Mirion Technologies NGM 202L User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
NGM 202L
Particulate, Gas & Iodine Monitors
NGM 202L
The NGM 202L monitor forms part of the RAMSYS
product line. It has been developed to continuously
sample the volumetric activity of noble gas and
tritium in a radioactive ef uent gaseous sample. This
monitor can operate as a stand alone device or in
conjunction with a particulate monitor (ABPM 201),
iodine monitor (IM 201) and with a high range noble
gas monitor (NGM 203) to form a very wide range
monitoring system.
APPLICATIONS
• Ef uent release monitoring
Ventilation ducts
Stacks, etc.
• Radioprotection of workers
RELATED MONITORS
• NGM 203S: accident range noble gas monitor
• NGM 204L: low range noble gas monitor
• NGM 204S: low range noble gas monitor
FEATURES
Dynamic gamma radiation compensation
Compact and reliable
Minimum periodic maintenance
Tritium measurement capability
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
NGM 202L
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
Radiation detected: alpha, beta, gamma and tritium
Detector: dual large differential ionization chamber
Sampling chamber: 10 l (10000 cc) each
Energy range: 5 keV to 8 MeV
Typical measurement range: 10
+4
to 3.7 10+9 Bq/m3
(2.7 10-7 to 10-1 µCi/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature: +10°C to +40°C
(+50°F to +104°F)
Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
TID (processing unit): 25 Gy (2.5 10
+3
rad)
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 35 l/min (1.24 scfm)
Pressure drop: according to the  lter dust loading
MECHANICAL CHARACTERISTICS
Dimensions: 1354 mm x 825 mm x 440 mm
(53.3 in x 32.5 in x 17.3 in)
Weight: ~ 96 kg (~ 212 lb)
Color: gray RAL 7030 (decontaminable paint)
Inlet tube connection: Ø 12 mm OD (1/2 in)
Outlet tube connection: Ø 25.4 mm OD (1 in)
ELECTRICAL CHARACTERISTICS
Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING (Applicable to LPDU only)
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60761
EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
230 Vac or 120 Vac
LPDU or LPU
Pumping unit
With or without RS485 junction box
Monitor wall or  oor mounting plate
ACCESSORIES
Calibration tools
Software
USB converters
Dust  lter holder
PIS sampler
www.mirion.com