Mirion Technologies LM 212S User Manual

LM 212S
Liquid Monitors
LM 212S
Off-line Gamma Liquid Monitor
been developed to sample liquid from a pipe, a tank
or a pool. A 1¼"x1" NaI scintillation detector face
a liquid sampler inside a 4 π lead shielding for the
gamma emitting isotope monitoring.
A radioactive source
allows compensation of potential drift whenever tem-
perature changes occur.
The spectrometry capability, based on a 1024 chan-
nel spectrum analysis, allows immediate and easy
radio isotope identi cation in case of alarm.
241
Am built into the NaI crystal
FEATURES
Energy spectrum and temperature changes
compensation
Available with or without display or local signaling
Seismically quali ed
The hinge mounted cover of the lead shielding makes
maintenance easier on the detector
Available under 10 CFR 50 App.B, ASME NQA-1 and
IEC61226 programs for safety related applications
APPLICATIONS
• Ef uent release monitoring
• Operational process monitoring
RELATED MONITORS
• LM 211S: version with 5 cm (2 in) lead shield
radiation monitoring systems
A Mirion Technologies Division
Featuring:
LM 212S
Liquid Monitors
PHYSICAL CARACTERISTICS
Radiation detected: gamma
Detector: 1¼” x1” NaI(TI) scintillator + PMT with
241
source
Lead thickness: 4 π/10 cm (4 π/4 in)
Energy range: 100 keV to 2.2 MeV
Measurement range: 3.7 10
+3
to 3.7 10+9 Bq/m3
(10-7 to 10-1 μCi/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature: +10°C to +40°C
(+50°F to +104°F)
Periodic temperature: -5°C to +55°C
(+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
TID (processing unit): 100 Gy (10
TID (detector): 1.5 10
+4
Gy (1.5 10+6 rad)
+4
rad)
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 10 l/min (0.57 scfm)
Beaker volume: 4.5 l
MECHANICAL CHARACTERISTICS
Dimensions: 1303 mm x 750 mm x 675 mm (51.3 in x
29.5 in x 26.6 in)
Weight: ~ 800 kg (~ 1763 lb)
Color: gray RAL 7030 (decontaminable paint)
Am
SIGNALING (Applicable to LPDU only)
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60861
Environmental: IEC60780, IEEE323
Seismic: IEC60980, IEEE344
EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac
+ 400 Vac 3Ø
LPDU or LPU
Valve plate
Pumping system
ACCESSORIES
Calibration tools
Software
USB converters
Seismic quali ed wall mounting bracket for LP(D)U
ELECTRICAL CHARACTERISTICS
Power supply: refer to possible versions
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
www.mirion.com
144399EN-E
Mirion Commercial (Beijing) Co., Ltd. Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.