RE-2000S Semi-automatic TLD-Reader
Passive Dosimetry
RE-2000S
Semi-automatic TLD Reader
RE-2000S is a universal, low cost and high
precision TLD-reader for processing of
dosimeter cards, and single chips.
WinTLD Light application software running on
a separate PC provides the easy-to-learn and
-operate reader control and user interface for the
reader.
FEATURES
evaluation of single TL-elements or TL-dosimeter •
cards
high measurement reproducibility with standard •
industrial N2 heating
fully programmable pre-heat, measurement and •
anneal cycles
hole code or bar code indenti cation of the do-•
simeter card
The reader can automatically process up to 20
dosimeter cards or 80 single TL-elements at one
load. The maximum processing speed is 180
single elements or 100 two element cards per
hour. The photon counting method used over the
whole measurement range provides an
excellent signal to noise ratio for the
measurement. Cooled PMT and built in self
diagnostics guarantee high measurement
stability and an error free operation.
With the optional cassette feeder, processing
capacity is 200 dosimeter cards or 800 single
elements.
health physics
A Mirion Technologies Division
Featuring:
TECHNICAL SPECIFICATIONS:
Physical Characteristics Dimensions: (HxWxD) 40 x 57 x 34 cm
Weight: 33 kg
Functional Characteristics Capacity: 20 personnel dosimeters or 80 single elements per loading•
Processing speed: •
- 100 ea. two element cards per hour
- 50 ea. four element cards per hour
- 180 ea. single elements per hour
Element types:•
- round pellets 4.5 mm ø
- square chips 3.2 x 3.2 x 0.9 mm
Dynamic range: 7 decades (9 decades with neutral lter)
Signal measurement: photon counting with max. count rate of 100 MHz•
Linearity: < 1% deviation•
System stability:•
- dose: < 1 µSv (standard deviation)
- high voltage: negligible while using photon counting
Reference light: high stability temperature controlled reference light source, •
short term stability < 0.5%
Heating method: contactless hot nitrogen, typically 5 l/min•
Time temperature pro le:•
- hot blast gas heating
- temperature range 60 - 400 °C; stability ±1 °C
pre heat, readout and post heat time: adjustable up to 140 seconds•
Dark current: negligible while using photon counting - variation in the bgr. •
count rate < 1 µSv
137
Cs equivalent dose
User interface: WinTLD Light Software running on a separate PC•
RE-2000S Semi-automatic TLD-Reader
Passive Dosimetry
Environmental Characterstics operating temperature: from +10 to +40 °C•
Electrical Characterstics voltage: 100 - 250 VDC 50/60 Hz•
Options Upgrade kit for upgrading to fully automatic reader RE-2000A•
storage temperature: from -10 to +50 °C•
consumption: 150 VA at 50 Hz•
data interface: RS-232 (9pin D-connector) or LAN (RJ-45 optional)•
WinTLD Pro Management software•
www.mirion.com
20996041_RE2000S_EN_A
MGP Instruments Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
MGP Instruments SA
BP 1
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
RADOS Technology Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 4684 600
F +358 2 4684 601
RADOS Technology GmbH
Ruhrstrasse 49
D-22761 Hamburg
Germany
T +49 40 85193 0
F +49 40 85193 256