7.1. SpectraMax M2 and M2e Performance Specifications
SpectraMax M2 & SpectraMax M2e Multi-mode Plate Readers User Guide — 0112-0102 Rev. D
47
7. Specifications
LUMINESCENCE PHOTOMETRIC PERFORMANCE
Detection limit (384-well microplate) 10 amol/well alkaline phosphatase, 200 µL
Wavelength range 250–850 nm
TIME-RESOLVED FLUORESCENCE PERFORMANCE
Detection limit (384-well microplate) 0.5 fmol/well Eu-Ch
Wavelength range 250–850 nm
PHOTOMETRIC ANALYSIS MODES
Front Panel Operation • Single wavelength Absorbance, %Trans-
mittance, Fluorescence reading of the
cuvette (or test tube)
Using SoftMax Pro • Express data as Absorbance, %Transmit-
tance, Fluorescence, Luminescence
• Single wavelength reading of microplate
and/or cuvette
• Multiple wavelength (up to four)
reading of microplate or cuvette
• Kinetic and kinetic graphics of
microplate and/or cuvette
• Spectral scan (190–1000 nm) of
microplate and/or cuvette
• Well scan of microplate using absorbance
or fluorescence intensity
MEASUREMENT TIME (CALIBRATION OFF)
Microplate read time (endpoint),
Standard read
• 96 wells in 24 seconds (single
wavelength, absorbance)
• 96 wells in 15 seconds (single
wavelength, fluorescence intensity)
• 384 wells in 1:57 minutes (single
wavelength, absorbance)
• 384 wells in 45 seconds (single
wavelength, fluorescence intensity)
Microplate read time (endpoint),
Standard read with PathCheck
• 96 wells in 2:07 minutes (single
wavelength, absorbance)
• 384 wells in 7:19 minutes (single
wavelength, absorbance)