MJ Research PTC-200 DNA Engine, CFD-3200 Opticon Operation Manual

DNA Engine Opticon® System
For Continuous Fluorescence Detection
PTC-200 DNA Engine® Cycler
Operations Manual
Supports Software Version 1.08
Detector
06678 revC.A
DNA Engine Opticon® System
For Continuous Fluorescence Detection
PTC-200 DNA Engine® Cycler CFD-3200 Opticon
Detector
Operations Manual
Supports Software Version 1.08
ii Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Copyright ©2004, Bio-Rad Laboratories, Incorporated. All rights reserved. Reproduction in any form, either print or electronic, is prohibited without written permission of Bio-Rad Laboratories, Inc.
Chill-out, DNA Engine, DNA Engine Opticon, Hard-Shell, Microseal, MiniCycler, MJ Research and the helix logo, Multiplate, Opticon, Opticon Monitor and PTC-100 are trademarks belonging to Bio-Rad Laboratories, Inc.
Amplifluor is a trademark of Intergen Company. DyNAzyme is a trademark of Finnzymes Oy. Scorpions is a trade­mark of DXS Ltd. SYBR is a trademark of Molecular Probes, Inc. TaqMan is a trademark of Roche Molecular Systems, Inc. Windows is a trademark of Microsoft Corporation.
Practice of the patented polymerase chain reaction (PCR) process requires a license. The DNA Engine Opticon system includes an Authorized Thermal Cycler and may be used with PCR licenses available from Applied Biosystems. Its use with Authorized Reagents also provides a limited PCR license in accordance with the label rights accompanying such reagents .Some applications may also require licenses from other third parties.
This instrument includes an Authorized Thermal Cycler, Serial No __________________. Its purchase price includes the up-front fee component of a license under United States Patent Nos. 4,683,195, 4,683,202 and 4,965,188, owned by Roche Molecular Systems, Inc., and under corresponding claims in patents outside the United States, owned by F. Hoffmann-LaRoche Ltd, covering the Polymerase Chain Reaction ("PCR") process, to practice the PCR process for internal research and development using this instrument. The running royalty component of that license may be purchased from Applied Biosystems or obtained by purchasing Authorized Reagents. This instrument is also an Authorized Thermal Cycler for use with applications licenses available from Applied Biosystems. Its use with Authorized Reagents also provides a limited PCR license in accordance with the label rights accompanying such reagents. Purchase of this product does not itself convey to the purchaser a complete license or right to perform the PCR process. Further information on purchasing licenses to practice the PCR process may be obtained by contacting the Director of Licens­ing, Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California, 94404, USA.
No rights are conveyed expressly, by implication or estoppel to any patents on real-time PCR.
Applied Biosystems does not guarantee the performance of this instrument.
06678 revCA
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com iii
Table of Contents
Explanation of Symbols .......................................................................................... iv
Safety Warnings .................................................................................................... iv
Safe Use Guidelines................................................................................................. v
Electromagnetic Interference .................................................................................... v
FCC Warning .......................................................................................................... v
1. Introduction ..................................................................................................... 1-1
2. Layout and Specifications ................................................................................. 2-1
3. Installation and Operation ................................................................................ 3-1
4. Compatible Chemistries, Sample Vessels, and Sealing Options ........................... 4-1
5. Introduction to Opticon Monitor™ Software ...................................................... 5-1
6. Experimental Setup and Programming .............................................................. 6-1
7. Run Initiation and Status ................................................................................... 7-1
8. Data Analysis................................................................................................... 8-1
9. Maintenance ....................................................................................................9-1
10. Troubleshooting ............................................................................................ 10-1
Appendix A .........................................................................................................A-1
Appendix B .......................................................................................................... B-1
Appendix C.......................................................................................................... C-1
Appendix D ......................................................................................................... D-1
Appendix E .......................................................................................................... E-1
Index ..................................................................................................................In-1
Declarations of Conformity ...............................................................................
DoC
-1
iv Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Explanation of Symbols
CAUTION: Risk of Danger! Wherever this symbol appears, always consult note
in this manual for further information before proceeding. This symbol identifies com­ponents that pose a risk of personal injury or damage to the instrument if improperly handled.
CAUTION: Risk of Electrical Shock! This symbol identifies components that pose a risk of electrical shock if improperly handled.
CAUTION: Hot Surface! This symbol identifies components that pose a risk of per­sonal injury due to excessive heat if improperly handled.
Safety Warnings
Warning:Warning:
Warning:Warning:
Warning: Operating the DNA Engine Opticon system before reading this manual can
constitute a personal injury hazard. Only qualified laboratory personnel trained in the safe use of electrical equipment should operate this instrument.
Warning:Warning:
Warning:Warning:
Warning: Do not open or attempt to repair the Opticon tower or base. Doing so will
void your warranties and can put you at risk for electrical shock. Return the DNA Engine Opticon system to the factory (US customers) or an authorized distributor (all other customers) if repairs are needed.
Warning:Warning:
Warning:Warning:
Warning: The sample block can become hot enough during the course of normal opera-
tion to cause burns or cause liquids to boil explosively. Wear safety goggles or other eye protection at all times during operation.
Warning:Warning:
Warning:Warning:
Warning: The DNA Engine Opticon system incorporates neutral fusing, which means that
live power may still be available inside the machines even when a fuse has blown or been removed. Never open the Opticon base; you could receive a serious electrical shock. Opening the base will also void your warranties.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com v
Safe Use Guidelines
The DNA Engine Opticon system is designed to operate safely under the following conditions:
Indoor use
Altitude up to 2000m
Ambient temperature 15˚–25˚C
Maximum relative humidity 80%, noncondensing
Transient overvoltage per Installation Category II, IEC 664
Pollution degree 2, in accordance with IEC 664
Electromagnetic Interference
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This device has been tested and found to comply with the EMC standards for emissions and susceptibility established by the European Union at time of manufacture.
This digital apparatus does not exceed the Class A limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Depart­ment of Communications.
LE PRESENT APPAREIL NUMERIQUE N'EMET PAS DE BRUITS RADIOELECTRIQUES DEPASSANT LES LIMITES APPLICABLES AUX APPAREILS NUMERIQUES DE CLASS A PRESCRITES DANS LE REGLEMENT SUR LE BROUILLAGE RADIOELECTRIQUE EDICTE PAR LE MINISTERE DES COMMUNICATIONS DU CANADA.
FCC Warning
Warning: Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radiofrequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be re­quired to correct the interference at his own expense.
1-1
1. Introduction
Meet the DNA Engine Opticon System, 1-2 Using This Manual, 1-2 Important Safety Information, 1-3
1-2 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Meet the DNA Engine Opticon® System
Thank you for purchasing a DNA Engine Opticon continuous fluorescence detection sys­tem from MJ Research, Incorporated. Designed by a team of molecular biologists and engineers, the Opticon™ system will meet your needs for a sensitive, easy-to-use, and com­pact continuous fluorescence detection system. Some of the DNA Engine Opticon system’s many features include:
A DNA Engine
®
Peltier thermal cycler delivers superior thermal accuracy and well-to-
well thermal uniformity.
A 96-well sample block accepts standard consumables (96-well, low-profile micro­plates and low-profile 0.2ml strip tubes).
An integrated heated lid permits oil-free cycling.
Long-lived LEDs excite fluorescent dyes in the 450-495nm range.
Sensitive optics detect fluorophores with emission spectra in the 515-545nm range (SYBR Green, FAM).
Intuitive Opticon Monitor
software facilitates experimental setup, run initiation, run
status, and data analysis.
Dual modes of temperature control include calculated control for maximum speed and accuracy, or block control for adapting protocols optimized in other cyclers.
Compact footprint measuring 47cm deep x 34cm wide x 60cm high, allows the Opticon unit to fit comfortably on any lab bench.
The Opticon detector is available separately as an upgrade for existing DNA Engine thermal cyclers.
Using This Manual
This manual contains instructions for operating your DNA Engine Opticon system safely and productively:
Chapter 2 acquaints you with the physical characteristics of the Opticon system.
Chapter 3 presents the basics of installing and operating the Opticon system.
Chapter 4 discusses the chemistry and sample vessel compatibilities of the Opticon system.
Chapters 5-8 step you through the use of the Opticon Monitor software includ- ing how to enter and run protocols, and analyze collected data.
Chapter 9 explains the proper maintenance of the Opticon system.
Chapter 10 offers troubleshooting information for the Opticon system.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 1-3
Introduction
Important Safety Information
Safe operation of the DNA Engine Opticon system begins with a complete understand­ing of how the instrument works. Please read this entire manual before attempting to op­erate the DNA Engine Opticon system. Do not allow anyone who has not read this manual to operate the instrument.
Warning: The DNA Engine Opticon system can generate enough heat to inflict
serious burns and could deliver strong electrical shocks if not used ac­cording to the instructions in this manual. Please read the safety warnings and guidelines at the beginning of this manual on pages iv and v, and exercise all precautions outlined in them.
2-1
2. Layout and Specifications
Front View, 2-2 Back View, 2-2 Specifications, 2-3 Gradient Specifications, 2-4 Computer Specifications, 2-4
2-2 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Back View
(Figure 2-2)
Front View
(Figure 2-1)
Power cord jack (some models)
Power switch (fuses, some models)
DAQ (data acquisition)
cable port
Serial cable port
Power module, left configuration (standard)
Power cord jack
Power switch (fuses)
Power module, right configuration (some models)
Optical tower
Cycler drawer
Air exhaust vents
(also on other side)
Air intake vents
(also on other side)
Blue protocol­indicator light
Blue trigger handle (door mechanism)
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 2-3
Layout and Specifications
Specifications
Thermal range: 0˚ to 105˚C, but not more than 30˚C
below ambient temperature
Accuracy: ±0.3˚C of programmed target @ 90˚C,
NIST-traceable
Thermal homogeneity: ±0.4˚C well-to-well within 30 seconds of
arrival at 90˚C
Ramping speed: Up to 3.0˚C/sec
Sample capacity: 96-well microplate (low-profile) or
96 x 0.2ml strip tubes (low-profile)
Line voltage: 100-240VAC
Frequency: 50-60Hz
Power: 850W maximum
Fuses: Two 6.3A, 250V Type S505, fast acting
(user changeable) Two 8.0A, 250V Type S505, fast acting (inaccessible)
Weight: 27kg (excluding computer and
monitor)
Size: 47cm deep x 34cm wide x 60cm high
(excluding computer and monitor)
Fluorescence Excitation Range: 450-495nm
Fluorescence Detection Range: 515-545nm
2-4 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Gradient Specifications
Accuracy: +0.3°C of target at end columns within
30 seconds (NIST-traceable)
Column uniformity:
+0.4°C, in column, well–to–well, within 30 seconds of target attainment
Calculator accuracy:
+0.4°C of actual column temperature (NIST-traceable)
Lowest programmable 30°C temperature:
Highest programmable 105°C temperature:
Gradient range: from 1°C up to 24°C (temperature differential)
Computer Specifications
(minimum specifications for the computer provided with the Opticon system)
Processor: 2.4GHz processor
Operating System: Windows XP Pro
Display: 15 inch flat-screen monitor
Memory: 256 MB RAM
Storage: 40GB hard drive
Data Acquisition Board: National Instruments PCI-6036E
200kS/s (samples per second)
3-1
3. Installation and Operation
Unpacking the Opticon Unit, 3-2 Packing Checklist, 3-2 Setting Up the DNA Engine Opticon System, 3-3 Environmental Requirements, 3-3 Power Supply Requirements, 3-4 Air Supply Requirements, 3-4
Ensuring an Adequate Air Supply, 3-4 Ensuring That Air Is Cool Enough, 3-4
Troubleshooting Air Supply Problems, 3-5 Turning the Opticon Unit and Computer On and Off, 3-5 Opening and Closing the Cycler Drawer, 3-6 Loading Sample Vessels into the Block, 3-6
3-2 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Unpacking the Opticon™ Unit
Please follow these instructions for unpacking the Opticon unit to reduce the risk of per­sonal injury or damage to the instrument.
Important: DO NOT lift the instrument out through the top of the box.
Important: DO NOT use the blue handle to lift the instrument at any time.
Cut the band securing the cardboard cover to the support base.
Open the top of the cardboard cover.
Remove the top foam insert.
Remove the accessory box (contents listed below).
Lift the cardboard cover up and off of the instrument.
Firmly grasp the sides of the instrument from beneath to support the weight of the cycler and the optical tower. Carefully lift the instrument off of the shipping sup­port. Do not lift the instrument by the blue handle or the cycler drawer.
Packing Checklist
After unpacking the DNA Engine Opticon® continuous fluorescence detection system, check to see that you have received the following:
1. One DNA Engine Opticon unit (Opticon detector with DNA Engine
®
thermal cycler)
2. One computer with keyboard, mouse, monitor, cables, & installed software (Opticon Monitor and Windows XP pro)
One serial cable for connecting the Opticon unit’s serial port (figure 2-2) to the
computer serial port
One data acquisition cable for connecting the Opticon unit’s DAQ port (figure 2-
2) to the data acquisition card in the computer
3. One Opticon accessory pack including:
One power cord for the Opticon unit
Two spare fuses
DNA Engine Opticon® Continuous Fluorescence Detection System Operations Manual
(this document)
Opticon Monitor
software CD ROM
Consumables samples including 0.2ml low-profile strip tubes in opaque white
(MJ Research catalog no. TLS-0851), optical flat caps for 0.2ml tubes and plates (MJ Research catalog no. TCS-0803), and low-profile Multiplate
96-well micro-
plates in opaque white (MJ Research catalog no. MLL-9651)
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 3-3
Installation and Operation
If any of these components are missing or damaged, contact MJ Research, Incorporated or the authorized distributor from whom you purchased the DNA Engine Opticon system to obtain a replacement. Please save the original packing materials in case you need to return the DNA Engine Opticon system for service. See Appendix C for shipping instructions.
Setting Up the DNA Engine Opticon System
The Opticon system requires a location with three power outlets to accommodate the Opticon unit, the computer, and the monitor. A location with network access (Ethernet 10/100BaseT) is recommended if you wish to transfer setup and analysis files between the computer running the Opticon unit and other computers.
The DNA Engine Opticon system requires only minimal assembly. Insert the power cord plug into its jack at the back of the instrument, just below the power switch (see figure 2­2 for the location of the jack). Then, plug the power cord into a standard 110V or 220V electrical outlet. The Opticon unit will accept 220V automatically, as does the monitor. However, you must set the voltage on the computer. See the “Power Supply Requirements” section below for more information.
Before launching the Opticon Monitor software (see Chapter 5), be sure that the Opticon unit is connected to the computer. There are two cables that connect the Opticon unit to the computer. Connect the serial cable to the serial cable port on the Opticon unit (see figure 2-2) and serial port #2 on the computer. Connect the data acquisition cable to the DAQ port on the Opticon unit (see figure 2-2) and the data port on the computer.
Note: The DAQ cable has high-density connectors; take care not to bend any of the pins.
Environmental Requirements
For reasons of safety and performance, ensure that the area where the DNA Engine Opticon system is installed meets the following conditions:
Nonexplosive environment
Normal air pressure (altitude below 2000m)
Ambient temperature 15˚–31˚C
Relative humidity above 10% and up to 80%
Unobstructed access to air that is 31˚C or cooler (see below)
Protection from excessive heat and accidental spills. (Do not place the DNA Engine Opticon system near such heat sources as radiators, and protect it from danger of having water or other fluids splashed on it, which can cause electrical short circuits.)
3-4 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Power Supply Requirements
The DNA Engine Opticon unit requires 100-240VAC, 50-60Hz and a grounded outlet. The DNA Engine Opticon unit can use current in the specified range without adjustment, so there is no voltage-setting switch. The monitor can also accept either 110 or 220V power without adjustment.
Important! For 220V operation of the computer, the red voltage-setting
switch located on the back of the computer, near the power cord jack, must display 230V rather than 115V.
The Opticon unit is equipped with a power-entry module that accepts cordsets with an IEC 60320-1 type C13 connector (this is the same standard configuration used by many computer manufacturers for their equipment). All cordsets used with the Opticon unit must be rated to carry at least 10A at 125V or 250V, the latter specification depending upon the supply voltage used. Additionally, the cordset must meet all other applicable national standards—thus at a minimum, the cordset should carry the mark of a nationally recog­nized testing agency appropriate to your nation.
Note: Do not cut the supplied 120V power cord and attach a different connector. Use a one-
piece molded connector of the type specified above.
Air Supply Requirements
The DNA Engine Opticon unit requires a constant supply of air that is 31˚C or cooler in order to remove heat from the heat sink. Air is taken in from the lower vents located on the sides of the instrument and exhausted from the upper vents on both sides (see figure 2-1). If the air supply is inadequate or too hot, the instrument can overheat, causing per­formance problems and even automatic shutdowns.
Ensuring an Adequate Air Supply
Do not block air intake vents (see figure 2-1).
Position the DNA Engine Opticon unit at least 10cm from vertical surfaces and other thermal cyclers or heat-generating equipment (greater distances may be required; see below).
Do not allow dust or debris to collect in the air intake vents.
Ensuring That Air Is Cool Enough
Do not position two or more DNA Engine Opticon units (or other instruments) so that hot exhaust air blows directly into the air intake vents.
Confirm that the DNA Engine Opticon unit receives air that is 31˚C or cooler by measur­ing the temperature of air entering the machine through its air intake vents.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 3-5
Installation and Operation
Place the DNA Engine Opticon unit where you plan to use it, and turn it on. Try to reproduce what will be typical operating conditions for the machine in that location, particularly any heat-producing factors (e.g., nearby equipment running, window blinds open, lights on). Run a typical protocol for 30 minutes to warm up the DNA Engine Opticon unit, then measure the air temperature at the air intake vents. If more than one machine is involved, measure the air temperature for each.
If the air intake temperature of any machine is warmer than 31˚C, consult Table 3-1 for possible remedies. After implementing possible remedies, verify that the temperature of the air entering the air intake vents has been lowered, using the procedure outlined above.
Table 3-1 Troubleshooting Air Supply Problems
Cause Possible Remedies
Air circulation is poor. Provide more space around instrument or adjust room
ventilation.
Ambient air temperature Adjust air conditioning to lower ambient air is high. temperature.
Instrument is in warm part Move instrument away from, or protect instrument from, of room. such heat sources as radiators, heaters, other equip-
ment, or bright sunlight.
Instruments are crowded. Arrange machines so that warm exhaust air does not
enter intake vents.
Turning the Opticon Unit and Computer On and Off
Locate the power switch on the back, left-side of the Opticon unit (back, right-side on some models) just above the power cord (see figure 2-2). To turn the Opticon unit on, press the switch so that the side marked “1” is depressed. The thermal cycler requires several minutes to warm up after the Opticon unit is powered up. To turn the Opticon unit off, depress the “0” side of the power switch.
Be sure that the Opticon unit is connected to the computer and turned on prior to launch­ing the Opticon Monitor software. The blue protocol-indicator light on the front of the Opticon unit (see figure 2-1) is illuminated only during a protocol run.
Press the power button on the front of the computer once to turn the computer on. Select
Shutdown
from the
Start
menu to turn the computer off. Press the power button on the
front of the monitor once to turn it on, and press it again to turn the monitor off.
3-6 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Opening and Closing the Cycler Drawer
To gain access to the Opticon unit’s thermal cycling block, first squeeze the blue trigger handle (1) and allow the spring-loaded door to lift up (2). Then, use the hand hold on the drawer front to slide the cycler drawer out toward you exposing the 96-well thermal cy­cler block (3).
To return the Opticon unit to the closed position, slide the cycler drawer back into the instrument and lower the blue handle to secure the cycler drawer. It is not necessary to squeeze the trigger handle.
Note: Do not open the cycler drawer while the blue protocol-indicator light is illuminated.
Opening the door, particularly during a scan of the plate, may interrupt the software’s control of the protocol.
Loading Sample Vessels into the Block
Important! Do not use full height 0.2ml tubes or full height unskirted
microplates. Refer to the “Selecting the Correct Sample Vessel” section
in Chapter 4 for tube and microplate recommendations.
To ensure uniform heating and cooling of samples, sample vessels must be in complete contact with the block. Adequate contact is ensured by doing the following:
Ensure that the block is clean before loading samples (see Chapter 9 for cleaning
instructions).
Firmly press strips of 0.2ml low-profile tubes, or a 96-well, low-profile microplate into
the block wells (see the “Selecting the Correct Sample Vessel” section in Chapter 4).
MJR strongly recommends that oil not be used to thermally couple sample vessels to
the block.
Tip: Spin down reactions in tubes or microplates prior to loading into the thermal-cycler
block. Air bubbles in samples or liquid on the plate deck can adversely affect results.
1.
2.
3.
4-1
4. Compatible Chemistries, Sample Vessels, and Sealing Options
Optical System, 4-2 Compatible Chemistries, 4-2
SYBR Green I, 4-2 Molecular Beacons, 4-3 Hydrolysis Probes (TaqMan Probes), 4-3 Scorpions Probes, 4-4 Amplifluor Universal Detection System, 4-4
Selecting the Correct Sample Vessel, 4-5
Vessels Optimized for Fluorescence Detection and Thermal Cycling, 4-5
Sealing Sample Vessels, 4-5
Sealing with Optical Caps and the Heated Lid, 4-6
Sealing with Chill-out™ 14 Liquid Wax, 4-6 Sample Vessel and Sealing Selection Chart for Optical Assays, 4-7 Reaction Volume Recommendations, 4-8
4-2 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Optical System
The Opticon™ detector uses an array of 96 blue LEDs to sequentially illuminate each of the 96 wells in the cycler block. The LEDs efficiently excite fluorescent dyes with excita­tion spectra in the 450 to 495nm range. The Opticon detector is optimized to detect dyes with emission spectra in the 515 to 545nm range, such as SYBR Green and FAM.
The Opticon detector is calibrated at the factory and requires no calibration before use. See Chapter 10, “Troubleshooting” for instructions on testing detector calibration and recalibrating.
Compatible Chemistries
The Opticon detector is compatible with popular dye chemistries including SYBR Green I, molecular beacons, hydrolysis probes (TaqMan probes), Scorpions probes, and the Amplifluor system. In addition to performing real-time quantification and DNA melting profiles, the Opticon system is also useful as a temperature-controlled fluorimeter for a number of applications including ligand binding and protein structure studies. If you have questions regarding the compatibility of a particular chemistry with the Opticon detector, contact MJ Research technical support at 888-652-9253.
SYBR Green I
SYBR Green I (available from Molecular Probes, Inc. of Eugene, Oregon) is a dsDNA binding dye thought to bind in the minor groove. The fluorescence of SYBR Green I is greatly enhanced upon binding dsDNA. This characteristic makes it ideal for detection of amplification products. The maximum absorbance of SYBR Green I is ~497nm and the emission maximum is ~520nm*.
SYBR Green I has several advantages for detection of nucleic acids in real time. Because SYBR Green I binds to all dsDNA, it does not have to be customized for individual tem­plates thereby providing the advantages of quick protocol adaptation and relatively low cost. Further, SYBR Green I is very sensitive because multiple dye molecules bind to a single amplification product. However, because SYBR Green I binds to all dsDNA, false positive signals from primer-dimers, secondary structure, or spurious priming can inter­fere with accurate quantification. Measuring fluorescence at elevated temperatures may help reduce the detection of nonspecific products
1
. Performing a melting curve to ana­lyze product homogeneity can also aid in analyzing quantification results obtained with SYBR Green I.
MJR recommends using buffers containing 5% dimethyl sulfoxide (DMSO) with a concen­tration of 1X or less SYBR Green I with the Opticon detector. For additional information on optimizing protocols using SYBR Green I with thermostable enzymes available from MJ Research, contact MJ Research technical support at 888-652-9253.
1
Morrison, T.B., J.J. Weis and C.T. Wittwer. 1998. Biotechniques 24:954-962.
*Molecular Probes, Inc.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 4-3
Compatible Chemistries, Sample Vessels, and Sealing Options
Molecular Beacons
Molecular beacons are dual-labeled oligonucleotide probes designed to form stem-loop structures in the absence of target. In the hairpin configuration, the fluorophore at one end of the molecule is brought into close proximity with a quenching moiety at the other end of the molecule. When the fluorophore is excited in this configuration, it transfers energy to the quencher rather than emitting that energy as light, in a process known as fluorescence resonance energy transfer (FRET). A “dark” quencher is often used, so the energy transferred from the fluorophore is emitted in the infrared as opposed to the vis­ible range. If a second fluorophore is used as a quencher, the transferred energy is emit­ted as light at the quenching fluorophore’s characteristic wavelength.
Molecular beacons are designed such that the loop, which is usually 15-30 nucleotides in length, is complimentary to the target sequence. The arms flanking the loop, which are usually 5–7 nucleotides in length, are designed such that they are complementary and favor formation of a stem structure. A fluorophore is attached to the end of one arm, and a quencher is attached to the other. Molecular beacons must be carefully designed such that at the annealing temperature of the reaction hairpins form in the absence of template, but that in the presence of template, the annealing of the loop sequence to the target is ener­getically favorable. When the loop of a molecular beacon hybridizes to the target sequence, the conformational change of the probe separates the fluorophore and the quencher. When the fluorophore is excited, it now emits light at its characteristic wavelength.
One advantage of molecular beacons is that unlike SYBR Green, molecular beacons spe­cifically detect the target of interest. Great sensitivity, including detection of single nucleotide polymorphisms (SNPs), is possible with carefully designed molecular beacons and optimized reaction conditions (temperature, buffer). However, each probe must be carefully and uniquely designed for the detection of a specific target.
Molecular beacons are a technology patented by the Public Health Research Institute of New York, NY and are available from a number of licensed suppliers. When designing molecular beacons for use with the Opticon detector, fluorophores with excitation and emission spectra falling within the Opticon detector’s excitation (450-490nm) and detec­tion (515-545nm) ranges, such as FAM, can be used. Either dark quenchers or a quench­ing fluorophore may be used with the Opticon detector. However, because the Opticon detector is a single-color detection system, light from a quenching fluorophore can not be separately monitored. Dark quenchers tend to give cleaner signal because there is no overlapping signal from light emitted by the quenching fluorophore.
Hydrolysis Probes (TaqMan Probes)
TaqMan probes are a patented technology available from a number of licensed suppli­ers. They are oligonucleotide probes whose fluorescence is dependent on the amplifica­tion of a target sequence. TaqMan probes are designed to anneal to the target sequence between the forward and reverse primers. A reporter fluorophore is attached to the 5’ end of the probe and a quencher to the 3’ end.
4-4 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
When the intact probe anneals to the target sequence, excitation of the reporter is quenched because of its proximity to the 3’ quencher. However, as extension proceeds, the 5’ exonu­clease activity of the polymerase cleaves the probe, separating the reporter from the quencher. TaqMan probes work well with enzymes derived from Thermus species, such as DyNAzyme™ II DNA polymerase from Thermus brockianus, available from MJ Research, Inc. Liberated reporter molecules accumulate as the number of cycles increases, such that the increase in fluorescence is proportional to the amount of amplified product.
One advantage of TaqMan probes, particularly for quantification, is that fluorescence is dependent not only on the presence of a specific target, but also on amplification of that target. However, like molecular beacons, TaqMan probes must be individually designed for specific targets. See the “Molecular Beacons” section above for recommendations on the use of specific fluorophores and quenchers with the Opticon detection system.
Scorpions Probes
Scorpions probes (available from licensed suppliers) contain both an amplification primer and a target-specific probe separated by an amplification blocker. The probe portion is flanked by complementary sequences favoring formation of a stem structure which brings a fluorophore and a quencher into close proximity.
During amplification, extension of the target sequence proceeds from the primer portion of the Scorpions probe. As the reaction cools following denaturation, a uni-molecular rearrangement occurs such that the Scorpions probe sequence binds to the amplified tar­get sequence, separating the complementary stem sequences and thus the fluorophore and quencher. Since the Scorpions probe is integrated into the product, there is a direct relationship between the number of targets generated and the amount of fluorescence.
See the “Molecular Beacons” section above for recommendations on the use of specific fluorophores and quenchers with the Opticon detection system.
Amplifluor Universal Detection System
The Amplifluor system (available from Intergen Company of Purchase, NY) makes use of a universal primer that emits a fluorescence signal only following incorporation of the primer into an amplification product. The universal primer consists of a 18 base primer tail ("Z sequence") coupled to a hairpin sequence labeled with a fluorophore and a quencher. First, the target is amplified using target-specific primers, one of which has the Z sequence added to its 5' end. In the following round of amplification, the complement to the Z sequence is incorporated into the product. The universal primer then anneals to the complement of the Z sequence and extension proceeds. In the next cycle, extension proceeds through the universal primer incorporating it into the amplification product. In the process, the hairpin is unfolded separating the fluorophore and quencher and emit­ting a fluorescence signal that is proportional to the amount of amplified product.
See the “Molecular Beacons” section above for recommendations on the use of specific fluorophores and quenchers with the Opticon detection system.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 4-5
Compatible Chemistries, Sample Vessels, and Sealing Options
Selecting the Correct Sample Vessel
Important! Do not use full height 0.2ml tubes or full height unskirted
microplates. Full height 0.2ml tubes and most unskirted microplates do
not provide sufficient clearance between the sample block and lid-heater assembly. Do not force the cycler drawer closed.
For proper clearance in the Opticon unit, the distance from the bottom of a tube/plate to the cap rim can not exceed 17.5mm. In general, fully-skirted 96-well microplates, such as MJ Research Microseal
®
and Hard-Shell® microplates, provide sufficient clearance when sealed with either domed or flat optical caps (see the "Sample Vessel and Sealing Selec­tion Chart for Optical Assays" below). If unskirted microplates are used, low-profile plates, such as the MJ Research MLL-series Multiplate
microplates, are required.
Low-profile 0.2ml strip tubes, such as MJ Research TLS-series tubes, are recommended for small numbers of samples. Full-height 0.2ml tubes do not provide sufficient clearance.
Vessels Optimized for Fluorescence Detection and Thermal Cycling
For optimal sensitivity in fluorescence-detection assays, we recommend thin-walled 0.2ml tube strips and microplates with opaque-white wells. MJ Research, Inc. offers microplates and tubes with opaque white or clear wells designed for fluorescence detection assays and optimized to ensure a precise fit in the cycler block (see the "Sample Vessel and Sealing Selection Chart for Optical Assays" below).
Microplates and tubes with black wells may be useful in applications requiring very low levels of background. However, signal strength is significantly reduced when plates and tubes with black wells are used.
Note: In-factory calibration of the Opticon detector is performed with opaque-white plates. If you are using natural (clear) or black plates, refer to Chapter 10 for instructions on performing a calibration test and recalibrating.
Sealing Sample Vessels
Steps must be taken to prevent the evaporation of water from reaction mixtures during ther­mal cycling so as to avoid changing the concentration of reactants. Only a layer of oil or wax, such as Chill-out liquid wax, will completely prevent evaporation from sample ves­sels. However, an adequate degree of protection can be achieved by sealing with optical caps, then cycling the samples using the heated lid to prevent condensation/refluxing.
4-6 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Sealing with Optical Caps and the Heated Lid
The heated inner lid maintains the upper part of sample vessels at a higher temperature than the reaction mixture. This prevents condensation of evaporated water vapor onto the vessel walls, so that solution concentrations are unchanged by thermal cycling. The heated lid also exerts pressure on the tops of vessels loaded into the block, helping to maintain a vapor-tight seal and to firmly seat tubes or microplates in the block for the most efficient transfer of heat to and from the samples.
Optical caps must be used along with the heated lid to prevent evaporative losses. Ultra­clear, optical cap strips (available from MJ Research, Inc.) provide high light transmis­sion for fluorescence detection and vapor-tight sealing. Tight-fitting caps do the best job of preventing vapor loss.
Note: When tubes are cooled to below-ambient temperatures, a ring of condensation may
form in tubes above the liquid level but below the top of the sample block. This is not a cause for concern since it occurs only at the final cool-down step when thermal cycling is finished.
Sealing with Chill-out™ Liquid Wax
Clear Chill-out liquid wax (available from MJ Research, Inc.) may be used to seal sample vessels for optical assays. Clear Chill-out liquid wax is the same easy-to-use alternative to oil as the standard, red-colored Chill-out wax. However, clear Chill-out wax provides ex­cellent light transmission for optimal performance in optical assays. Chill-out liquid wax provides 100% prevention of condensation and vapor loss. At room temperature and above, this overlay is transparent and can be applied by pipet. Chill-out liquid wax so­lidifies below 14°C. Use only a small amount of Chill-out liquid wax; 1-3 drops (15-50µl) are usually sufficient. (Include this volume in the total volume when setting up a calcu­lated-control protocol.) Be sure to use the same amount of wax in all samples vessels to ensure a uniform thermal profile.
Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com 4-7
Compatible Chemistries, Sample Vessels, and Sealing Options
Sample Vessel and Sealing Selection Chart for Optical Assays
The following sample vessels and sealing options are recommended for use with the DNA Engine Opticon® system and are available from MJ Research, Inc. To place an order, call 888-729-2165 or fax 888-729-2166.
tcudorP
golataChcraeseRJM
rebmuN
sthgilhgihtcudorP
slesseV
eliforP-woL
sebutlm2.0
8fospirts
)etihweuqapo(1580-SLT
)raelc(1080-SLT
fosrebmunllamsroflaedI
selpmas
mumixamofsllewetihwesU
langis
mrofinurofsllewraelcesU
gniweivelpmasdnalangis
etalpitluMeliforP-woL
detriksnu
llew-69
setalporcim
)etihweuqapo(1569-LLM
)raelc(1069-LLM
selpmas69nahtrewefroflaedI
ezisottucebnacsetalporcim---
mumixamofsllewetihwesU
langis
mrofinurofsllewraelcesU
gniweivelpmasdnalangis
llew-69®laesorciM
s
etalporcimdetriks
)etihweuqapo(1569-PSM
)raelc(1069-PSM
mumixamofsllewetihwesU
langis
mrofinurofsllewraelcesU
gniweivelpmasdnalangis
llew-69®llehS-draH
setalporcimdetriks
sllewetihw/llehsetihw5569-PSH sllewetih
w/llehskcalb5669-PSH
sllewetihw/llehsder5169-PSH
sllewetihw/llehswolley5269-PSH
sllewetihw/llehseulb5369
-PSH sllewetihw/llehsneerg5469-PSH
sllewraelc/llehsetihw1069-PSH sllewraelc/llehskcalb1669-PSH
sllewraelc
/llehsder1169-PSH
sllewraelc/llehswolley1269-PSH
sllewraelc/llehseulb1369-PSH
sllewraelc/llehsneerg1469-
PSH
selpmas69roflaedI
talfyletulosbasniameretalP
rof,gnilcyclamrehtgnirud
noitcellocthgilmrofinu
rofdoog
erasllehsderoloC
gnidoc-roloc
snoitpOgnilaeS
.spactalflacitpO
8fospirts
3080-SCT
thgilhgihrofraelc-artlU
noissimsnart
5>semulovgnilcyclamrehT
diuqiL
tuO-llihC
edarg-lacitpo,xaW
1141-OHC
yalrevoliolarenimsecalpeR
noissimsnartthgilhgiH
2>semulovgnilcyc
-lamrehT
4-8 Tech Support: (888) 652-9253 • Sales: (888) 735-8437 • tech@mjr.com • www.mjr.com
Opticon System Operations Manual
Reaction Volume Recommendations
Reaction volumes of 20-100µl are recommended for most applications. However, it is beneficial to empirically optimize reagent concentrations and sample volumes with the Opticon detector as the sensitivity of the optical system often allows a cost-saving reduc­tion in reagent concentrations. Volumes as low as 10µl can be used, though sensitivity is slightly reduced.
The maximum recommended sample volume is 100µl. Volumes exceeding 100µl do not maintain adequate contact with the wells of the sample block resulting in nonuniform heating and cooling within the sample.
The reaction volume is used to calculate the temperature of the samples during a calcu­lated-control run (see the “Temperature Control Method” section in Chapter 6). Therefore, thermal accuracy is optimized when all samples contain identical volumes.
Loading...
+ 96 hidden pages