Mirion Technologies IM 201M User Manual

Page 1
Featuring:
radiation monitoring systems
A Mirion Technologies Division
IM 201M
Particulate, Gas & Iodine Monitors
IM 201M
The IM 201M monitor forms part of the RAMSYS
product line.
It has been developed to continuously measure
the gamma volumetric activity of radioactive iodine
sample, in both molecular and organic forms (methyl
iodine), contained in air drawn from stacks, ventilation
ducts or working areas.
An NaI scintillation detector faces the activated
charcoal cartridge in which radioactive iodine is
trapped. The proximity of the detector and the
cartridge, enclose within a 4 π/5 cm (4 π/2 in) lead
shielding, serves to optimize detection ef ciency. A
radioactive
241
Am source built into the NaI
christal allows compensation of temperature and
aging related drifts. The spectrometry capability,
based on a 1024 channel spectrum analysis, allows
radio iodine isotope localization.
APPLICATIONS
• Ef uent release monitoring
• Radioprotection of workers
○ Containment atmosphere
○ Control room air, etc.
• Operational process monitoring
RELATED MONITORS
• IM 201L: “light” version
• IM 201S: seismically quali ed version
• PING 206S: version with particulate and noble gas
• PIM 206S: version with particulate
FEATURES
● Embedded
241
Am source for energy spectrum
stabilization against temperature changes and aging
● 1024 channels spectrum analysis
● Ef uent trapping of both molecular and organic forms
of iodine
● Can be used as temporary bypass for IM 201S or
IM 201L to maintain full monitoring capability during
maintenance
Page 2
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
IM 201M
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
● Radiation detected: gamma
● Detector: 1¼”x1” NaI(TI) scintillator + PMT
● Iodine cartridge: 57.7 mm (2.27 in)
● Energy range: 100 keV to 3 MeV
● Typical energy windows: 314 - 414 keV
(
131
I, Eγ 364.5 keV)
● 1024 channels spectrum
● Typical measurement range: 3.7 to 3.7 10
+6
Bq/m3
(10
-10
to 10-4 µCi/cc)
ENVIRONMENTAL CHARACTERISTICS
● Long term temperature: +10°C to +40°C
(+50°F to +104°F)
● Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
● MTBF: > 20 000 hours, with preventive maintenance
● TID: 100 Gy (10
+4
rad)
PNEUMATIC CHARACTERISTICS
● Standard  ow rate: 35 l/min (1.24 scfm)
● Pressure drop: according to the  lter dust loading
MECHANICAL CHARACTERISTICS
● Dimensions: 1406 mm x 520 mm x 700 mm
(55.4 in x 20.5 in x 27.6 in)
● Weight: ~ 200 kg (~ 441 lb)
● Color: gray RAL 7030 (decontaminable paint)
● Inlet tube connection: Ø 12 mm OD (1/2 in)
ELECTRICAL CHARACTERISTICS
● Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
● Data link outputs: 1 RS232 and 2 isolated RS485
● Alarm relays: 3 SPDT relays
● I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING
● Alphanumeric display: measurement, status…
● Sound alarm: buzzer 90 dBA at 1 meter
● Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
● Nuclear: IEC60761, IEC61171
● EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
● 230 Vac or 120 Vac
● With or without dust  lter holder
● With or without output dust  lter
ACCESSORIES
● Calibration tools
● Software
● USB converters
www.mirion.com
144105EN-D
Loading...