Mirion Technologies PIM 206S User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
PIM 206S
Combination Monitors
PIM 206 S
The PIM 206S monitor forms part of the RAMSYS
product line.
It has been developed to continuously measure the
particulate and iodine volumetric activities in stacks,
ventilation ducts or working areas. It integrates all the
functions and performances of the ABPM 201 and IM
201 monitors into a single monitor.
APPLICATIONS
• Effl uent release monitoring
Stacks, etc.
• Radioprotection of workers
Containment atmosphere
Control room air, etc.
• Post-accident operations (in combination with PIS
203 and NGM 203)
Containment atmosphere
Stacks, etc.
RELATED MONITORS
• ABPM 201S: particulate monitoring only
• IM 201S: iodine monitoring only
• PING 206S: adding noble gas monitor
• PIS 203S: accident range particulate and iodine
sampler
• NGM 203S: high range noble gas monitor
FEATURES
Particulate monitoring with static and dynamic
compensation of the radon and thoron solid progenies
Iodine monitoring for both molecular and organic forms
1E quali cation and embedded safety related software
Available under 10 CFR 50 App.B, ASME NQA-1 and
IEC 61226 programs for safety related applications
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
PIM 206S
Combination Monitors
PHYSICAL CHARACTERISTICS
Particulate (ABPM 201):
Radiation detected: alpha, beta and gamma Detector: dual large area silicon Filter type: FSLW2 (MILLIPORE) Typical energy windows:
- Alpha: 2 MeV to 10 MeV
- Beta: 80 keV to 2.5 MeV
- Gamma: 80 keV to 2.5 MeV
Typical measurement range:
- Alpha: 10-2 to 3.7 10+6 Bq/m3 (2.7 10
-13
to 10-4 µCi/cc)
- Beta: 1 to 3.7 10+6 Bq/m
3
(2.7 10
-11
to 10-4 µCi/cc)
Iodine (IM 201):
Radiation detected: gamma Detector: 1¼”x1” NaI(TI) Iodine cartridge: 57.7 mm (2.27 in) Energy range: 100 keV to 3 MeV Typical energy window: 314 - 414 keV (
131
I, Eγ 364.5 keV)
1024 channels spectrum Typical measurement range: 3.7 to 3.7 10
+6
Bq/m3
(10
-10
to 10-4 µCi/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature: +10°C to +40°C (+50°F to +104°F)
Periodic temperature: -5°C to +55°C (+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
TID (processing unit): 100 Gy (10
+4
rad)
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 35 l/min (1.24 scfm)
Pressure drop: 100 to 350 mbar (1.45 to 5.07 psi)
MECHANICAL CHARACTERISTICS
Dimensions: 1614 mm x 1535 mm x 690 mm
(63.5 in x 60.4 in x 27.1 in)
Weight: between 670 kg (1477 lb) and 720 kg (1587 lb),
depending on versions
Color: gray RAL 7030 (decontaminable paint)
Inlet tube connection: Ø 25.4 mm OD (1 in)
Outlet tube connection: Ø 12 mm OD (1/2 in)
ELECTRICAL CHARACTERISTICS
Power supply: refer to possible versions
Data link outputs: 1 RS232 and 5 isolated RS485
Alarm relays: 6 SPDT relays and 5 DPDT relays
I/O: 6 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING
Graphic display: measurement, historical trend, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60761, IEC61171, IEC61172, IEC61578
Environmental: IEC60780, IEEE323, RG 1.97
Seismic: IEC60980, IEEE344
EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4
VERSIONS
230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac + 400 Vac 3Ø
Solenoid check sources for ABPM 201 and IM 201
PIS particulate and iodine samplers
Second pump for redundancy
ACCESSORIES
Local & remote display units
Calibration tools
Software
USB converters
www.mirion.com
144109EN-D
Loading...