Mirion Technologies ABPM 201L User Manual

ABPM 201L
Particulate, Gas & Iodine Monitors
ABPM 201L
Alpha Beta Particulate Monitor
product line.
It has been developed to sample air extracted from
ventilation ducts or stacks. A double silicon detector
performs the gamma compensation and a radial  n
grid limits the scattering of the alpha particles (static
compensation) which facilitates the compensation of
the radon and thoron solid progenies by the
processing algorithms (dynamic compensation).
Operating costs are minimised by the autonomous
operation through automatic  lter advance
management.
FEATURES
Static and dynamic compensation of the radon and
thoron solid progenies
Natural and arti cial alpha volumetric activity indication
On-line spectrometry
Up to 6 months  lter cassette autonomy
APPLICATIONS
• Effl uent release monitoring
Stacks, etc.
• Radioprotection of workers
Containment atmosphere Control room air, etc.
RELATED MONITORS
• ABPM 201M: mobile version
• ABPM 201S: seismically quali ed version
• PING 206S: version with iodine and noble gas
• PIM 206S: version with iodine
• PNG 206S: version with noble gas
radiation monitoring systems
A Mirion Technologies Division
Featuring:
ABPM 201L
Particulate, Gas & Iodine Monitors
PHYSICAL CHARACTERISTICS
Radiation detected: alpha, beta and gamma
Detector: dual large area silicon
Filter type: FSLW2 (MILLIPORE)
Typical energy windows: Alpha: 2 MeV to 10 MeV Beta: 80 keV to 2.5 MeV Gamma: 80 keV to 2.5 MeV
Typical measurement range: Alpha: 10
(2.7 10
Beta: 1 to 3.7 10
(2.7 10
-2
to 3.7 10+6 Bq/m
-13
to 10-4 μCi/cc)
-11
to 10-4 μCi/cc)
+6
Bq/m
3
3
ENVIRONMENTAL CHARACTERISTICS
Long term temperature:
+10°C to +40°C (+50°F to +104°F)
Periodic temperature:
-5°C to +55°C (+23°F to +131°F)
MTBF: > 20 000 hours, with preventive maintenance
TID: 100 Gy (10
+4
rad)
PNEUMATIC CHARACTERISTICS
ELECTRICAL CHARACTERISTICS
Power supply: 230 Vac – 50 Hz or 120 Vac – 60 Hz
Data link outputs: 1 RS232 and 2 isolated RS485
Alarm relays: 3 SPDT relays
I/O: 2 isolated analog outputs and 1 isolated analog
input (0/4-20 mA)
SIGNALING (Applicable to LPDU only)
Alphanumeric display: measurement, status…
Sound alarm: buzzer 90 dBA at 1 meter
Visual alarm: 3 lights (red, yellow, green)
REFERENCE STANDARDS
Nuclear: IEC60761, IEC61172, IEC61578
EMC: 2006/95/CE and 2004/108/CE, IEC61000-6-2
and IEC61000-6-4
VERSIONS
230 Vac or 120 Vac
LPDU or LPU
Pumping unit
Monitor wall or  oor  xing plate
Standard  ow rate: 35 l/min (1.24 scfm)
Pressure drop: 100 to 350 mbar (1.45 to 5.07 psi)
MECHANICAL CHARACTERISTICS
Dimensions: 784 mm x 725 mm x 440 mm
(30.9 in x 28.5 in x 17.3 in)
Weight: ~ 160 kg (~ 352 lb)
Color: gray RAL 7030 (decontaminable paint)
Inlet tube connection: Ø 18 mm OD (3/4 in)
Outlet tube connection: Ø 12 mm OD (1/2 in)
Mirion Technologies (MGPI) SA
Route d’Eyguières FR-13113 Lamanon France
T +33 (0) 4 90 59 59 59 F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway Suite 150 Smyrna, GA 30082 USA T +1 770 432 2744 F +1 770 432 9179
ACCESSORIES
Calibration tools
Software
USB converters
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a DE-80687 Munich Germany
T +49 (0) 89515 13-0 F +49 (0) 89515 13 169
www.mirion.com
144098EN-E
Mirion Commercial (Beijing) Co., Ltd. Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu Zhabei District, Shanghai 200436 PR of China T +86 21 6180 6920 F +86 21 6180 6924
Since norms, speci cations and designs are subject to occasional change, please ask for con rmation of the information given in this publication.
Mirion, Mirion Technologies and MGP Instruments are trademarks and/or registered trademarks of Mirion Technologies, Inc. and/or its a liates in the United States and/or other countries. All other trademarks are the property of their respective owners.