Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières
FR-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna, GA 30082
USA
T +1 770 432 2744
F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a
DE-80687 Munich
Germany
T +49 (0) 89515 13-0
F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu
Zhabei District, Shanghai 200436
PR of China
T +86 21 6180 6920
F +86 21 6180 6924
IM 201L
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
● Radiation detected: gamma
● Detector: 1¼”x1” NaI(TI) scintillator + PMT
● Iodine cartridge: 57.7 mm (2.27 in)
● Energy range: 100 keV to 3 MeV
● Typical energy windows: 314 - 414 keV
(
131
I, Eγ 364.5 keV)
● 1024 channels spectrum
● Typical measurement range: 3.7 to 3.7 10
+6
Bq/m3
(10
-10
to 10-4 µCi/cc)
ENVIRONMENTAL CHARACTERISTICS
● Long term temperature: +10°C to +40°C
(+50°F to +104°F)
● Maximum periodic ambient temperature:
-5°C to +55°C (+23°F to +131°F)
● MTBF: > 20 000 hours, with preventive maintenance
● TID: 100 Gy (10
+4
rad)
PNEUMATIC CHARACTERISTICS
● Standard ow rate: 35 l/min (1.24 scfm)
● Pressure drop: according to the lter dust loading
MECHANICAL CHARACTERISTICS
● Dimensions: 864 mm x 725 mm x 440 mm
(34 in x 28.5 in x 17.3 in)
● Weight: ~ 220 kg (~ 495 lb)
● Color: gray RAL 7030 (decontaminable paint)
● Inlet tube connection: Ø 12 mm OD (1/2 in)
● Outlet tube connection: Ø 12 mm OD (1/2 in)
ELECTRICAL CHARACTERISTICS
● Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
● Data link outputs: 1 RS232 and 2 isolated RS485
● Alarm relays: 3 SPDT relays
● I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING (Applicable to LPDU only)
● Alphanumeric display: measurement, status…
● Sound alarm: buzzer 90 dBA at 1 meter
● Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
● Nuclear: IEC60761, IEC61171
● EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
● 230 Vac or 120 Vac
● LPDU or LPU
● With or without dust lter holder
● Pumping unit
● Monitor wall or oor xing plate
ACCESSORIES
● Calibration tools
● Software
● USB converters
www.mirion.com
144104EN-D