Mirion Technologies PIS 203S User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
PIS 203S Samplers
PIS 203S
The PIS 203S sampler forms part of the RAMSYS
product line. It has been developed to sample air
ef uents under accident conditions conforming to the
requirements set forth by RG 1.97. One sampler for
batch sampling or dual samplers for continuous sam-
pling are available with or without heat tracing. The
sample  ow rate through the sampler is measured
and totalized. The collected activities of particulates
and iodine are periodically analyzed in a laboratory as
needed.
The PIS 203S is typically used in conjunction with
noble gas monitor and an aeraulic skid that automa-
tically control the operation from standby to active
sampling as applicable.
APPLICATIONS
• Ef uent release monitoring
• Post-accident operations
RELATED MONITORS
• PIS 204L: « light » version
• PIS 205L: « light » version with  ow rate control
• NGM 203S: low range noble gas monitor
• NGM 204S: high range noble gas monitor
• PING 206S: particulate, iodine and noble gas
monitor
FEATURES
Available for batch or continuous sampling acquisition
version (single or dual PIS 203S units)
Automatic start up in accident con guration
Heat tracing available when required
Provided with manipulation tools and lead shielded
transport cart
Heavily shielded for ALARA radiation considerations
Available under 10 CFR 50 App.B, ASME NQA-1 and
IEC61226 programs for safety related application
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
PIS 203S Samplers
PHYSICAL CHARACTERISTICS
Particulate:Filter type:  berglassEf ciency: > 99.95%
Iodine:Cartridge type: silver zeoliteEf ciency: > 99.99% (for methyl iodine)
Lead shielding: 4 π/15 cm (4 π/5.9 in)
Concentration on  lter: 3.7 10
+7
to 3.7 10
+12
Bq/m3
(10-3 to 10+2 μci/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature:
+10°C to +40°C (+50°F to +104°F)
Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
Pressure: 860 to 1060 hPa
MTBF: > 20 000 hours, with preventive maintenance
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 1 l/min
MECHANICAL CHARACTERISTICS
Dimensions: 1214 mm x 650 mm x 740 mm
(48 in x 25.6 in x 29.1 in)
Weight: ~ 750 kg (~ 1650 lb)
Color: gray RAL 7030 (decontaminable paint)
Inlet tube connection: Ø 6.3 mm OD (1/4 in)
Outlet tube connection: Ø 6.3 mm OD (1/4 in)
ELECTRICAL CHARACTERISTICS
Power supply: refer to possible versions
SIGNALING
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Environmental: RG 1.97
EMC: 2006/95/CE, 2004/108/CE
VERSIONS
230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac + 400
Vac 3Ø
Single module or dual unit
Heat tracing with heat control unit, temperature
sensors, thermal insulation and heating cartridge
ACCESSORIES
Shielded transport cart
Sampling handling tool
Filter paper
www.mirion.com
145019EN-C