RAD-62 System Dosimeter
Active Dosimetry
RAD-62
System Dosimeter
The RAD-62 system dosimeter is an
integrated solid state dosimeter with full range of
functions for use in the RADOS
automatic dose management systems.
It forms a highly cost effective package with
its outstanding performace and is handy and
convenient to use even by people wearing
protective clothing.
FEATURES
enhanced EMI immunity •
ruggedized clip xing •
improved wear-out and decontamination properties•
increased buzzer volume•
compact and lightweight•
standard AAA alkaline battery for 1800 h of operation•
solid state detector•
compatible with all Rados automatic dose•
management systems
health physics
A Mirion Technologies Division
Featuring:

TECHNICAL SPECIFICATIONS:
Radiological characteristics Radiation detected:•
RAD-62 System Dosimeter
Active Dosimetry
gamma and X-rays
Detectors:•
energy compensated Si-Diode
Measurement range:•
- integrated dose: 1 µSv - 10 Sv or 0.1 mrem - 1000 rem
- dose rate: 0.005 - 3000 mSv/h or 0.001 - 300 rem/h
Calibration:•
better than ±5 % (Cs-137, 662 keV, at 1.4 mSv/h, 140 mrem/h), Hp(10)
Energy response:•
Hp(10), 55 keV - 3 MeV, better than ±25 %, up to 6 MeV, better than ± -25
%...+35
Dose rate linearity:•
better than ±15 %, up to 3 Sv/h (300rem/h)
Dose rate linearity:•
better than ±10 % for 0.005 mSv/h - 1 Sv/h (1mrem/h - 100 rem/h)
Audible alarms:•
- eight separate alarms, sound level typically better than 85 dBA at 30 cm
- integrated dose
- dose rate
- dose over ow
- dose rate over ow at 3 Sv/h or 300 rem/h
- low battery 1 and 2
- defect
Alarm thresholds:•
two freely selectable values for integrated dose and one for dose rate
Push button functions:•
- toggle between dose and dose rate display
- display alarm level
Mechanical characteristics Dimensions: 80 x 67 x 22 mm (3.1x2.6x0.8 in)•
Envirionmental Characterstics Temperature range:•
Electrical characteristics Power supply:•
MGP Instruments Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
MGP Instruments SA
BP 1
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Weight: 80 g (2.8 oz) including battery•
- from -20 oC to +50 oC (-4 oF to 122 oF ) operational
- from -20 oC to +70 oC (-4 oF to 158 oF) storage
- humidity up to 90 % rH, non-condensed
one AAA alkaline cell, life typically 1800 h in background eld (dose mode)•
Reader communication:•
by infrared through bottom•
RADOS Technology Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 4684 600
F +358 2 4684 601
RADOS Technology GmbH
Ruhrstrasse 49
D-22761 Hamburg
Germany
T +49 40 85193 0
F +49 40 85193 256
www.mirion.com
20996028_RAD62_EN_A