conjugates. Kinetic analysis is capable of providing improved dynamic range, precision, and sensitivity relative to endpoint analysis.
Spectrum Read
Spectral analysis measures optical density or % Transmittance across a spectrum of wavelengths 200nm to 1000nm. For fluorescence or luminescence mode,relative fluorescence units (RFU) or relativeluminescenceunits (RLU) values are reported.
All spectrum readings are made using the scanning monochromatorsof the instrument.
A well scan read takes one or more readings of a single well of amicroplate at single or multiple wavelengths. Every option available for endpoint reads is available for well scans.
Some applications involve the detection of whole cells in large-area tissue culture plates. Wellscan mode can be used with such microplates to allow maximum surface area detection in whole-cellassays.Since many cell lines tend to grow as clumps or in the corners of microplate wells, you can choosefrom several patterns and define the number of points to be scanned to work best with your particular application.
Values reported are optical density, % Transmittance, relativefluorescence units (RFU), or relative luminescence units (RLU).
Automix
The Automix function permits automaticlinear shaking along the long axis of the microplate at preset intervals, thereby mixing the contents within each well. Automix must be selected before beginninga reading. The actions associated withtheAutomix settingdependon the readmode chosen:
•Endpoint mode: Automix shakes the plate for a definable number
of seconds and then readsat all selected wavelengths.
•Kinetic mode: two types of Automix can be enabled: Automix can
shake the plate for a definable number of seconds before the initial reading, and/or for a definable number of seconds before each subsequent reading.
Use of Automix is strongly recommended for ELISAs and other solid-phase, enzyme-mediated reactions to enhance accuracy.
Computer Control
The SpectraMax Multi-Mode MicroplateReaders are equipped with an 8-pin DIN RS-232 serial port through which the computer communicates with the instrument using the SoftMax Pro Software. Different types of cables are available for connecting to different typesof computers. See
0112-0115 F31
AppendixB: Cables and Accessories on page77.
Principles of Operation
32 0112-0115 F
Installation
WARNING!Always make sure the power switch on the instrument is in the OFF position and remove the power cord from the back of the instrument prior to any installation or relocation of the instrument.
WARNING!Do not operate the instrument in an environment where potentially damaging liquids or gases are present.
CAUTION!Do not operate the instrument in a coldroom with a
temperature below 15°C.
3
CAUTION!Do not touch or loosen any screws or parts other than those
specifically designated in the instructions. Doing so might cause misalignment and voids the instrument warranty.
0112-0115 F33
Installation
Unpacking
Note:In this user guide, references to the SpectraMax® readers
include the M3, M4,M5, and M5e models. When a feature or capability applies to only certain readers, this exception is noted.
The SpectraMax Multi-Mode Microplate Readers are packed in a specially designed carton. Please retain the carton and the packing materials. If the unit should needto be returned for repair, you must use the original packing materials and carton for shipping. If the carton has been damaged in transit, it is particularly important that you retain itfor inspection by the carrier in casethere has also been damageto the instrument.
WARNING!The SpectraMax Multi-Mode Microplate Reader weighs approximately 36 pounds (16.4 kg) and should be lifted with care. It is recommended that two people lift the instrument together, taking the proper precautions to avoid injury.
After examining the carton, place iton a flat surface in the upright position. Open the top of the box and lift the accessory kit out. Open the accessory kit box and check that all parts are accounted for:
•Plate adapter (purple)
•Reader dustcover
•Test tube cover
•Hex wrench,3/32˝, ball drive, L
•Mouse pad, SpectraMax instrument
•Cable, PC–SpectraMax, 9 pin–8 pin mini
•Country-specific Power cord
•Fuses, 4-amp (2 ea.)
•SpectraMax User Guide
•Applications guide to microplate systems
Makesure all these items are present before proceeding.Remove the cardboard divider from the top of the SpectraMax. Liftthe
reader up and out of the shipping box and set it down carefully.
1.Place the instrument on a level surface, away from direct
sunlight, dust, drafts, vibration, and moisture.
2.Turn the instrumentaround so that theback of theinstrument isfacing you as shown in
of a reader. on page18
3.Insert the round end of the serial cable into the RS-232 serialport on the back panel of the instrument. (A Keyspan USB adapter is necessary for a Macintosh computer or a Windows computer without a serial port; see information on adapter cables.)Attach the other end to your computer.
4.Insert the female end of the power cord into the power receptacle at the rear of the instrument. Connect the male end to a grounded power outlet of the appropriate voltage. Molecular Devices recommends that you use a surge protector between the power cord and the grounded power outlet.
5.Turn the instrument around sothat the control panel now faces you. Ensure no cables run beneaththeinstrument. Leave at least three inches between the back of the instrument and the nearest objects or surfacesto ensure proper ventilation and cooling.
6.Remove the tape from the cuvette door.
7.Turn on the power to the instrument, wait for the microplate
drawer to open, and remove the tape and protective covering from the drawer subplate.
Figure1-6 Schematicof the back panel
.
AppendixB for more
0112-0115 F35
Installation
DRAWER
Installing the Drawer Adapter
CAUTION!Incorrect insertion or removal of the adapter may cause
damage to the microplatedrawer of the SpectraMax instrument. Thecorner cutoutmust be in the lower left corner where the plate pusher is located.
If you are reading standard 96-well or 384-wellmicroplates from the top, you need to installthe drawer adapter.
1.Power on the instrument using the switch on the back panel.
2.Press the button on the front panel or select the
Control > Open Drawer command in SoftMax® Pro Software.
3.Hold the adapter so that the label is on the front side facing up.
4.Place the top back (RowA) portion of the adapter into the
drawer first. The corner cutout mustbe in the lower left cornerwhere the plate pusher is located. While pushing against the back edge of the adapter,lower the front of the adapter into the drawer.
If the adapter is in the drawer and you are either reading from the bottom or using “high profile” (6-well, 12-well, 24-well, or 48-well) plates, you need to remove the adapter.
Incorrect insertion or removal of the adaptermay cause damageto themicroplate drawer of the SpectraMax instrument.
1.Power on the instrument using the switch on the back panel.
2.Press the button on the front panel or select the
Control > Open Drawer command in SoftMax Pro Software.
3.Remove the adapter plate.
Figure 3-2The microplate drawer without adapter.
0112-0115 F37
Installation
38 0112-0115 F
Operation
TEMP on/off
REF
READ CUVETTE
Cuvette Read—Quick Overview
If you are an experienced userof this instrument, the following steps provide a quick reminder of the basic operating procedures required to read a cuvette using a SpectraMax Multi-Mode Microplate Reader:
1.Turn on the power switch(locatedon the back panel). The microplate drawer opens automatically.
2.If you wantto regulate the temperature inside the chamber,touch the (incubator) key to turn the incubator on
and bring the chamber to the default temperature of 37.0°C. The microplate drawercloses.
3.If the incubator is on, the LCD shows the current temperature of the cuvette chamber along withthe temperature set point. Tochangethe set point (to any setting from ambient +2°C to60°C), press the up or down arrow keys.
4.Select the desired measurement wavelength by pressing the up or down arrownear
5.Load the prepared cuvette into the chamber, being sure that the clear sides are left and right (when facing the instrument).
6.Press the or key.
.
4
0112-0115 F39
Note:In this user guide, references to the SpectraMax® readers include the M3, M4, M5, and M5e models. When a feature or capability applies to only certain readers, this exception is noted.
Operation
TEMP on/off
Microplate Read—Quick Overview
If you are an experienced userof this instrument, the following steps provide a quick reminder of the basic operating procedures required to read a microplate using a SpectraMax Multi-Mode Microplate Reader:
1.Turn on the power switch(locatedon the back panel). The microplate drawer opens automatically.
2.If you wantto regulate the temperature inside the chamber,touch the (incubator) key to turn the incubator on
and bring the chamber to the default temperature of 37.0°C. The microplate drawercloses.
3.If the incubator is on, the LCD shows the temperature set point,but the currenttemperature in the microplate chamber is reported in SoftMax® Pro only—the LCDreports the cuvette chamber temperature. Tochange the set pointto any setting fromambient +2°Cto 60°C, press the up or down arrow keys.
4.Select the desired instrument settings (read mode, type ofanalysis, template, etc.) using SoftMax Pro Software on the external computer.
5.If you are performing kinetic analysis, add substrate at thistime.
6.Load the prepared microplate into the drawer, being sure to match well A1 with the A1 mark on the upper left-hand corner ofthe drawer.
7.Using SoftMaxPro Software,startthereading by selecting the Control > Read commandor clicking the Read button on the Plate
section tool bar.
Preparing for a Cuvette or Microplate Reading
40 0112-0115 F
Turn the Instrument and Computer On
The power switch is located on the back panel. Press the rocker switch to the ON position.
The instrument automatically performsdiagnostic checks to ensure that itisfunctioning correctly. Turn the computer on atthistime also and start the SoftMax Pro Software program.
Toset the temperature within the microplate or cuvette chamber, you should turn on theincubator first,allowing enough time for the temperature to reach the set point before performing a reading. When you first turn the instrument on, up to 60minutes may be required for the temperature within the chamber to reach the set point. Turning on the incubator and choosing a temperature set point can be done using the software or the frontpanelofthe instrument (described below). However, only the cuvette temperature is reported on the front panel;the SoftMax Pro program reports the current microplate chamber temperature, which lags very slightly behind the cuvette temperature.
Temperature cannot be regulated at a set pointthat is lower than2°C above the ambient temperature.
Toenable the incubator
1.Press the incubator key.
2.The LCD display indicates that temperature control is on and
shows the set point, and current temperature of the cuvette chamber only.
To change the temperature set point, press the up or down arrow keysuntil the desired temperature set point isshown in the display.
Tocontrol the temperature from the SoftMax Pro Software, use the Control > Incubator dialog box both toenablethe incubator and to setthe temperature.
The chamber temperature is maintained at the set point until you disable temperature control by touching the incubator key again. When the incubator is off, the temperature within the chamber gradually returns to ambient.
Shouldyou turn the incubator backon after a momentary shutdown, allow about ten minutes for the control algorithmto fully stabilize the chamber temperature.
0112-0115 F41
Operation
Select the Wavelength
The absorbance wavelength, or the excitation or emission wavelengths for a fluorescence or luminescence read, can be selected for cuvette reading using either the control panel or the SoftMax Pro Software. Thewavelengths mustbe selected using the software when reading microplates.
To select the wavelength using the control panel
1.Select the desired measurement wavelength by pressing the up or down arrownear .
2.Scroll through the wavelengths shown on the LCD to incrementor decrement the wavelength setting (in1-nm increments) until the desired measurement wavelength is reached.
3.If A or%T type is chosen, only the Abs/Ex wavelengthcan be set.
To select the wavelength using SoftMax Pro Software
1.Select the Plate > Settings command, or click the Settingsbutton in the appropriate Plate or CuvetteSet section.
2.In the Settings dialog box, select Wavelengths from the list of settings on the left-hand side.
3.Specify the number of wavelengthsto be read in each wellor cuvette.
4.Specify the excitation and emission wavelengthsto use.
5.Select Auto-cutoff if you would likethereader softwareto
automatically select the emission cut-off filter.
42 0112-0115 F
Read the Cuvette
REF
READ CUVETTE
1.Insert the cuvette intothe chamber,making sure that the clear
sides are to the left and right (facing the instrument). Do not touch the clear surfaces of the cuvette.
2.Make sure the cuvette is completely seated in the chamber and
close the cuvette door.
3.If the cuvette contains a blank (typically this solventcontains
everything that the samples contain except for analyte), press the key to acquire the reference reading from the cuvette.
The instrument automatically calibrates in less than two seconds, closes the microplate drawer (if it is open), and reads the cuvette according to the selected instrument settings.
4.If the cuvette contains a sample, touch the key to
acquire the sample reading from the cuvette.
5.When the reading is complete, removethe cuvette.
Read the Microplate
Note:The underside of the microplate must be dry prior to placing it in
the drawer. If the microplate has fluidon the underside, dry it using a paper towel (or equivalent) before placing it in the drawer.
1.Insert the filled microplate into the drawer, matching well A1
withpositionA1 in the drawer. Make sure the microplate is flatagainst the drawer bottom (for 6-, 12-, 24-, or 48-well microplates) or against the adapter (if using topread for 96- or 386-well plates-see for more information).
2.You must have SoftMax Pro Softwarerunning on a computer
connectedto the instrument. Open a SoftMax Pro data file orprotocol file that contains the
appropriate experiment settings for the plate read. Alternatively,create new settings by selecting the Plate section in the SoftMax Pro program andconfiguring the instrument using the Plate >Settings dialog box.
Installing the Drawer Adapter on page36
Operation
DRAWER
3.Select the Control > Read command or press the Read button in SoftMax Pro Software to start the plate read.
4.When reading is complete, the drawer of the instrument opens, allowing you to remove the microplate. If the incubator is on, the drawer closes again after approximately 10 seconds.
5.If you return to the SpectraMax instrument and find the drawer closed after a reading has finished, press the
When the drawer opens, you can remove the microplate.
For more information about configuring the software for platereading, please consult the SoftMax®Pro Software User Guide.
Optimizing Fluorescence Assays
The wavelengthof the transmitted light can be adjusted in 1-nm incrementsbetween 200nm and 1000nm. Guke also allows reading up to four wavelengths per plate. This enables reference wavelength readings such as A260 and A280 for nucleic determination.
An appropriate plate blank should be applied. Unless the user suspects that there issignificant well-to-wellvariability due to the thickness and optical propertiesof the plate, the use of Pre-Read Plate in the SoftMax Pro program isnot required. Instead, we recommend using appropriate plate blanks or group blanks inthe Template dialog box of the Plate section in the SoftMax Pro program. Fordiscussionof the different types of blanking, pleaserefer to the SoftMax®Pro Software User Guide.
If desired, the PathCheck Technology feature inSoftMax Pro program can be activated to normalize the data to a 1-cm pathlength.
key.
Optimizing Absorbance Assays
The optimum instrument settings for detection of a particular fluorophore depend on a number of different factors. Settings that can be adjusted for assay optimization include the excitation and emission wavelengths,emission cutoff filter, readings per well, the PMT voltage,and the temperatureofthe reading chamber.
Another important factor that is independent of the instrument butwhich affect assaysoptimizationisthe Stokes’ shift. When the Stokes' shift is very small, optimizing the excitation and emission wavelengths and correct cutoff filter choices are very important.
Excitationand Emission Wavelengths
The excitation and emission wavelengths may beset in1-nmincrements between 250nm and 850nm. A procedure to optimizeexcitation and emission wavelengths for a given assay is outlined below.
The emission cutoff filters assist in reducing background. Sources of background include stray excitation light and native fluorescence ofplate materials, sample constituents, and solvents (including water).The default setting allows the instrument and SoftMax Pro Software todetermine which cutoff filter should beused (see Table for default settings) in endpoint and kinetic modes. The spectral scan mode default uses no cutoff filter.
Readings Per Well
The number of readings per well may vary between 1 (used for a quick estimate) and 100 (for very precise measurements). The default number of readings per well varies withthe read mode: for fluorescence, the defaultis 6, and for luminescence the display shows 1 read per well.
PMT Voltage
The voltage of the photomultiplier tube may be set to low (for higher concentration samples), medium, or high (for lower concentration samples)in all read modes.In endpoint and spectrummode, there is an additional setting, automatic, in which the instrument automaticallyadjusts the PMT voltage for varyingconcentrationsof sample in the plate.
Temperature Control
The chamber of the SpectraMax Multi-Mode Microplate Reader isisothermal at ambient as well asat elevated temperatures. The temperature in the reading chamber may beadjustedfrom 2°C aboveambient to 60°C.
Note that assay optimization requires the use of a computer and SoftMax Pro Software.
0112-0115 F45
Operation
Using Spectral Scanningto Optimize Excitation and Emission Wavelengths for Fluorescence Assays
Put 200µL ofsample that includesthe fluorophore and 200µL of a buffer controlinto separate wells ofa microplate.
1.Perform the excitation scan:
UsingSoftMax Pro Software, set up a Plate section for a
fluorescence read, spectrum mode, Em Fixed/Ex Scan, withno cutoff filter (default), and medium PMT.
Set the emission wavelengthbased on the tentative value
from the literature (or from a customary filter set used to measure your fluorophore). If the emission wavelengthis not known, select a tentative emission wavelength about 50 nanometers greater than the absorbance maximum of the fluorophore.If necessary, the absorbance maximum can be determined by performing an optical density spectral scan first.
Set the excitation scan to start/stop approximately 50nm
below/above the tentative excitation value obtained from the literature (or the customary excitation filter).
Set the step increment to 2 or 3nm. (You may choose to do
a preliminary scan with a 10-nm increment to determinetheapproximate peak location, and then repeat the scan over a narrower wavelength rangewith a 2-nmor3-nm increment.)
Perform the scan and view the results as a plot of emission
fluorescence vs. excitation wavelength. Note the excitationwavelength at the emission peak and the maximum RFU value.
If an error message reporting missingdata points occurs, it may be due to possible saturation reported by the SoftMax Pro program at the end of the spectral scan. Reset the PMT to “low” and rescan the sample(scan the buffer blank with the PMT set to“medium” or “high”).If the error occurs after scanning with the PMT set to “low,” it may be necessary to dilutethe sample.
If the excitation scan showsno apparent peak, change the PMTsetting to “high” and rescan the sample. If the spectral scan still shows no apparent peak, adjust the Y-scale of thezoom plot so that the plot fills the graph.
Select the optimal excitation wavelength.If the excitation
peak wavelength and emission wavelength are separated by more than 80nm, use the excitation peak wavelengthvalue. If the excitation and emission wavelengths are less than80nm apart, use the shortestexcitation wavelength thatgives 90% maximal emission. (Follow the plot to the left of the peak until the RFU value falls toapproximately 90% ofthe maximum, and then drop a line from the 90% point on the plot to the x-axis—see Figure4-1.)
RFU at 90% of max λ
max λ
RFU
Wavelength
90% of max λ
Figure 4-1Plotof RFU vs.wavelength.
2.Perform emission scan #1:
In the SoftMax Pro program, set up a second plate section
for a fluorescence read, spectrum mode, Ex Fixed/Em Scan, withno cutoff filter (default), and medium PMT.
Set the excitation wavelength to the value determined in
Step1w above.
Set the emission scan to start/stop approximately 50nm
below or above the tentative emission value obtainedfrom the literature (or existing filter pair). Note: If the Stokesshift is less than 50nm, then start the emission scan abovetheexcitation wavelength.
Set the step incrementto 2–3nm (or do a preliminary scan
with a 10-nm increment to determine the approximate peak location and then repeat the scan over a narrowerwavelength range using a 2–3nm increment.)
Perform the scan and view the results as a plot of
fluorescence vs. emission wavelength.
0112-0115 F47
Operation
3.Choose the emission filter:
Select an emission cutoff filterthat blocks as much of the
residual excitation light as possible without unduly reducing the fluorescence signal. The cutoff wavelengthchoices are 325, 420, 435, 475, 495, 515, 530, 550, 570, 590, 610,630, 665, or 695nm. The cutoff value should be near the maximum emission wavelength (preferably between the excitation wavelength and the maximal emission wavelength) but at least 10nm less than the emission wavelength. If you havequestions about this procedure pleasecontact MDC Technical Support and ask to speak to an applications scientist.
4.Perform emission scan #2:
In the SoftMax Pro Software, set up a third plate section for
an emission scan as specified in Step2 above, except selecting Manual Cutoff Filter and setting the wavelengthtothat determined in Step3.
Perform the scan and view the results as a plot of
fluorescencevs. emissionwavelength. Note the wavelengthgiving the maximum emission (the optimalemission wavelength).
Compare the spectra ofthe sample containing the
fluorophore to the spectra of the buffer blank to get an estimate of the signal-to-noise ratio. If there is significantbackground interference, repeat Steps 3 and 4 with another choice of cutoff filter.
5.ResultsThe optimal excitation and emission wavelengths are those
determined inSteps1 and 4, above.
6.Comments
In endpoint or kinetic fluorescence modes, the “Autofilter”
feature generally selects the same cutoff filter wavelength asthe above optimization method. If desired, however, you may specify the cutoff filters manually.
For emission wavelengths less than 325 nanometers,
experimental iteration is usually the best method of determining the optimal emission and excitation wavelengths. Begin optimization by performing Steps 1–4 above. Try emission and excitation wavelength combinations withthe 325 cutoff or withno cutoff filter. Similarly,for excitation wavelengths greater than 660 nanometers, try emission and excitation wavelength combinationswith the 695 cutoff or with no cutoff filter.
Figure 4-2Effects of cutoff filters on fluorescein. Emission was scanned from 490 to 560nm; excitation was fixed at 485nm.
Figure4-2 shows the effects of different cutoff filters on a scan of fluorescein where excitation was fixed at 485nm and emission wasscanned from 490nm to 560nm (buffer blanks are not shown in this plot). Tablefollowing lists default settings for the emission cutofffilters.
Table4-1Emission cutoff filter default settings.
#
1None< 415
2420415–434
3435435–454
4455455–474
5475475–494
6495495–514
7515515–529
0112-0115 F49
Automatic Cutoff SelectionEndpoint and KineticModes
Wavelength (nm)Emission Wavelength (nm)
Operation
Table4-1Emission cutoff filter default settings.
#
8530530–549
9550550–569
10570570–589
11590590–609
12610610–629
13630630–664
14665665–694
15695695–850
Automatic Cutoff SelectionEndpoint and Kinetic Modes
Wavelength (nm)Emission Wavelength (nm)
For spectrum mode, the default is “manual”(no automaticcutoff).
Optimizing Time-resolved Fluorescence Assays
Time-resolved fluorescence assays on a SpectraMax M4, M5, and M5ereadermay be read from the top or bottom of a microplate. Solid whiteplates are recommended for top time-resolved fluorescence reads, and white plates withclear bottoms are recommended for bottom reads.
If the time-resolved fluorescence assay you are using has low signalor givesresults with high%CV, use 100 readings per well. If a faster read speed is required, be sure Settling Time is “Off” in the SoftMax Pro Plate Settings dialog box, and experiment with fewer flashes per welluntil acceptable precision and speed are achieved.
Important settings for obtaining the best results in TRF assays are integration delay and integration time:
•The integration delay is the amount of time that elapses
between the flash of the lamp (excitation) and the beginning ofdata acquisition from the well.
•The integration time is the amount of time the well is read.
Delay and integration time are usually specified in the package insert of commercially available TRF reagent kits. If a kit is not used, start with a delay of 50µs and try different delays up to 400µs witha fixed integration time of 400µs. Once the optimumdelay is chosen (based on the highest ratio of a well containing a fluorophore divided bywells containing only buffer) optimize the integration time, which is usually between 400µs and 1000µs.
Fluorescencepolarization for SpectraMax M5 and M5e readers may only be read from the top of a microplate. The plastic from a microplate willaffect the light polarization, precluding bottom reads and reading a covered plate.
Solid black plates are recommended for fluorescence polarization reads. If the assay components seem to bind to the microplate, as evidenced by poor mP dynamic range (small difference between bound and unbound tracer), we suggest using plates treated to minimize binding, or polypropylene plates and/or adding a very smallamount ofdetergent, such asTween-20, to the assay buffer.
Background wells, containing all assay components minus the fluorophore, should be tested. If the signal inthe background wells is more than 1/10 the signal in the wells containing fluorophore, thenbackground wells should berun on eachassay plate. The average raw signal from the background's parallel and perpendicular readings should be subtracted from the raw parallel and perpendicular readingsof eachsample well before the mP calculation is performed.See the
SoftMax
in fluorescence polarization.For best precision in assays using a low amount of fluorophore (for
example, <5nm fluorescein), set the PMT sensitivity to High and the number of readings to 100. If faster read speed is required, be sure Settling Time is “Off” inthe SoftMax Pro Plate Settings dialog box, and experiment withfewer flashes per well until acceptable precision and speed are achieved.
®
Pro Software User Guide for set-up of background subtraction
0112-0115 F51
Operation
Optimizing Luminescence Assays
Luminescence may beread from the top or the bottom of a microplate or the cuvette. Solid white plates or white plates with clear bottoms are recommended for luminescence reads.
For standard luminescence a separate light pathwithout monochromators carries the emitted light toa dedicated PMT. The optimum emission wavelength is between 360 and 630nm. Under reader set-up the emission says “All”.
For wavelength-selectable luminescence, the emission monochromator is used to differentiate the wavelengths being emitted from the well. Up tofour emission wavelengths between 250nm and 850nm may be specified.If readingonly one luminescentevent in the well, best sensitivity should be achieved using the standard luminescence measurement, without a wavelength selected.
Luminescence read times are not designated by multiple reads per well, but rather by choosing the total integration time desired between 1ms and 1,500ms. Typical luminescence assays require between 500ms and 1,000ms integration.
If wells have been incubating for a long period of time, it is a good idea to mix the plate before reading. This can be done using Automix in the reader.
If it appears that the signal is always higher in the first wells read (for example, column A), the plate may need to be “dark adapted” to reduce the auto-luminescence of the white plastic. The auto-luminescence decreases quickly,somanually load the plate from the control panel and wait for 1–2 minutes before initiating the read and determine if the read-out is more consistent across the plate.
52 0112-0115 F
Maintenance
Technical Support
Molecular Devices is a leading worldwide manufacturer anddistributor of analytical instrumentation. Weare committed tothe quality of our products and to fully supporting our customers with the highest possible levelof technical service. In order tofully benefit from our technical services, please complete the registration card and return it to the address printed on the card.
Note:In this user guide, all references to SpectraMax Multi-ModeMicroplate Readers include the M3, M4, M5, and M5e models. When a feature or capabilityapplies to only certain readers, this exception isnoted.
If you have any problems using your SpectraMax Multi-Mode Microplate Reader, in the U.S.,contact our Technical Services groupat1-800-635-5577; elsewhere contact your local representative.
WARNING!BIOHAZARD: It is your responsibility to decontaminate the instrument, as well as any accessories, before requesting service by Molecular Devices representatives and before returning the instrument or any components to Molecular Devices.
5
WARNING!All maintenance procedures described in this manual can be safely performed by qualified personnel. Maintenance not covered in this manual should be performed only by an Molecular Devices representative.
WARNING!Removal of protective covers that are marked with the High Voltage warning symbol shown below can result in a safety hazard.
0112-0115 F53
Maintenance
WARNING!Always turn the power switch off and disconnect the power cord fromthe main power source before performingany maintenance procedure that requires removal of any panel or cover or disassembly of any interior instrument component.
WARNING!Never perform any operation on the instrument in an environment where liquids or potentially damaging gases are present.
WARNING!Risk of electrical shock. Refer servicing to qualified personnel.
CAUTION!Use of organic solvents (such as dichloromethane) may
cause harm to the optics in the instrument. Extreme caution isadvised when using organic solvents.Always use a plate lid and avoid placing a plate containing these materials in the reading chamber for prolongedperiods of time. Damage caused by the use of incompatible or aggressivesolvents is NOT covered by the instrument warranty
CAUTION!Never touch any of the optic mirrors, filters, or cables or their housing, or manifold. Theoptics are extremely delicate, and criticalto the function of the instrument.
CAUTION!Do not touch or loosen any screws or parts other than those specifically designated in the instructions. Doing so could cause misalignment and possibly void warranty.
If you need to relocate a SpectraMax Multi-Mode Microplate Reader, follow these steps.
WARNING!The instrument weighs approximately 36 pounds (16.4 kilograms). To avoid injury, it is recommended that twopeople lift the instrument together, using proper lifting techniques.
1.Remove any microplate from the drawer and then close the
drawer.
2.Turn off the power switch and unplug the power cord from the
source and from the receptacle on the back of the instrument.
3.Depending on the distance that you are moving the instrument,
youmaywant to repackage the instrument in itsoriginalshipping carton. Otherwise, carry the instrument or place it on a rolling cart to transport it.
4.Ensure that the new location meets the proper specifications as
described in “SettingUp the Instrument”.
General
Keep the drawer closedwhen theinstrument is not in use. The drawer can be opened by pressing the button. Always close the drawer immediately prior to switching the instrument off.
0112-0115 F55
Maintenance
Cleaning
WARNING!BIOHAZARD: Wear gloves during any cleaning procedure that could involve contact with either hazardous or biohazardous materials or fluids.
CAUTION!Never clean the inside of the instrument. Cleaning the
interior may cause damage to the instrument.
Periodically, you should clean the outside surfaces of the instrument using a cloth or sponge thathas been dampened with water:
•Do not use abrasive cleaners.
•If required, clean the surfaces using a mild soap solution diluted
with water or a glass cleaner and then wipe with a dampcloth or sponge to remove any residue.
•Do not spray cleaner directly onto the instrument.
If needed,clean the microplate drawer using a cloth or sponge that has been dampened with water.
Should fluidsspill in the drawer area (when the drawer isout), they are directed toa tray at the bottom of the instrument, from which they exitto the bench or counter beneath the instrument. Wipe up any spills immediately.
Do not allow excess water or other fluidsto drip inside the instrument.
The fan filter on the bottom of the instrument requires periodiccleaning. The frequency of cleaning depends on how dusty yourparticular lab is and could range from once a month to once every six months.
1.Turn power to the instrument OFF and then remove the power
cord and cables from the back of the instrument.
2.Removeany plate or adapter from the instrument drawer.Turn
the instrument overso that itrests flaton the bench.
3.Pop the black fan cover off and removethe filter.
4.Clean the filter by blowing clean, canned air through it or by
rinsing it—first with water and then with alcohol—and allowing itto dry completely.
5.Place the clean, dry filter overthe fan and replace the black
cover.
6.Turn the instrument back over. Reconnect the power cord and
cables to the instrument.
Changing the Fuses
Fuses burn out occasionally and must be replaced. If the instrument doesnotseem to be getting power after switching it
on (the LCD shows no display):
•Check to see whether the power cord issecurely plugged in to a
functioning power outlet and to the receptacle at the rear of the instrument.
If power failed while the instrument was already on:
•Check that the power cord is not loose or disconnected and that
power to the power outlet is functioning properly.
If these checks fail to remedy the loss of power, follow the steps listed below to replace the fuses. Spare fuses (two U.S. and two metric) are shipped with the instrument.TheU.S. and metric fuses are identical except for physical size. They may betaped to the back of the instrument.
If you no longer have spare fuses, youmay obtain new ones fromMolecular Devices (part numbers: 4601-0013 for U.S., 4601-0014 for metric) or froma local hardware store. Make sure fuses are rated SLOWBLOW (U.S.:4-amp time-delay; metric:4-amp, 5x20mm, time-delay).
0112-0115 F57
Maintenance
To change fuses
1.Switch power tothe instrument off and then remove the power cord from the outlet and from the instrument power cord receptacle.
2.Remove the computer cable (if connected) from the back of the instrument.
3.Turn the instrument around foreasy access totherear panel.
4.On the left-hand side of the rear panel (viewed from the back) is
the power switch, fuse box, and power cord receptacle. As shown in the figures below, press to the left of the black plasticcover of the fusebox to release it.Pull the fusebox cover awayfrom the instrument. The fuse box will begin to slide forward.
5.Continue gently pulling the fuse box forward until it is free of the instrument.
Figure 5-1Prying open the fuse box cover.
6.When removed, the fuse assembly contains two fuses. Once the
fuse box is out,you will see a holder inside containingtwo fuses. Pull the fuseholder outof the box (see Figure5-2).
7.It is possible that onlyone of the fuses may have blown.
However, Molecular Devices recommends that you replace both fuses to ensure continued properoperation. Pull both fuses out of the holder and discardthem.
Figure 5-2The fuse box and holder withfuses removed.
8.Insert new SLOWBLOW-rated fuses into the fuse holder. Either
endof the fuse may be forward.
9.Insert the fuse holder intothe fuse box, making sure that the
fuses face toward the right (toward the tongue on the cover) as youinsert it.Slidethefuse holder all the way intothebox.
10.Insert the fuse box into the opening in the instrument, making
sure that the fuses are on the left side (toward the power receptacle). Press the fusebox into place, making sure the cover snaps closed.
11.Reconnect the power cord to the instrument and to the wall
outlet and reconnect other cables previously disconnected.
0112-0115 F59
Maintenance
60 0112-0115 F
Troubleshooting
This chapter lists error codes that may occur while using the instrument, followed bytheir most likely causes and remedies.
Maintenance procedures are described in Chapter1.
Note:In this user guide,SpectraMax® refers to several SpectraMax Multi-ModeMicroplate Readers including the M3, M4, M5, and M5e. When a feature or capability applies to only certain readers, thisexception is noted.
For problems withSpectraMax Multi-Mode Microplate Readers that are not listed here, inthe U.S., contact Molecular Devices Technical Services group at 1-800-635-5577; elsewhere,callyour local representative.
WARNING!BIOHAZARD: It is your responsibility to decontaminate the instrument, as well as any accessories, before requesting service by Molecular Devices representatives and before returning the instrument or any components to Molecular Devices.
6
Opening the Drawer Manually
If an error occurs while the drawer is closed and you need to removea microplate, press the DRAWER key.
If the drawer doesnot open, turn power to the instrument off and then on again. If the drawer still remains closed,turn the power off and using your thumbnail, locate the groove in the upper left side wall of the door. Open the door, and with your index finger, pull the microplate drawer out of the instrument (do not force the drawer)and remove the microplate. This action will not harm the instrument, butshouldonly be taken if the first two options havefailed to open the drawer.
If you are still unable to open the drawer, contact your local Molecular Devices representative.
0112-0115 F61
Troubleshooting
Error Codes and Probable Causes
If a problem occurs duringoperation that causes an unrecoverable error, the instrument will stop and an error code number will be shown in the display on the front panel. To correct the problem, call your local Molecular Devices representative for assistance.
Error Messages
The LCD displays Fatal Error codes when a situation arisesthat requires attention. Any reading in progress will stop.
Warning messages do not stop a reading but are logged inthe error buffer; they indicate a situation that requires attention but is notsufficient to stop or prevent a reading. Examples ofsituations thatmight cause warning messages are low memory, entries being out ofrange, or operations that could result in loss of data. These messages are generally self-explanatory.
For assistance regarding warning messages, contact your localMolecular Devices representative.
Table6-1SpectraMax
®
error code ranges.
Error Code Numbers
100-199Errors possibly caused by unrecognized commands
200-299Errors probably due to a main board failure or an error
300-399Instrument errors due to either a main board failure or
400-499Errors causedby a motor motion failure. Most of these
500-599Errors dueto failure or improper initialization of the
Possible Causes
being sent from the computer to the instrument.
in the firmware code. Most of these errors require the assistance of Technical Support.
other system failure. Most of these errors require the assistance of Technical Support.
errors require the assistance of Technical Support.
instrumentsnon-volatile memory (NVRAM). All of these errors require the assistance of Technical Support.
Some errors areconsidered fatal in that if they are detected duringpower up, the instrument aborts the power up sequence and displays “FATAL ERROR”on the LCD panel.
Check the following list to see ifthere is something that you can do to change the condition ofthe instrument to prevent the fatal error. (forexample, closing the cuvette door during the power up sequence preventserrors 111, 219, 302, and 310).
After correcting the problem, leave the instrument onfor about fiveminutes, turn it off and then back on.
If you continue to get the fatal error message on power up, record the error message number and contact Molecular Devices Technical Support or your local representative for assistance.
If the instrument is functioning normally whenusing SoftMax Pro Software, no errors should be inthe buffer (except error number 100).
For all other errormessages (codes not listed here), please contact your local Molecular Devicesrepresentative for assistance.
Table6-2Error codes, error messages, and notes about the errors
ERROR
CODE
ERROR MESSAGENOTES
100-199: Operator Errors
100command not foundCommand string not recognized.
101invalid argumentCommand Argument not recognized.
102too many argumentsToo manyarguments after command.
103not enoughargumentsMissing arguments.
104input line too longToo manycharacters in the input line.
105command invalid,
system busy
106command invalid,
measurement in progress
107no data to transferInputting transfer when there's no data
108data buffer fullToo manydata sets in the buffer. Can
109error buffer overflowMore than65 errors in the buffer, clear
110stray light cuvette, door
open?
111invalid read settings
200assert failedFirmware error.
Instrument could not perform the givecommand because it was busy doing another task. Example: Request a wavelength while the monochromator is in motion.
Instrument could not perform command because a measurement was in progress
in thebuffer
be caused by setting up a long kinetic and disconnecting computer, or SoftMax Pro is preempted by another application.
the buffer.
Cuvette door open while doing a read.
0112-0115 F63
Troubleshooting
Table6-2Error codes, error messages,andnotes about the errors
muchdata overserial port and ignoring flow control.
serial data.
much dataand ignoring the flow control signal.
Firmware error.
300-399: Hardware Errors
300thermistor faultyUnable to read a reasonable thermistor
301safe temperature limit
exceeded
302low ref lightNot enough light detected to makean
303unable to cal dark
current
304signal level saturationDuring a cuvette read, could be due to
value. Thermistor faulty or disconnected, Main board problem, or ambient temperature out of range.
A temperature of over 60°C detectedon one or moreof the 4 thermistors. Temperature will be shut off and remain off until a successful completion of power-up reset.
accurate measurement. If doing a cuvette read, the cuvette door may be open.
Too much stray light detected on power-up, faulty or disconnected pre-amp boards.
Thermal specifications for microplates used intheSpectraMax®Multimode Microplate Reader apply to flat-bottom microplates with isolated wells. All other microplate specificationsapply tostandard 96-well polystyrene flat-bottom microplates.
Note:In this user guide, all references to SpectraMax Multi-ModeMicroplate Readers include the M3, M4, M5, and M5e models. When a feature or capabilityapplies to only certain readers, this exception isnoted.
Performance specifications forcuvette readings apply only to aqueous solutions having solute molal concentrations less than 0.4M.
When pathlength compensation is applied tomicroplate absorbance measurements, agreement with cuvette absorbance measurements for the same solutionrequiresthat the solutionvolume inthemicroplatewell is between 100µL and 300µL.
Technical specifications are subjectto change without notice.
TableA-1Technical Specifications
ABSORBANCE PHOTOMETRIC PERFORMANCE
Wavelength range 200–1000nm
Wavelength selectionMonochromator tunable in 1-nm
Wavelength bandwidth4.0nm full width half maximum
Wavelength accuracy ± 2.0nm across wavelength range
Wavelength repeatability± 0.2nm
Photometric range 0.0 to 4.0OD
Photometric resolution 0.001OD
Photometric accuracy/linearity, 0–2.0 OD
Photometric precision (repeatability), 0–2.0 OD
0112-0115 F67
increments
< ± 1.0% and± 0.006 OD
< ± 1.0% and± 0.003 OD
Specifications
TableA-1Technical Specifications (cont’d)
Stray light 0.05% at 230nm
Photometric stabilization Instantaneous
Photometric driftNone — continuous referencing of
monochromatic input
Calibration Automatic before first kinetic read and
beforeevery endpoint reading
Optical alignment None required
Light sourceXenon flash lamp (50Watts)
Average lamp lifetime1 billion flashes
PhotodetectorsSilicon photodiode
Endpoint baseline noise (cuvette)±0.003 OD @190, 405, 850nm
Endpoint kinetic noise (cuvette)±0.003 OD @190, 405, 850nm
Temperature regulation diagnostics Temperature regulation system is
Evaporation Plate lid required to minimize
Recommended microplate Flat-bottom microplates with isolated
ControlFront panel reports cuvette chamber
continuously monitored and updated
evaporative cooling
wells and lid
temperature only (temperature for microplate chamber reported in SoftMax® Pro Software)
AUTOMIX WITH SOFTMAX PRO
Platemixing modesSelectable: off, once prior to any
Platemixing duration Selectable: 0 to 999 seconds
reading, and once prior to andbetween kinetic readings
(three-second default)
COMPATIBILITY
MicroplatesStandard 6- to 384-well flat-bottomed
Cuvettes Standard height (45mm) cells with
Test tubes12x 75 mm test tubes can beused in
microplates. Polystyrene plates for absorbance wavelengths above 340nm; UV-transparent plates for absorbance readings above 220nm; quartz plates for absorbance readings above 200nm;low-volume 384-well plates. Use purple adapter plate only with 96- and 384-wellplates.
10mm pathlength (12.5mm x
12.5mm outside) with minimum inside width of 4mm (typicalfor 3mL volumecells). See more information.
the cuvette chamber with the test tube cover.
Cuvettes on page78 for
GENERAL INSTRUMENT
Display2-x-20-character backlit LCD
Operating panel 11-key membrane keypad
Self-diagnosisContinuous on-board diagnostics
0112-0115 F71
Specifications
TableA-1Technical Specifications (cont’d)
Spill controlDrawer mechanism and reading
Computer interface8-pin DIN RS-232 serial
Printer interfaceParallel25-pin to Centronics
Microplates supportedAll 6-to 384-well and strip-well
chamber assembly protected from accidental spillage by drainage ports
(double shielding required)
(double shielding required)
microplates, including lids
ROBOTICS AND AUTOMATION
Robot compatible drawerPositioning and plate gripping
Integrated automation interface SoftMax® Pro Software automation
as drawer closes
interface integrated with robot partners. SpectraMax® Multi-Mode Microplate Readers and SoftMax Pro Software are the #1choiceof robotic partners and robots. Please visit the Molecular Devices web site for more information: www.moleculardevices.com.
ENVIRONMENTAL
Operating temperature15°C to 60°C
Operating humidity 0 to 70%, non-condensing
Storage temperature-20°C to 65°C
PHYSICAL
Size (h x w x d) 8.6" (220mm) x 22.8" (580mm)
Weight36lb (16.4kg)
Power consumption<420W
Line voltage and frequency100–240 VAC autoranging, 3.5A, 50/60
Values in the following tablesare based on the literature. You mustscan the fluorochromeof interest in the SpectraMaxM5orM5e reader to determine the optimalexcitation and emission wavelengths for your application. Excitation and emission wavelengths listed by fluorochrome manufacturers are generally determined in methanoland do notreflect actual values, due to changes in pH, salt content, etc.
Molecular Devices recommends that you use high-quality, double-shielded cables to connect your SpectraMax® readerto the computer. Choose cables that meet the requirement described inthis appendix.
B
Note:In this user guide, all references to SpectraMax Multi-Mode
Microplate Readers include the M3, M4, M5, and M5
a feature or capability applies toonly certain readers, this exception is noted.
e
models. When
Serial Interface Cable
The serial interface cable used to connect the instrument tothe computer is a custom cable designedand built by Molecular Devices. Please use the cable supplied by Molecular Devices, or contact Molecular Devices for specific pin-out requirements:
Male DB8 to Female DB9 (custom cable made by Molecular Devices, PN 9000-0149)
USB Adapter Cable
Macintosh computers, and many newer Windows-based computersdo not have a serial port. You can connect a serial cable between these computers and the instrument using a USB-to-serial adapter.
Molecular Devices has tested many third-party serial-to-USB adapter cables and has found the Keyspan USA-19HS (Molecular Devices, PN 9000-0938) to be the mostreliable. It is the only one we recommend.
Figure B-1Molecular Devices’ custom serialcable (left) and a serial-to-USB converter (right).
0112-0115 F77
Cables and Accessories
Accessories
DescriptionPart #
SpectraTest ABS1 Absorbance Validation Test Plate0200-6117
SpectraTest FL1 Fluorescence Validation Test Plate0200-5060
Power Cord (US, Canada, Japan,Mexico, India)4400-0002
Power Cord, EC1 (Germany, France, Scandinavia, Italy,
Korea)
Power Cord,EC2(UK, Indonesia, Singapore, Malaysia)4400-0037
Power Cord, AP1 (Australia, HongKong,China)4400-0038
SpectraMax Mouse Pad9000-0133
Cable, RS-232, 8-pin DIN to 8-pin DIN (instrument to
pre-G3 Macintosh)
Cable, RS-232, 9-pin DIN to 8-pin DIN (instrument to
PC serial port)
Adapter USB-Serial High-Speed (KeySpan adapter;
instrument to USB-only instrument)
Test Tube Cover2300-0277
4400-0036
9000-0091
9000-0149
9000-0938
Cuvettes
The guidelines for cuvette use in the SpectraMax Multi-Mode Microplate Readers are the same that apply to any high-quality spectrophotometer. The user mustensure that the meniscus iscomfortably above the light beam instandard cuvettes and that the samplechamber in a microcuvette is aligned properly with the beam. The light beam is 0.625in (15.87mm) above the cuvette bottom.
Below are some cuvettes that have been tested. All have an optical pathlength of 1cm(10mm) and standard external dimensions (12.5mm x 12.5mm). Their fill volumes differ only because of their different internal width and chamber height dimensions.