Mirion Technologies IM 201S User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
IM 201S
Particulate, Gas & Iodine Monitors
IM 201S
The IM 201S monitor forms part of the RAMSYS
product line. It has been developed to continuously
measure the gamma volumetric activity of radioactive
iodine sample, in both molecular and organic forms
(methyl iodine), contained in air drawn from stacks,
ventilation ducts or working areas.
An NaI scintillation detector faces the activated
charcoal cartridge in which radioactive iodine is
trapped. The proximity of the detector and the
cartridge, enclose within a 4 π/5 cm (4 π/2 in) lead
shielding, serves to optimize detection ef ciency.
A radioactive
241
Am source built into the NaI
christal allows compensation of temperature and
aging related drifts. The spectrometry capability,
based on a 1024 channel spectrum analysis, allows
radio iodine isotope localisation.
APPLICATIONS
• Ef uent release monitoring
Stacks, etc.
• Radioprotection of workers
Containment atmosphere
Control room air, etc.
• Post-accident operations
Containment atmosphere
Stacks, etc.
RELATED MONITORS
• IM 201L: “light” version
• IM 201M: mobile version
• PING 206S: version with particulate and noble gas
• PIM 206S: version with particulate
FEATURES
Embedded
241
Am source for energy spectrum
stabilization against temperature changes and aging
1024 channels spectrum analysis
1E quali cation and embedded safety related
software
Available under 10 CFR 50 App.B, IEC61226 and
ASME NQA-1 programs for safety related application
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
IM 201S
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
Radiation detected: gamma
Detector: 1¼”x1” NaI(TI) scintillator + PMT +
241
Am
source
Iodine cartridge: 57.7 mm (2.27 in)
Energy range: 100 keV to 3 MeV
Typical energy windows: 314 - 414 keV
(
131
I, Eγ 364 keV)
1024 channels spectrum
Typical measurement range: 3.7 to 3.7 10
+6
Bq/m3
(10
-10
to 10-4 µCi/cc)
ENVIRONMENTAL CHARACTERISTICS
Long term temperature:
+10°C to +40°C (+50°F to +104°F)
Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
TID: 100 Gy (10
+4
rad)
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 35 l/min (1.24 scfm)
Pressure drop: according to the  lter dust loading
MECHANICAL CHARACTERISTICS
Dimensions: 1280 mm x 830 mm x 680 mm (50.4 in x
32.7 in x 26.8 in)
Weight: ~ 300 kg (~ 661 lb)
Color: gray RAL 7030 (decontaminable paint)
Inlet tube connection: Ø 12 mm OD (1/2 in)
Outlet tube connection: Ø 12 mm OD (1/2 in)
ELECTRICAL CHARACTERISTICS
Power supply: refer to possible versions
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60761, IEC61171
Environmental: IEC60780, IEEE323
Seismic: IEC60980, IEEE344
EMC: 2006/95/CE and 2004/108/CE, EPRI 102323,
MIL STD 461 E, IEC61000-6-2 and IEC61000-6-4
VERSIONS
230 Vac or 230 Vac + 400 Vac 3Ø or 120 Vac
+ 400 Vac 3Ø
Solenoid check source
With or without dust  lter holder
Gas grab sampler ports
ACCESSORIES
Calibration tools
Software
USB converters
Ethernet
ESD board
www.mirion.com
144106EN-D
Loading...