Filling the water container
Although the mains voltage is isolated from exposed wiring in the device, it is
recommended that:
i) The water container / reservoir (1) is removed from the device for filling and that
the lid is always in place when the water container is in-situ on the device.
ii) The water container is not overfilled.
iii) The water used to fill or clean the water container does not exceed normal
body temperature.
Fitting the disposable “one use” jet tip applicators
The Propulse II Ear Irrigator is supplied with ten disposable “one use” jet tip applicators
(2) – use one disposable “one use” jet tip per patient.
To remove a jet tip applicator pull the jet tip applicator directly from the jet tip holder
body, without using a sideways movement.
To fit, or change, a jet tip applicator reverse the above procedure.
Disposable “one use” angled jet tips are available in boxes of 100 individually
wrapped tips from your normal Propulse supplier or from Mirage direct.
USING THE PROPULSE II EAR IRRIGATOR
The clinical criteria for use of the Propulse II should be noted before treatment
is commenced.
1) Follow the setting up procedure.
2) Make sure the Foot Switch plug is fully inserted into the foot switch socket (9).
3) Set the Pressure/Flow control knob (6) to it’s lowest setting (fully anti-clockwise).
4) Switch the Operating Switch (7) from to .
5) Press the Foot Switch (10).
6) Turn the Pressure/Flow control knob slowly clockwise from the MIN position until
the pressure/flow of the water is compatible with the treatment requirements and
patient comfort.
7) Follow the Cleaning Guidelines.
Page 7
c) the water jet tip holder (3) - is attached to the main body of the device by a
coiled tube. This tube is not detachable.
The water jet tip holder accommodates disposable “one use” jet tip applicators.
Only disposable “one use” jet tip applicators designed for the P
ropulse II Ear
Irrigator must be used. The manufacturer cannot accept responsibility for
damage or injury, however caused, through the use of non-approved jet
tip applicators.
The water jet tip holder incorporates a water jet stop button (5).
This button will stop the flow of water when it is pressed fully in.
The pressure/flow set by the pressure/flow control knob will be resumed upon
release of the water stop button, or by pressing or depressing the footswitch.
d) the foot switch (10) - attached to the main body by a detachable socket
connection. The device will only function if the foot switch is connected and
the operating switch is in the ON ( ) position.
Pressing the footswitch causes the device to produce water at the water jet tip
applicator at the pressure/flow set by the pressure/flow control knob.
SETTING - UP PROCEDURE
Initial cleaning
Although the device has been thoroughly cleaned after manufacture it is recommended
that it is cleaned and sterilised prior to use. (See Cleaning Guidelines).
The Propulse II Ear Irrigator must only be used with a mains power supply that is the
same as that indicated on the data panel on the base of the device.
The manufacturer cannot accept responsibility for damage or injury, however
sustained, from connecting the device to a non-compatible power supply.
Adapters should not be used when connecting the device to the power supply.
Caution: The product is not permanently disconnected from the mains in the
off ( ) position. Please remove the plug from mains when not in use to
prevent overheating.
Page 6