Mirion Technologies DIS-1 User Manual

health physics
A Mirion Technologies Division
DIS-1 Direct Ion Storage Dosimeter
Passive Dosimetry
FEATURES
direct measurement of Hp(10) and Hp(0.07) over • the entire energy range instant readout• extremely easy readout process• passive operation• not sensitive to EM and RF interference• operation at high dose rates• operation at pulsed  elds• PTB approved model available
DIS-1
Direct Ion Storage Dosimeter
The RADOS DIS-1 personal dosimeter is
based on an ionisation chamber combined
with a modern electronic Direct Ion Storage
(DIS) memory cell. The Ion chamber is widely
used as a reference detector in radiation
detection and is now available in everyday
dosimetry applications.
The DIS-1 dosimeter could be described as a pas-
sive electronic TLD or Film badge, which can be
read instantly and non-destructively without any
loss of dose information. This unique feature al-
lows the user of the DIS-1 to read his/her accumu-
lated doses whenever necessery.
The DIS-1 dosimeter has a small, lightweigth, rug-
ged and watertight construction, which makes the
DIS-1 dosimeter reliable and easy to use.
The radiological range of the DIS-1 covers the
entire Hp (10) and Hp (0.07) photon and beta
energies without any compromises. The wide
dose and energy range, the ability to operate in
pulsed  elds and the performance at high dose
rates make DIS-1 an ideal device for all kinds
of radiation dosimetry applications. The DIS-1
allows for the detection of heavy, high-energy
ions and its immunity to any external
interference is unequalled. There are no devia-
tions in the dose readings even at very high
EM or RF  elds.
www.mirion.com
Mirion Technologies (MGPI) Inc 5000 Highlands Parkway Suite 150 Smyrna Georgia 30082 USA T +1.770.432.2744 F +1.770.432.9179
Mirion Technologies (MGPI) SA
BP 1 F-13113 Lamanon France T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506 FIN-20101 Turku Finland T +358 2 4684 600 F +358 2 4684 601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49 D-22761 Hamburg Germany T +49 40 85193 0 F +49 40 85193 256
DIS-1 Direct Ion Storage Dosimeter
Passive Dosimetry
20996029_DIS1_EN_C
TECHNICAL SPECIFICATIONS:
Physical Characteristics
detector type: three
TM
DIS (Direct Ion Storage)detectors and two MOSFET detectors sensitive to gamma, X-ray and beta radiation• insensitive to neutrons (<5 %)• instant readout of ICRU dose equivalents:
- Hp(10) 1 μSv to 40 Sv (0.1 mrem to 4000 rem)
1)
- Hp(0.07) 10 μSv to 40 Sv (1 mrem to 4000 rem)
1)
calibration accuracy:• ±5 % at 1 mSv
137
Cs Hp(10)
±10 % at 10 mSv
137
Cs Hp(0.07) energy response in the dose range up to 1 Sv:• photons
- Hp(10) ±30 % from 15 keV to 9 MeV
2)
- Hp(0.07) ±30 % from 6 keV (to 9 MeV) beta
- Hp(0.07) -20...+35% from 0.24 MeV to 0.80 MeV (E
mean
)
angular response for photons:
- Hp(10) ±20 % up to 60° at 65 keV
- Hp(0.07) ±20 % up to 60° at 65 keV
1)
When calibrated after every 10 Sv of accumulated dose
2)
PTB approved model up to 7 Mev
Functional Characteristics
recording of o cial Hp(10) and Hp(0.07) doses memory:
- calibration date
- dose reset dates
- user identi cation
Mechanical Characteristics
size: 41 x 44 x 12 mm, with holder 47 (95 with strap) x 49 x 13 mm (1.61 x 1.73 x 0.47 in , with • holder 1.85 (3.74 with strap) x 1.93 x 0.51 in) weight: 25 g (0.88 oz) , with holder 43 g (1.52 oz)• beta window: aluminized PI (app. 7 mg/cm2)• holder: anodized aluminium
Environmental Characteristics temperature range from - 10°C to +50 °C ( 14 °F to 122 °F)
enclosure class: IP 67 (waterproof)
The DBR-1 and DBR-2 Readers are designed to read DIS-1 dosimeters assembled in the DDH-2 Snap-in Dosimeter Holders. To obtain the most recent dose value, the user simply plugs the dosimeter into the reader and the dose values are displayed in a few seconds.
Typical energy response of personal dose equivalent Hp(10)
Typical energy response of personal dose equivalent Hp(0.07)
Loading...