Mirion Technologies KG 80 PED-PEF User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
KG 80 PED/PEF
Detectors
KG 80 PED/PEF
The ionization chambers series KG 80 PED/PEF have
been developed for gamma dose rate monitoring
under extreme conditions. There is a check volume
integrated in the detector housing which enables
remote testing without access to the location of the
detector.
Gamma radiation produces charged particles in the
gas  lling of the ionization chamber.
The transportation of these charged particles in the
electrical  eld and their discharge on the electrodes
causes charge pulses.
The charge pulses are integrated to a DC current,
which can be measured in the external circuit.
APPLICATIONS
• Radioprotection of workers
Measurement of medium and high dose rates
for harsh environment and locations with dif cult
access
• Operational process monitoring
RELATED PRODUCTS
• KG 220/221: ionization chambers for low and
medium dose rates
• KG 50/51 SEC: ionization chambers for post
accident conditions
• TK 250: digital signal processing channel
FEATURES
Ionization chamber with check volume for
functional check at non accessible locations
Wide dynamic range up to 100 kGy/h
Energy range: 80 keV to 3 MeV
Output signal: DC current starting at 1 pA
Operates up to 180°C (356°F) and 7 bar
Robust and reliable
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
KG 80 PED/PEF
Detectors
TECHNICAL DATA
www.mirion.com
144972EN-B
348 235 EN-C
GAMMA IONIZATION CHAMBER WITH CHECK VOLUME
The KG 80 ionization chambers have a second electrode system within the pressurized housing.
The second system is irradiated by a built-in beta source, which generates a continuous detector current for test purposes.
This test current can be added to the current of the measuring system by an external switch unit, which is part of the NV 102 P c/f-converter.
Properties KG 80 PED KG 80 PEF Acceptable Tolerance
Measuring range 1e–5 Gy/h … 1e+3 Gy/h
1e+3 Gy/h … 1e+5 Gy/h
± 20%; 800V acceptable deviation of sensitivity ± 40%; 1200V acceptable deviation of sensitivity
Sensitivity Energy range
1.0e–8 A/(Gy/h)
80 keV … 3 MeV
± 15% ± 20% acceptable deviation of sensitivity
Detector housing Gas  lling Check source in the sensitive volume Check source in the check volume
Stainless steel 1.4571
Argon, 6 bar Sr-90, 3.7 MBq Sr-90, 37 MBq
Operating voltage Basic current in the sensitive volume Basic current in the check volume
800 … 1200V
Approx. 2 ... 5e-13 A
Approx. 1.2e-9 A
Temperature range Short term temperature range, for max 1 h Environmental pressure Dimensions (Ø × L, without cable)
Weight (without cable)
-25 … 180°C (-13 ... 356°F) 190°C (374°F)
0.7 … 7 bar 120 mm × 500 mm (4.7 in x 19.6 in)
6.3 kg (13.8 lb)
-25 … 100°C (-13 ... 212°F) 120°C (248°F)
0.7 … 7 bar 120 mm × 500 mm (4.7 in x 19.6 in)
6.5 kg (14.3 lb)
Cable
Connectors (KG 80 PEF only)
Coaxial cable (e.g. Raychem 5031B; the maximum length (capacity) of cable is determined by signal processing)
Coaxial cable (e.g. Raychem 5031B; the maximum length (capacity) of cable is determined by signal processing)
Fischer size 105
Interfacing of the KG 80 with TK 250 signal processing