Mirion Technologies DMC 2000GN User Manual

FEATURES
DMC 2000GN
Active Dosimetry - Dosimeters
DMC 2000GN
Personal Electronic Dosimeter
The DMC 2000GN takes advantage of a unique
single diode. The full energy range of thermal, interme-
diate and high energy neutrons is covered with a high
sensitivity and very good gamma rejection, tested up to
6 MeV.
In addition, the gamma detection technology, identical
to the proven solution of the DMC 2000S, allows to
measure the gamma dose from 50 keV to over 6 MeV
gamma range with a unique linearity to over 10 Sv/h.
The DMC 2000GN is a gamma and neutron detection
dosimeter featuring dose rate and programmable alarms.
The DMC 2000GN is user friendly, lightweight and
waterproof.
Full neutron energy range coverage
Very good gamma rejection in neutron channel
Audible and visual alarms
Very large autonomy
Hand free communication, pass-by exchange
Fully compatible with DMC 2000 hand free readers and
dosimetry software
RELATED PRODUCTS
MGP Instruments offers a range of products which can
be used with the DMC 2000GN to create integrated
dosimetry systems including:
LDM 220, LDM 230 proximity readers
LDM 2000 pass-by data exchange
DOSISERV dosimetry centralization and access
control software
DOSIMASS dosimeter con guration software
DOSICARE and DOSIFAST operational dosimetry
software
IRD 2000 irradiator for dosimeters
health physics
A Mirion Technologies Division
Featuring:
DMC 2000GN
Active Dosimetry - Dosimeters
PHYSICAL CHARACTERISTICS
Compliant to IEC 61526 Ed2 for gamma and neutron
Display units: mSv, μSv or mrem
Neutron measurement:
dose display: 1 μSv to 10 Sv (0.1 mrem to 1000 rem) dose rate display: 0.1 mSv/h to 10 Sv/h measurement range: 1 μSv/h to 10 Sv/h energy range: 0.025 eV to 15 MeV
Gamma measurement:
dose display: 1 μSv to 10 Sv (0.1 mrem to 1000 rem) rate display: 10 μSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h) or from 1 μSv/h / 0.1 mrem/h (extended option) measurement range: 0.1 μSv/h to 10 Sv/h energy range: 50 keV to 6 MeV linearity:
<± 20% up to 1 Sv/h (100 rem/h) ­<± 30% up to 10 Sv/h (1000 rem/h) -
accuracy: <10% (
137
Cs, ~ 25 mSv/h, including ± 5% of
extended uncertainty K=2)
ENVIRONMENTAL CHARACTERISTICS
Temperature range: -10°C to 50°C (14°F to 122°F)
Humidity: < 90 % at 42°C (108°F) Storage: -30°C to 71°C (-22°F to 160°F) Shock, vibration and drop resistant Waterproof IP67
Neutron dose display
Neutron dose rate display
ELECTRICAL CHARACTERISTICS
Standard calculator battery LiMnO
CR2450
2
Autonomy: 9 months typical (8h per day in run mode)
MECHANICAL CHARACTERISTICS
Dimensions: 87 x 48 x 31 mm (3.4 x 1.9 x 1.2 in) with clip 87 x 48 x 21 mm (3.4 x 1.9 x 0.8 in) without clip Weight: 80 g (2.8 oz) with battery
Front view
Rear view with the clip
Health Physics Division
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna Georgia 30082 USA T +1.770.432.2744 F +1.770.432.9179
Mirion Technologies (MGPI) SA
BP 1 F-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506 FIN-20101 Turku Finland
T +358 2 468 4600 F +358 2 468 4601
The histogram enables neutron and gamma doses, events, to be recons­tructed in detail (3800 steps of 10 s, 1 min, 10 min, 1 h or 24 h). Radiological supervisors can then analyze the data surrounding an incident. All parameters change (alarm threshold, calibration factors, ...) are memorized in the histogram. Histogram available with DOSIMASS software
www.mirion.com
144272EN-C
Mirion Technolgies (RADOS) GmbH
Ruhrstrasse 49 DE-22761 Hamburg Germany
T +49 (0) 40 851 93-0 F +49 (0)40 851 93 256
Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.
Loading...