Mirion Technologies SG 65-66 User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
PRELIMINARY
SG 65/66
Detectors
NaI Scintillators
The SG 65 and SG 66 scintillators have been deve-
loped for safety related monitoring of activity according
to KTA 1502. In the scintillator crystal, ionizing radia-
tion generates light pulses, which are converted and
ampli ed into charge pulses by the photo multiplier.
The height of the pulse is proportional to the energy
of the gamma radiation. An adjustable pulse ampli er
offers calibration of the sensitivity and matching of the
output pulses to the impedance of the detector cable.
APPLICATIONS
• Monitoring for gamma activity from in air and
liquids
FEATURES
Reliable
Robust construction
Type tested
High sensitivity
Compact with 2” × 2” sodium iodide crystal
Built-in pulse ampli er and buffer
Output signal: low-impedance voltage pulses
Temperature range until 80°C (SG 66)
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
SG 65/66
Detectors
TECHNICAL DATA
(1) U
discr
= -100 mV, U = 1000 V, T = 25°C
(2) without connector
EXAMPLE OF APPLICATIONS
www.mirion.com
145025EN-2
Properties SG 65 M SG 66
Background (1)
Sensitivity (1)
Energy resolution (50%) Measuring range
without screen with 10 cm lead Test device TKA 04 with 185 kBq Cs-137 for Cs-137 with DPK 251
65 s-1 approx. 4 s
-1
approx.
2e4 s
-1
approx.
< 8%
0.5 … 5e5 S
-1
43 s-1 approx.
2.8e4 s-1 approx.
< 12%
0.5 … 5e5 s
-1
Output signal Pulse width High voltage Supply voltage
0 ... -5V at 50 Ohms
1.2 μs approx.
-1 mV/keV 1000 V, 0.2 mA approx.
14.25 ... 25 V, 30 mA approx.
0 ... -5V at 50 Ohms
0.5 μs approx.
-1 mV/keV 1000 V, 0.06 mA approx.
14.25 ... 25 V, 30 mA approx.
Scintillator crystal Detector housing Magnetic shielding Connectors Cable
Ø × length NaI (Tl); 2” × 2”
Aluminium, Brass μ-metal Fischer S105 A 06093 TKK 05
NaI (Tl); 2” × 2” Aluminium, Brass μ-metal Fischer S105 A 06093 TKK 05
Temperature range
Temperature change rate Ambient pressure Dimensions Weight
Ø x length (2) (2)
0 … +50°C (+32 ... 122°F)
< 5 K / h
0.7 … 1.2 bar 65 × 340 mm (2.5 x 13.4 in) 2 kg (4.4 lb) approx.
0 … +80°C (+32 ... +176°F), 100°C (212°F) for <10h < 2 K / min
0.7 … 1.2 bar 65 × 395 mm (2.5 x 15.5 in)
2.5 kg (5.5 lb) approx.
NaI scintillator detector SG 65 M with digital signal processing channel DPK 251
Detector assembly for liquids with SG 65 M and RSD 221
Remote check source TKA 51.1
Measuring vessel RSD 221
Lead shielding RSA 468
Check source in test position
Detector SG 65 M with 2 ” x2 ” NaI