KB 100 PEF/SEF
Detectors
TECHNICAL CHARACTERISTICS
www mirion com
144952EN-B
348 204 EN-D
Properties KB 100 PEF KB 100 SEF
Beta-sensitivity Kr-85
Xe-133
Measuring range Kr-85
Xe-133
Beta energy range
5.6e-21 A/(Bq/m3)
1.7e-21 A/(Bq/m3)
2e7 … 5e13 Bq/m
3
1e8 … 1.5e14 Bq/m
3
70 keV … 3 MeV
5.6e-21 A/(Bq/m3)
1.7e-21 A/(Bq/m3)
1e8 … 3e16 Bq/m
3
70 keV … 3 MeV
Gamma sensitivity
Gamma energy range
1.2e-11A/(mGy/h)
30 keV … 3 MeV
1.2e-11 A/(mGy/h)
30 keV … 3 MeV
Gas lling
Detector housing
Detector window
Check source
Argon, 5.5 bar
Stainless steel 1.4571
Titanium,15 µm
Strontium-90, 37 MBq
Argon, 5.5 bar
Stainless steel 1.4571
Titanium,15 µm
—
Operating voltage
Check current from check volume
800V
1e-9 A approx.
800V
—
Temperature range
Ambient pressure
Dimensions Ø ( ange) x L
Weight (without connector)
0 … 135°C (32 ... 275°F)
0 … 5 bar
115 (139) × 220 mm (6.1 (5.4) x 8.6 in)
6.5 kg approx. (14.3 lb)
0 … 80°C (32 ... 176°F)
0 … 5 bar
115 (139) × 114 mm (6.1 (5.4) x 4.5 in)
4.6 kg approx. (10.1 lb)
Beta-sensitivity and measuring range are in combination with measuring vessel RSD 211.
(in the main direction
through the window)
BETA IONIZATION CHAMBER WITH CHECK VOLUME
The KB 100 PEF ionization chamber has a second
electrode system within the pressurized housing.
The second system is irradiated by a built-in beta source,
which generates a continuous detector current for test
purposes.
This test current can be added to the current of the
measuring system by an external switch unit, which is
part of the NV 102 P c/f-converter.
Interfacing of the KB 100
with TK 250 signal
processing