Mintek LV-T, LV User Manual

Stratos
Family of Cardiac Resynchronization
Therapy Pacemakers
Technical Manual
Stratos CRT-Ps
Implantable Cardiac Resynchronization Therapy Pacemakers
Stratos
Radiopaque Identification
A radiopaque identification code is visible on standard x-ray, and identifies the pulse generator:
Stratos LV/LV-T
SV
CAUTION
Lead / CRT-P Compatibility – Because of the numerous
available 3.2-mm configurations (e.g., the IS-1 and VS-1 standards), lead/ CRT-P compatibility should be confirmed with the CRT-P and/or lead manufacturer prior to the implantation of the system.
IS-1, wherever stated in this manual, refers to the international standard, whereby leads and generators from different manufacturers are assured a basic fit. [Reference ISO 5841-3:1992(E)].
©2008 BIOTRONIK, Inc., all rights reserved.
Stratos LV/LV-T Technical Manual i
Contents
1. General ..............................................................................1
1.1 Device Description ........................................................1
1.2 Indications .....................................................................2
1.3 Contraindications ..........................................................3
1.4 Note to Physician ..........................................................3
1.5 Warnings and Precautions ............................................3
1.5.1 Interactions with Other Medical Therapy...............4
1.5.2 Storage and Sterilization .......................................6
1.5.3 Lead Connection and Evaluation ..........................7
1.5.4 Programming and Operation.................................8
1.5.5 Home Monitoring ................................................. 11
1.5.6 Electromagnetic Interference (EMI) ....................13
1.5.7 Home and Occupational Environments...............13
1.5.8 Cellular Phones ................................................... 14
1.5.9 Hospital and Medical Environments....................15
1.5.10 Device Explant and Disposal............................... 15
1.6 Potential Effects of the Device on Health.................... 16
1.7 Clinical Studies............................................................17
1.7.1 Stratos LV Clinical Study – AVAIL CLS/CRT ......17
1.7.2 Stratos LV Clinical Study – OVID study ..............39
1.7.3 AVAIL and OVID Combined Primary
Endpoint-Complication-free Rate (Safety) ..........47
1.7.4 Tupos LV/ATx Clinical IDE Study -
OPTION CRT/ATx...............................................49
1.7.5 Conclusions Drawn from Studies ........................ 61
2. Programmable Parameters............................................63
2.1 Pacing Modes..............................................................63
2.1.1 Rate-adaptive Modes ..........................................63
2.1.2 DDD.....................................................................63
2.1.3 DDI ......................................................................67
2.1.4 DVI....................................................................... 67
2.1.5 VDD ..................................................................... 67
2.1.6 AAI and VVI .........................................................68
2.1.7 AAI, VVI ............................................................... 68
2.1.8 AOO, VOO...........................................................68
2.1.9 DOO ....................................................................68
2.1.10 VDI.......................................................................68
ii Stratos LV/LV-T Technical Manual
2.1.11 OFF (ODO) ..........................................................69
2.2 Biventricular Synchronization of the Stratos CRT-Ps..69
2.3 Timing Functions .........................................................70
2.3.1 Basic Rate ...........................................................70
2.3.2 Rate Hysteresis ...................................................71
2.3.3 Scan Hysteresis...................................................72
2.3.4 Repetitive Hysteresis...........................................74
2.3.5 Night Mode ..........................................................75
2.3.6 Refractory Periods...............................................76
2.3.7 Atrial PMT Protection...........................................77
2.3.8 Ventricular Refractory Period ..............................78
2.3.9 AV Delay..............................................................78
2.3.10 VES Discrimination after Atrial Sensed Events ...80
2.3.11 Sense Compensation ..........................................80
2.3.12 Ventricular Blanking Period .................................81
2.3.13 Safety AV Delay ..................................................81
2.4 Pacing and Sensing Functions....................................82
2.4.1 Pulse Amplitude and Pulse Width .......................82
2.4.2 Sensitivity ............................................................83
2.4.3 Lead Polarity........................................................83
2.5 Automatic Lead Check ................................................84
2.6 Antitachycardia Functions: ..........................................86
2.6.1 Upper Rate and UTR Response .........................86
2.7 Wenckebach 2:1..........................................................86
2.8 Mode Switching ...........................................................88
2.9 PMT Management....................................................... 89
2.9.1 Protection ............................................................89
2.9.2 PMT Detection.....................................................90
2.10 Adjustment of the PMT Protection Window.................91
2.11 Atrial Upper Rate.........................................................92
2.12 Preventive Overdrive Pacing (Overdrive Mode) .........92
2.13 AES Detection and Pacing..........................................94
2.13.1 AES Detection .....................................................94
2.13.2 Post AES Stimulation ..........................................95
2.14 Parameters for Rate-Adaptive Pacing ........................95
2.14.1 Rate-Adaptation...................................................95
2.14.2 Sensor Gain.........................................................96
2.14.3 Automatic Sensor Gain .......................................96
2.14.4 Sensor Threshold ................................................97
2.14.5 Rate Increase ......................................................97
Stratos LV/LV-T Technical Manual iii
2.14.6 Maximum Activity Rate ........................................ 98
2.14.7 Rate Decay ..........................................................98
2.15 Sensor Stimulation ......................................................98
2.16 Rate Fading................................................................. 99
2.17 Home Monitoring (Stratos LV-T)................................100
2.17.1 Transmission of Information ..............................102
2.17.2 Patient Device ...................................................103
2.17.3 Transmitting Data ..............................................103
2.17.4 Types of Report Transmissions ........................105
2.17.5 Description of Transmitted Data........................107
2.18 Statistics ....................................................................109
2.18.1 Timing ................................................................109
2.18.2 Arrhythmia .........................................................109
2.18.3 Sensor ...............................................................109
2.18.4 Sensing..............................................................109
2.18.5 Pacing................................................................109
2.18.6 General Statistical Information ..........................110
2.19 Interrogating and/or Starting Statistics ......................110
2.20 Timing Statistics ........................................................ 111
2.20.1 Event Counter.................................................... 111
2.20.2 Event Episodes.................................................. 111
2.20.3 Rate Trend.........................................................112
2.20.4 Atrial and Ventricular Rate Histogram ...............112
2.21 Arrhythmia Statistics.................................................. 113
2.21.1 Tachy Episode Trend ........................................113
2.21.2 AF Classification................................................113
2.21.3 AES Statistics ....................................................114
2.21.4 VES Statistics ....................................................115
2.22 Sensor Statistics........................................................ 116
2.22.1 Sensor Rate Histogram .....................................116
2.22.2 Activity Report ...................................................117
2.22.3 Sensor Optimization ..........................................117
2.23 Sensing Statistics ...................................................... 117
2.24 Pacing Statistics ........................................................ 118
3. Follow-up Procedures.................................................. 119
3.1 General Considerations ............................................119
4. Real-Time IEGM ............................................................121
4.1 IEGM Recordings......................................................121
iv Stratos LV/LV-T Technical Manual
5. Battery, Pulse and Lead Data ......................................123
5.1 Threshold Test - Testing the Pacing Function ...........123
5.2 P/R Measurement - Testing the Sensing Function....124
5.3 Testing for Retrograde Conduction ...........................125
5.4 Non-Invasive Programmed Stimulation (NIPS).........125
5.4.1 Description.........................................................125
5.4.2 Burst Stimulation ...............................................126
5.4.3 Programmed Stimulation...................................126
5.4.4 Back up Pacing..................................................126
5.4.5 NIPS Safety Features........................................127
6. Other Functions/Features............................................129
6.1 Temporary Programming...........................................129
6.2 Patient Data Memory.................................................130
6.3 Safe Program Settings ..............................................130
6.4 Magnet Effect ............................................................131
6.5 Position Indicator.......................................................131
6.6 Pacing When Exposed to Interference .....................132
7. Product Storage and Handling....................................135
7.1 Sterilization and Storage ...........................................135
7.2 Opening the Sterile Container...................................136
7.3 Pulse Generator Orientation .....................................138
8. Lead Connection ..........................................................139
8.1 Lead Configuration....................................................139
8.2 Lead Connection .......................................................140
9. Elective Replacement Indication (ERI).......................145
10. Explantation ..................................................................149
11. Technical Data...............................................................151
11.1 Available Pacing Modes............................................151
11.2 Pulse- and Control Parameters................................. 151
11.3 Diagnostic Memory Functions...................................156
11.4 Home Monitoring (Stratos LV-T)................................156
11.5 Additional Functions..................................................157
11.6 Programmers.............................................................158
11.7 Default Programs ......................................................159
11.8 Materials in Contact with Human Tissue...................159
11.9 Electrical Data/Battery...............................................159
11.10 Mechanical Data ..................................................160
Stratos LV/LV-T Technical Manual v
12. Order Information.........................................................161
Appendix A ..........................................................................163
vi Stratos LV/LV-T Technical Manual Stratos LV/LV-T Technical Manual 1
1. General
1.1 Device Description
The Stratos LV and Stratos LV-T CRT-Ps are rate adaptive pacemakers designed to provide Cardiac Resynchronization Therapy (CRT). The Stratos CRT-Ps provide all standard bradycardia pacemaker therapy with the additional capabilities of biventricular pacing for CRT. Biventricular pacing in the Stratos CRT-Ps can be programmed to initially pace in either the right or left ventricular chambers with separately programmable outputs for both left and right channels. Sensing of cardiac signals only occurs in the right ventricular chamber.
The Stratos CRT-Ps can also provide single and dual chamber pacing in a variety of rate-adaptive and non-rate adaptive pacing modes. Pacing capability is supported by an extensive diagnostic set. For motion-based rate-adaptation, the Stratos CRT-Ps are equipped with an internal accelerometer. This sensor produces an electric signal during physical activity of the patient. If a rate-adaptive (R) mode is programmed, then the accelerometer sensor signal controls the stimulation rate.
The Stratos LV-T additionally also employs BIOTRONIK’s Home Monitoring™ technology, which is an automatic, wireless, remote monitoring system for management of patients with implantable devices. With Home Monitoring, physicians can review data about the patient’s cardiac status and CRT-P’s functionality between regular follow-up visits, allowing the physician to optimize the therapy process. Stratos CRT-Ps are also designed to collect diagnostic data to aid the physician’s assessment of a patient’s condition and the performance of the implanted device.
The bipolar IS-1 connections are used for pacing and sensing (right atrial and ventricle) and the additional IS-1 connection is used for pacing in the left ventricle in either a bipolar or unipolar configuration depending on the left ventricular lead. The pulse amplitude and pulse width of each of the three channels is separately programmable.
2 Stratos LV/LV-T Technical Manual
Stratos CRT-Ps are designed to meet all indications for Cardiac Resynchronization Therapy in CHF patients as well as those for bradycardia therapy as exhibited in a wide variety of patients. The Stratos family is comprised of two CRT-Ps that are designed to handle a multitude of situations.
Stratos LV
Stratos LV-T
Throughout this manual, specific feature and function descriptions may only be applicable to the Stratos LV-T and those features will be referenced as such. Otherwise, reference to Stratos CRT-Ps refers to both devices.
Triple chamber, rate-adaptive, unipolar/bipolar pacing CRT-P
Triple chamber, rate-adaptive, unipolar/bipolar pacing CRT-P with Home Monitoring
1.2 Indications
The Stratos LV and Stratos LV-T Cardiac Resynchronization Therapy Pacemakers (CRT-Ps) are indicated for patients who have moderate to severe heart failure (NYHA Class III/IV), including left ventricular dysfunction (EF 35%) and QRS 120 ms and remain symptomatic despite stable, optimal heart failure drug therapy.
Stratos LV/LV-T Technical Manual 3
1.3 Contraindications
Use of Stratos LV and Stratos LV-T CRT-Ps are contraindicated for the following patients:
Unipolar pacing is contraindicated for patients with an
implanted cardioverter-defibrillator (ICD) because it may cause unwanted delivery or inhibition of ICD therapy.
Single chamber atrial pacing is contraindicated for
patients with impaired AV nodal conduction.
Dual chamber and single chamber atrial pacing is
contraindicated for patients with chronic refractory atrial tachyarrhythmias.
1.4 Note to Physician
As with any implantable pulse generator, there are certain infrequent risks associated with Stratos CRT-Ps. Section 1.6 lists the adverse events that have been observed or may potentially occur with these Cardiac Resynchronization Therapy Pacemakers. The warnings and precautions listed in Section 1.5 should be taken under serious consideration in order to aid in avoiding device failures and harm to the patient.
Regular monitoring of the patient and their implanted device should be conducted to identify performance concerns and ensure appropriate therapy is being administered to the patient. Please communicate any performance concerns to BIOTRONIK and to FDA.
All explanted devices should be returned to the manufacturer for testing to help understand device reliability and performance. Refer to Section 10 for recommended procedures for handling explanted devices.
1.5 Warnings and Precautions
Certain therapeutic and diagnostic procedures may cause undetected damage to a Cardiac Resynchronization Therapy Pacemakers, resulting in malfunction or failure at a later time. Please note the following warnings and precautions:
4 Stratos LV/LV-T Technical Manual
Magnetic Resonance Imaging (MRI) – Avoid use of magnetic resonance imaging as it has been shown to cause movement of the CRT-Ps within the subcutaneous pocket and may cause pain and injury to the patient and damage to the CRT-P. If the procedure must be used, constant monitoring is recommended, including monitoring the peripheral pulse.
Rate Adaptive Pacing – Use rate adaptive pacing with care in patients unable to tolerate increased pacing rates.
NIPS – Life threatening ventricular arrhythmias can be induced by stimulation in the ventricle. Ensure that an external cardiac defibrillator is accessible during tachycardia testing. Only physicians trained and experienced in tachycardia induction and reversion protocols should use non-invasive programmed stimulation (NIPS).
High Output Settings – High output settings combined with extremely low lead impedance may reduce the life expectancy of the Stratos CRT-Ps. Programming of pulse amplitudes, higher than 4.8 V, in combination with long pulse widths and/or high pacing rates may lead to premature activation of the replacement indicator.
1.5.1 Interactions with Other Medical Therapy
Before applying one of the following procedures, a detailed analysis of the advantages and risks should be made. Cardiac activity during one of these procedures should be confirmed by continuous monitoring of peripheral pulse or blood pressure. Following the procedures, CRT-P function and stimulation threshold must be checked.
Therapeutic Diathermy Equipment Use of therapeutic diathermy equipment is to be avoided for pacemaker patients due to possible heating effects of the CRT-P and at the implant site. If diathermy therapy must be used, it should not be applied in the immediate vicinity of the CRT-P or leads. The patient's peripheral pulse should be monitored continuously during the treatment.
Stratos LV/LV-T Technical Manual 5
Transcutaneous Electrical Nerve Stimulation (TENS) –
Transcutaneous electrical nerve stimulation may interfere with CRT-P function. If necessary, the following measures may reduce the possibility of interference:
Place the TENS electrodes as close to each other as
possible.
Place the TENS electrodes as far from the CRT-P/lead
system as possible.
Monitor cardiac activity during TENS use.
Defibrillation – The following precautions are recommended to minimize the inherent risk of CRT-P operation being adversely affected by defibrillation:
The paddles should be placed anterior-posterior or along
a line perpendicular to the axis formed by the CRT-P and the implanted lead.
The energy setting should not be higher than required to
achieve defibrillation.
The distance between the paddles and the CRT-P/leads
should not be less than 10 cm (4 inches).
Radiation – The CRT-P’s internal electronics may be damaged by exposure to radiation during radiotherapy. To minimize this risk when using such therapy, the CRT-P should be protected with local radiation shielding.
Lithotripsy – Lithotripsy treatment should be avoided for CRT-P patients since electrical and/or mechanical interference with the CRT-P is possible. If this procedure must be used, the greatest possible distance from the point of electrical and mechanical strain should be chosen in order to minimize a potential interference with the CRT-P.
Electrocautery – Electrocautery should never be performed within 15 cm (6 inches) of an implanted CRT-P or leads because of the danger of introducing fibrillatory currents into the heart and/or damaging the CRT-P. Pacing should be asynchronous and above the patient’s intrinsic rate to prevent inhibition by interference signals generated by the cautery. When possible, a bipolar electrocautery system should be used.
6 Stratos LV/LV-T Technical Manual
For transurethral resection of the prostate, it is recommended that the cautery ground plate be placed under the buttocks or around the thigh, but not in the thoracic area where the current pathway could pass through or near the CRT-P system.
1.5.2 Storage and Sterilization
Storage (temperature) – Recommended storage temperature
range is 5° to 55°C (41°-131°F). Exposure to temperatures outside this range may result in CRT-P malfunction (see Section 7.1).
Low Temperatures – Exposure to low temperatures (below 0°C) may cause a false elective replacement indication to be present. If this occurs, warm the device to room temperature and reset the ERI with magnet application (see Section 7.1).
Handling – Do not drop. If an unpackaged CRT-P is dropped onto a hard surface, return it to BIOTRONIK (see Section 7.1).
FOR SINGLE USE ONLY - Do not re-sterilize the CRT-P or accessories packaged with the CRT-P, they are intended for one-time use.
Device Packaging – Do not use the device if the packaging is wet, punctured, opened or damaged because the integrity of the sterile packaging may be compromised. Return the device to BIOTRONIK.
Storage (magnets) – Store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic interference (EMI) to avoid damage to the device.
Temperature Stabilization – Allow the device to reach room temperature before programming or implanting the device. Temperature extremes may affect the initial device function.
Use Before Date Do not implant the device after the USE BEFORE DATE because the device may have reduced longevity.
Stratos LV/LV-T Technical Manual 7
1.5.3 Lead Connection and Evaluation
Lead Check –
Feature Description
: Lead Check is a feature that, when activated, automatically measures the lead impedance with every pace. Based on these measurements, the lead configuration will be set to either unipolar or bipolar. Refer to Section 2.5 for more details regarding this feature.
Caution
: Lead check will not lead to disabling of cardiac resynchronization therapy. It limits the use of the resynchronization features.
1. Lead check is possible only when the right ventricle is paced first.
2. Lead check works only when the pacing voltages are programmed between 2.4 and 4.8 V. The lead check feature can be programmed OFF in patients that require cardiac resynchronization therapy.
Care should be taken when programming Stratos CRT-Ps with Lead Check ON as the device may switch from bipolar to unipolar pacing and sensing without warning. This situation may be inappropriate when using a Stratos CRT-P for patients with an Implantable Cardioverter Defibrillator (ICD). The following associated message appears when programming this feature:
“Lead check may result in a switch to unipolar pacing and sensing, which may be inappropriate for patients with an ICD.”
Additionally, Lead Check should be programmed OFF before lead connection as the feature will automatically reprogram the device to unipolar in the absence of a lead.
Lead / CRT-P Compatibility – Because of the numerous available 3.2-mm configurations (e.g., the IS-1 and VS-1 standards), lead/ CRT-P compatibility should be confirmed with the CRT-P and/or lead manufacturer prior to the implantation of the system.
8 Stratos LV/LV-T Technical Manual
IS-1, wherever stated in this manual, refers to the international standard, whereby leads and generators from different manufacturers are assured a basic fit. [Reference ISO 5841­3:1992(E)].
Lead Configuration – The polarity of the implanted lead dictates what lead configuration can be programmed for the CRT-P. Pacing will not occur with a unipolar lead if the lead configuration of the respective channel is programmed to bipolar (see Section 8).
Setscrew Adjustment – Back-off the setscrew(s) prior to insertion of lead connector(s) as failure to do so may result in damage to the lead(s), and/or difficulty connecting lead(s).
Cross Threading Setscrew(s) – To prevent cross threading the setscrew(s), do not back the setscrew(s) completely out of the threaded hole. Leave the torque wrench in the slot of the setscrew(s) while the lead is inserted.
Tightening Setscrew(s) – Do not overtighten the setscrew(s). Use only the BIOTRONIK supplied torque wrench.
Sealing System – Be sure to properly insert the torque wrench into the perforation at an angle perpendicular to the
connector receptacle. Failure to do so may result in damage to the plug and its self-sealing properties.
1.5.4 Programming and Operation
IEGM – Due to the compression processes that the signals
undergo, the IEGM recordings are not suitable for making some specific cardiac diagnoses, such as ischemia; although, these tracings may be useful in diagnosing arrhythmias, device behavior or programming issues.
Post AES - Before activating post-AES, check whether the selected program can cause Pacemaker Mediated Tachycardia (PMT) and whether post-AES pacing results.
Overdrive Pacing Mode - When programming the overdrive pacing mode, check whether the selected program can cause PMT, and whether atrial over drive pacing would result. Corresponding to the measured retrograde conduction time, the PMT protection interval must be programmed to a correct value.
Stratos LV/LV-T Technical Manual 9
AV Hysteresis – If the AV hysteresis is enabled along with the algorithm for recognizing and terminating PMTs (PMT management), the AV delay for recognizing and terminating a PMT has a higher priority than the AV hysteresis.
Sensing – The Stratos CRT-Ps do not sense in the left ventricle.
AV Conduction – In patients with intact AV conduction, the
intrinsic atrial tachycardia is conducted to the ventricle 1:1. With the resynchronization mode activated, spontaneous rate of the right ventricle mode is synchronized for a rate up to 200 ppm in the left ventricle. For this reason, biventricular pacing mode should be turned OFF in such cases.
Unipolar/Bipolar – If the pacing or sensing function is to be programmed to bipolar in the atrial channel, it must be verified that bipolar leads have been implanted in that chamber. If the atrial lead is unipolar, unipolar sensing and pacing functions must be programmed in that chamber. Failure to program the appropriate lead configuration could result in patient experiencing entrance and/or exit block.
In addition, if the atrial lead polarity setting within the Patient Data Memory has been set to bipolar, the polarity of the corresponding implanted lead must be confirmed to be bipolar.
Safe Program – Activating the “Safe Program” is a way of quickly programming the device to multiple settings in the event of an emergency. These settings include unipolar pacing with pacing output OFF in the left ventricular channel. Refer to Section 6.3 for further details.
Programmers – Use only BIOTRONIK’s ICS 3000 programmer equipped with appropriate software to program Stratos CRT-Ps. Do not use programmers from other manufacturers.
Pulse Amplitude – Programming of pulse amplitudes, higher than 4.8 V, in combination with long pulse widths and/or high pacing rates can lead to premature activation of the replacement indicator. If a pulse amplitude of 7.2 V or higher is programmed and high pacing rates are reached, output amplitudes may differ from programmed values.
10 Stratos LV/LV-T Technical Manual
Pacing thresholds – When decreasing programmed output (pulse amplitude and/or pulse width), the pacing threshold must first be accurately assessed to provide a 2:1 safety margin.
EMI – Computerized systems are subject to (Electromagnetic Interference (EMI) or “noise”. In the presence of such interference, telemetry communication may be interrupted and prevent programming of the Stratos CRT-P.
Programming Modifications – Extreme programming changes should only be made after careful clinical assessment. Clinical judgment should be used when programming permanent pacing rates below 40 ppm or above 100 ppm.
Short Pacing Intervals – Use of short pacing intervals (high pacing rates) with long atrial and/or ventricular refractory periods may result in intermittent asynchronous pacing and, therefore, may be contraindicated in some patients.
OFF Mode – The OFF mode can be transmitted as a temporary program only to permit evaluation of the patient’s spontaneous rhythm. (see Section 2.1.11)
Myopotential Sensing – The filter characteristics of BIOTRONIK implantable devices have been optimized to sense electrical potentials generated by cardiac activity and to reduce the possibility of sensing skeletal myopotentials. However, the risk of CRT-P’s operation being affected by myopotentials cannot be eliminated, particularly in unipolar systems. Myopotentials may resemble cardiac activity, resulting in inhibition of pacing, triggering and/or emission of asynchronous pacing pulses, depending on the pacing mode and the interference pattern. Certain follow-up procedures, such as monitoring CRT-P performance while the patient is doing exercises involving the use of pectoral muscles, as well as Holter monitoring, have been recommended to check for interference caused by myopotentials. If sensing of myopotentials is encountered, corrective actions may include selection of a different pacing mode or sensitivity setting.
Muscle or Nerve Stimulation – Inappropriate muscle or nerve stimulation may occur with unipolar pacing when using a non­coated Stratos CRT-P.
Stratos LV/LV-T Technical Manual 11
Atrial Sensitivity – In dual chamber systems, the atrial sensitivity of 0.1 mV should only be programmed in conjunction with a bipolar lead configuration.
Programmed to Triggered Modes – When programmed to triggered modes, pacing rates up to the programmed upper limit may occur in the presence of either muscle or external interference.
Triggered Modes – While the triggered modes (DDT, DVT, DDTR/A, DDTR/V, DDI/T, VDT, VVT, and AAT) can be programmed permanently, these modes are intended for use as temporary programming for diagnostic purposes. In triggered pacing modes, pacing pulses are emitted in response to sensed signals, and therefore the pacing pulse can be used as an indicator, or marker of sensed events for evaluating the sensing function of the pulse generator using surface ECG. However, real-time telemetry of marker channels and/or intracardiac electrogram via the programmer and programming wand is recommended over the use of a triggered pacing mode in the clinical setting. A triggered pacing mode may be preferred in situations where positioning the programming head over the pulse generator would be impossible or impractical (i.e., during exercise testing or extended Holter monitoring).
Another possible application of triggered modes is to ensure pacing as a short term solution during a period of inhibition of pacing by extracardiac interference, mechanical noise signals, or other sensing abnormalities. Because triggered modes emit pacing pulses in response to sensed events, this may result in unnecessary pacing during the absolute refractory period of the myocardium, inappropriate pacing in response to oversensing of cardiac or extracardiac signals. The risks associated with triggered pacing include excessive pacing, arrhythmias due to the R-on-T phenomenon, and early battery depletion. Therefore, it is important that the triggered modes are not used for long term therapy, and that the CRT-P is always returned to a non­triggered permanent program.
1.5.5 Home Monitoring
Patient’s Ability - Use of the Home Monitoring system requires
the patient and/or caregiver to follow the system instructions and cooperate fully when transmitting data.
12 Stratos LV/LV-T Technical Manual
If the patient cannot understand or follow the instructions because of physical or mental challenges, another adult who can follow the instructions will be necessary for proper transmission.
Electromagnetic Interference (EMI) – Precautions for EMI interference with the Stratos CRT-Ps are provided in Section 1.5.6. Sources of EMI including cellular telephones, electronic article surveillance systems, and others are discussed therein.
Use in Cellular Phone Restricted Areas - The mobile patient device (transmitter/receiver) should not be utilized in areas where cellular phones are restricted or prohibited (i.e., commercial aircraft).
Event Triggered Report - A timely receipt of the event report cannot be guaranteed. The receipt is also dependent on whether the patient was physically situated in the required coverage range of the patient device at the time the event information was sent.
Patient-Activated Report - The magnet effect must be programmed “synchronous” if the [Patient Report] function is activated.
Not for Conclusive Diagnosis - Because not all information available in the implant is being transmitted, the data transmitted by Home Monitoring should be evaluated in conjunction with other clinical indicators (i.e., in-office follow-up, patient symptoms, etc.) in order to make a proper diagnosis.
Frequency of Office Follow-Ups When Using Home Monitoring - The use of Home Monitoring does not replace
regular follow-up examinations. When using Home Monitoring, the time period between follow-up visits may not be extended.
Stratos LV/LV-T Technical Manual 13
1.5.6 Electromagnetic Interference (EMI)
The operation of any implanted device may be affected by certain environmental sources generating signals that resemble cardiac activity. This may result in inhibition of pacing and/or triggering or in asynchronous pacing depending on the pacing mode and the interference pattern. In some cases (i.e., diagnostic or therapeutic medical procedures), the interference sources may couple sufficient energy into a pacing system to damage the device and/or cardiac tissue adjacent to the leads.
BIOTRONIK CRT-Ps have been designed to significantly reduce susceptibility to electromagnetic interference (EMI). However, due to the variety and complexity of sources creating interference, there is no absolute protection against EMI. Generally, it is assumed that EMI produces only minor effects, if any, in CRT-P patients. If the patient may be exposed to one of the following environmental conditions, then the patient should be given the appropriate warnings.
1.5.7 Home and Occupational Environments
The following equipment (and similar devices) may affect normal CRT-P operation: electric arc welders, electric melting furnaces, radio/television and radar transmitters, power-generating facilities, high-voltage transmission lines, electrical ignition systems (also of gasoline-powered devices) if protective hoods, shrouds, etc., are removed, electrical tools, anti-theft devices at retail stores and electrical appliances, if not in proper condition or not correctly grounded and encased.
Patients should exercise reasonable caution in avoidance of devices which generate a strong electric or magnetic field. If EMI inhibits pacing or causes a reversion to asynchronous pacing or pacing at magnet rate, moving away from the source or turning it off should allow the CRT-P to return to its normal mode of operation. Some potential EMI sources include:
High Voltage Power Transmission Lines – High voltage power transmission lines may generate enough EMI to interfere with CRT-P operation if approached too closely.
14 Stratos LV/LV-T Technical Manual
Home Appliances – Home appliances normally do not affect CRT-P operation if the appliances are in proper condition and correctly grounded and encased. There are reports of CRT-P disturbances caused by electrical tools and by electric razors that have touched the skin directly over the CRT-P.
Communication Equipment – Communication equipment such as microwave transmitters, linear power amplifiers, or high­power amateur transmitters may generate enough EMI to interfere with CRT-P operation if approached too closely.
Commercial Electrical Equipment – Commercial electrical equipment such as arc welders, induction furnaces, or resistance welders may generate enough EMI to interfere with CRT-P operation if approached too closely.
Electrical Appliances – Electric hand-tools and electric razors (used over the skin directly above the CRT-P) have been reported to cause pacemaker disturbances. Home appliances that are in good working order and properly grounded do not usually produce enough EMI to interfere with implanted device operation.
Electronic Article Surveillance (EAS) – Equipment such as retail theft prevention systems may interact with the CRT-Ps. Patients should be advised to walk directly through and not to remain near an EAS system longer than necessary.
Radio-Frequency Identification (RFID) – RFID tags may interact with the CRT-Ps. Patients should be advised to avoid leaving a device containing such a tag within close proximity to the CRT-P (i.e., inside a shirt pocket).
1.5.8 Cellular Phones
Recent studies have indicated there may be a potential interaction between cellular phones and pacemaker operation. Potential effects may be due to either the radio frequency signal or the magnet within the phone and could include inhibition or asynchronous pacing when the phone is within close proximity (within 6 inches [15 cm]) to the CRT-P.
Stratos LV/LV-T Technical Manual 15
Based on testing to date, effects resulting from an interaction between cellular phones and the implanted pacemakers have been temporary. Simply moving the phone away from the implanted device will return it to its previous state of operation. Because of the great variety of cellular phones and the wide variance in patient physiology, an absolute recommendation to cover all patients cannot be made.
Patients having an implanted CRT-P who operate a cellular phone should:
Maintain a minimum separation of 6 inches (15 cm) between a hand-held personal cellular phone and the implanted device. Portable and mobile cellular phones generally transmit at higher power levels compared to hand held models. For phones transmitting above 3 watts, maintain a minimum separation of 12 inches (30 cm) between the antenna and the implanted device.
Patients should hold the phone to the ear opposite the side of the implanted device. Patients should not carry the phone in a breast pocket or on a belt over or within 6 inches (15 cm) of the implanted device as some phones emit signals when they are turned ON but not in use (i.e., in the listen or standby mode). Store the phone in a location opposite the side of implant.
1.5.9 Hospital and Medical Environments
Refer to Section 1.5.1 for information regarding CRT-P interaction with the following medical procedures / environments:
Electrosurgical Cautery
Lithotripsy
External Defibrillation
High Radiation Sources
1.5.10 Device Explant and Disposal
Device Incineration - Never incinerate a CRT-P. Be sure the
CRT-P is explanted before a patient who has died is cremated. (see Section 10)
Explanted Devices – Return all explanted devices to BIOTRONIK.
16 Stratos LV/LV-T Technical Manual
1.6 Potential Effects of the Device on Health
The following possible adverse events may occur with this type of CRT-P based on implant experience including:
Potential Adverse Events
Air embolism
Allergic reactions to
contrast media
Arrhythmias
Bleeding
Body rejection
phenomena
Cardiac tamponade
Chronic nerve damage
Damage to heart valves
Elevated pacing
thresholds
Extrusion
Fluid accumulation
Infection
Keloid formation
Lead dislodgment
Lead fracture / insulation
damage
Lead-related thrombosis
Local tissue reaction / fibrotic tissue formation
Muscle or nerve stimulation
Myocardial damage
Myopotential sensing
Pacemaker mediated
tachycardia
Pneumothorax
Pocket erosion
Hematoma
Device migration
Thromboembolism
Undersensing of
intrinsic signals
Venous occlusion
Venous or cardiac
perforation
Stratos LV/LV-T Technical Manual 17
1.7 Clinical Studies
The subsequent sections summarize the following three clinical studies that were used to support the safety and effectiveness of the Stratos LV/LV-T CRT-Ps.
The AVAIL CLS/CRT clinical study
The OVID clinical study (OUS)
The OPTION CRT/ATx clinical study
Two of the studies, AVAIL CLS/CRT and OVID, collected significant safety data supporting use of the Stratos LV/LV-T CRT-P system. The third study, OPTION CRT/ATx, supports the effectiveness of cardiac resynchronization therapy (CRT). The OPTION CRT/ATx study was conducted on a device that delivers CRT but, in addition, also offers defibrillation therapy (CRT-D).
1.7.1 Stratos LV Clinical Study – AVAIL CLS/CRT
Study Design
The AVAIL CLS/CRT was a multi-center, prospective, randomized, blinded clinical study designed to support approval for cardiac resynchronization therapy for a Heart Failure (HF) patient population not requiring back up defibrillation and that are indicated for an ablate and pace procedures. All patients enrolled into the clinical study were randomly assigned to one of three groups using a 2:2:1 ratio for randomization.
18 Stratos LV/LV-T Technical Manual
Patients assigned to Group 1 received biventricular pacing with CLS-based rate adaptive pacing using BIOTRONIK’s Protos DR/CLS, which is a dual-chamber pulse generator with CLS-based rate adaptive pacing. During this study, the Protos DR/CLS devices were implanted with two ventricular leads: the right ventricular lead was connected to the ventricular port, and the left ventricular lead was connected to the atrial port. Protos DR/CLS was included in this study to evaluate biventricular pacing with a different type of rate adaptive sensor technology.
Patients assigned to Group 2 received biventricular pacing with accelerometer-based rate adaptive pacing using the Stratos LV.
Patients assigned to Group 3 (control group) received right ventricular pacing with accelerometer-based rate adaptive pacing using the Stratos LV. Therefore, 60% of the patients received a Stratos LV device.
Primarily, the study evaluated and compared the functional benefits of CRT between the three randomized groups using a composite endpoint consisting of a six-minute walk test (meters walked) and quality of life measurement (assessed using the Minnesota Living with Heart Failure Questionnaire). Relevant measurements were completed twice for each patient: once at the Baseline evaluation (prior to implant and ablation) and again at a six-month follow-up evaluation. The data collected during this clinical study was used to demonstrate superiority of CRT to RV only pacing. This study also evaluated the safety of both the Protos DR/CLS and Stratos LV devices through an analysis of the complication-free rate through six months. Secondarily, the study also evaluated the superiority of CRT with CLS rate adaptation compared to CRT with accelerometer rate adaptation.
Stratos LV/LV-T Technical Manual 19
Clinical Inclusion Criteria
To support the objectives of this investigation, patients were required to meet the following inclusion criteria prior to enrollment:
Meet the indications for therapy
Persistent (documented for more than 7 days),
symptomatic AF with poorly controlled rapid ventricular rates or permanent, (documented for more than 30 days with failed cardioversion, or longstanding AF of 6 months or more) symptomatic AF with poorly controlled rapid ventricular rates.
Eligible for AV nodal ablation and permanent pacemaker implantation
NYHA Class II or III heart failure
Age 18 years
Understand the nature of the procedure
Ability to tolerate the surgical procedure required for
implantation
Give informed consent
Able to complete all testing required by the clinical
protocol
Available for follow-up visits on a regular basis at the investigational site
20 Stratos LV/LV-T Technical Manual
Clinical Exclusion Criteria
To support the objectives of this investigation, the exclusion criteria at the time of patient enrollment included the following:
Meet one or more of the contraindications
Have a life expectancy of less than six months
Expected to receive heart transplantation within six
months
Enrolled in another cardiovascular or pharmacological clinical investigation
Patients with an ICD, or being considered for an ICD
Patients with previously implanted biventricular pacing
systems
Patients with previously implanted single or dual chamber pacing system with > 50% documented ventricular pacing
Patients with previous AV node ablation
Six-minute walk test distance greater than 450 meters
Any condition preventing the patient from being able to
perform required testing
Presence of another life-threatening, underlying illness separate from their cardiac disorder
Conditions that prohibit placement of any of the lead systems
Follow-Up Schedule
At the enrollment screening, the physician evaluated the patient to verify that all inclusion/exclusion criteria have been met in accordance to the protocol and the patient has signed the informed consent. After successful enrollment, all patients were implanted with either a Stratos LV CRT-P or Protos DR/CLS device. Evaluations at the Four Week, Three and Six Month follow-ups included NYHA classification, medications, and percentage of ventricular pacing.
Clinical Endpoints
Stratos LV/LV-T Technical Manual 21
Primary Endpoint: Complication-free Rate (Safety) The safety of the Stratos LV was evaluated based on complications (adverse events that require additional invasive intervention to resolve) related to the implanted CRT system which includes the Stratos LV, the right ventricular, the left ventricular lead, lead ventricular lead adapters (if used) and the implant procedure. The target complication-free rate at six months is 85%.
Primary Endpoint: Six Minute Walk Test & QOL (Effectiveness) The purpose of Primary Endpoint 1 was to evaluate the effectiveness of the CRT (Groups 1 and 2) compared to RV only (Group 3) pacing as measured by the average composite rate of improvement in six minute walk test and QOL.
Accountability of PMA Cohorts
After randomization and enrollment, 23 patients (8 in Group 1, 8 in Group 2 and 7 in Group 3) did not receive an implant. The reasons for patients not receiving an implant are outlined in
Figure 1
. Two additional patients in Group 1 had an unsuccessful first implant attempt (unable to implant the LV lead), but follow up data was not received.
22 Stratos LV/LV-T Technical Manual
Enrolled and Randomized Patients
Group 1 43 Group 2 50 Group 3 25
Implant Attempted
Group 1 39 Group 2 44 Group 3 21
Successful implant
Group 1 33 Group 2 42 Group 3 18
Completed 6-Month Follow-up
Group 1 23 Group 2 30 Group 3 15
No im pla nt Atte mpte d
Withdrawal of Consent
Not Meeting Inclusion Criteria
Unsuccessful implant
Withdrawal of IC before 2nd Attempt
Follow-up to failed implant data pending
Ablation/Abla tion Data Pending
6-Month Fol low-up Da ta
Patient Death before 6-Month
Withdrew before 6-Month FU
Not Reached 6-Month FU or Data Pending
6-month FU Not Completed
Figure 1: Patient Accountability
Demographics and Baseline Parameters
Group 1 2 Group 2 4 Group 3 3
Group 1 2 Group 2 2 Group 3 1
Group 1 4 Group 2 2 Group 3 3
Group 1 2 Group 2 0 Group 3 0
Group 1 3 Group 2 0 Group 3 0
Group 1 0 Group 2 2 Group 3 0
Group 1 1 Group 2 1 Group 3 0
Group 1 6 Group 2 8 Group 3 3
Group 1 0 Group 2 1 Group 3 0
Table 1
provides a summary of the patient demographics at enrollment. There were no statistical differences in enrollment demographics between the 3 groups.
Loading...
+ 142 hidden pages