Your Thermolyne MIRAK™ stirrer, hot plate, or stirring hot plate has been designed with function,
reliability, and safety in mind. It is your responsibility to install it in conformance with local electrical codes.
Alert Signals
WARNING
Warnings alert you to a possibility of personal injury.
CAUTION
Cautions alert you to a possibility of damage to the equipment.
NOTE
Notes alert you to pertinent facts and conditions.
Warnings
WARNING
To avoid electrical shock, always:
1. Use a properly grounded electrical outlet of correct voltage and current handling capacity.
2. Disconnect from power supply before servicing.
To avoid personal injury:
1. Do not use in the presence of flammable or combustible materials; fire or explosion may result.
This device contains components which may ignite such materials.
2. Keep top surface clean. Use a non-abrasive cleaner. Alkali spills, hydrofluoric acid spills or
phosphoric acid spills may damage top and lead to thermal failure. Unplug unit and remove spills
promptly. Do not immerse unit for cleaning.
3. Replace the top immediately if damaged by etching, scratching or chipping. A damaged top can
break in use.
4. Do not remove or modify grounded power plug. Use only properly grounded outlets to avoid
shock hazard. Not rated for use in hazardous atmospheres.
5. Use caution when heating volatile materials; top surface and element can reach the “Flash Point
Temperature” of many chemicals. These hot plates and stirring hot plates are not explosion proof.
Fire or explosion may result. Unit contains components which may ignite such materials.
6. Use appropriate hand and eye protection when handling hazardous chemicals.
7. "Caution: Hot Surface. Avoid Contact." The hot plate will remain hot without any visual indication
for some time after power has been removed.
8. Do not use in highly corrosive atmospheres; corrosive fumes and spill may damage top and
internal components, creating shock hazard.
9. Refer servicing to qualified personnel.
CAUTION
Space unit 12 inches away from combustible materials under any conditions.
Gross weight of items placed on top of Hot Plates should not exceed 25 lbs. on the 7" X 7"
and 30 lbs. on the 12" X 12" models.
3
Barnstead |Thermolyne • 800-553-0039
Please note the following WARNINGS:
WARNING
This warning is presented for compliance with California Proposition 65 and other
regulatory agencies and only applies to the insulation in this product. This product
contains refractory ceramic, refractory ceramic fiber or fiberglass insulation,
which can produce respirable dust or fibers during disassembly. Dust or fibers
can cause irritation and can aggravate pre-existing respiratory diseases.
Refractory ceramic and refractory ceramic fibers (after reaching 1000°C) contain
crystalline silica, which can cause lung damage (silicosis). The International
Agency for Research on Cancer (IARC) has classified refractory ceramic fiber
and fiberglass as possible carcinogenic (Group 2B), and crystalline silica as
carcinogenic to humans (Group 1).
The insulating materials can be located in the door, the hearth collar, in the chamber of
the product or under the top plate top. Tests performed by the manufacturer indicate
that there is no risk of exposure to dust or respirable fibers resulting from operation of
this product under normal conditions. However, there may be a risk of exposure to
respirable dust or fibers when repairing or maintaining the insulating materials, or
when otherwise disturbing them in a manner which causes release of dust or fibers
therefrom. By using proper handling procedures and protective equipment you can
work safely with these insulating materials and minimize any exposure. Refer to the
appropriate Material Safety Data Sheets (MSDS) for information regarding proper
handling and recommended protective equipment. For additional MSDS copies, or
additional information concerning the handling of refractory ceramic products, please
contact the Customer Service Department at Barnstead|Thermolyne Corporation at
1-800-553-0039.
REFER SERVICING TO QUALIFIED PERSONNEL.
WARNING
4
Introduction
Please read all the information in this manual before operating the unit.
Your Thermolyne MIRAK™ hot plate, stirrer or stirring hot plate is a general purpose heating and/or stirring
plate designed for laboratory procedures requiring precise control of temperature and/or stirring speed.
Each MIRAK model includes a digital display for monitoring actual temperature or stirring speed, and the
stirring hot plate models have dual displays so temperature and stirring speed can be monitored
simultaneously. The hot plate is capable of producing accurately controlled top plate temperatures from
40°C through 540°C on all 7" x 7" models and the 208 volt and 240 volt models of the 12" x 12" hot plates.
(12" x 12" 100 volt and 120 volt models have a temperature range of 40°C to 370°C.) The temperature
may be controlled at the plate surface by an internal Type K thermocouple sensor, or the solution
temperature may be controlled by utilizing an accessory ungrounded Type K thermocouple probe. A 6" or
10" general purpose stainless steel immersion probe, or a chemical-resistant 7" solid Teflon® immersion
probe may be ordered separately. The stirrer will accurately maintain stirring speeds from 60 rpm up to
1200 rpm. On stirring hot plate models, the accessory probe may also be utilized to display the actual
temperature of heat-sensitive solutions during stirring-only functions. The 7" x 7" or 12" x 12" top plate on
the MIRAK units is solid ceramic, and is suitable for use with glass or metal vessels.
Your MIRAK hot plate, stirrer or stirring hot plate may be used for general purpose heating applications
and/or general laboratory mixing of solutions, including sample preparation, heating reagents, melting
paraffin, warming resinous chemicals, content analysis, solvent evaporations, digestions, media
preparation and sterilization, titrations, sand baths, and microscale chemistry applications.
Unpacking
Remove your MIRAK hot plate, stirrer or stirring hot plate from the carton. Inspect to ensure that the unit
has not been damaged during shipment. If the unit appears to have sustained shipping damage contact
the distributor from whom you purchased this product or Barnstead/Thermolyne Customer Service at 800553-0039. Remove the packaging from underneath the top plate by gently pulling the packaging material
toward you.
NOTE
A special 20-amp cord set is included on the CSA-approved models of
the 12" x 12" 120 volt MIRAK hot plate and stirring hot plate
(HP72935, SP73035). This 20-amp cordset can only be plugged into
a non-standard 20-amp, 120 volt receptacle. If you do not have the
appropriate receptacle, contact a qualified electrician for installation of
a 20-amp 120 volt outlet.
A 12" x 12" 120 volt MIRAK hot plate and stirring hot plate (HP7293560, SP73035-60) is available with a standard 15-amp, 120 volt cordset.
These models are not CSA-approved.
5
Barnstead |Thermolyne • 800-553-0039
20-amp, 120 volt
15-amp, 120 volt
6
Installation
WARNING
Set the unit on a flat stable surface at least 12" away from combustible materials, and plug the cordset into
a properly grounded electrical outlet of correct voltage and current handling capacity. Use your right hand
to locate the ON/OFF power switch which is on the side, near the front right corner as you face the unit.
Press the black ON/OFF power rocker switch into the ON position. The red LED display(s), and the green
power light to the left of the display(s) should illuminate. The unit is ready for operation.
Use a properly grounded electrical outlet of correct voltage and current handling capacity.
Do not remove or modify grounded power plug. Use only properly grounded outlets to avoid
shock hazard. Not rated for use in hazardous atmospheres.
Do not use in the presence of flammable or combustible materials; fire or explosion may
result. This device contains components which may ignite such materials.
Do not use in highly corrosive atmospheres; corrosive fumes and spill may damage top and
internal components, creating shock hazard.
NOTE
Some misalignment of the motor bearings in this product may have occurred during shipping.
Prior to using this product, run the stirrer at maximum speed for ten minutes to realign the
bearings.
Principles of Operation
Each MIRAK unit utilizes the latest in microprocessor technology to deliver the most reliable, accurately
controlled, ceramic top hot plate, stirrer and stirring hot plate on the market.
Your MIRAK hot plate or stirring hot plate has an electronic closed-loop feedback control which will
accurately maintain temperature setpoints from 40°C through 370°C or 540°C, depending on the model.
The top plate surface temperatures are calibrated and verified at the factory using the latest in infrared
temperature measurement. The measurements are made with an infrared camera mounted approximately
26" above the top surface of the hot plate or stirring hot plate. If the temperature measurement of the
ceramic top is made with measurement devices other than infrared, the error of the measuring technique
may be greater than the error of the unit.
The MIRAK units use a Type K thermocouple for measuring both the top plate temperature, and the
solution temperature when using the probe. Both top plate surface and probe temperature are measured
with independent analog circuits. This provides a significant increase in reliability due to the redundancy of
the circuit. Because of the dual sensors it is possible to heat certain types of metal vessels and sandbaths
on the MIRAK without damaging the ceramic top. However, we recommend that you do not control the
temperature with the probe when heating a metal vessel or a sandbath.
The electronic stirring speed control will maintain the speed setpoint when the unit is loaded at ±2 RPM as
displayed by the LED display on the unit. The motor in the MIRAK stirrers and stirring hot plates produces
maximum stirring torque under normal laboratory load conditions, and is combined with a powerful magnet
to provide exceptional magnetic coupling with a stirbar.
7
Barnstead |Thermolyne • 800-553-0039
Operation
WARNING
Removing the Clear Plastic Cover
Your MIRAK unit has a protective plastic cover over the front control panel. This plastic cover is designed
to protect the control knobs and the display panel from spills, and should remain in place when the unit is
being used. To use the knob(s) while the plastic cover is in place, spin the knob(s) from the bottom where
the knob projects below the plastic cover. This plastic cover can be easily removed for cleaning by
grasping the cover at each bottom corner and gently pulling the sides of the plastic cover away from the
case while lifting the cover toward you. To replace the cover simply slide it back onto the case.
Use caution when heating volatile materials; top surface and element can reach the “Flash
Point Temperature” of many chemicals. These hot plates and stirring hot plates are not
explosion proof. Fire or explosion may result. Unit contains components which may ignite
such materials.
Use appropriate hand and eye protection when handling hazardous chemicals.
"Caution: Hot Surface. Avoid Contact." The hot plate will remain hot without any visual
indication for some time after power has been removed.
CAUTION
To avoid damage to the top plate or heating element, always keep a vessel filled with liquid on
the top plate of a hot plate or stirring hot plate when the unit is heating or cooling.
The Power Switch
Your MIRAK unit has a separate ON/OFF power switch which is a black rocker switch located on the lower
front corner of the right side as you're facing the unit. This separate power switch allows you to turn your
MIRAK hot plate, stirrer or stirring hot plate on or off without disturbing your temperature and/or speed
setting. Your last setpoint will be held in memory even if the unit is unplugged and will reappear on the
display as soon as the unit is powered up again. Whenever the power is on, a green power light illuminates
on the front control panel.
Setting the Stirring Speed
Your MIRAK stirrer or stirring hot plate has an electronic feedback speed control which will maintain a
precise speed setpoint from 60 rpm through 1200 rpm at ±2 rpm. (Maximum speed is dependent on
viscosity of the solution.) The MIRAK stirrers and stirring hot plates are equipped with a strong magnet and
high torque motor which will draw a vortex in up to 1800 ml of water stirred in a 2 liter flask with a new 2" stir
bar.
To set the speed, spin the SPEED CONTROL knob until your desired setting is indicated on the digital
display. As you move the knob, the SET light next to the RPM display will illuminate. Turn the knob
clockwise to increase the speed or counterclockwise to decrease the speed. The unit will not start to stir
until you press the speed control knob. Once the unit begins stirring the display will register ACTUAL
speed and the SET light will no longer be illuminated. To check your setpoint, move the knob
counterclockwise one position. You will feel a click, the SET light will illuminate, and your setpoint will be
displayed again for five seconds. After five seconds the display will automatically return to indicating actual
speed and the SET light will go out. To stop the stirring action without changing your setpoint press the
knob once. To resume stirring at the set speed, press the knob again.
8
If you want to ensure that the unit will not start to stir if the knob is inadvertently pressed, turn the speed
control knob until the display indicates the word OFF.
Stirring Speed Guide
Amount of
UnitFlask SizeStirbar Size
7" x 7" MIRAK1 liter2"800 mlwater1200 rpm
7" x 7" MIRAK2 liter2"1800 mlwater1200 rpm
7" x 7" MIRAK1 liter2"1000 mlCooking oil700 rpm
7" x 7" MIRAK500 ml2" 350 ml30 weight
*Maximum speed obtained without the stir bar loosing coupling with the magnet.
The 12" X 12" MIRAK stirrers and stirring hot plates will produce stirring speed results similar to those
provided above for the 7" X 7" units. For speed stability and display accuracy please refer to the Operating
Specifications on page 13.
SolutionViscosity
motor oil
Maximum
Speed*
1100 rpm
Setting the Temperature
Your MIRAK hot plate or stirring hot plate has an electronic closed-loop feedback control which will
accurately maintain temperature setpoints from 40°C through 370°C or 540°C, depending on the model.
An unloaded hot plate will heat to maximum temperature in just 8 minutes. A HOT light on the front panel
will illuminate whenever the top surface temperature exceeds 50°C. The temperature may be controlled
either at the top plate by the internal Type K thermocouple sensor, or in the solution using an accessory
ungrounded Type K thermocouple probe. (A 6" or 10" stainless steel general purpose immersion probe,
or a chemical-resistant 7" Teflon® immersion probe are available through your distributor, or by contacting
Barnstead/Thermolyne, 800-553-0039). Your MIRAK hot plate, stirrer or stirring hot plate will accept any
ungrounded Type K thermocouple probe, however the accuracy of probes other than the ones we've
tested and specified may exceed the probe accuracy rating we've designated.
Your MIRAK hot plate or stirring hot plate will display the temperature in °C. The display will always indicate
'OFF' or your last setpoint, and the 'SET' light will be illuminated until you press the heat control knob to
initiate operation. Once the heat control knob has been pressed the ACTUAL temperature of the top
plate or the probe will be displayed, and the 'SET' light will no longer be illuminated. If you want to check
your setpoint, you do not have to press the knob again, simply turn the knob counterclockwise one click.
Your setpoint will be displayed for approximately five seconds until the display returns to indicating the
actual temperature.
NOTE
If you want to ensure that the unit will not heat if the knob is inadvertently pressed, turn the temperature
control knob until the display indicates the word 'OFF'.
The temperature display will not indicate a temperature less than 40°C. If the actual
temperature is less than 40°C, the word 'LO' will be indicated until the actual temperature
reaches 40°C.
9
Barnstead |Thermolyne • 800-553-0039
Loading...