Mirion Technologies UltraRadiac-Plus User Manual

Page 1
Radiation Safety. Amplified
.
Personal Radiation
Monitor
7068914D User’s Manual
Page 2
UltraRadiac™-Plus
Personal Radiation
Monitor
7068914D User's Manual
Page 3
Body: 7068916D
Copyright 201 7, Mirion Techn olo gi es ( Canb er ra), Inc. The material in this document, including all information,
pictures, graphics and text, is the property of Mirion Technologies (Canberra), Inc. and is protected by U.S. copyright laws and international copyright conventions.
Mirion Technologies (Canberra) expressly grants the purchaser o f this prod uct the right t o copy an y material in thi s document for the p urchaser’s own use, including as pa rt of a submission to regulatory or legal authorities pursuant to the purchaser’s legitimate business needs.
No material in this document may be copied by any third party, or used for any commercial purpose or for any use other than that granted to the purchaser without the written permissi on of Miri on Technolo gies (Ca nbe rr a), Inc.
Mirion Technologies (C anberra), Inc, 107 Unio n Valley Road, Oak Ridge, TN 37830 Tel: 865-220-6300 FAX: 865-483-0406
http://www.canberra.com/
.
The information in this document describes the product as accurately as possible, but is subject to change without notice.
Printed in the United States of America.
UltraRadiac is a trademark of Mirion Technologies, Inc. and/or its affiliates in the United States and/or other countries.
2
Page 4
Table of Conte nt s
Preface ............................................................ v
1. Introduction .............................................. 1
2. Controls and Indicators ........................... 5
The Function Keys ............................................. 5
The Battery Life Indicator .................................. 7
3. Basic Operation ........................................ 9
Powering the Unit .............................................. 9
Turning Off an Alarm ...................................... 10
Checking Yo ur Stay Times .............................. 12
Checking the Accumulated Dose ..................... 12
Clearing the Accumulated D ose ....................... 13
Reading the Alarm Threshold .......................... 13
Changing the Alarm Indicators ........................ 14
Using the Backlight .......................................... 15
4. Operational Verification Tests................16
Turning the Unit On ......................................... 17
Turning the Unit Off ........................................ 19
Instrument Test ................................................ 20
Display Test Sequence ................................ 20
Ending the Display Test ................................... 23
Backlight Test .................................................. 24
Sleep Mod e Test ............................................... 24
i
Page 5
Completing the Test ......................................... 25
GM Tube Failure Indication ............................. 26
Operatio n at Low Temperatures ....................... 26
Error Displays .................................................. 27
5. Setting Up the Unit ..................................28
The Rate Mode ................................................. 29
The Rate Alar m........................................... 29
Reading the Rate Alarm Threshold ............. 31
Changin g t he Rate Alarm Indicators ........... 31
Changing the Rate Alarm Thresholds ......... 32
The Dose Mode ................................................ 35
The Dose Alarms ........................................ 36
Reading the Dose Alarm Threshold ............ 37
Changing the Dose Alarm Indicators .......... 38
Changing the Dose Alarm Thresholds ........ 39
The Source Finder Mode .................................. 41
Enabling the Source Finder Mode .............. 41
Exiting the Source Finder Mode ................. 42
The Sleep Mode ............................................... 42
Enabling the Sleep Mode ............................ 43
When the Unit Wakes Up ........................... 43
Exiting the Sleep Mode ............................... 44
The Data Logging Mode .................................. 44
Capturin g Data Pairs ................................... 45
ii
Page 6
Viewing th e D ata Log ................................. 45
Clearing the Data Log ................................. 47
6. Maintenance.............................................48
Cleaning the Unit ............................................. 48
The Batteries .................................................... 49
Installing or Replacing the Batteries ........... 49
Charging the Batteries ................................ 51
The Battery Compartment ................................ 53
Latching the Compartment Door ................ 54
Attaching the Belt Clip..................................... 55
The Carrying Case ........................................... 55
A. Specifications ..........................................58
Features ............................................................ 58
Detector ............................................................ 59
Display ............................................................. 60
Alarms .............................................................. 60
Power ............................................................... 61
Environmental .................................................. 61
Physical ............................................................ 63
Qualification Testing ........................................ 63
Ordering Information ....................................... 63
iii
Page 7
B. Default Alarm Settings ............................65
C. Display Units............................................66
D. Glossary ...................................................68
E. Disposing of This Equipment .................70
iv
Page 8
Preface
The UltraRadiac­Plus is a hand­held pers ona l radiation monitor designed for the needs of first responders.
Measures and displays radiation dose
rate and tot al dose
Presett able two level au dio, visual an d
vibrating alarms for rate and dose.
Designed to operate in ext remes of
temperature, shock, humidity, dust, immersion and radiati on
Large, easy-to-read backlit LCD display
Intuitive six button user interface
Operable/ r eadable by personnel in fire
fighting or HAZMAT protective gear
Can be powered by NiMH rechargeable
batteries ( optional DC charger)
v
Page 9
Ultra-small size (12.8 in.3) and weight
(9.5 oz)
Low life cycle costs due to calibration
stability and automatic self calibration
Mirion Technologies’s unique time-to-
count technology for wider dynamic range
Data logging capability to 999 points
“Stay time” di splay shows ti me
remaining to dose alarm
Source finder mode
Derived from military qualified
MRAD115 R adiac Set
vi
Page 10
1. Introduction
The UltraRadiac-Plu s is a du r abl e, portable, personal radiation monitor. Once set up, it is preconfigured to detect certain types of radiation, and then alert you using an audible and/or vibrating alarm and a visual data display. It can log (save a reco r d of) radiation data for further evaluation by your Administrator.
READ THIS
If you don’t read anything else
in this manual, read the
chapters
Controls and
Indicators on pag e 5 and Basic Operations on page 9
. These cover the basics of using the UltraRadiac-
Plus Personal
Radiation Monitor.
Note: Operational pro cedures are the
responsibility of the UltraRadiac­Plus’s users. Instructions provided in this manual are “how to’s”, not procedural recommendations.
1
Page 11
Chapter 1 Introduction
Your UltraRadiac-Plu s can measure and display both the instantaneous radiation dose rate (Rate), t he amount of radiation bein g measured at t his moment, and the total accumulated radiation dose receiv ed (Dose) since the dose memory was last cleared.
The Alarms
The UltraRadiac-Plus’s presettable alarms for both instantaneous R ate and cumul ative Dose can alert you to hazardous conditions. When an enabl ed alarm’s threshold is exceeded, a visual indicator will begin flashing and, optionally, an LED will blink and/or an audible alarm will sound. All units also include a vibrator alar m th at can be independently set.
The Radiation Units
Depending on the model, your UltraRadiac­Plus is factory-set to display the Rate and the Dose in one of these units of radiation measurement:
R (roentgen) – a measure of radiation
exposure.
Sv (sievert) – a measure of absor bed
dose equival ent.
2
Page 12
Most illustrations in this manual show the radiation units as µR (microroentgens, one millionth of a roentgen). The units are defined in the appendix Default Alarm Settings on
page 65.
Equipment Check
If your organization’s procedures require periodic equipment checks, refer to the chapter Operational Verification Tests on page 16.
Setting Up the Unit
The UltraRadiac-Plu s has five modes: R ate, Dose, Source Finder, Sleep, and Data Logging. The chapter Setting Up the Unit on page 28, describ es them and tells you how to set and en able the Rate alar ms and the Dos e alarms.
Maintenance Procedures
The chapter Maintenance on page 48, lists the UltraRadiac-Plus's pr eventive maintenance checks and describes how to install, charge, or replace th e batteries. Th e U ltraRadiac-Plus uses four AAA batteries and can accept either alkaline or rechargeable Ni MH .
3
Page 13
Chapter 1 Introduction
Note: The battery recharging d evi ce should
only be used with NiMH batteries. Connecting the charger with Alkaline batteries may result in severe damage to the batteries or the device.
4
Page 14
The Function Keys
2. Controls and Indicators
This is a brief overview of the UltraRadiac­Plus’s controls and indicators and its protective case. Each of the controls and indicators is covered in detail in the remaining sections of the manual.
Figure 1 The UltraRadiac-Plus Front Panel
The Function Keys
The primary function of each of the function keys is briefly described below. The programming functions of these keys are covered in the chapter Setting Up the Unit on page 28.
5
Page 15
Chapter 2 Controls and Indicators
ON/OFF
Press and hold to turn the unit on or off .
RATE
Press to change to the Rate Mode.
DOSE
Press to change to the Dose Mode.
ALARM
Press to see t he Stay Time, the number of minutes you can safely sta y in the area at the current Dose Rate. Specific information on Stay Time is given in the Glossary on page 68.
WARNING:
If the Dose Rate goes up,
your remaining Stay Time
will go down. You should check your Stay
Time frequently when you
a
re in the presence of
radiation.
6
Page 16
The Battery Life Indicator
LIGHT
In the dark, press to
illuminate the display for about five seconds. Note that the backlight may not be visible in daylight or normal room lighting.
CLR/TEST
In the Rate Mode, press
and hold to enable the Display Test Sequence
The Battery Life Indicator
The UltraRadiac-Plus has two battery life indicat ors. See The Batteries on page 49 for instruct ions on chang i ng the ba t teries .
If a blinking 8 is displayed, the unit has
stopped functioning. Replace the batteries before the unit’s next use.
If a blinking BAT is seen in the top-left
corner of the display, the unit’s batteries have 10 hours or less of useful life. Replace the batteries as soon as possible.
If the display is blank, the batteries are
dead. Replace the batter ies before th e unit’s next use.
7
Page 17
Chapter 2 Controls and Indicators
Remaining Battery Life
In the Rate Mode with the BAT indicator blinking , pr e s s the CLR/TEST key. A t hree­digit number indicating the approximate remaining battery life, in minutes, will be displayed.
8
Page 18
Powering the Unit
3. Basic Operation
The first time you take your Ul traRadiac-Plus out of its shipping carton, you’ll have to install its batteries. For detailed instruction see The Batteries on page 49.
Powering the Unit
To turn the unit on, press ON/OFF
until 000 is displayed.
Release ON/OFF. In a few seconds, the instantaneous
rate will be displayed, with the word RATE at the top left o f the screen.
If a 8 or a blinking BAT indicator
appears in the display or if the display is blank, refer to The Battery Life Indicators on page 7.
To turn the unit off, press ON/OFF
until – – – is displayed.
9
Page 19
Chapter 3 Basic Operation
Turning Off an Alarm
The UltraRadiac-Plu s has two alarms, Low Level and High Level for each of two modes, Rate and Dose.
If any alarm sounds, the next four sections tell you how you turn it off.
Clearing the Low Rate Alarm
Press the CLR/TEST key to turn off
the vibrator and the audio alert.
The visual alerts will continue until
the rate falls below the Low Rate Alarm threshold.
Clearing the High Rate Alarm
Press the CLR/TEST key to turn off
the vibrator.
The audio (if enabled) and visual
alerts will continue until the Rate falls below th e High Rate Alarm threshold.
10
Page 20
Checking Your Stay Times
At that time, the unit may pass into
the Low Rate Alarm region depending on the set point. See above instructions for clearing this alarm.
Clearing the Low Dose A larm
Press the CLR/TEST key to turn off
the vibrator and the audio alert.
The visual alerts will continue until
the int egrated Dose is reset to 0.
To turn off the visual alerts, you’ll
have to clear t he dose memory. See Clearing the Accumulated Dose on page 13.
Clearing the High Dose Alarm
Press the CLR/TEST key to turn off
the vibrator.
The visual alerts will continue until
the int egrated Dose is reset to 0.
To turn off the alerts, you’ll have to
clear the dose memory. See Clearing the Accumulated Dose on page 13.
11
Page 21
Chapter 3 Basic Operation
Checking Your Stay Times
Press ALARM to see the number of minutes you can safely stay in the area at the current Dose Rate. Specific information on Stay Time is given in the Glossary on page 68.
WARNING:
If the Dose Rate goes up, your remaining Stay Time will go down.
You should check your Stay Time frequently when you
are in the presence of
radiation.
Checking the Accumulated Dose
Press DOSE t o see the accumulated
dose (the word DOSE is at at the top right of the screen).
To return to the Rate Mode, press the
RATE key.
12
Page 22
Clearing the Accumulated Dose
Clearing the Accumulated Dose
Pres s a nd hold DOSE + CLR/TEST. The display will flash for a few
seconds, then clear the accumulated dose.
To return to the Rate Mode, press the
RATE key.
Reading the Alarm Threshold
Press ALARM + RATE or ALARM
+ DOSE.
You’ll s ee an H, for High Level
Alarm, displayed.
To read the Lo w Level Alarm setting,
press the RATE (DOSE) key a gain; the display will show an L.
Press the CLR/TEST key; the
flashing display will show the selected (H or L) alarm threshold setting.
13
Page 23
Chapter 3 Basic Operation
If the Vibrator Alarm has b een
activated, the display will alternate between the t hreshold value and a vertically scrolling horizontal bar (– – –).
Changing the Alarm Indicators
Press the ALARM + RATE keys or
the ALARM + DOSE keys.
You’ll s ee an H, for High Level
Alarm, displayed.
To change the Low Level indicators,
press the RATE (DOSE) key again; the display will show an L.
Press the CLR/TEST key to access
the selected (H or L) indi cators.
Repeatedly pres s the ALARM key
until you see the desired combination of indicators: AUDio, VISual, vibrator, or no indicators.
14
Page 24
Using the Backlight
Using the Backlight
In dim light, you can enable the unit’s backlight t o read the display.
Press the LIGHT key.
The display’s backlight will turn on.
To conserve b attery power, t he
backlight will automatically turn off in about five seconds.
15
Page 25
Chapter 4 Operational Verificati on Tests
4. Operationa l Verificat ion Tests
These Operational Verification Tests should be performed periodically, as required by your departmental procedures.
Installing the Batteries
The first time you take your Ul traRadiac-Plus out of its shipping carton, you’ll have to install its batteries. See The Batteries on page 49 for instructions.
CAUTION:
Use of non-rechargeable batteries or rechar geab le batteries other than those supplied or recommended by Mirion Technologies (Canberra) may cause an explosion or other fire hazard resulting in personal injury. Mirion Technologies (Canberra) is not liable for personal injury or damage resulting from the use of incompatible batteries.
16
Page 26
Turning the Unit On
Turning the Unit On
Pres s a nd hold the ON/OFF key until the display appears, then release the key.
The unit will display the numbers 0, 1,
2, 3, 4, in sequence.
The Rate Mode display should appear, with the word RATE shown at the top left of the screen ( Figure 2).
Figure 2 The Rate Display
If 8 or a blinking BAT indicato r appear in the display or if the display is blank, refer to The Battery Life Indicators on page 7.
17
Page 27
Chapter 4 Operational Verificati on Tests
AUD and VIS indicate that the Audio and Visual alarms are both enabled.
The unit will start counting and displaying the instantaneous Rate. Naturally occurring background radiation will cause the unit to display a low reading
1
.
To switch to the Dose Mode, press the DOSE key.
Press the RATE key to return to the Rate Mod e.
The rate units indicator
hr
Rµ
will
blink slowly to show that the Rate Mode is active.
Note: Because the UltraRadiac automatically
adjusts its display to accommodate the amount of data, the units (R, for instance) may be displayed with a prefix. For instance, at natural
background levels, you’ll s ee µR, microroentgens.
1
Naturally occurring background radiation typically
measures less than 20 mR/h or 0.20 mSv/h.
18
Page 28
Turning the Unit Off
At higher exposure rates, the display
will change to mR (milliroentgens). With large exposure rates, it will show just R. It’s important to understand these units because each one is 1000 times greater th an the pr evi ous one.
Turning the Unit Off
To turn the unit off:
Pres s a nd hold the ON/OFF key.
OFF will be displayed.
Then you’ll see – – –.
Release the ON/OFF key; the unit will
turn its power off.
All data collected up to this time will be stored in memory.
19
Page 29
Chapter 4 Operational Verificati on Tests
Instrument Test
While in the Rate Mode, press and hold the CLR/TEST key until you see the test display in Figure 3, then rel eas e the key
2
.
Figure 3 The Test Display
Display Test Sequence
A set of numbers in the following order will appear. C heck that all of the numbers are exactly as shown below.
2
The 'GN' located in the upper right corner of the
display is used only by maintenance technicians. It's shown here so you can verify that the entire display is operating properly.
20
Page 30
Instrument Test
00.0 µR
1 1 1 µR
2.22 mR
33.3 mR
444 mR
5.55 R
66.6 R
777 R
888 R
999 R
00.0 µR
9 (blinking)
The blinking 9 indicates that the unit passed all tests. A blinking 0 indicates a failure; contact your local Mirion Technologies (Canberra) Service Center.
Press the CLR/TEST key to return to
Rate Mode (or wait about 10 seconds).
21
Page 31
Chapter 4 Operational Verificati on Tests
Checking the Alarm
At any time dur ing the Clr/Test sequence of numbers:
Press the RATE key – the audio alarm should sound and the Rate LED, the left and one below the display, should turn on.
Press the DOSE key – the audio alarm should sound and the Dose LED, the right hand one below the display, should turn on.
Press the ALARM key – the audio alarm should sound and both LEDs should turn on.
If the optional vibrator alarm is installed, the unit will vibrate when any of these three keys is pressed.
Checking the Function Keys
When the 000 display appears during the “Testing the Display” sequ ence, press each function key in turn, except the ON/OFF key.
22
Page 32
Ending the Display Test
The unit will beep and display the indications shown in the following table. Any other numerical indication indicates a fault in that key’s circuit.
Function Indication
DOSE
002
ALARM
010
RATE
001
LIGHT
020
CLR/TEST
004
Ending the Display Test
At the end of the test sequence, the blinking 9 will appear, indicating that the Rate Mode is operational. A blinking 0 indicates a failure; contact you r local Mirion Technologies (Canberra) Service C enter.
The unit will return to the Rate Mode in about ten seconds.
23
Page 33
Chapter 4 Operational Verificati on Tests
Backlight Test
Press the LIGHT key; the display’s backlight will turn on for about five seconds. This function is meant to be used in total darkness; the backlight may not be visible in daylight or normal room lighting.
Sleep Mode Test
Pres s a nd hold the RATE key, then press the ON/OFF key.
The characters SLP will be displayed
briefly.
When you see – – – , release both keys.
You’ll s ee th e SLEEP indicato r at t he
top of the screen (Figure 4)
24
Page 34
Completing the Test
Figure 4 The Unit is Asleep
To wake the unit, press and hold the
ON/OFF key until the SLEEP indicator disappears.
The unit will cycle through the 0–4 startup s equence, th en return to the Rate Mode.
Completing the Test
When the Ul traRadiac h as passed all tests , the unit is ready for operation.
If the unit fails any of these Operational Verification Tests, contact your local Mirion Technologies (Canberra) Service Center.
25
Page 35
Chapter 4 Operational Verificati on Tests
GM Tube Failure Indication
If the unit’s Geiger-Mueller tube should stop counting, the display will flash, showing three rows of three d as hed lin es each (Figure 5).
Figure 5 The GM Tube Has Failed
Contact your local Mirion Technologies (Canberra) Service Center for repairs .
Operation at Low Temperatures
At temperatures below –22 °F (–30 °C), it takes somewhat longer for characters to form on the display. The UltraRadiac-Plus automatically corrects for this b y sensi ng the ambient temperature and increasing the display refresh time from two to five seconds. Operatio n of the UltraRadiac-Plus is otherwise unchanged.
26
Page 36
Error Displays
Error Displays
Above 200 R/hr (2.0 Sv/h), and up to 600 R/h (6.0 Sv/h), the upper three -segment bar (– – –) on the display will begin flashing to indicate that:
The dose rate exceeds the ran ge of the UltraRadiac-Plus.
The error migh t be greater than 20%.
The value of the displayed d ose rate
(and consequently of the calculated dose) is not accurate.
The “true rate” might be significantly higher than the displayed value.
Above 600 R/h (6.0 Sv/h) the unit will not increase dose-rate measurement and will keep displaying 600 R/h (6.0 Sv/h).
27
Page 37
Chapter 5 Setting Up the Unit
5. Setting Up t he Unit
This chapter has been written for the System Administrator, the person responsible for your organizati on’s Ult r aRadiac-Plus units . It describes each of the unit’s functions and gives detailed instructions on how to set and enable the alarms. This is ap plicabl e only if the functions were not disabled by the optional Mo del S900 UltraRadiac Setup Software.
CAUTION:
To make your changes permanent, you must go through the entire setup sequence outlined in this
chapter. Exiting the sequence before it
is complete, such as by
changing only one parameter then waiting 10 seconds for the display to return to the
Rate Mode, will cause all
changes to be lost when the unit is turned off.
28
Page 38
The Rate Mode
The Rate Mode
To turn the UltraRadiac-Plus on, press and hold ON/OFF key until you see t he Rate display in Figure 6.
Figure 6 The Rate Display
The Rate Alarm
There are two Rat e Alarm Thresholds: Low Rate, for a min or hazard, and High Rate, for a severe hazard.
The alarm indicators will remain enab led as long as you remain in the radiation area that triggered the alarm.
29
Page 39
Chapter 5 Setting Up the Unit
High Threshold
There are several i ndicators for the High Rate Alarm Threshold (Figure 6).
Visual Enabled – The entire display and the left LED will flash quickly.
Audio Enabled – The audio will beep quickly.
Optio nal Vi brator Enabled – The unit will vibrate.
Low Threshold
There are several i ndicators for the Low Rate Alarm Threshold (Figure 6).
Visual Enabled – The RATE and ALARM indicato r s and the left LED
will flash slowly.
Audio En abled – The audio will beep slowly.
Optio nal Vi brator Enabled – The unit will vibrate.
The optional Vibrator Alarm can be used in addition to the audio and visual alarms when background noise makes the audio alarm hard to hear.
30
Page 40
The Rate Mode
Reading the Rate Alarm Threshold
To display the High level (H) or Low level (L) rate alarm thr es holds:
Press the ALARM and RATE keys.
You’ll s ee an H, for High Level Al arm,
displayed.
To read the Lo w Level Alarm setting, press the RATE key again; the display will show an L.
Press the CLR/TEST key; the flashing display will show the selected (H or L) alarm threshold setting.
If the optional Vibrator Alarm has been activated, the display will alternate between the threshold value and a vertically scrolling horizontal bar (– – –).
Changing the Rate Alarm Indicators
To change th e H igh level (H) o r Low level (L) alarm indicator settings:
Press the ALARM and RATE keys.
You’ll s ee an H, for High Level Al arm,
displayed.
31
Page 41
Chapter 5 Setting Up the Unit
To change the Low Level indicators, press the RATE key again; the display will show an L.
Press the CLR/TEST key to access the
selected (H or L) indicators.
Repeatedly pres s the ALARM key until
you see the desired combination of indicators, AUDio, VISual, vibrator (– – –), or no indicators.
The unit will automatically return to the
Rate Mode display about 10 seconds after the last keypress.
Changing the Rate Alarm Thresholds
To change th e H igh level (H) o r Low level (L) rate alarm setti ngs:
Press the ALARM and RATE keys.
You’ll s ee an H, for High Level Al arm,
displayed.
To access the Low Level Alarm settings, press th e RATE key again; the
display will show an L.
Press the CLR/TEST key; the flashing
display will show the selected alarm level (H or L) settings.
32
Page 42
The Rate Mode
If the optional Vibrator Alarm has been activated, t he display will alternate between the threshold value and a vertically scrolling horizontal bar (– – –).
Changing the Alarm Indicators
To choose one or more of the alarm indicat ors for the selected alarm level:
Repeatedly pres s the ALARM key until you see the desired combination of indicators, AUDio, VISual, vibrator, or no indicators.
Setting the Alarm Threshold
To set the threshold for the selected alarm level (Figure 7):
Figure 7 Changing the Rate Alarm Threshold
33
Page 43
Chapter 5 Setting Up the Unit
Press the CLR/TEST key; the decimal point and the threshold unit indicator will begin to flash.
Repeatedly pres s the RATE key until the desired decimal point location and threshold unit indication are displayed.
Press the CLR/TEST key. This action
will lock in the selected decimal point location and threshold unit indication.
The left-most digit will begin to flash.
Repeatedly pres s the RATE key until
the desired value of the digit is displayed.
Press the CLR/TEST key to lock in the
digit’s value. The next digit will begin to flash.
Repeat the above two steps to select and
lock in each of the two remaining digits.
When the las t digit has b een set , the
entire threshold value and unit will flash.
Press the CLR/TEST key to return to
the Rate Mode or wait about 10 seconds.
34
Page 44
The Dose Mode
The Dose Mode
When the unit is on, dose data is continuously accumulated and updated every two secon ds.
To enter the Dose Mode, pr es s the DOSE key.
The Dose Mode display (Figure 8) will show you the dose that has accumulated since the unit's memory was last reset to zero.
Figure 8 The Dose Display
The AUD and VIS display indicators will be seen only if the corresponding Dose Alarm mode and its indicator have been enabl ed. See Changi ng the Dose Alarm Indicator on page
38.
35
Page 45
Chapter 5 Setting Up the Unit
The Dose Alarms
There are two Dose Alarm threshold s : Low Dose, for a minor hazard, and High Dose, for a severe hazard.
The optional Vibrator Alarm can be used in addition to the audio and visual alarms when background noise makes the audio alarm hard to hear.
High Threshold
There are several i ndicators for the High Dose Alarm Threshold (Figure 8).
Visual Enabled – The DOSE and ALARM indicators and the right LED will flash quickly.
Audio Enabled – The aud io will beep quickly.
Optio nal Vi brator Enabled – The unit will vibrate.
36
Page 46
The Dose Mode
Low Threshold
There are several i ndicators for the Low Dos e Alarm Threshold (Figure 8).
Visual Enabled – The ALARM indicators and the right LED will flash slowly.
Audio Enabled – The audio will beep slowly.
Optio nal Vi brator Enabled – The unit will vibrate.
Reading the Dose Alarm Threshold
To display the High level (H) or Low level (L) dose alarm thresholds:
Press the ALARM and DOSE keys.
You’ll s ee an H, for High Level Al arm,
displayed.
To read the Lo w Level Alarm setting, press the DOSE key again; t he display
will show an L.
37
Page 47
Chapter 5 Setting Up the Unit
Press the CLR/TEST key; the flashing display will show the selected (H or L) alarm threshold setting.
If the optional Vibrator Alarm has been activated, t he display will alternate between the threshold value and a vertically scrolling horizontal bar (– – –).
Changing the Dos e Alarm Indicators
To change th e H igh level (H) o r Low level (L) alarm indicator settings:
Press the ALARM and DOSE keys.
You’ll s ee an H, for High Level Al ar m,
displayed.
To change the Low Level indicators, press the DOSE key again; t he display
will show an L.
Press the CLR/TEST key to access the
selected (H or L) indicators.
Repeatedly pres s the ALARM key until
you see the desired combination of indicators, AUDio, VISual, vibrator (– – –), or no indicators.
38
Page 48
The Dose Mode
The unit will automatically return to the Dose Mode display about 10 seconds after the last keypress.
Changing the Dos e Alarm Thresholds
To change th e H igh level (H) o r Low level (L) rate alarm settings:
Press the ALARM and DOSE keys.
You’ll s ee an H, for High Level Al arm,
displayed.
To access the Low Level Alar m settings, press th e DOSE key again; the
display will show an L.
Press the CLR/TEST key; the flashing
display will show the selected alarm level (H or L) settings.
If the optional Vibrator Alarm has been
activated, the display will alternate between the threshold value and a vertically scrolling horizontal bar (– – –).
39
Page 49
Chapter 5 Setting Up the Unit
Setting the Alarm Threshold
To set the threshold for the selected alarm level:
Press the CLR/TEST key; the decimal point and the threshold unit indicator will begin to flash.
Repeatedly pres s the DOSE key until the desired decimal point location and threshold unit indication are displayed.
Press the CLR/TEST ke y. T h is action will lock in the selected decimal point location and threshold unit indication.
The left-most digit will begin to flash.
Repeatedly pres s the DOSE key until
the desired value of the digit is displayed.
Press the CLR/TEST key to lock in the digit’s value. The next digit will begin to flash.
Repeat the above two steps to select and lock in each of the two remaining digits.
When the las t digit has b een set , the entire threshold value and unit will flash.
40
Page 50
The Source Finder Mode
Press the CLR/TEST key to return to the Rate Mode or wait about 10 seconds.
The Source Finder Mode
The Source Finder Mode allows the unit to locate an i s olated sou r ce. In this mode, th e unit will start clicking if the unit is near radioactive material. As you move clos er to the radioactivity (or further away), the unit will click faster (or slower).
For best results, aim the top of the unit at the suspected source of the radiation.
The Rate and Dose modes are active when t he Source Finder Mode is enabled. This means that you can r ead the higher Rate or Dose without exiting the Source Finder Mode.
Enabling the Source Finder Mode
To enable the Source Finder Mode:
Pres s a nd hold the RATE key until the display changes to a flashing 1.
Release the RATE key.
41
Page 51
Chapter 5 Setting Up the Unit
Exiting the Source Finder Mode
To exit the Source Finder Mode:
Pres s a nd hold the RATE key until the display changes to a flashing 0.
Release the RATE key.
The Sleep Mode
When the Sleep Mode is enabled, most of the unit’s circuits will be turned off to conserve battery power. Every 5 minutes, the unit will wake up for 4–6 seconds and check the current R ate.
The battery life time for th e U ltraRadiac-Plus in this mode is approximately 2000 hours.
WARNING:
Do not place the unit in
Sleep Mode before
responding! Doing so can expose you to
radiation for as long as five
minutes before you are
alerted.
42
Page 52
The Sleep Mode
Enabling the Sl eep Mode
To enable the Sleep Mode:
Pres s a nd hold the RATE key, then press the ON/OFF key.
The characters SLP will be displayed briefly.
When you see – – – , release both keys.
You’ll s ee th e SLEEP indicator at the
top of the screen (Figure 9)
Figure 9 The Unit is Asleep
When the Unit Wakes Up
It compares th e current Rate with the Rate Alarm th r es holds.
If a Rate Alarm is triggered, the unit will s tay awake.
43
Page 53
Chapter 5 Setting Up the Unit
If no Rate Alarm thresholds have been exceeded, the unit will go back to sleep for five minutes.
This wake-sleep c ycle will repeat continuously until the Sleep Mode is cancelled.
Note: The unit does not accumulate Dose
data while asleep.
Exiting the Sleep Mode
To exit the Sleep Mode:
Pres s a nd hold the ON/OFF key until the unit starts cycling through the 0–4 startup s equence.
The unit will then exit the Sleep mode and ent er R ate Mode.
The Data Logging Mode
The Data Log lets you record up to 999 individual dose/rate data pairs , each wit h an index number, for later retrieval. This is useful for performing a survey or for cataloging readings.
44
Page 54
The Data Logging Mode
Capturing Data Pairs
To capture data pairs, the unit must be ON and in the Rat e Mode.
Press the RATE and DOSE keys together.
Hold both buttons down until the displayed number stops blinking (in about three seconds),
The next sequential rat e/dose pair in dex number will appear.
Release both buttons. The unit will save the data pair and its index, then return to the Rate M ode. This process takes about 10 seconds.
To record another reading, point the unit at the location to be recorded and repeat the process.
Viewing the Data L o g
To view the stored data, the unit must be OFF.
Press the RATE and OFF/ON keys together.
45
Page 55
Chapter 5 Setting Up the Unit
When the unit turns on, the last sequence number recorded will be displayed.
Press the RATE button to scroll down, or the DOSE button to scroll up, through the memory.
The first number shown in the display will be the index number, scrolling up (DOSE button) will next display the Rate data and scrolling up once again will display the Dose data.
Scrolling down (RATE bu tton) will reverse the order of sequ ence numbers but will still display first Rate data then Dose data.
Scrolling will wrap around the memory, from first index entry to last, or last to first.
To return to normal operation, the unit must be turned off and then on again (normal turn on sequence).
46
Page 56
The Data Logging Mode
When the 999th sequence number has been recorded , the unit will cycle back to the first sequence number and any new data records will overwrite the previously recorded data in that location.
Clearing the Data Log
To clear all stored data, you must be in the Viewing th e D ata Log mode. (See the previous section.)
While pressing both the RATE and ALARM keys, press and hold the CLR/TEST key for two seconds.
The first index number (0) will appear
This ind icates that al l d ata has been
cleared.
To return to the Rate/Dose mode of operation, turn the unit OFF, then back ON.
47
Page 57
Chapter 6 Maintenance
6. Maintenance
Prevent ive maintenance consists of routine checks of the eq uipment b efore and after each mission, or as required by procedure.
Routine Checks Include:
Cleani ng the uni t.
Inspecti ng the battery compartment ’s
door gasket.
Inspecti ng the battery contacts for corrosion.
Cleaning the Unit
Remove dust, moisture and loose dirt from the outside surfaces of the Ultr aRadiac-Plus with a clean soft cloth. If necessary, the unit may be cleaned with a mild solution of ordinary detergent and water, rinsed, and thoroughly dried.
48
Page 58
The Batteries
The Batteries
If you’re installing batteries for the first time, the following caution can be disregarded. If you’re replaci ng the bat teries, be sure to turn the unit off first.
Installing o r Replacing the Batteries
To open the battery compartment door, turn its fastener a qu arter-turn counterclockwise, swing the door open and slide out all four batteries.
Install fresh batteries in the order shown in Figure 10.
CAUTION:
Be sure to observe the
polarity next to each of the four battery slots. The end of the battery facing out should
match the slot’s marked
polarity, as shown in Figure 10 and Figure 11.
With the batteries correctly installed, close and latch the door. See Latching the Compartment Door on pa g e 54.
49
Page 59
Chapter 6 Maintenance
Figure 10 Location of the Batteries
Figure 11 The Battery Polarity Indicators
50
Page 60
The Batteries
Charging the Batteries
The UltraRadiac-Plus is equipped with the capability to use rechargeable AAA batteries as well as standard alkaline batteries. Nickel Metal Hydride (NiMH) batteries can be inserted in to the unit and used directly. To recharge the NiMH batteries insert them into the UltraRadiac-Plus and set the battery charger switch to ON position. In this setting, the charger will continuously charge the batteries while they are in the unit. In the POWER position, the UltraRadiac-Plus can be powered directly off of the charger without the batteries present. It can run in this mode indefinitely, although it can not recharge batteries in this mode. See Figure 12 and Figure 13.
Figure 12 UltraRadiac-Plus Battery Charger
51
Page 61
Chapter 6 Maintenance
Figure 13 Right Side View of UltraRadiac-
Plus
Note: Do not attempt to charge standard
alkaline batteries. Damage can occur.
52
Page 62
The Battery Compartment
The Battery Compartment
To open the battery compartment door, turn its latch a quarter-turn counterclockwise and swing the door ope n.
Inspect the four battery contacts on the inside of the door (Figure 14). If any corrosion is seen, cl ean it off with a pencil eras er .
Inspect the s eal on the inside of the
door (Figure 14); it should be whole and ent ire. If the seal is cracked or broken, the unit may no longer be watertight.
Inspect t he O-ring seal around the door latch’s post (Figure 14). If the seal is cracked or broken, the unit may no longer be watertight.
For repair, co nt act your local Mirion
Technologies (Canberra) Service Center.
When the inspection is complete, clo s e
and latch the door. See the following section Latching the Com par tm e nt Door.
53
Page 63
Chapter 6 Maintenance
Figure 14 The Inside of the Battery Door
Latching the Compartment Door
Swing the b attery cover closed, press it firmly against th e unit and turn the fastener a quarter­turn clockwise.
Note: It may be necess ary to turn the fastener
slightly to line it up with the latch post before latch will engage.
When the door is properly closed, it will be possible to lay the fastener’s D-ri ng down in the cover’s fastener recess.
Note: When the door is latched, the unit may
turn itself on in the Sleep Mode. To return to normal operation:
54
Page 64
Attaching the Belt Clip
Press and hold the ON/ OFF key for about 2 seconds.
The unit will cycle through the
display startup sequence, then enter the Rate Mode.
Attaching the Belt Clip
To attach the supplied belt clip, screw the clip’s two captive retaining screws into the threaded inserts at the top of the unit’s rear panel.
Note: To prevent cross-threading damage,
insert each s cr ew loosely, then tighten them alternately a few turns at a time. Do not over-tighten.
The Carrying Case
If the belt clip is installed, remove it from the back of the Ult raRadiac-Plus by unscrewing its two captive retaining screws and lifting the clip off the unit.
55
Page 65
Chapter 6 Maintenance
Now put the UltraRadiac-Plus into the carrying case with the display toward the bottom (closed) end of the case. To close the case, refer to Figure 15 and...
Fold the top flap down over the front of the case.
Pull the narrow strap across the top flap.
Secure the narrow strap to both the top flap and the side of the case.
Thread your belt through the slots in the belt loop.
56
Page 66
The Carrying Case
Figure 15 Using the Carrying Case
57
Page 67
Appendix A Specifications
A. Specifications
Features
DOSE RATE – URAD-PLUS-S: 0.01 µSv/h to 2 Sv/h; URAD-PLUS-R: 1.0 µR/h to 200 R/h. Meets ANSI-N42.33-2006 standard above 1 µSv/h (100 µR/h). Dose rat e alarm will function to 5.0Sv/h (50 0 R/h).
DOSE – URAD-PLUS-S: 0.001 µSv to 999 Sv; URAD-PLUS-R: 0.1 µR to 999 R.
PRESETTABLE AUDIBLE AN D VISUAL ALARMS – User-defined and -set dose and dose rate al ar ms.
INITIALIZATION TIME – Operational in less than five seconds.
SETUP TIME – Less than one minute for all checks and alarms.
ACCURACY – Within ±30%. fro m 1 µSv/h (100 µR/h) to 2.0 Sv/h (200 R/h).
58
Page 68
Detector
DATA RECORDING – Local data logging to 999 data points; data downloadable via optical (IR) communications port.
SELF MONITORING – Continual self­monitoring for the instrument’s state of health.
CASE – Die-cast al uminum; available in black or yellow.
EMI SUSCEPTIBILITY – Will not be affected, or cause other equipment to be affected, by its use.
OPERABLE AND READABLE – By persons wearing protective clothing.
Detector
DETECTOR – Energy compensated GM tube.
GAMMA ENERGY DEPENDENCE – ±40%, 60 keV to 1.5 MeV.
TOTAL (ACCUMULATED) DO SE READOUT – Will not be erased when r ead; resettable t o zero as des ired.
59
Page 69
Appendix A Specifications
RESPONSE TIME – Achieves 90% of final reading in one second; subject to a step change from 1mR/h (10 µSv/h) to 55mR/h (550 µSv/h); updates display every second.
Display
LCD – Readable at 3 ft (1 m), updated every two seconds; can be backlit for night use.
UNITS – URAD-PLUS-S provides data in units of µSv, mSv, Sv, µSv/h, mSv/h and Sv/h; URAD-PLUS-R provides data in units of µR, mR, R, µR/h, mR/h and R/h.
Alarms
SELECTABLE ALARMS – Has selectable Visual and Audible indicators for day or night use. Alarm levels are settabl e over ent ire dynamic range.
ALARM TYPES – Alarm on dose rate and total cumulative dose.
ALARM LEVELS – Two alarm levels available for each type to i ndicate minor or severe hazard.
60
Page 70
Power
AUDIBLE ALARM – >85 dBA at 30 cm.
VIBRATION ALARM – Included.
Power
BATTERIES – Four AAA 1.5 V alkaline batteries.
MINIMUM BATTERY LIFE – 10 hours of battery life remaining allows display of remaining time.
LOW BATTERY INDICATION – Triggered when the remaining bat tery life is approximately 10 hours; allows display of remaining time in minutes.
Accept standard off the shel f rechargeable NiMH batteries with DC charger to URAD­PLUS power connector.
Environmental
OPERATING TEMPERATURE: –22°F to 141°F (–30°C to +61°C). Vibrating and audible alarms functional to -50°C.
61
Page 71
Appendix A Specifications
At operating temperatures below -22°F (­30°C), it takes somewhat longer for characters to form on the display. The UltraRadiac-Plus automatically corrects for this b y sensi ng the ambient t emperature and increasing display refresh time from two to five second s. Operatio n of the UltraRadiac-Plus is otherwise unchanged. Measurement is still accurate and audible and vibrating alar ms are still functional even with the slower display refresh rate at low temperatures.
STORAGE/TRANSPOR T TEMPE RATURE: –40°F to 158°F (–40°C to +70°C).
HUMIDITY – Functions correctly over RH 40% to 93% at 95 °F (35°C).
WATER IMMERSION (including salt water) – 3 ft (1 m) for at least 2 hr.
SAND/DUST – Operates in winds with exposure to fine dust and sand particles.
FUNGUS – Built from fungus resistant materials.
VIBRATION AND SHOCK – Withstands vibration associated with transport and shocks of dropping while in use.
62
Page 72
Physical
ALTITUDE – 40 000 ft (12 000 m).
Physical
DIMENSIONS – 3.94 x 2.62 x 1.24 in. (100 x 66 x 31 mm) (HWD).
WEIGHT – 8.8 oz (249 g) including batteries. Pouch weight is 2.4 oz (68 g).
VOLUME – 12.8 in.
3
(210 cc).
Qualification Testing
UltraRadiac-Plus has been type-tested to meet ANSI-N42.33-2006 with minor exceptions: Sec. 6.4, Sec. 5.12 and S ec. 5.7 (option al).
Ordering Information
URAD-PLUS-S/Y – UltraRadiac-Plus, Yellow, Sv unit, alkaline batteries.
URAD-PLUS-S/B – UltraRadiac-P lu s , Black, Sv unit, alkaline batteries.
URAD-PLUS-R/Y – UltraRadiac-Plus, Yellow, R unit, alkaline batteries.
63
Page 73
Appendix A Specifications
URAD-PLUS-R/B – UltraRadiac-Plu s , Black, R unit, alkaline batteries.
URAD-PLUS-DCCHARG – DC ch arger with cigarette ligh ter conn ector (needs NiMH rechargeable batteries to operate).
URAD-PLUS-MOUNT – Vehicle mount for URAD-PLUS (allows the use of DC charger).
64
Page 74
B. Default Alarm Settings
The Dose and Rate alarm settings (thresholds) for the Ultr aRadiac-Plus are factory set to the values in the following tables.
R Low Setting High Setting
Rate 500 µR/hr 2 R/hr
Dose 100 mR 10 R
Sv Low Setting High Setting
Rate 1 mSv/hr 100 mSv/hr
Dose 1 mSv 100 mSv
The factory settings can b e changed manual ly (refer to the chapter Setting Up the Unit on page 28), or via the optional Model S900 UltraRadiac Setup Software.
CAUTION:
The System Administrator is
responsible for determining and setting the proper
operational alarm thresholds for each UltraRadiac-Plus.
65
Page 75
C. Displ a y Units
Your unit’s display will show R or Sv*3. Any unit can have larger and smaller subunits, which are indicated by prefixes. As the Rate and Dose change, the Ult raRadiac-P lus will automatically adjust its displayed un its.
For example: The Rate may start with µR/h (natural background radiation is less than 20 µR/h). If the Rate increases beyond 999 µR/h, the display will change to mR/h.
micro (µ) µR – one millionth of a
roentgen
milli (m)
mR – one thousandth of a roentgen (or µR times
1000)
centi (c)**
cR – one hundr e d t h of a roentgen (or mR times
10)
the base unit
R – one roentgen (cR times 100)
3
The display units are set at the factory; they are not
user chan geable.
66
Page 76
** The ‘centi’ prefix will be seen only on
UltraRadiac's displaying sievert units.
67
Page 77
D. Glossary
These brief exp lanations are included to define terms used in this manual.
Dose
The total amount of radiation received. Also called Accumul ated Dose.
Dose Rate
The average rat e ( in time) of radi ation; e.g., Roentgen per hour. Also called Rate.
Rate
The amount of radiation measured by the UltraRadiac-Plus every 2 seconds, then extrapolated to and displayed as units per hour.
Roentgen (R)
Unit of measure of radiation exposure; monitors the amount of radiation received.
Sievert (Sv)
Unit of measure of absorbed radiation; monitors the biological danger of radiation.
68
Page 78
Stay Time
The Stay Time is how much ti me r emains, at the current Dose Rate, before the High Dose Alarm is triggered. Sta y Time is calculat ed by dividing (the High Dose Alarm Setpoint minus the al r eady accumulated Dose) b y the current Dose Rate, then converting the result to minutes. If the display reads 999, the maximum, your Stay Time is at least 16.5 hours.
WARNING:
If the Dose Rate goes up, your remaining Stay Time will go down.
You should check your
Stay Time frequently when you are in the presence of radiation.
69
Page 79
E. Disposing of This
Equipment
Electrical and electronic equipment contain hazardous substances that, if disposed of improperl y, can have a negati ve effect on the environment and on human healt h.
Users / owners of this equipment have the responsibility to ensure that this equipment does not pose a threat to the environment or to humans when it becomes obsolete and requires disposal.
The symbol below, also found on your equipment, indicates that this equipment should not be di sposed of in unsorted municipal waste.
Therefore, following the provisions of COUNCIL DIRECTIVE 2002/96/EC on waste electrical and electronic equi pment (WEEE), we ask that you contact your nearest Mirion Technologies office for instructions on the proper disposal of this equipment.
70
Page 80
Warranty
Mirion Technologies (Canberra), Inc., (we, us, our) warrants to the customer (you, your) that for a period of ninety (90) days from the date of shipment, software provided by us in connection with equipment manufactured by us shall operate in accordance with applicable specifications when used with equipment manufactured by us and that the media on which the software is provided shall be free from defects. We also warrant that (A) equipment manufactured by us shall be free from defects in materials and workmanship for a period of one (1) year from the date of shipment of such equipment, and (B) services performed by us in connection with such equipment, such as site supervision and installation services relating to the equipment, shall be free from defects for a period of one (1) year from the date of performance of such services.
If defects in materials or workmanship are discovered within the applicable warranty period as set forth above, we shall, at our option and cost (A) in the case of defective software or equipment, either repair on a return to factory basis or replace the software or equipment, or (B) in the case of defective services, re-perform such services.
LIMITATIONS
EXCEPT AS SET FORTH HEREIN, NO OTHER WARRANTIES OR REMEDIES, WHETHER STATUTORY, WRITTEN, ORAL, EXPRESSED, IMPLIED (INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE) OR OTHERWISE, SHALL APPLY. IN NO EVENT SHALL WE HAVE ANY LIABILITY FOR ANY SPECIAL, EXEMPLARY, PUNITIVE, INDIRECT OR CONSEQUENTIAL LOSSES OR DAMAGES OF ANY NATURE WHATSOEVER, WHETHER AS A RESULT OF BREACH OF CONTRACT, TORT LIABILITY (INCLUDING NEGLIGENCE), STRICT LIABILITY OR OTHERWISE. REPAIR OR REPLACEMENT OF THE SOFTWARE OR EQUIPMENT DURING THE APPLICABLE WARRANTY PERIOD AT OUR COST, OR, IN THE CASE OF DEFECTIVE SERVICES, REPERFORMANCE AT OUR COST, IS YOUR SOLE AND EXCLUSIVE REMEDY UNDER THIS WARRANTY.
Page 81
EXCLUSIONS
Our warranty does not cover damage to equipment which has been altered or modified without our written permission or damage which has been caused by abuse, misuse, accident, neglect or unusual physical or electrical stress, as determined by our Service Personnel.
We are under no obligation to provide warranty service if adjustment or repair is required because of damage caused by other than ordinary use or if the equipment is serviced or repaired, or if an attempt is made to service or repair the equipment, by other than our Service Personnel without our prior approval.
Our warranty does not cover detector damage due to neutrons or heavy charged particles. Failure of beryllium, carbon composite, or polymer windows or of windowless detectors caused by physical or chemical damage from the environment is not covered by warranty.
We are not responsible for damage sustained in transit. You should examine shipments upon receipt for evidence of damage caused in transit. If damage is found, notify us and the carrier immediately. Keep all packages, materials and documents, including the freight bill, invoice and packing list.
Revised 04 Nov 16
Radiation Safety. Amplified
Loading...