ABPM 203M
Mobile Alpha Beta Particulate Monitor
line.
Its small and light weight extendable sensor allows
this monitor to function locally next to the respiratory
tract of workers. A dual silicon detector (pips)
performs the gamma compensation and a radial  n
grid limits the scattering of the alpha particles (static
compensation) which facilitates the compensation of
the radon and thoron solid progenies by the
processing algorithms (dynamic compensation).
Operating costs are minimised through unattended
operation, by the use of a continuous  lter and the
on-line spectroscopy capability. All these features
make the ABPM 203M an ef cient diversi ed and
cost effective tool.
FEATURES
Static and dynamic compensation of the radon and
thoron solid progenies
Perfectly adapted for alpha and beta measurement of
particulates in environment with high rate of radon
Optimized alpha measurement for high energies
238
(PU
Real time alpha spectrometry
Up to 6 months  lter cassette autonomy
, PU
239
)
APPLICATIONS
• Radioprotection of workers
Containment atmosphere Control room air, etc.
RELATED MONITORS
• ABPM 203P: portable version
• ABPM 204M: uranium optimized version
radiation monitoring systems
A Mirion Technologies Division
Featuring:
ABPM 203M
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
Radiation detected: alpha, beta and gamma
Detector: dual large area silicon
Filter type: FSLW2 (MILLIPORE)
Typical energy windows:
Alpha: 2 MeV to 10 MeV Beta: 80 keV to 2.5 MeV Gamma: 80 keV to 2.5 MeV
Typical measurement range:
Alpha: 10
(2.7 10
Beta: 1 to 3.7 10
(2.7 10
-2
to 3.7 10+6 Bq/m
-13
to 10-4 μCi/cc)
-11
to 10-4 μCi/cc)
+6
Bq/m
3
3
ENVIRONMENTAL CHARACTERISTICS
Long term temperature:
+10°C to +40°C (+50°F to +104°F)
Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
of the pump
TID: 100 Gy (10
+4
rad)
SIGNALING
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 2 lights (red, yellow)
REFERENCE STANDARDS
Nuclear: IEC60761, IEC61172, IEC61578
EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
230 Vac or 120 Vac
Detection sub-assembly available with or without
shielding
Hoses length: 1.5 m (5 ft), 3 m (10 ft), 10 m (33 ft) or
20 m (66 ft)
ACCESSORIES
Calibration tools
Software
USB converters
Radiomodem (either customer speci ed or WRM2)
PNEUMATIC CHARACTERISTICS
Standard  ow rate: 35 l/min (1.24 scfm)
Pressure drop: 100 to 350 mbar (1.45 to 5.07 psi)
MECHANICAL CHARACTERISTICS
Dimensions: 1270 mm x 360 mm x 303 mm
(50 in x 14.2 in x 12 in)
Weight: ~ 26 kg (~ 57 lb)
Color: gray RAL 7030 (decontaminable paint)
ELECTRICAL CHARACTERISTICS
Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs (0/4-20 mA)
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
www.mirion.com
144101EN-F
Mirion Commercial (Beijing) Co., Ltd. Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.