Mirion Technologies SPN User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
SPN DETECTORS
Neutron Flux Instrumentation
SPN DETECTORS
SPN-detectors are designed for the detection of
thermal neutrons within the core of the reactor. These
detectors generate a current proportional to the
neutron  ux density without applying an external
voltage source.
In general, the detector current of these detectors is
caused by the emission of beta radiation (electrons)
from the target material which gets activated by
neutrons. Therefore, besides the name SPN-detector
(Self-Powered Neutron detector), also the name n,ß-
detector (neutron-beta-detector) is in use.
APPLICATIONS
• Operational process monitoring
Detection and measurement of the thermal
neutrons inside the reactor core for neutron  ux
distribution monitoring and monitoring of neutron
 ux peaks
RELATED PRODUCTS
• KNK/KNU 50: neutron ionization chambers
• Proportional counter tubes BF3  lled
• TK 250: digital signal processing channels for
start-up and power range
FEATURES
Detection of thermal neutrons within the reactor core
WL-23293 with co-emitter: short response time
Small dimensions, proven construction
Large measuring range up to 1e15 nv
Output signal: DC-current
Very low gamma sensitivity
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
SPN DETECTORS
Neutron Flux Instrumentation
www.mirion.com
145028EN-B
348 244 EN-C
PREFERRED TYPES OF THE IST-PROGRAM
Notice: 1 nv = 1/cms
COMMON DATA
Operating temperature: 400°C max. (752°F)
Diameter of detector and cable: 1.6 mm
Connector: BNC-Plug
Material of detector and cable sheath: stainless steel
Insulation: Al2O3 or MgO
Manufacturer: Mirion Technologies (IST) Corporation,
Horseheads, NY/USA
Type Emitter
Material
Neutron
Sensitivity
(A/nv)
Gamma
Background
Measuring
Range
(nv)
Response
Time
(s)
Detector
Length L1
(mm)
Total
Length L2
(mm)
WL-23215
WL-23226
WL-23283
Rhodium
Vanadium
Cobalt
1.0e-20
1.1e-20
1.4e-21
2%
2%
5%
1e9 ... 1e15
1e9 ... 1e15
1e10...1e15
60
300
1e-10
76
1.321
864
8 543
9 144
4 877
Loading...