Mirion Technologies UltraRadiac-Plus User Manual

Radiation Safety. Amplified
.
Personal Radiation
Monitor
7068914D User’s Manual
UltraRadiac™-Plus
Personal Radiation
Monitor
7068914D User's Manual
Body: 7068916D
Copyright 201 7, Mirion Techn olo gi es ( Canb er ra), Inc. The material in this document, including all information,
pictures, graphics and text, is the property of Mirion Technologies (Canberra), Inc. and is protected by U.S. copyright laws and international copyright conventions.
Mirion Technologies (Canberra) expressly grants the purchaser o f this prod uct the right t o copy an y material in thi s document for the p urchaser’s own use, including as pa rt of a submission to regulatory or legal authorities pursuant to the purchaser’s legitimate business needs.
No material in this document may be copied by any third party, or used for any commercial purpose or for any use other than that granted to the purchaser without the written permissi on of Miri on Technolo gies (Ca nbe rr a), Inc.
Mirion Technologies (C anberra), Inc, 107 Unio n Valley Road, Oak Ridge, TN 37830 Tel: 865-220-6300 FAX: 865-483-0406
http://www.canberra.com/
.
The information in this document describes the product as accurately as possible, but is subject to change without notice.
Printed in the United States of America.
UltraRadiac is a trademark of Mirion Technologies, Inc. and/or its affiliates in the United States and/or other countries.
2
Table of Conte nt s
Preface ............................................................ v
1. Introduction .............................................. 1
2. Controls and Indicators ........................... 5
The Function Keys ............................................. 5
The Battery Life Indicator .................................. 7
3. Basic Operation ........................................ 9
Powering the Unit .............................................. 9
Turning Off an Alarm ...................................... 10
Checking Yo ur Stay Times .............................. 12
Checking the Accumulated Dose ..................... 12
Clearing the Accumulated D ose ....................... 13
Reading the Alarm Threshold .......................... 13
Changing the Alarm Indicators ........................ 14
Using the Backlight .......................................... 15
4. Operational Verification Tests................16
Turning the Unit On ......................................... 17
Turning the Unit Off ........................................ 19
Instrument Test ................................................ 20
Display Test Sequence ................................ 20
Ending the Display Test ................................... 23
Backlight Test .................................................. 24
Sleep Mod e Test ............................................... 24
i
Completing the Test ......................................... 25
GM Tube Failure Indication ............................. 26
Operatio n at Low Temperatures ....................... 26
Error Displays .................................................. 27
5. Setting Up the Unit ..................................28
The Rate Mode ................................................. 29
The Rate Alar m........................................... 29
Reading the Rate Alarm Threshold ............. 31
Changin g t he Rate Alarm Indicators ........... 31
Changing the Rate Alarm Thresholds ......... 32
The Dose Mode ................................................ 35
The Dose Alarms ........................................ 36
Reading the Dose Alarm Threshold ............ 37
Changing the Dose Alarm Indicators .......... 38
Changing the Dose Alarm Thresholds ........ 39
The Source Finder Mode .................................. 41
Enabling the Source Finder Mode .............. 41
Exiting the Source Finder Mode ................. 42
The Sleep Mode ............................................... 42
Enabling the Sleep Mode ............................ 43
When the Unit Wakes Up ........................... 43
Exiting the Sleep Mode ............................... 44
The Data Logging Mode .................................. 44
Capturin g Data Pairs ................................... 45
ii
Viewing th e D ata Log ................................. 45
Clearing the Data Log ................................. 47
6. Maintenance.............................................48
Cleaning the Unit ............................................. 48
The Batteries .................................................... 49
Installing or Replacing the Batteries ........... 49
Charging the Batteries ................................ 51
The Battery Compartment ................................ 53
Latching the Compartment Door ................ 54
Attaching the Belt Clip..................................... 55
The Carrying Case ........................................... 55
A. Specifications ..........................................58
Features ............................................................ 58
Detector ............................................................ 59
Display ............................................................. 60
Alarms .............................................................. 60
Power ............................................................... 61
Environmental .................................................. 61
Physical ............................................................ 63
Qualification Testing ........................................ 63
Ordering Information ....................................... 63
iii
B. Default Alarm Settings ............................65
C. Display Units............................................66
D. Glossary ...................................................68
E. Disposing of This Equipment .................70
iv
Preface
The UltraRadiac­Plus is a hand­held pers ona l radiation monitor designed for the needs of first responders.
Measures and displays radiation dose
rate and tot al dose
Presett able two level au dio, visual an d
vibrating alarms for rate and dose.
Designed to operate in ext remes of
temperature, shock, humidity, dust, immersion and radiati on
Large, easy-to-read backlit LCD display
Intuitive six button user interface
Operable/ r eadable by personnel in fire
fighting or HAZMAT protective gear
Can be powered by NiMH rechargeable
batteries ( optional DC charger)
v
Ultra-small size (12.8 in.3) and weight
(9.5 oz)
Low life cycle costs due to calibration
stability and automatic self calibration
Mirion Technologies’s unique time-to-
count technology for wider dynamic range
Data logging capability to 999 points
“Stay time” di splay shows ti me
remaining to dose alarm
Source finder mode
Derived from military qualified
MRAD115 R adiac Set
vi
1. Introduction
The UltraRadiac-Plu s is a du r abl e, portable, personal radiation monitor. Once set up, it is preconfigured to detect certain types of radiation, and then alert you using an audible and/or vibrating alarm and a visual data display. It can log (save a reco r d of) radiation data for further evaluation by your Administrator.
READ THIS
If you don’t read anything else
in this manual, read the
chapters
Controls and
Indicators on pag e 5 and Basic Operations on page 9
. These cover the basics of using the UltraRadiac-
Plus Personal
Radiation Monitor.
Note: Operational pro cedures are the
responsibility of the UltraRadiac­Plus’s users. Instructions provided in this manual are “how to’s”, not procedural recommendations.
1
Chapter 1 Introduction
Your UltraRadiac-Plu s can measure and display both the instantaneous radiation dose rate (Rate), t he amount of radiation bein g measured at t his moment, and the total accumulated radiation dose receiv ed (Dose) since the dose memory was last cleared.
The Alarms
The UltraRadiac-Plus’s presettable alarms for both instantaneous R ate and cumul ative Dose can alert you to hazardous conditions. When an enabl ed alarm’s threshold is exceeded, a visual indicator will begin flashing and, optionally, an LED will blink and/or an audible alarm will sound. All units also include a vibrator alar m th at can be independently set.
The Radiation Units
Depending on the model, your UltraRadiac­Plus is factory-set to display the Rate and the Dose in one of these units of radiation measurement:
R (roentgen) – a measure of radiation
exposure.
Sv (sievert) – a measure of absor bed
dose equival ent.
2
Most illustrations in this manual show the radiation units as µR (microroentgens, one millionth of a roentgen). The units are defined in the appendix Default Alarm Settings on
page 65.
Equipment Check
If your organization’s procedures require periodic equipment checks, refer to the chapter Operational Verification Tests on page 16.
Setting Up the Unit
The UltraRadiac-Plu s has five modes: R ate, Dose, Source Finder, Sleep, and Data Logging. The chapter Setting Up the Unit on page 28, describ es them and tells you how to set and en able the Rate alar ms and the Dos e alarms.
Maintenance Procedures
The chapter Maintenance on page 48, lists the UltraRadiac-Plus's pr eventive maintenance checks and describes how to install, charge, or replace th e batteries. Th e U ltraRadiac-Plus uses four AAA batteries and can accept either alkaline or rechargeable Ni MH .
3
Chapter 1 Introduction
Note: The battery recharging d evi ce should
only be used with NiMH batteries. Connecting the charger with Alkaline batteries may result in severe damage to the batteries or the device.
4
The Function Keys
2. Controls and Indicators
This is a brief overview of the UltraRadiac­Plus’s controls and indicators and its protective case. Each of the controls and indicators is covered in detail in the remaining sections of the manual.
Figure 1 The UltraRadiac-Plus Front Panel
The Function Keys
The primary function of each of the function keys is briefly described below. The programming functions of these keys are covered in the chapter Setting Up the Unit on page 28.
5
Chapter 2 Controls and Indicators
ON/OFF
Press and hold to turn the unit on or off .
RATE
Press to change to the Rate Mode.
DOSE
Press to change to the Dose Mode.
ALARM
Press to see t he Stay Time, the number of minutes you can safely sta y in the area at the current Dose Rate. Specific information on Stay Time is given in the Glossary on page 68.
WARNING:
If the Dose Rate goes up,
your remaining Stay Time
will go down. You should check your Stay
Time frequently when you
a
re in the presence of
radiation.
6
The Battery Life Indicator
LIGHT
In the dark, press to
illuminate the display for about five seconds. Note that the backlight may not be visible in daylight or normal room lighting.
CLR/TEST
In the Rate Mode, press
and hold to enable the Display Test Sequence
The Battery Life Indicator
The UltraRadiac-Plus has two battery life indicat ors. See The Batteries on page 49 for instruct ions on chang i ng the ba t teries .
If a blinking 8 is displayed, the unit has
stopped functioning. Replace the batteries before the unit’s next use.
If a blinking BAT is seen in the top-left
corner of the display, the unit’s batteries have 10 hours or less of useful life. Replace the batteries as soon as possible.
If the display is blank, the batteries are
dead. Replace the batter ies before th e unit’s next use.
7
Chapter 2 Controls and Indicators
Remaining Battery Life
In the Rate Mode with the BAT indicator blinking , pr e s s the CLR/TEST key. A t hree­digit number indicating the approximate remaining battery life, in minutes, will be displayed.
8
Powering the Unit
3. Basic Operation
The first time you take your Ul traRadiac-Plus out of its shipping carton, you’ll have to install its batteries. For detailed instruction see The Batteries on page 49.
Powering the Unit
To turn the unit on, press ON/OFF
until 000 is displayed.
Release ON/OFF. In a few seconds, the instantaneous
rate will be displayed, with the word RATE at the top left o f the screen.
If a 8 or a blinking BAT indicator
appears in the display or if the display is blank, refer to The Battery Life Indicators on page 7.
To turn the unit off, press ON/OFF
until – – – is displayed.
9
Chapter 3 Basic Operation
Turning Off an Alarm
The UltraRadiac-Plu s has two alarms, Low Level and High Level for each of two modes, Rate and Dose.
If any alarm sounds, the next four sections tell you how you turn it off.
Clearing the Low Rate Alarm
Press the CLR/TEST key to turn off
the vibrator and the audio alert.
The visual alerts will continue until
the rate falls below the Low Rate Alarm threshold.
Clearing the High Rate Alarm
Press the CLR/TEST key to turn off
the vibrator.
The audio (if enabled) and visual
alerts will continue until the Rate falls below th e High Rate Alarm threshold.
10
Checking Your Stay Times
At that time, the unit may pass into
the Low Rate Alarm region depending on the set point. See above instructions for clearing this alarm.
Clearing the Low Dose A larm
Press the CLR/TEST key to turn off
the vibrator and the audio alert.
The visual alerts will continue until
the int egrated Dose is reset to 0.
To turn off the visual alerts, you’ll
have to clear t he dose memory. See Clearing the Accumulated Dose on page 13.
Clearing the High Dose Alarm
Press the CLR/TEST key to turn off
the vibrator.
The visual alerts will continue until
the int egrated Dose is reset to 0.
To turn off the alerts, you’ll have to
clear the dose memory. See Clearing the Accumulated Dose on page 13.
11
Chapter 3 Basic Operation
Checking Your Stay Times
Press ALARM to see the number of minutes you can safely stay in the area at the current Dose Rate. Specific information on Stay Time is given in the Glossary on page 68.
WARNING:
If the Dose Rate goes up, your remaining Stay Time will go down.
You should check your Stay Time frequently when you
are in the presence of
radiation.
Checking the Accumulated Dose
Press DOSE t o see the accumulated
dose (the word DOSE is at at the top right of the screen).
To return to the Rate Mode, press the
RATE key.
12
Clearing the Accumulated Dose
Clearing the Accumulated Dose
Pres s a nd hold DOSE + CLR/TEST. The display will flash for a few
seconds, then clear the accumulated dose.
To return to the Rate Mode, press the
RATE key.
Reading the Alarm Threshold
Press ALARM + RATE or ALARM
+ DOSE.
You’ll s ee an H, for High Level
Alarm, displayed.
To read the Lo w Level Alarm setting,
press the RATE (DOSE) key a gain; the display will show an L.
Press the CLR/TEST key; the
flashing display will show the selected (H or L) alarm threshold setting.
13
Chapter 3 Basic Operation
If the Vibrator Alarm has b een
activated, the display will alternate between the t hreshold value and a vertically scrolling horizontal bar (– – –).
Changing the Alarm Indicators
Press the ALARM + RATE keys or
the ALARM + DOSE keys.
You’ll s ee an H, for High Level
Alarm, displayed.
To change the Low Level indicators,
press the RATE (DOSE) key again; the display will show an L.
Press the CLR/TEST key to access
the selected (H or L) indi cators.
Repeatedly pres s the ALARM key
until you see the desired combination of indicators: AUDio, VISual, vibrator, or no indicators.
14
Using the Backlight
Using the Backlight
In dim light, you can enable the unit’s backlight t o read the display.
Press the LIGHT key.
The display’s backlight will turn on.
To conserve b attery power, t he
backlight will automatically turn off in about five seconds.
15
Chapter 4 Operational Verificati on Tests
4. Operationa l Verificat ion Tests
These Operational Verification Tests should be performed periodically, as required by your departmental procedures.
Installing the Batteries
The first time you take your Ul traRadiac-Plus out of its shipping carton, you’ll have to install its batteries. See The Batteries on page 49 for instructions.
CAUTION:
Use of non-rechargeable batteries or rechar geab le batteries other than those supplied or recommended by Mirion Technologies (Canberra) may cause an explosion or other fire hazard resulting in personal injury. Mirion Technologies (Canberra) is not liable for personal injury or damage resulting from the use of incompatible batteries.
16
Loading...
+ 56 hidden pages