RD-02 & RD-02L Intelligent Detectors
Detection & Identi cation
20996043_RD02_02L_EN_A
RD-02 Intelligent Detector RD-02L Intelligent Detector
Physical characterstics
* Background shielded
50 mm Pb + 3 mm Al
radiation detected: gamma and x-rays, according •
to ambient dose equivalent H*(10)
energy range: 50 keV - 1.3 MeV for dose rate •
range 0.01 µSv/h - 10 mSv/h, and 80 keV - 1.3 MeV
for dose rate range 10 mSv/h - 10 Sv/h
detector type: two halogen quenched, energy •
compensated GM tubes
dose measurement range: 0.01 µSv - 10 Sv•
dose rate measurement range: 0.01 µSv/h - 10 •
Sv/h
calibration accuracy: ± 5% of the reading in 3 •
mSv/h Cs-137 exposure at +20oC
energy response: ± 30% over energy range 50 keV •
- 1.3 MeV, from background to 10 mSv/h,
±30% over energy range 50 keV - 1.3 MeV,
above 10 mSv/h
±20% over energy range 80 keV - 1.3 MeV,
above 10 mSv/h
dose rate linearity: ± 15% + 1 digit over the range •
0.10 µSv/h - 10 Sv/h
directional response: ± 15% within 180•
o
horizontally from the calibration direction,
± 30% within ± 45o vertically from the calibration
direction
GM tube volumes: 7 cm (low dose rate) and 0.13 •
cm (high dose rate)
GM tube sensitivity: 200 cps (low dose rate) and 3 •
cps (high dose rate) @ 100 µSv/h
GM tube dead time: 110 µs for low dose rate and •
13 µs for high dose rate tube
self count rate: 10 cpm for low dose rate tube and •
1 cpm for high dose rate tube *
radiation detected: gamma and x-rays, according to •
ambient dose equivalent H*(10)
energy range: 50 keV - 1.3 MeV for dose rate range •
0.01 µSv/h - 2 mSv/h, and 80 keV - 1.3 MeV for dose
rate range 2 mSv/h - 10 Sv/h
detector type: two halogen quenched, energy •
compensated GM tubes
dose measurement range: 0.01 µSv - 10 Sv•
dose rate measurement range: 0.01 µSv/h - 10 Sv/h•
calibration accuracy: ± 5% of the reading in 3 mSv/h •
Cs-137 exposure at +20oC
energy response: ± 30% over energy range 50 keV - •
1.3 MeV, from background to 10 mSv/h,
± 30% over energy range 50 keV - 1.3 MeV,
above 10 mSv/h
±20% over energy range 80 keV - 1.3 MeV,
above 10 mSv/h
dose rate linearity: ± 15% ± 1 digit over the range •
0.03 µSv/h - 10 Sv/h
directional response: ± 15% within ± 180•
o
horizontally from the calibration direction
± 30% within ± 45•
o
vertically from the calibration
direction
GM tube volumes: 60 cm (low dose rate) and 0.13 •
cm (high dose rate)
GM tube sensitivity: 1400 cps (low dose rate) and 3 •
cps (high dose rate) @ 100 µSv/h
GM tube dead time: 160 µs for low dose rate and 13 •
µs for high dose rate tube
self count rate: 40 cpm for low dose rate tube and 1 •
cpm for high dose rate tube *
Electrical characterstics
power requirement: 12 VDC, 10mA (at background level)•
connections: 8-pin lockable MIL-connector •
integration time: adjustable from 1 minute to 99 minutes •
memory: capacity for 864 dose rate records•
interface: RS-232 (1200- 9600 baud)•
compatible with: RADOS AAM-95 Environmental Monitoring SW•
Mechanical characterstics
casing: impact resistant casing of anodized aluminum, shielded against RF-interference and NEMP•
RF resistance: Up to 1 GHz @ 100 V/m•
EMP resistance: Emax = 50 kV, rise time 10 nsec, fall time 1 µsec•
enclosure class: IP 67 (short term)•
ength: 260 mm, cylinder diameter 60 mm•
weight: 510 g•
length: 560 mm, cylinder diameter 60 mm•
weight: 990 g•
Environmental characterstics
temperature range: operation -40•
o
C to +70oC,
storage -40oC to +70oC
temperature range: operation: -40•
o
C to +70oC,
storage: -40oC to +70oC