Instruction for use and hygiene
MK-dent Air Motor AM1116
Important Details
The MK-dent Air Motor (AM1116)
- is only intended for dental treatment of patients.
- is according to relevant national regulations as a medical device.
These provisions indicate, that
- applicable accident prevention regulations and health and safety regulations must be observed.
- only a competent user is allowed to use the product according to this instruction.
- only faultless work equipment is to be used.
- any contamination from the product must be avoided.
- patients, users and third parties are to be protected from danger.
Technical Facts
MK-dent Air Motor (AM1116)
Input speed: 5.000 - 15.000 rpm
Air input: 2,2 - 2,8 bar
Air use: up to 60Nl/ min.
Connection: 4-hole standard connection, INTRAmatic connector
Meaning of Symbols
Item No. Serial No. Caution! Sterilizable in an autoclave
up to max. 135°C/ 275°F
Suitable for spray- and
wipe disinfection
Connection to the Treatment Unit
Before rst use and after each use, sterilize the Air Motor. Always use dry, cleaned and decontaminated compressed
air. Premature wear from contaminated and humid compressed air can shorten the lifetime of the product.
Place the connection of the turbing on the Air Motor and
turn the thread until stop.
Spray Regulation
Regulate the amount of water by turning the regulation ring
at the back of the Air Motor.
Checking the Compressed Air with a Manometer
Install the manometer between tubing and Air Motor. The following values should be displayed:
Forced Air (F.A./ T.R.) = 2,2 - 2,8 bar
Return Air (R.A./ R.L.) < 1,1 bar
Water (W.) = 0,8 - 2,5 bar
Spray Air (S.A./ Spr.L.) = 1,0 - 4,0 bar
Connecting to the INTRAmatic-connection of Handpieces
Place the handpiece on the Air Motor until it audibly locks in place.
In order to avoid damage to the driver, do not attach or release the Air Motor while the handpiece is in movement.
Rev: A/09_2015 File: Manual_AM1116_EN.indd