MITECH MFD500 Operating Manual

Page 1
MFD500
Ultrasonic Flaw Detector
Operating Manual
Mi
tech Inc. Ltd.
1
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 2
1 Introduction···································································································4
1.1 Features of the Instrument..................................................................4
1.2 Specifications......................................................................................6
1.3 Base Instrument Package...................................................................6
2 Understanding the Keypad, Menu System and Displays······························7
2.1 Structure Feature................................................................................7
2.2 Screen Display....................................................................................8
2.3 Keys and Rotary Knob Features.......................................................10
2.4 Menus and Functions.........................................................................11
2.5 Alarm Lights.......................................................................................11
3 Initial Start-up································································································13
3.1 Power Supply....................................................................................13
3.2 Connecting a Probe..........................................................................13
3.3 Starting the Instrument......................................................................14
4 Operation······································································································15
4.1 Adjusting the Display Range.............................................................15
4.2 Setting the Material Velocity..............................................................15
4.3 Setting the Display Delay..................................................................15
4.
4 Selecting Probe Test Mode...............................................................16
4.5 Selecting a Rectification Mode..........................................................16
4.6 Setting the A-Scan Reject Level.......................................................18
4.7 Setting the Gain................................................................................18
4.8 Changing the Gain-Adjustment Increment (DB STEP).....................18
4.9 Auto Gain Feature.............................................................................19
4.10 Configuring the Gates.....................................................................19
4.11 Setting the Damping Level..............................................................22
4.12 Setting the Pulse Energy Level.......................................................22
4.13 Setting the Pulse Width...................................................................23
4.14 Adjusting the Pulse Repetition Frequency (PRF)............................23
4.15 Specifying the Probe Frequency.....................................................23
4.16 Specifying the Piezo Crystal Size...................................................23
4.17 Setting the Probe X-Value...............................................................23
4.18 Probe Delay Calibration..................................................................24
4.19 Setting the Angle of Incidence........................................................24
4.20 Magnifying the Contents of a Gate..................................................24
4.21 Freezing the A-Scan Display...........................................................24
4.22 Setting the Display Grid..................................................................24
4.23 Selecting Units................................................................................25
4.24 Scale Setting...................................................................................25
4.25 Fill function......................................................................................25
4.26 Setting the Display Brightness........................................................26
4.27 Setting the Display Color................................................................26
4.28 Setting the A-Scan Color.................................................................26
4.
29 Key Sound......................................................................................27
4.30 Date and Time Setting....................................................................27
4.31 Display the System Information......................................................27
4.32 Resetting the Instrument.................................................................27
4.33 Connecting to a Computer..............................................................28
5 Calibration and Measurement·······································································29
5.1 Calibration with Straight- and Angle-Beam Probes...........................29
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
2
Page 3
5.2 Calibration with Dual-Element (TR) Probes......................................30
5.3 Distance Amplitude Curve.................................................................32
5.4 Measuring with AVG/AVG.................................................................34
5.5 Curved Surface Correction...............................................................37
5.6 Flaw Sizing Feature..........................................................................37
5.7 Crack Height Measuring Feature......................................................38
5.8 Envelope Function............................................................................38
5.9 Peak Hold Feature............................................................................39
5.10 B-Scan Feature...............................................................................39
5.11 Data Storage...................................................................................40
6 Maintenance and Care··················································································42
6.1 Care of the Instrument......................................................................42
6.2 Care of the Batteries.........................................................................42
6.3 Maintenance.....................................................................................42
6.4 Warranty...........................................................................................42
6.5 Tips on Safety...................................................................................42
Appendix··········································································································43
Appendix A: Charging the battries..........................................................43
Ap
pendix B: Menu Structure...................................................................45
3
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 4
1 Introduction
The MFD350 is an advanced digital ultrasonic flaw detector featuring a multi-color TFT LCD and a host of new features to meet challenging inspection requirements. It combines powerful flaw detection and measurement capabilities, extensive data storage, and the ability to transfer detailed inspection data to the PC via its high-speed USB port. The instrument incorporates many advanced signal processing features including a 10MHz RF bandwidth to permit testing of thin materials, narrowband filters to improve signal to noise in high gain applications and a spike pulser to optimize penetration on thick or highly attenuating materials. The instrument can be widely used in locating and sizing hidden cracks, voids, disbands, and similar discontinuities in welds, forgings, billets, axles, shafts, tanks and pressure vessels, turbines, and structural components.
1.1 Features of the Instrument
Th
e instrument extends the performance and range of applications that are
apable of being satisfied by a portable instrument. The quality, portability,
c d
urability, and dependability that you have come to expect from the popular
Mitech MFD Series of instruments remain.
Display
Hi-resolution (320 × 240 pixels) multi-color TFT LCD Display with nine user-selectable color display schemes and brightness control provides high contrast viewing of the waveform from bright, direct sunlight to complete
darkness. The hi-resolution color LCD display with fast 60 Hz update gives an analog look to the waveform providing detailed information that is critical in many applications including nuclear power plant inspections.
Range
Up to 6000 mm in steel. Suitable for use on large workpieces and in high-resolution measurements.
Pulser
Pulse Energy selectable among Low, Medium and High Pulse Repetition Frequency adjustable from 10 Hz to 1 KHz in 1 Hz increments. Damping selectable among 100Ω、200Ω and 400Ω for optimum probe
performance Test Modes include Pulse echo, dual and thru-transmission
Receiver
Sampling10 digit AD Converter at the sampling speed of 160 MHz RectificationPositive Halfwave, Negative Halfwave, Fullwave and RF Analog Bandwidth: 0.5MHz to 10MHz capability with selectable frequency
ranges (automatically set by the instrument) to match probe for optimum performance. Gain0 dB to 110 dB adjustable in selectable steps 0.1 dB, 2 dB, 6 dB, and locked.
4
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 5
Gates
Two fully independent gates offer a range of measurement options for signal height or distance using peak triggering. The echo-to-echo mode allows accurate gate positioning for signals which are extremely close together. Gate Start: Variable over entire displayed range Gate Width: Variable from Gate Start to end of displayed range Gate Height: Variable from 0 to 99% Full Screen Height Alarms: Threshold positive/negative
Memory
Memory of 100 channel files to store calibration set-ups Memory of 1000 wave files to store A-Scan patterns and instrument settings. All the files can be stored, recalled and cleared.
Functions
l Semiautomatic two point calibration: Automated calibration of transducer
ze
ro offset and/or material velocity
l Flaw LocatingLive display Sound-path, Projection (surface distance),
Depth, Amplitude,
l Flaw sizing: Automatic flaw sizing using AVG/AVG or DAC, speeds reporting
of defect acceptance or rejection.
l Digital Readout and Trig. Function: Thickness/Depth can be displayed in
digital readout when using a normal probe and Peam path, Surface Distance and Depth are directly displayed when angle probe is in use.
l Both the DAC and the AVG method of amplitude evaluation are available. l Curved Surface Correction Feature l Crack Height Measure function l Magnify gatespreading of the gate range over the entire screen width l Video Recording and play l Auto-gain function l Envelope: Simultaneous display of live A-scan at 60 Hz update rate and
envelope of A-scan display
l Peak Hold: Compare frozen peak waveforms to live A-Scans to easily
interpret test results.
l A Scan FreezeDisplay freeze holds waveform and soundpath data l B Scan display feature
Real Time Clock
The instrument clock keeps running tracking the time.
Communication
High speed USB2.0 port. The optional DataPro software helps manage and format stored inspection data for high-speed transfer to the PC. Data can be printed or easily copied and pasted into word processing files and spreadsheets for further reporting needs. New features include live screen capture mode and database tracking.
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
5
Page 6
Battery
Internal rechargeable Li-ion battery pack rated 7.2V at 6600 mAh 8 hours nominal operating time depending on display brightness 8-10 hours typical recharge time
Knob
Operating adjustments are easily and quickly made using the rotary knob.
1.2 Specifications
l Range: 0 to 6000 mm, at steel velocity l Material Velocity: 1000 to 9999m/s l Display Delay: -20 to 3400 µs l Probe Delay/Zero Offset : 0 to 99.99µs l Sensitivity: 110 dB max in selectable resolution 0.1, 1.0, 2.0, 6.0 dB and
locked.
l Test Modes: Pulse echo, dual element and thru-transmission l Pulse Repetition Frequency ranges from 10 Hz to 1000 Hz l Pulse Energy: Low, Medium and High l Damping: 100, 200, 400 ohms l Bandwidth (amplifier bandpass ): 0.5 to 10 MHz l Gate Monitors: Two independent gates controllable over entire sweep range l Rectification: Positive halfwave, negative halfwave, fullwave, RF l System Linearity: Horizontal: +/-0.2% FSW, Vertical: 0.25% FSH, Amplifier
Accuracy +/-1 dB.
l Reject (suppression): 0 to 80% full screen height l Units: Inch or millimeter l
Transducer Connections: BNC
l Power Requirements: AC Mains 100-240 VAC, 50-60 Hz l Dimensions: 263H×170W×61D mm
l Operating Temperature: -10 to 50 l Storage Temperature: -30 to 50
1.3 Base Instrument Package
Portable Ultrasonic Flaw Detector with Hi-Resolution Color LCD Display Straight Beam Transducer (2.5 MHz, Φ20)
Angle Beam Transducer (2.5 MHz, 13 mm×13 mm, 63.5°) Interconnect Cable for the transducer (Q9-Q9) Rechargeable Li-Ion Battery Package, 6.6 amp hour
Power supply/charger unit Supporting pillar Operating Manual in English
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
6
Page 7
2 Understanding the Keypad, Menu System and Display
2.1 Structure Feature
The right figure gives an overview of the instrument system.
1The Main Unit 2、Probe (Transducer)
The Main Unit
1, Belt 2, LCD Display 3, LOGO 4. Function Keys 5 Menu Keys 6, Alarm LED 7, Power LED 8, Keypad 9, Rotary Knob 10, Product Label 11, Support Pillar Screw 12, Battery Case 13, Power adapter port 14, Battery Switch 15, USB Socket 16, Probe Cable Port (Transmit) 17, Probe Cable Port(Receive)
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
7
Page 8
2.2 Screen Display
The instrument displays are designed to be easy to interpret.
The Instrument has a digital screen for the display of A-Scan. Top figure shows the main screen.
The names of the menus or submenus are displayed at the bottom of the screen. The selected menu or submenu is highlighted.
Indicated at the right of the display, next to the A-Scan, are the functions of the corresponding menu. The display of the functions disappears on the home screen.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
8
Page 9
The measurement line below the
.
mode;
wave display shows values of range settings and measured values.
: Amplitude height. : Sound path
: Projection. :Depth A 61.3 means gate mode is single; selected gate is gate A; and peak echo position: sound path 61.3 mm. AB 24.0 means Echo-Echo selected gate is gate A; soundpath:
24.0 mm.
The setting line below the measurement line shows the channel file name, the beam angle, the velocity and the gain.
The status area at the bottom-right corner shows the system status. The following status will be showed:
: Indicates the battery capacity
: Straight beam probe
: Angle beam probe
: Single element probe
: Dual element probe : Through-transmit mode : Envelope function on
: Peak hold function on
: USB online; : A-Scan freezed
The echo position area at the top-right corner indicated shows the projection and the depth of the peak echo.
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
9
Page 10
2.3 Keys and Rotary Knob Features
The instrument is designed to give the user quick access to all of the instruments functions. Its easy-to-use menu system allows any function to be accessed with no more than three key presses. To access any function: l Press one of the five menu keys (F1~F5) to select a menu. The menus
across the bottom of the display will immediately be replaced with the submenus contained in the selected menu.
l Press a menu key (F1~F5) again to select the submenu containing the
desired function.
l Up to four functions will be
displayed in the function bar on the right side of the display. Select the desired function, by pressing one o
f the four function keys (S1~S4). Change the value listed in
l
the
function box with the rotary knob. Some values can also be adjusted with repeatedly pressing the
orresponding function key.
c
PWR
ANGL
DAC
DAC AVG
AVG
ALRM
Special keys such as Gain, Gate, Range, Probe Zero-Offset Calibration, Angle Calibration, Freeze, Save Waveform, Auto-Gain, Envelop, Peak Hold and etc. are grouped together for easy thumb control. This design allows direct access to important instrument set-up parameters and provides easy and
fast operating in difficult
inspection environments
A/B
AUTO
GAIN
Flaw Detector
ENVE PEAK LOPE
Ultrasonic
CAL
REC
Home key immediately returns the instrument to the main screen Freeze key freezes the A-Scan Magnify key expands the gate range to full screen
Alarm key turns the gate alarm on or off
For turning the instrument on or off
Db-Step key selects the gain submenus or switch the db step
A/B
Gate key - Gate functions selecting gate functions and Gate A/B switch
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
10
Page 11
Range key - Range function selecting
CAL
Probe-Delay key – Probe Delay calibration
ANGL
Probe-Angle key selects the angle calibration functions
Save key performs a data-storage of the A-Scan pattern
AUTO GAIN
Auto-Gain key starts or stops the Auto-Gain function
ENVE LOPE
Envelope key turns the Envelope function on and off
PEAK
REC
Peak-Hold key turns the Peak-Hold function on and off.
Video key starts or stops recording a segment of the display
DAC AVG
DAC-AVG key selects the menus of DAC/AVG
searches next echo towards left, or move the cursor to left when inputting
digit numbers.
searches next echo towards right, or move the cursor to right when inputting
digit numbers.
confirms or switches current selection
Rotary knob: For direct setting the currently selected function. A single key press on the rotary knob will function as pressing .
2.4 Menus and Functions
The instrument menu system consists of several menus, submenus, and functions. It allows the operator to select and adjust various features and instrument settings. l Available menus are accessed via the Home Menu. Note that the menus
visible on your particular instrument depend on which options are installed.
l Each menu contains several submenus. l Menus and submenus are selected by pressing the key below the desired
item (F1 to F5).
l When a submenu is selected, the functions contained in that submenu are
listed in the Function Bar down the right-hand side of the display screen.
l Functions are then selected by pressing the adjacent Function Key (S1 to
S4).
l Turning the Function Knob, and in some cases continuing to press the
adjacent Function Key, will change the value shown in the selected functions box.
Note that some functions, like RANGE, have both coarse and fine adjustment m
odes. Coarse and fine modes are selected by pressing the adjacent Function Key more than once. An icon of when the function is in fine adjustment mode. When the function is in coarse adjustment mode, turning the function knob will produce large changes in the selected functions value. When the function is in fine adjustment mode, turning the function knob will change the value by smaller amounts.
will appear on the left of the function name
2.5 Alarm Lights
Two alarm lights appear at the top-right corner front of the instruments keypad. One alarm light is marked as ALRM, which is assigned to the gate alarm. When a gate alarm is triggered, this light will illuminate. The other light is marked as PWR indicating power and battery status.
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
11
Page 12
Status ALRM LED PWR LED Gate alarm Red on × Battery charging × Flash red/green Battery charged with
× Green on external power connected External power connected
× Green on No battery installed No external power × off
NDTzone • Non-Destructive Testing Equipment Specialists
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
12
Page 13
13
3 Initial Start-up
3.1 Power Supply
The instrument can be operated with an external power adapter or with batteries. You can connect the instrument to the mains supply system even if it carries batteries. A discharged battery is charged in this case, viz. parallel to the instrument operation.
Operation Using the Power Supply Unit
Connect the instrument to the mains socket-outlet using the power supply unit. The plug receptacle is at the top left of the instrument. Push the plug of the power supply unit into the plug receptacle until it snaps into place with a clearly audible click. The PWR LED on the keypad of the instrument will light in green color if the connection is properly aligned
Operation Using Batteries
Use a lithium-ion battery pack provided with the instrument for the battery operation. The battery compartment is situated at the instrument back. The lid is fastened with 4 attachment block. To insert the battery pack l Move the four attachment block of the battery compartment downward in
order to loosen them.
l Lift the lid off upward. At the bottom of the battery compartment, you will see
two springs.
l
Insert the battery into the battery compartment. Make sure that the two
ntacts on the back of the battery pack are connected with the springs in
co th
e battery compartment.
l Close the battery compartment and fasten the attachment blocks. l Check the battery switch on the top of the instrument. Make sure to switch it
on before internal charging. Note: l The battery switch on the top of the instrument must be set to ON when
operating with the battery pack or charging it. l When not using the instrument, set the switch to OFF to save power.
3.2 Connecting a Probe
To prepare the instrument for operation, you have to connect a probe to it. The instrument is available with the probe connectors BNC. When connecting a probe to the instrument, its not only important that the probes physical connection be properly made. Its also important that the instrument is properly configured to work with the installed probe. The Instrument will operate with one or two single-element probes or with a dual-element probe. To install a single-element probe, connect the probe cable to either of the two ports on the front of the instrument. When two probes, or a dual-element probe is connected to the instrument, the Receive probe connector should be installed in the right port and the Transmit’ probe connector in the left port.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 14
14
Connect one single element probe to either port. Connect leads from a Dual Element Probe to both ports. For through-transmission, connect two single element probes to the transmit
(left, labeled as ) and receive (right, labeled as ) ports.
3.3 Starting the Instrument
To start the instrument, press the switch-on key . If it operates on the internal battery pack, make sure to set the battery switch to ON position before starting. The start display of the instrument appears; here you will also see the current software version and the serial number of the instrument. The instrument carries out a self-check and then switches over to stand-by mode. The settings of all function values are the same as before switching-on of the instrument. The instrument will shut off automatically when the battery capacity level is too low.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 15
15
4 Operation
4.1 Adjusting the Display Range
The function group BASE enables you to make the basic adjustment of the display range. The display on the screen must be adjusted for the material to be tested and for the probe used. Calibration requires the use of calibrated standard made of the same material as the test piece. Prior to calibrating the instrument/probe combination, the A-Scan display-screen range (the material thickness value represented by the full horizontal width of the screen) will normally be set to a value equal to or slightly larger than the calibrated standard.
- Activate the BASE submenu (located in the BASIC menu) by pressing the menu key below it. Three functions will appear down the right side of the display screen.
- Press the function key next to the selection titled RANGE. Youll note that RANGE has both coarse and fine adjustment modes. Coarse and fine modes are selected by repeatedly pressing the corresponding function key next to it.
- To change the range, turn the knob. Note that the maximum of range setting is 6000 mm.
- The display’s horizontal range will remain as set.
R
ange is selectable in fixed steps
or continuously variable.
4.2 Setting the Material Velocity
Use VELOCITY to set the sound velocity within the test object. Always ensure that the function VELOCITY is correctly set. The instrument calculates all range and distance indications on the basis of the value adjusted here. Velocity range1000m/s9999m/s
Coarse adjustment, in steps as follows :
2260m/s 2730m/s 3080m/s 3230m/s 4700m/s 5920m/s 6300m/s
Fi
ne adjustment: 1000~9999 in steps of 1 m/s
.3 Setting the Display Delay
4
Here you can choose whether to display the adjusted range (for example 300 mm) starting from the surface of the test object, or in a section of the test object
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 16
16
starting at a later point. This allows you to shift the complete screen display and consequently also the display zero. If the display should for example start from the surface of the test object, the value in D-DELAY must be set to 0. To set the display delay
- Activate the BASE submenu (located in the BASIC menu) by pressing the menu key below it. Three functions will appear down the right side of the display screen.
- Press the function key next to the selection titled D-DELAY.
- To change the display delay, turn the knob. Youll note that the displayed echoes shift to the left or right.
Range: -20~3400 µs Coarse adjustment: 10 pixel space (in µs) Fine adjustment: 1 pixel space (in µs)
4.4 Selecting Probe Test Mode
You can use the function PRB MODE to activate the pulser-receiver separation. The following modes are available:
l SINGLE- For single-element operation ( will be displayed); the probe
connection sockets are connected in parallel.
l DUAL – For the use with dual-element (TR) probes ( will be displayed);
the right-hand socket is connected with the amplifier input whereas the initial pulse is available at the left-hand socket.
l THRU – Through-transmission mode for the use with two
single-element probes ( will be displayed); the receiver is connected with right, the pulser is connected with left.
- Activate the RCVER submenu (located in the P/R menu) by pressing the menu key below it.
- Press the function key next to the selection titled PRB MODE.
- To change the probe mode, turn the knob. Each available probe mode is represented by an icon thats displayed near the down-right corner of the display whenever that probe mode is indicated.
- The probe mode will be set to the last one displayed.
4.5 Selecting a Rectification Mode
You can select the rectification mode of the echo pulses according to your application in the function RECTIFY. Rectification effects the orientation of the A-Scan on the display screen. The A-Scan represents the pulse (echo) thats returned from the material being tested to the instrument. The series of echoes looks like the Radio Frequency (RF) signal.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 17
17
Note that the RF signal has a negative component below the axis, and a positive component above the axis. RF rectification is useful when evaluating a probe. RF rectification is forbidden when displaying the DAC/AVG curves.
Positive Half Rectification means that only the upper (positive) half of the RF signal is displayed. Negative Half Rectification means that only the bottom (negative) half of the RF signal is displayed. In the figure above, note that even though it’s the negative half of the RF signal, its displayed in the same orientation as a positive component. This is only to s
implify viewing. The signal displayed
he view identified as Negative
in t Rectification is the negat
ive
component of the RF signal.
Full-Wave Rectification combines the positive and negative rectified signals together, and displays both of them in a positive orientation. Full-wave rectification is recommended for most inspections.
Use the following procedure to select a rectification mode
- Activate the RCVER submenu (located in the P/R menu) by pressing the menu key below it. Three functions will appear down the right side of the display screen.
- Press the function key next to the function titled RECTIFY. Youll note that there are four options:
l POS- Shows the positive component of the RF signal l NEG- Shows the negative component of the RF signal but displays
it in a positive orientation
l FULL-Shows the positive and negative halves of the RF wave, but
both are oriented in the positive direction.
l RF- Shows the echo with no rectification
- Select the rectification method.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 18
18
4.6 Setting the A-Scan Reject Level
The function REJECT allows you to suppress unwanted echo indications, for example structural noise from your test object. To omit a portion of the A-Scan, you must define the percentage of full-screen height you wish to omit. You may omit A-Scan up to 80% of the screen height. To Set a reject percentage
- Activate the RCVER submenu (located in the P/R Menu) by pressing the menu key below it.
- Press the function key next to the selection REJECT.
- Set the required percentage value by turning the knob.
You should handle this function with great caution, as it may of course happen that you suppress echoes from flaws as well. Many test specifications expressly forbid using the reject function.
4.7 Setting the Gain
Instrument gain, which increases and decreases the height of a displayed A-Scan, is adjusted with the gain functions.
The gain is displayed as: A+B+C dB
. It includes the basic gain, the offset gain and the sensitivity correction. The maximum of the total gain (A+B+C) is 110 dB.
The offset gain is for the purpose of flaw evaluation. The DAC/AVG curves remain the same when adjusting the offset gain. Sensitivity correction: Compensate for the transfer losses in the material under test. This correction is necessary if test object and reference block have different surface qualities. You have to find out the adjustment value for the compensation of transfer losses by experiments. The gain is varied accordingly in this correction, while the curve line remains the same. To set the sensitivity correction, select the function T-CORR and then adjust its value according to the experiment. W
hen DAC is active, adjusting the basic gain setting will result in an equal adjustment of the DAC curve position. To increase the instrument gain without changing the DAC curve position, use the T-CORR function to compensate for
arying couping/surface conditions.
v
4.
8 Changing the Gain-Adjustment Increment (DB STEP)
When adjusting the A-Scan Gain, each click of the adjustment knob increases
decreases the gain level by a dB increment equal to the DB STEP. Several
or values can be specified for DB STEP. Available increments include: 0.1dB, 1dB, 2dB, 6dB and 0dB. Note that setting the DB STEP to 0dB prevents adjustment of the instrument gain and prevents any changes using the knob.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 19
19
To select one of the existing DB STEP values:
- Activate the GAIN submenu (located in the BASIC menu) by pressing the menu key below it.
- Press the function key next to the selection titled DB STEP.
- Change the DB STEP value by repeatedly pressing the corresponding function key or by turning the knob.
Note that DB STEP value can be quickly changed by repeatedly pressing the key .
4.9 Auto Gain Feature
Use the Auto Gain function to automatically set the basic gain, so as to adjust the peak echo to a target height.
- Set the target echo height using the AUTO function located in the FUNC submenu.
- Press the screen.
- The Auto Gain function will end when the echo height reaches the target value. Or you can stop the function by pressing
AUTO GAIN
to start the Auto Gain function. AUTO-XX% will be shown on
AUTO GAIN
again.
4.10 Configuring the Gates
The gates monitor the range of the test object where you expect to detect a flaw. If an echo exceeds or falls below the gate, an alarm signal is sent out via the ALRM LED. The gate chooses the echo for the digital time-of-flight or amplitude measurement. The instrument has two gates: A gate and B gate. Gate A and B are independent of one another. Setting the position and characteristics of the A­and B-Gates is the first step to configuring the instrument for flaw detecting or ma
terial-thickness measurement. The GATE menu controls not only the location of the A and B-Gates, but also the alarms and other features activated when an A-Scan signal crosses a specific gate. A-Scan signals crossing the A or B-Gate are evaluated for the purposes of flaw detection and material-thickness evaluation. When the signal crosses the A or
Gate, the maximum point (peak) of the signal (in the specific gate) is used for
B­evaluation purposes. The measured value is indicated in the measurement l
Selecting a Gate
lecting a gate by using the GATE
Se SE
L function.
ou can also repeatedly press
Y
A/B
to switch the selected gate between Gate A and Gate B. The selected gate will be shown in solid line. The unselected gate will be shown in dashed line.
ine.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 20
20
Positioning Gates
Use the following procedures to set the vertical and horizontal position of the A and B-Gates. Remember that gate position has the following effects on instrument performance: l A-Scan echoes on the right side of the display screen represent features
that occur at a greater depth from the test-material surface than those on the left of the display screen. Therefore, moving a gate to the right means that the gate is evaluating a deeper portion of the test material.
l A wider gate will simply span the equivalent of more test-material depth l Increasing the vertical height (called threshold) of a gate means that only
reflected signals of sufficiently large amplitude will cross the gate.
Setting a Gate’s Starting Point
You can fix the starting point of the gates A or B within the display range.
- Activate the GATE submenu (located in the GATE menu).
-
Select the gate to be positioned using the GATE SEL function.
- Select the gate START function and adjust the starting point by turning the knob. Increasing and decreasing the value of the starting point moves the gate to the right and left, respectively.
- The gate starting point will remain as set, even when width adjustment is made.
Coarse adjustment: 10 pixel space
ne adjustment: 1 pixel space
Fi Wh
en the gate exceeds the display
nge, as shown in right figure, the
ra m
easured value will be shown as *
in the measurement line.
You can also set the selected gates starting point using the Quik Search Function. Quick Search Function is achieved by simply pressing the and .
Note that the instrument will only search the echo whose peak crosses the selected gate in this function.
Adjusting a Gates Width
You can determine the gate width within the display range.
- Activate the GATE submenu (located in the GATE menu).
- Select the gate to be positioned using the GATE SEL function.
- Select the gate WIDTH function and adjust its value by turning the knob.
Coarse adjustment: 10 pixel space Fine adjustment: 1 pixel space
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 21
21
Setting a Gate’s Threshold (Vertical Position)
You can determine the threshold value of the gates within the range of 0 to 99% screen height for triggering the LED alarm if this value is exceeded or not reached, depending on the setting of the LOGIC function.
- Activate the GATE submenu (located in the GATE menu).
- Select the gate to be positioned using the GATE SEL function.
- Select the THRESH function and adjust the vertical height by turning the knob. Increasing and decreasing the value of the threshold moves the gate up and down, respectively.
D
efining B-Gate Alarm Logic
This function allows you to choose the method for triggering the B-Gate alarm. Gate ala (P
OSITIVE logic) or when no echo crosses the gate (NEGATIVE logic) within
rms can be set to trigger when an A-Scan echo crosses the gate the displayed range. A-Gate is locked to POSITIVE logic. And
-Gate can be set to either POSITIVE or
B N
EGATIVE. The appearance is different between the POSITIVE gate and the NEGATIVE gate when displayed. This is shown in right figures.
Note that the alarm and measurement function of the gates is only active within the display range.
Use the following procedure to specify B-Gate logic:
- Activate the LOGIC submenu (located in the GATE menu) by pressing the menu key below it.
- Select the LOGIC function and choose the gate-alarm triggering logic.
Turning the Audible Alarm On or Off
When a gate alarm is activated, one or more of the following will occur:
l An alarm indication light on the front of the instrument will illuminate l An audible alarm (HORN) will sound
Use the following procedure to turn the horn off or on:
- Activate the LOGIC submenu (located in the GATE menu) by pressing the menu key below it.
- Select the ALARM function and turn the audible alarm ON or OFF.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 22
22
Setting the Gate Work Mode
Two options for the gate work mode: SINGLE and E-E. SINGLE: measures the distance from the surface to the peak echo. E-E (Echo-Echo mode): measure the distance between two peak echoes (the two peak echoes should be selected by two gates). It is very useful in thickness measuring. Use the GATE MODE function in the LOGIC submenu to select the gate work mode.
4.11 Setting the Damping Level
This function serves for matching the probe. You can use it to adjust the damping of the probes oscillating circuit and to consequently change the height, width and resolution of the echo display.
- Activate the PULSER submenu (located in the P/R menu) by pressing the menu key below it.
- Press the function key next to the selection titled DAMPING.
- To change the specified damping level and optimize the A-Scan signal appearance, continue pressing the function key or turn the knob. You’ll note that the following damping levels are available: 100、200、400.
- The damping level will be set to the one last displayed.
4.12 Setting the Pulse Energy Level
Use the function ENERGY to set the pulser voltage. The relative energy with which the pulser fires can be set to LOW, MEDIUM or HIGH. The setting MEDIUM is recommended for most inspections. HIGH is used for inspections in which maximum sensitivity is import, e.g. for the detection of small flaws. Choose the setting LOW for broadband probes or if narrow echoes are
required (better lateral resolution).
To set the pulser energy
- Activate the PULSER submenu (located in the P/R menu) by pressing the menu key below it.
- Press the function key next to the function titled ENERGY.
- Select the energy level by turning the knob or continuing to press the function key.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 23
23
4.13 Setting the Pulse Width
This function is Not Available for MFD350.
4.14 Adjusting the Pulse Repetition Frequency (PRF)
The pulse repetition frequency indicates the number of times an initial pulse is triggered per second. You can determine whether you need the highest possible PRF value, or whether you are satisfied with a low value. The PRF value ranges from 10 Hz to 1000 KHz. The larger your workpiece, the smaller PRF values are needed in order to avoid phantom echoes. In the case of smaller PRF values, however, the A-Scan update rate becomes lower, for this reason, high values are required if a workpiece should be scanned fast. The best way to determine the suitable PRF value is by experimenting: start from the highest step and reduce the value until there are no more phantom echoes. To set the PRF level
- Activate the PULSER submenu (located in the P/R menu) by pressing the menu key below it.
-
Press the function key next to the function titled PRF.
- Change the PRF value by turning the knob.
4.15 Specifying the Probe Frequency
In this function, you can specify the probe frequency according to the frequency of your probe. The instrument will automatically utilize a built-in filter to match the probe frequency.
- Activate the PROBE submenu (located in the P/R menu) by pressing the menu key below it.
- Press the function key next to the selection titled XTAL FREQ.
- To change the specified frequency, turn the knob.
- The probe frequency will be set to the last one displayed.
4.16 Specifying the Piezo Crystal Size
Use EFF DIAM function to set the Piezo Crystal Size of your probe. This value is required when programming the AVG.
4.17 Setting the Probe X-Value
The function X-VALUE enables you to set the X-Value (distance between the probes leading face and probe index/sound exit point) of the probe used. This value is required for the automatic calculation of the reduced projection distance.
Adjustment range: 0~50 mm
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 24
24
4.18 Probe Delay Calibration
Every probe has a delay line between the transducer element and the coupling face. This means that the initial pulse must first pass through this delay line before the sound wave can enter the test object. You can compensate for this influence of the delay line in the function P-DELAY. If the value for P-DELAY is not known, read Chapter 5 in order to determine this value.
4.19 Setting the Angle of Incidence
The ANGLE function enables you to adjust the angle of incidence of your probe for the material used. This value is required for the automatic calculation of the flaw position. The angle of incidence for the straight-beam probe is fixed to 0°.
Adjustment range: 0°~80°
4.20 Magnifying the Contents of a Gate
Whenever an A-Scan is active, pressing the key enlarges the displayed portion of the A-Scan contained in the selected gate. Any of the available gates may be specified as the selected gate. The width of the magnified gate determines the level of magnification since the display is magnified until the gate width equals the full-screen width. The display will contain the magnified
view until pressing again.
Before Magnifying
After Magnifying
4.21 Freezing the A-Scan Display
Whenever an A-Scan is active, pressing freezes the A-Scan display. The active A-Scan will remain as it appeared when was pressed and the display will remain frozen until is pressed again.
4.22 Setting the Display Grid
- Activate the SYS submenu (located in the CFG menu) by pressing the menu key below it. Three functions will appear down the right side of the display screen.
- Press the function key next to the selection titled GRID.
- To change the grid type, continue pressing the function key or turn the knob. Each grid style is shown in the display screens A-Scan window as
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 25
25
it is selected. Youll note that the following styles are available:
l G1-Ten major horizontal and five vertical divisions. l G2-Two major horizontal and vertical divisions. l G3-No on-screen grid. Only display-edge marks are visible.
- The grid style will be set to the last one displayed.
4.23 Selecting Units
In the function UNITS you can choose your favorite units between mm or inch.
- Activate the SYS submenu (located in the CFG menu) by pressing the menu key below it. Three functions will appear down the right side of the display screen.
- Press the function key next to the selection titled UNITS. Youll note that the following options are available:
l METRIC-metric unit system. l IMPERIAL-Imperial unit system
- To change the units of measurements, repeatedly pressing the corresponding function key or by means of the knob.
- The unit of measurement will be set to the choice last displayed.
4.24 Scale Setting
As an alternative to the measured values, the instrument enables to display a scale in the measurement line. The scale gives you an overview of the position of echoes. You have a choice between a dimensionless ten-division scale and a scale showing the real position of the echoes.
The following settings are possible: l
S-PATH: Display of sound path scale
l P-D: Display of projection distance
scale
l DEPTH: Display of depth distance
scale
l DIV: Display of a dimensionless scale l RANGE: Display of Range scale
Select the function SCALE. Then use the knob to set the required display mode. As an alternative, you can repeatedly press to switch the scale.
4.25 Fill function
The function FILL toggles between the filled and the normal echo display mode. The filled echo display mode improves the echo perceptibility due to the strong contrast, especially in cases where workpieces are scanned more quickly.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 26
26
4.26 Setting the Display Brightness
Use the function BRIGHT to set the display brightness. You can choose between four brightness options: 25%, 50%, 75% and 100%.
- Activate the LCD submenu (located in the CFG menu) by pressing the menu key below it. Four functions will appear down the right side of the display screen.
- Press the function key next to the selection titled BRIGHT.
- To change the brightness level, continue pressing the function key or turn the knob.
-
The display brightness will remain at the level last displayed. Note: For the 25% option, the instrument consumes less current and consequently increases the operating time in battery operation.
4.27 Setting the Display Color
The SYS COLOR value determines the color of the background, the gate and the DAC/AVG curves. All color schemes are recommended for indoor operation while S0 and S3 are best suited for outdoor operation.
- Activate the LCD submenu (located in the CFG menu) by pressing the menu key below it. Four functions will appear down the right side of the display screen.
- Press the function next to the selection titled SYS COLOR. There are nine preset color schemes.
- To change the displays color scheme, continue pressing the function key or turn the knob.
- The display color will remain at the scheme last displayed.
4.28 Setting the A-Scan Color
- Activate the LCD submenu (located in the CFG menu) by pressing the menu key below it.
- Press the function key next to the selection titled WAV COLOR. There are eight A-Scan color options.
- To change the A-Scans color, continue pressing the function key or turn the knob.
- The A-Scan echo will remain the color last displayed.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 27
27
4.29 Key Sound
The key sound can be turned on or off using the KEY SOUND function located in the HORN submenu.
4.30 Date and Time Setting
For a correct documentation you should always make sure that you are using the correct date and time setting.
- Activate the CLOCK submenu (located in the CFG menu) by pressing the menu key below it. Two functions will appear down the right side of the display screen.
- Press the function key next to the function Y-M-D. The date is displayed in Year-Month-Day format. Note that the first time you press the function key, the year item is highlighted. The next time you press the function key, the month item is h
ighlighted. Finally, pressing the function key again will cause the day item to be highlighted.
- To change the year, month or days, turn the knob while the desired item is highlighted.
- When complete, press the function one more time. The current date will be set to the date displayed.
Time is displayed in Hour-Minute format. And the time setting procedure is similar to that of the date. Once set, the internal clock of the instrument will maintain the current date and time.
4.31 Display the System Information
You can enter the INFO submenu to see the software version and the serial number of the instrument. The VERSION and the S/N values are read-only. They cannot be modified by the user.
4.32 Resetting the Instrument
In case the instrument can no longer be operated, or you need to make a basic initialization (factory setting), you can reset the instrument. The instrument can be reset by the SYS RESET function. All the memory data, including the wave files, the channel files and the video files, will be cleared during system reset. And the instrument settings will be reset to default. To reset the instrument
- Activate the RESET submenu (in the MEM menu) by pressing the menu key below it.
- Press the function key next to the SYS RESET function.
- Continue pressing the function key or turn the knob to trigger the reset
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 28
28
action. It will prompt out Reset system to original ?”. Press the menu key below YES to confirm the reset operation. Or press the menu key below NO to cancel the reset operation.
NOTE:
l The effects of resetting the instrument may not be reversed. l No key action should be performed during resetting process.
4.33 Connecting to a Computer
The instrument is equipped with a USB port at the upper left of the instrument. The PC can connect with the instrument via the USB port. A-Scan display, instrument settings, and videos stored in the memory of the instrument can be transferred to the computer through the USB port. Detailed information of the communication software and its usage refer to the software manual.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 29
29
5 Calibration and Measurement
Before working with the instrument, you have to calibrate the instrument: you have to adjust the material velocity and display range and allow for the probe delay depending on the material and dimensions of the test object. To ensure a safe and proper operation of the instrument, it is necessary that the operator be adequately trained in the field of ultrasonic testing technology. Below you will find some examples of common calibration methods for certain test tasks.
5.1 Calibration with Straight- and Angle-Beam Probes
Case A: With Known Material Velocity Calibration process
- Set the known material velocity in VELOCITY.
- Couple the probe to the calibration block.
- Set the required display range in RANGE. The calibration echo must be displayed on the screen.
- Position the gate on one of the calibration echoes until the sound path of the echo is indicated in the measurement line.
- After this, change the adjustment of the function P-DELAY until the correct sound path for the selected calibration echo is indicated in the measurement line.
Example:
You are carrying out the calibration for the calibration range of 300 mm via the function group BASE using the calibration block DB-P (thickness 225 mm) which is laid flatwise.
-
Set RANGE TO 300 mm.
- Set the known material velocity of 5920 m/s in VELOCITY.
- Set the gate so that it is positioned on the first calibration echo (from 225 mm).
- Read the sound path in the measurement line. If this value is not equal to 225 mm, change the adjustment for the function P-DELAY until it is at 225 mm.
225
This completes the calibration of the instrument to the material velocity of 5920 m/s with a calibration range of 300 mm for the straight-beam probe used.
Case B: With Unknown Material Velocity
Use the semiautomatic calibration function of the instrument via the function group CAL for this calibration case. The distances between 2 calibration echoes must be entered as default data. The instrument will then carry out a plausibility check, calculate the material v
elocity and the probe delay, and automatically set the parameters.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 30
30
Calibration Process
- Set the required display range in RANGE. The two calibration echoes selected must be displayed on the screen. Set the range so that the second calibration echo is located on the right edge of the screen.
- Select the function group PDELAY.
- Enter the distances of the two calibration echoes in S-REF1 and S-REF2.
- Position one gate on the first calibration echo.
- Position the other gate on the second calibration echo.
- Press the function key next to the selection CAL to trigger the calibration.
The correct calibration is confirmed by the message Calibration is finished. The instrument will now automatically determine the sound velocity and the probe delay and set the corresponding functions accordingly. The value of the function P-DELAY will be set to the correct value. If the instrument is not able to carry out any valid calibration on the basis of the input values and the echoes recorded, a corresponding error message is displayed. In that case, please check the values of your calibration lines and r
epeat the process of recording the calibration echoes.
Example:
- Enter the distances (thicknesses) of the two calibration lines S-REF1 (225 mm) and S-REF2 (450 mm).
- Position one gate on the first calibration echo.
- Position the other gate on the second calibration echo.
-
- Press the function key next to the function CAL to carry out the calibration.
The valid calibration is briefly confirmed and carried out.
5.2 Calibration with Dual-Element (TR) Probes
Dual-element (TR) probes are especially used for wall thickness measurement. The following peculiarities must be taken into account when using these probes:
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 31
31
Echo Flank. Most dual-element (TR) probes have a roof angle (transducer elements with inclined orientation toward the test surface). This causes mode conversions both at beam index (sound entry into the material) and at the reflection from the backwall, which can result in very jagged echoes. V-Path Error. Dual-element (TR) probes produce a v-shaped sound path from the pulser via the reflection from the backwall to the receiver element. This so-called V-path error affects the measuring accuracy. You should therefore choose two wall thicknesses that cover the expected thickness measurement range for the calibration. In this way, the V-path error can be corrected to the greatest possible extend.
Higher Material Velocity. Due to the V-path error, a higher material velocity than that of the test material is given during calibration, especially with small thicknesses. This is typical of dual-element (TR) probes and serves for compensation of V-path error. With small wall thicknesses, the above-described effect leads to an echo amplitude drop which has to be especially taken into account with thicknesses less than 2 mm. A stepped reference block having different wall thicknesses is required for calibration. The wall thicknesses must be selected so that they cover the expected readings.
C
alibration Process
We recommend to use the semiautomatic calibration function for the calibration with T/R probes.
- Set the required test range
- Increase the probe delay until the two calibration echoes selected are displayed within the range
- Set the pulser and receiver functions according to the probe used and the test application.
- Select the function group PDELAY.
- Enter the distances of the two calibration echoes in S-REF1 and S-REF2.
- Position one gate on the first calibration echo.
- Position the other gate on the second calibration echo.
- Press the function key next to the selection CAL to trigger the calibration.
The correct calibration is confirmed by the message Calibration is finished. The instrument will now automatically determine the sound velocity and the probe delay and set the corresponding functions accordingly. The value of the function P-DELAY will be set to the correct value. If necessary, check the calibration on one or several known calibration lines, e.g. using the stepped reference block.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 32
32
1.
5.3 Distance Amplitude Curve
The instrument is available with Multi-Curve Distance Amplitude Curve (DAC) function. Functions for the distance-amplitude curve are accessed through the
DAC Menu, which is located by pressing When displayed, the DAC curve visually represents a line of constant reflector peaks over a range of material depths. A new feature of the instrument is a multiple-curve option that displays three dB offset DAC curves simultaneously. Each curve represents constant reflector size at varying material depth. Remember that in DAC mode, the only deviation from traditional display and operation is the appearance of the DAC curve. All A-Scan echoes are displayed at their non-compensated height. A DAC curve is programmed using a series of same-reflector echoes at various depths covering the range of depths to be inspected in the test material. Because near field and beam spread vary according to transducer size and frequency, and materials vary in attenuation and velocity, DAC must be programmed differently for different applications. A DAC curve can be based on up to 16 data points (material depths).These points are recorded from the DAC me
nu as described below.
DAC AVG
.
Recording the DAC Curve
DAC Curve points are typically taken from a standard with equally sized reflectors (holes) located at various material depths. The primary echo from each of these points (for up to a total of 16 echoes) are recorded. When DAC is active, the instrument displays a curve that represents echo peaks for constant reflectors at varying material depth. Before starting to record a reference curve, the instrument must be correctly calibrated. To program the DAC Curve:
DAC AVG
- Access the DAC menu by pressing
, then select the DAC submenu
by pressing the menu key below it.
- Press the function key next to the PROGRAM function. Repeatedly press the PROGRAM function key or turn the knob to Trigger the DAC program. When DAC program is triggered, DAC and a digit number indicating the point index appears on the top right corner of the display.
- Couple the probe to the first reference point and adjust the gate so that it
is broken by the primary echo. If necessary, adjust the gain so that the echo crosses the gate and the highest peak in the gate is at approximately 80% of full-screen height. The highest peak must not be
higher than 100% full-screen height. Note that you can press
or
to search the next echo very quickly.
- While the gate is lined up over the first reference echo, press the key next to the RECORD function. Continue to press the key next to the RECORD function to record the first DAC Curve point. Note that the largest echo to cross the A-Gate will be treated as the reference echo. The gain value at which this point is recorded becomes the baseline gain value.
- Continue to record additional curve points up to a maximum of 16 points (note that at least two DAC Curve points are required). As soon as you have recorded at least two curve reference points, your DAC is already
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 33
33
active.
- When complete press the key next to FINISH and follow the on-screen prompts to exit the DAC program.
Note that stored DAC curve points can be edited as described in next section.
Editing DAC Curve
After reference points are recorded, their values may be manually adjusted, or some points may be manually deleted. To edit or delete points:
- With the DAC menu accessed, select EDIT submenu.
- Press the function key next to the EDIT function. Continue to press the function key or turn the knob to start the EDIT process. When DAC EDIT process is triggered, DAC and a digit number appear indicating the point index on the top right corner of the display.
- Press or to select the point to edit.
- To adjust a points height-Press the function key next to the ADJUST function. Turn the knob clockwise or anticlockwise to adjust the points echo height.
- To delete a point – Press the function key next to the DELETE function. Continue to press the function key or turn the knob to delete that point from the DAC records.
- When complete press the function key corresponding to FINISH and follow the on-screen prompts to finish the editing.
Creating DAC Offset Curves
When operating with DAC turned on, three DAC curves including RL, SL and EL are typically displayed. The three curves can be offset from the original DAC curve (ML) by a user-inputted amount.
- Select the function DAC-RL (or DAC-SL, DAC-EL) located in the OFFSET submenu.
-
Change the offset value by
means of the knob.
The offset range from -50dB to 50dB.
Specifying the DAC Reference Curve
Once the DAC curve is recorded and displayed (by turning DAC DISP on), echoes are automatically compared to the size reference curve, which can be specified in the SIZE REF selection. The options include RL, SL, EL, and ML. The dB equivalent height of the signal above or below the corresponding DAC Reference Curve amplitude is displayed as a comparison result.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 34
34
The comparison result is shown as right figure. Note that the result is like: RL-3.9
Deleting a DAC Curve
To delete a stored DAC curve
- With the DAC menu activated, select the DEL submenu.
- Press the function key next to the DEL DAC function.
- Press the function key next to the DEL DAC function a second time or turn the knob. Then confirm your selection.
Turning On/Off the DAC display
Turning DAC DISP On and Off causes the DAC curves to be displayed or removed.
- With the DAC menu activated, select the DISP submenu.
- Press the function key next to the DAC DISP function.
- Press the function key next to the DAC DISP function a second time or tu
rn the knob to turn On/Off the DAC display. Note that if the DAC curve is not created, it will prompt out No DAC curve found and DAC DISP will be set to OFF.
Echo Evaluation with DAC
In order to be able to evaluate a flaw indication by means of the DAC, certain conditions must be met.
l The DAC must already be recorded. l It only applies to the same probe that was used when recording the curve.
Not even another probe of the same type must be used.
l The DAC only apply to the material corresponding to the material of the
reference block.
l All functions affecting the echo amplitude must be set the same way as they
were when the curve was recorded.
5.4 Measuring with DGS/AVG
Using the AVG (the same as DGS) function, you can compare the reflecting power of a natural flaw in the test object with that of a theoretical flaw (circular disk-shaped equivalent reflector) at the same depth. Note: You are comparing the reflecting power of a natural flaw with that of a theoretical flaw. No definite conclusions may be drawn on the natural flaw (roughness, inclined position, etc.). The so-called AVG diagram forms the basis for this comparison of the reflecting
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 35
35
power. This diagram consists of a set of curves showing the correlation of three influencing variables:
l Distance between the probe and circular disk-shaped equivalent reflector l Difference in gain between various large circular disk-shaped equivalent
reflectors and an infinitely large backwall
l Size of the circular disk-shaped equivalent reflector. The influencing variable
always remains constant for one curve of the set of curves. The advantage of the AVG method lies in the fact that you can carry out reproducible evaluations of small discontinuities. The reproducibility is most of all important, for example, whenever you aim to carry out an acceptance test. Apart from the influencing variables already mentioned, there are other factors determining the curve shape: sound attenuation, transfer losses, amplitude correction value, probe and etc. The following probe parameters affect the curve shape:
l Element or crystal diameter l Frequency l Delay length l Delay velocity
Y
ou can adjust these parameters on the instrument in such a way that you can
use the AVG method with many different probes and on different materials. Note: Before setting the AVG function, the instrument must first be calibrated
because all functions affecting the AVG evaluation mode (VELOCITY, P-DELAY, DAMPING, ENERGY, P-WIDTH, XTL FREQ, EFF DIAM, RECTIFY) can no longer be changed after the reference echo has been recorded.
Settings for the AVG measurement
Before using the AVG feature to evaluate reflectors in test pieces, the characteristics of the attached probe must be specified, certain characteristics of the reference standard must be input, and a reference echo must be stored. You must select the REF submenu and input the characteristics for the probe youve connected including:
XTL FREQ- The probe’s frequency rating
EFF DIAM – The probe elements effective diameter rating
Record the Reference Echo that Defines the AVG Curve
Prior to generating the AVG curve, a test standard with a known reflector must be used to define a reference point. Acceptable test standards include these reference types:
l BW – Backwall echo with reference defect size defined as infinity l SDH – Side Drilled Hole with a reference defect size defined as the hole’s
diameter. l FBH – Flat Bottom Hole with a reference defect size equal to the holes
facial diameter. Follow these steps to record a reference echo:
- Select the REF submenu, then the REF TYPE function. This function a
llows you to select one of the three reference types described above.
- Se
lect the REF SIZE function, and specify the size of the known
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 36
36
standard’s reference flaw.
- Select the AVG submenu, then the PROGRAM function. Turn the knob to start the AVG program. When AVG program is triggered, AVG appears indicating the point index on the top right corner of the display.
- Couple the probe to the known standard, capture the reference flaw so that its reflected echo is displayed on the instruments A-Scan, and adjust selected Gates starting point to ensure that the resulting echo triggers the gate.
- Adjust the gain until the reference flaws A-Scan peak measures 80% of FSH (Full Screen Height).
- With the probe coupled to the standard, and the reference flaws echo captured by selected Gate, press the function key next to the RECORD function to record an AVG reference echo.
- After the AVG reference echo is recorded, its value may be manually adjusted using the ADJUST function.
- When complete press the key next to FINISH and follow the on-screen prompts to exit the AVG program.
N
ote that only one AVG reference echo can be stored at a time. To delete the
currently stored reference
- Access the DEL submenu
- Select DEL AVG, and then follow the on-screen prompts.
Once a reference echo has been recorded, the AVG curves (the AVG-UL, the AVG-ML and the AVG-DL) are automatically displayed using default settings.
Creating AVG Offset Curves
When operating with AVG turned on, three AVG curves including the AVG-UL, the AVG-ML and the AVG-DL are typically displayed. The three curves can be offset by user-inputted equivalent flaw size values in the AVG-UL, AVG-ML and AVG-DL selection. The offset values will usually depend on the largest acceptable flaw size.
Deleting AVG Curve
To delete stored AVG curve
- With the AVG menu activated, select the DEL submenu.
- Press the function key next to the DEL AVG function.
- Press the function key next to the DEL AVG function a second time or turn the knob. Then follow the on-screen prompts to confirm the deleting.
Turning On/Off the AVG display
Turning the AVG DISP On and Off causes the AVG curve to be displayed or removed.
- With the AVG menu activated, select the DISP submenu.
- Press the function key next to the AVG DISP function.
-
Press the function key next to the AVG DISP function a second time or
turn the knob to turn On/Off the AVG display. Note that if the AVG curve is not created, it will prompt out No AVG curve found and AVG DISP
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 37
37
D
i
s
t
a
n
c
e
value will be OFF. Also note that switching this value to OFF does not delete the curve - it simply removes the curve from the display and disables the AVG mode.
Evaluating Test Samples in AVG Mode
Once the curve is recorded and displayed (by turning AVG DISP on), echoes are automatically compared to the AVG curve and evaluated based on the recorded reference. The result is shown as ERS (Equivalent Reflector Size), which evaluates the reflected echo and calculates the equivalent reflector size.
This is shown in right figure. Note that the ERS is Φ2.1 mm.
5.5 Curved Surface Correction
When using angle beam transducers on a curved surface the instrument will calculate the surface distance and depth of defect, taking into ac
count the internal or external
adius.
r
THICKNESS
THICK: Use the THICK function to set the materials wall thickness. This value is required for the automatic calculation of the real depth. Adjustment range: 10~5000 mm O-DIAM (Outside diameter of the test object): You will need the CSC function for tests on circular curved surfaces, for example when testing longitudinally welded tubes. In order to make the instrument carry out the corresponding correction of (reduced) projection distance and depth, you should enter the outside diameter of your test object in this function. Adjustment range: 2~5000 mm
CSC: If you plan to carry out the flaw position calculation for the curved surface workpieces, the CSC function should be turned on. Otherwise turn off the CSC function.
Flaw Depth
Outside
Inside
5.6 Flaw Sizing Feature
Using the Flaw Sizing feature, you can calculate the equivalent size of the flaw
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 38
38
easily and quickly.
- Position the selected gate over the echo from the backwall of the workpiece.
- Activating the SIZE submenu locating in the GATE menu by pressing the corresponding key below it.
- Select the BACK GAIN function. Then press the corresponding function key next to it to trigger the back-echo recording and analysis process. The instrument will record the echo and calculate the required gain sensitivity automatically. And then the instrument will automatically adjust the gain to the calculation result.
- Position the selected gate over the echo from the flaw.
- Select the CALC function. Then press the corresponding function key next to it to calculate the equivalent size.
The calculation result (the equivalent size of the flaw) will be displayed in the CALC bar.
5.7 Crack Height Measuring Feature
This feature allows measuring the height of a crack found inside the workpiece. Before starting to use the Crack Height Measure function, the instrument and the probe must be correctly calibrated.
- Activate the CRACK submenu locating in the GATE
menu by pressing the corresponding key below it.
- Position the selected gate over the echo from the first point of the crack. Then record this position using the A-POINT function.
- Position the selected gate over the echo from the second point of the crack. Then record this position in the B-POINT function.
- The instrument will automatically calculate the crack height using the recorded parameters and displays the result in the HEIGHT bar.
Crack Height
5.8 Envelope Function
When the Envelope function is activated, the echoes with the highest amplitude within selected gate will be recorded and displayed statically on the screen additionally to the live A-Scan. And the icon will appear near the down right of the display.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 39
39
To activate the Envelope function
- Activate the FUNC submenu (located in the GATE menu) by pressing the menu key below it.
- Press the function key next to the selection titled ENVELOPE.
- To turn on/off the ENVELOPE function, continue pressing the function key or turn the knob.
ENVE
Also you can continue pressing
LOPE
to turn on/off the Peak Hold function.
Note that Peak Hold and Envelop cannot function simultaneously.
5.9 Peak Hold Feature
When Peak Hold is activated, the echo with the highest amplitude in the selected gate becomes the reference echo displayed statically on the screen additionally to the live A-Scan. And the icon will appear near the down right of the display. You can use this function to record and document the peaked, maximum echo display.
To activate the Peak Hold Feature
- Activate the FUNC submenu (located in the GATE menu) by pressing the menu key below it.
- Press the function key next to the selection titled PEAK HOLD.
- To turn on/off the Peak Hold function, continue pressing the function key or turn the knob.
PEAK
Also you can continue pressing
REC
to turn on/off the Peak Hold function.
Note that Peak Hold and Envelop cannot function simultaneously.
5.10 B-Scan Feature
In B-Scan mode, the instrument displays a graphical cross-section of the workpiece.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 40
40
Use the following procedure to switch to B-Scan mode.
- Select the B-SCAN function.
- Switch the value to ON by means of the knob.
- Use the BSCAN DIR function to set the scan-display direction.
5.11 Data Storage
The instrument is equipped with a data-storage system. A-Scans patterns, instrument settings, and video can be stored in data files for later retrieval. File data can be saved, recalled and deleted by using the MEM menu.
Wave Files
W
ave files are used to store A-Scan pattern, along with the instrument settings.
Data in the wave files can be stored, recalled and erased.
Wave STORE
ou can save the A-Scan pattern to a wave file using the STORE function.
Y
- Select the function FILE.
- Use the knob to set the file name where you would want to store the current A-Scan pattern (F000~F999).
- Select the function STORE
- Use the knob or repeatedly press the corresponding function key to trigger the STORE operation.
Note:
The asterisk (*) before a selected file name indicates that this file is already occupied. It is not possible to overwrite an occupied file; select another file n
ame which is still empty, or clear the occupied file. To avoid loss of data e.g. in
case of a software update you should save the file to a PC.
Wave RECALL
You can recall a stored wave file; your instrument will then display all the test-relevant technical features that existed at the moment of the setup. A frozen display of the stored A-Scan appears.
Wave CLEAR
An occupied file is marked with an asterisk (*) before the file name. You can clear the file if you no longer need them. Wave Files are cleared using the CLEAR function.
- Select the function FILE.
- Use the knob to set the file name to clear (F000~F999).
- Select the function CLEAR.
- Use the knob or repeatedly press the corresponding function key to trigger the CLEAR operation. It will then prompt: Clear wave file?
- Confirm by pressing the corresponding key one more time.
The file is now cleared; the asterisk preceding the file name is no longer there.
Channel Files
Channel files (CH00~CH99) are used to store a specific instrument setup configuration. This means that whenever you recall a stored channel file, your instrument is again set up exactly the same as it was at the moment when the file was stored. This makes each one of your tests reproducible. You will find the following functions:
Channel STORE
To store instrument settings in a channel file follow this procedure
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 41
41
- Activate the CHANNEL submenu (located in the MEM menu) by pressing the menu key below it.
- Press the function key next to the selection titled FILE. Then change the file name to the one you desire by turning the knob. Note that if the file is no empty, an icon of * will appear before the file name and data cant be saved to this file.
- Press the function key next to the selection titled STORE. Perform the store operation by pressing the function key once more or turning the knob. If the store operation is successful, it will prompt out: Data saved. Note that if the channel file is not empty, the store operation will be canceled and prompt out a message: Error: This channel is not empty. You should change the FILE to an empty file to store the instrument settings.
Channel RECALL:
When the channel file is recalled, all instrument settings are modified to match the settings stored in the channel file. Once a channel file is opened, instrument settings may be modified from their channel file configurations. If necessary, s
ave the current instrument setup to a new channel file before loading a
channel file.
Channel CLEAR:
Channel files are cleared using this procedure:
- Activate the CHANNEL submenu (located in the MEM menu) by pressing the menu key below it.
- Press the function key next to the FILE function. Turn the knob until the desired file name appears.
- Press the function key next to the CLEAR function. Continue pressing the function key or turn the knob to perform the file CLEAR function. It will prompt out :“Clear this channel?. Press the menu key below YES to confirm the operation, or press the menu key below NO to cancel the operation.
Note: The deleted file may not be retrieved. The current settings of the instrument will be reset to default after the channel clear operation.
Video Files
Video file (DV01) is used to store a segment of video. To record a segment of video
- Activate the video recording operation by pressing
. An icon will
appear in the status area when recording the video.
- To stop the recording, simply press once more.
Note that the recording process will automatically end when the video file is full. Use RECALL function to RECALL a video file. The recalled video file will be
shown at a fixed refresh rate. Press to stop the show. Or press to pause and restart the show.
RESET Function
You can clear all channels, all waves or all videos using the functions in the RESET submenu.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 42
42
6 Maintenance and Care
6.1 Care of the Instrument
Clean the instrument and its accessories using a moist cloth. Only use the following recommended instrument cleaners: water, a mild household cleaner or alcohol (no methyl alcohol).
Attention:
Do not use any methyl alcohol, solvents, or dye penetrant cleaners! The plastic parts can be damaged or embrittled by this.
6.2 Care of the Batteries
Capacity and life of batteries mainly depend on the correct handling. Please therefore observe the tips below. You should charge the batteries in the following cases:
l Before the initial startup l After a storage time of 3 months or longer l After frequent partial discharge
6.3 Maintenance
The instrument requires basically no maintenance.
Attention:
Repair work may only be carried out by authorized Service staff of us.
6.4 Warranty
When used in accordance with the manufacturers written instructions and u
nder normal operating conditions, the instrument is conditionally guaranteed to e free from defects in material and workmanship for a period of two years from
b date of shipment. This warranty shall not apply to equipment subjected to misuse or abuse, improper installation, alteration, neglect, or accident. Excluded from this warranty are expendable items such as transducers, interconnecting cables, and batteries. This warranty is limited to the original purchaser and is not transferable. No other warranty, expressed or implied, is made.
6.5 Tips on Safety
The design of the instrument meets safety standard. During the operation, it shall meet the specified external ambient condition, and the operator shall be fu
rnished with concerned technology background, so as to guarantee safe
operation.
Note:
1. This instrument is a non-destructive inspection instrument for inspecting material, it is not allowed to use as a medical instrument.
2. The instrument is limited to be used in lab and industrial environment.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 43
43
Appendix
Appendix A: Charging the Battries
The approximate level of remaining battery life is visually displayed by the battery indicator. When fresh batteries with full capacity are installed, the icon
will appear as “ ”. As the battery life is consumed, the icon will begin to “empty. When the battery indicator is almost empty as indicated by the symbol ,
charge the batteries as soon as possible. You may take a second battery pack or replacement battery pack with you if you cannot connect the instrument to mains. The instrument automatically shuts off when batteries are too weak for reliable operation. Settings are saved and restored when the instrument is turned on again. You can charge the lithium-ion battery pack either directly in the instrument or by
means of an external battery charger provided with the instrument.
Attention:
You should only use the batteries provided by us and the corresponding battery charger. An improper handing of the batteries and of the battery charger may cause explosion hazard.
Internal Charging
Check the battery switch on the top of the instrument. Make sure to switch it on before internal charging. If a battery pack is located in the instrument, the charging process is started automatically when you connect the plug-in power supply unit. You can carry out ultrasonic inspections and charge a battery at the same time. The charging time is 10 hours with a simultaneous ultrasonic inspection. If the instrument is not being used for ultrasonic inspections, the charging time is 8 hours. This charging time applies to ambient temperatures from 25 to 30.
Please take into consideration that the batteries are not charged to their full capacity at higher temperatures. The PWR LED display at the top-right corner of the keypad indicates the status of the charging process. It will flash red/green during the charging process. And the flash will stop when the charging process stops.
External Charging
Lithium-ion batteries can be charged by means of an external battery charger.
- Take out the battery pack from the battery compartment of the instrument.
- Plug the chargers power cord into a 100-240 VAC power source. Its Green indication lamp will illuminate when the charger is supplied with power.
- Push the adapter of the charger into the plug receptacle of the battery pack until it snaps into place. The green LED of the battery pack will light if the connection is properly aligned.
- When charging the battery pack, the battery pack’s RED indication lamp will light steadily.
- When charging is complete (typically after six to eight hours) the RED
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 44
44
indication lamp will shut off. You may remove the charger adapter from the battery pack.
- Reinstall the battery pack to the compartment of the instrument.
Note: l The battery switch on the top of the instrument must be set to ON for
internal charging
l The battery pack should be charged over a period of time, even not used.
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Page 45
Appendix B: Menu Structure
BASIC P/R GATE MEM CFG DAC AVG
BASE
GAIN
COMP
CSC
B­SCAN
RANGE DAMPING GATE SEL FILE GRID PROGRAM
PULSER
VELOCITY ENERGY START STORE UNITS RECORD RECORD D-DELAY P-WIDTH WIDTH RECALL SCALE ADJUST
GAIN PRB MODE GATE
RCVER
DB STEP RECTIFY B-LOGIC STORE BRIGHT ADJUST EFF DIAM OFFSET REJECT RECALL SYS
T-CORR XTL FREQ AUTO FILE KEY
PROBE
EFF DIAM ENVELOPE X-VALUE PEAK
CSC P-DELAY BACK
PDELAY
THICK S-REF1 CALC WAVES H:M O-DIAM S-REF2 VIDEOS
A-SCAN ANGLE HEIGHT VERSION
ANGLE B-SCAN HL-DIAM A-POINT S/N BSCAN DIR
PRF
GATE
THRESH
LOGIC
MODE
ALARM
W A V E
CLEAR FILE FILL EDIT XTL FREQ
C H A N N L
CLEAR
S Y S
D I S P
COLOR WAV
D
AVG A C
FINISH
E
REF D I T
DELETE REF TYPE FINISH
COLOR
DAC-RL AVG-UL
O F
OFF-
SET F S E T
SIZE REF
D
DEL E L
CAL
FUNC
SIZE
HOLD
GAIN
V I D
DAC-SL AVG-ML
E
RECALL DAC-EL AVG-DL
H O R N
SOUND
O
R E S E T
CLEAR CHANNELS
SYS
Y-M-D DEL DAC DEL AVG
C L O C K
RESET
CRACK
I N
HL-DEPTH B-POINT
F O
CAL
DAC DISP AVG DISP
D
DISP I S P
PROGRAM
FINISH
REF SIZE
45
Tel.: +1.877.363.4243 • eMail: sales@NDTzone.com • Web: www.NDTzone.com
NDTzone • Non-Destructive Testing Equipment Specialists
Loading...