Mirion Technologies DWK 250 User Manual

Featuring:
radiation monitoring systems
A Mirion Technologies Division
DWK 250
Neutron Flux Instrumentation
DWK 250
The DWK 250 digital wide range channel forms part
of the proTK™ product line. It has been designed for
neutron  ux monitoring combining pulse and inter-
mediate range in combination with only one detector,
e.g. in-core or ex-core wide range  ssion chamber.
Hardware and software of the DWK 250 are designed
and quali ed for applications at the level of the reac-
tor protection system.
APPLICATIONS
• Operational process monitoring
Safety related signal processing for neutron  ux
monitoring combining pulse range and intermediate
range
RELATED PRODUCTS
• Neutron detectors: e.g. wide range  ssion
chambers
• Digital signal processing channels: e.g. for pulse
range or power range monitoring
FEATURES
10 to 12 decades of measurement range with one
detector only
Calibration into neutron  ux units (nv, P/Pn)
Filtering of the signal with an adaptive time constant
Calculation of the relative  ux change rate (reciprocal of
the reactor period) for the whole wide range
Generation of analog and binary signals for the reactor
protection system
Remote test signal generators, even in the preampli er
The correct function of all signal pathes can be tested
before start-up
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
DWK 250
Neutron Flux Instrumentation
www.mirion.com
144970EN-C
DETECTORS
Wide range  ssion chambers, e.g. WSK 61 (AREVA
NP), or CFUG 08 (Photonis)
Cable length to the preampli er: up to 100 m (328 ft)
Detector supply: 0 ... 500V/4mA or 0 ... 800V/30mA
PREAMPLIFIER
Input and output impedance matched to the cable
impedance
Remote operated test signal generator
Short term operation (max. 3 h) up to 100°C (212°F)
DIGITAL SIGNAL PROCESSING
80C31/32 multi-processor system
Program memory: EPROM
Parameter memory: CMOS-RAM with integrated
Li-battery
Data interface: RS232 and/or RS485
Internal LC-display: 2 x 16 characters
OUTPUT SIGNALS
Linear pulse count rate, e.g. 0 ... 5e5 cps
Linear RMS signal with automatic range selection,
e.g. 0 ... 40/125 %Pn
Logarithmic wide range signal, e.g. 1e3 ... 1e13 nv
Neutron  ux change rate = 1 / reactor period,
e.g. -1.25 ... 0 ... 12.5 %/s corresponding to
-80 ... ∞ ... +8 s reactor period
Analog outputs: 0/4 ... 20mA/600 Ω, insulated
Binary outputs: insulated relays change overs
60V/0.5A or 125V/1A
OTHER CHARACTERISTICS
DC power supply: 18 ... 33 VDC, approx. 1.6A at 24 V
Optional: AC power supply: 230 VAC or 115 VAC
+10%/-15%, approx. 40 VA
Operating temperature: 0 ... 70°C (32 ... 158°F) for the
main electronics
Mechanical vibrations: < 5 g, 5 ... 100 Hz
19” modular system according to IEC 60297
Rack size (W×H×D): 483 mm × 133 mm × 280 mm
(19 in x 5.2 in x 11 in)
Plug-in boards: 100 mm × 160 mm (3.9 in x 6.3 in)
DWK 250 schematic diagram