Mirion Technologies Radview CT User Manual

DATA SHEET
LIT4391_1_US
RADVIEW™ CT VISUAL RADIATION DOSIMETER
Dose range color indicator
The RadView ™ CT from ISP, can be used as a QA tool to ensure the appropriate exposure is administered to a patient during a CT Scan. The RADView CT measures the entrance dose to a patient. The special lm inside the RADView CT badge darkens with x-ray exposure and gives instant results that can be visually read following a diagnostic procedure. The dose displayed can then be compared to the expected dose for the CT scan received by a patient.
FEATURES
• Quick independent assessment
• FDA class 1 device
• Instant reading
• No reader needed
• Cards can be scanned and saved as PDF les for record retention
• Cost effective
• Independently tested:
• 89% accuracy in determining dose range
• Precision of +/-15%
CT Imaging Systems are sophisticated devices equipped with precise technology and controls designed to calibrate and ensure the level of radiation output. Even with all of the controls in place and the sophistication of the CT Imaging Systems, it is important to have an overall system of quality assurance. RADView CT provides a view of actual output. This ability to correlate output is an essential component of quality healthcare services that can be lled be RADView CT.
RADView CT can be scanned such that an PDF le of the results can be preserved as an independent record of a particular CT Scan, for a particular CT Machine and, for that particular patient.
dosimetry services
A Mirion Technologies Division
Featuring:
HOW TO READ THE RADVIEW CT DOSIMETER
In the RADView™ CT dosimeter, calibrated dose threshold indicators surround the radiation sensitive lm in the center of each reference square. The color of each reference square is calibrated to the dose printed on that segment. The reference indications are permanently printed on the badge and do not change, providing a constant reference value for each dose level. If the sensitive lm in the center is darker than the background of a square, the dose is greater than the number on that segment (“dose is darker”).
Before Exposure After Exposure
the active element is lighter than the 20 mGy reference.
the active element is darker than the 10 mGy reference.
PLACEMENT INSTRUCTIONS
Place RADView CT on the patient or the phantom inside the scan slice to en-sure entire Colorimetric scale is utilized. The dosimeter should be positioned so that its long dimension is perpendicular to the axis of the scanner and the active area of the dosimeter is close to the central part of the scanned eld.
In the image on the left of an X-ray using the RADView CT Dosimeter (the circled portion), you will notice the dosimeter does not obstruct the image.
2652 McGaw Avenue Irvine, CA 92614 USA U.S./Canada +1.800.251.3331 U.K. 0170.629.9329 Worldwide +1.949.419.1000