Thank you for purchasing the MINOLTA FLASH METER VI.
The Minolta Flash Meter VI has the following features:
● The integrated exposure meter combines incident light measurement
and spot (reflected light) measurement in a single unit.
● For spot measurement, the Flash Meter VI uses a parallax-free optical
system. This eliminates the displacement of the measurement area that
varies with the distance from the subject.
● With the latitude display function, the Flash Meter VI can simultaneously
display the results of both incident light measurement and spot light
measurement. It provides a clear and simple graphical decision process
for determining the exposure suited to the nature of the photograph.
● With the analyze scale, you can determine the proportion of flash light
and ambient light in a single flash light measurement.
● The Flash Meter VI provides a memory function capable of storing up to
10 measured values; an averaging function that calculates an average
exposure from stored measurement data; and a brightness difference
function that displays deviation from the standard exposure.
● For spot measurement, both shadow-based and highlight-based
exposure calculation functions are provided.
● The Flash Meter VI provides a "custom setting (Alt) mode" that allows
you to customize the meter according to your preference. This feature
includes an exposure correction value setting function and a shutter
speed increment-setting function.
● Measurement results are shown on both the analog and digital displays
on the meter’s data panel. The clear and legible display eliminates
reading errors.
● The results of spot measurements are shown on the digital display in the
viewfinder and on the external data panel. The viewfinder features a
dioptric adjustment mechanism.
● In addition to displaying a conventional 10-level intermediate f-number
display, the Flash Meter VI provides an f-number direct reading display.
This enables the measured value to be applied to any camera with an fnumber direct reading display, eliminating the need for f-number
conversion.
Safety-related Icons
The following icons are used in this manual to alert you to important
information for preventing accidents due to improper handling of
equipment.
This denotes a safety-related caution. Read the caution carefully
to ensure safe use of the product.
This denotes actions to be strictly avoided.
Make sure to avoid these actions.
This denotes actions to be avoided.
Do not attempt to disassemble the product.
Safety Warnings and Cautions
To ensure proper use of the instrument, take special care to observe the
following handling instructions when using this instrument. Read this
instruction manual carefully and keep it securely in a place where you can
refer to it readily.
indicates a danger that improper use of the
WARNING
Do not use the instrument in a place where inflammable or
combustible vapors (e.g. gasoline) are present. Otherwise there is
a risk of causing a fire.
Do not throw batteries into fire. Do not recharge (nonrechargeable batteries), short circuit, heat or disassemble
batteries. Otherwise, there is a risk of causing fire or injury due to
an explosion or fluid leakage.
Never attempt to disassemble or modify the instrument yourself.
Otherwise there is a risk of causing fire or electric shock.
instrument will lead to the death or serious
injury of the user.
Never attempt to look directly at the sun through the viewfinder of
the meter. Doing so will damage your eyesight.
The instrument should not be operated if it is damaged, or smoke
or odd smells occur. Doing so may result in a fire. In such
situations turn off the power immediately, disconnect the AC
adapter, and contact the nearest authorized service facility.
indicates a danger that improper use of the
CAUTION
instrument will lead to injury to the user or to
property damage.
Do not use any batteries other than those designated for use with
the instrument. When fitting batteries, make sure to align them
according to the polarity shown on the instrument (plus "+" and
minus "-"). Otherwise there is a risk that the batteries may leak or
become damaged, leading to fire, injury or pollution of the
surrounding environment.
Do not walk around while looking into the viewfinder. Doing so
may result in a fall or other accident.
STATEMENT OF FCC COMPLIANCE
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct
the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
This Class B digital apparatus complies with Canadian ICES-003.
Table of Contents
Names of Parts and Displays ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・2
● Data panel displays ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・4
● Brightness difference function ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・52
* Measuring lighting ratio using the Flat Diffuser ・・・・・・・・・・・・・・・・・・・・・・58
* Using the Flash Meter VI as a simplified illuminance meter ・・・・・・・・・・・・63
* Using the Flash Meter VI as a simplified luminance meter ・・・・・・・・・・・・63
For the purpose of explanation, the diagram above shows
all indicators that light up on the LCD.
5. S/A/H indicator
6.indicator
1. Analog scale status indicator
The left (L) and right (R) analog scales are used for incident light
measurement and spot measurement, respectively.
2. Analog scale L
The display of the pointers corresponds to measurement data and
memory data for incident light measurement. It also corresponds to the
standard exposure or latitude for incident light
measurement or spot measurement.
The small digit to the right of the two-digit
reading (f-number) on the digital readout
indicates a fractional value between stops. The
value shown on the analog display is rounded
down or up to the nearest 0.5 stops. (Values of
0.2 or lower are rounded down to 0; those of
0.3 to 0.7 are rounded to 0.5; and those of 0.8
or greater are rounded up to 1.)
When a latitude range is indicated, all dots between the upper and lower
limits are lit.
3. Analog scale R
The display of the pointers corresponds to measurement data and
memory data for spot measurement.
The small digit to the right of the two-digit reading (f-number) on the
digital readout indicates a fractional value between stops. The value
shown on the analog display is rounded down or up to the nearest 0.5
stops. (Values of 0.2 or lower are rounded down to 0; those of 0.3 to 0.7
are rounded to 0.5; and those of 0.8 or greater are rounded up to 1.)
4. Measurement data status indicator
When a value measured with incident light measurement is displayed,
the indicator appears. When a value measured with spot
measurement is displayed, the indicator appears.
5. S/A/H indicator
Holding down the S/A/H button while a measured value is displayed
lights the S, A or H indicator corresponding to the currently selected
mode.
6.indicator
This indicator turns on when the LATITUDE button is pressed.
Upper limit
Standard
value
Lower limit
E5E4
7. Digital readout
8. Shutter speed/framingrate display
9. Analyze scale
10.Flash light measuring
indicator
11.Film speed display
12.Measuring mode
display
For the purpose of explanation, the diagram above shows
all indicators that light up on the LCD.
Names of Parts and DisplaysNames of Parts and Displays
7. Digital readout
When the measurement data display unit is set to "FNo." or "FNo. direct
reading," the f-number (FNo.) is displayed. When the display unit is set
to "EV," the exposure value (EV) is displayed in 0.1-stop increments. For
flash light measurement, only the FNo. display mode is available.
Holding down the incident light measuring button or spot measuring
button (which activates the brightness difference function) in latitude
display mode causes
displayed. When the measuring button is released, the standard
exposure is displayed.
8. Shutter speed/framing-rate display
Displays the shutter speed or frame rate specified with the Up/down dial.
When shutter speed is between 0.6 to 50 sec, s is displayed; between 1
min. and 30 min., m is displayed.
Setting range: Shutter speed: 30 min. to 1/16000 sec. (1, 1/2, 1/3
9. Analyze scale
Displays the proportion of flash light in the total exposure value obtained
from flash light measurement. For more information on the Light Ratio
Analyze function, see p. 36.
10. Flash light measuring indicator
This indicator appears together with the analyze scale in CORD and
NON CORD mode.
11. Film speed display
Displays the film speed setting.
Holding down the instant film ISO button displays the indicator.
Setting range: ISO 3 to ISO 8000
12. Measuring mode display
Displays one of the three measurement modes (AMBI, CORD or NON
CORD) according to the setting of the MODE button.
Repeatedly pressing the MODE button cycles the measurement modes
in the following sequence:
AMBI ➞ CORD ➞ NON CORD ➞ AMBI ➞
For the purpose of explanation, the above figure shows
all available indicators on the display.
13. Viewfinder indicator
Same as the indicator on the external data panel.
14. Viewfinder FNo./EV display
Displays an f-number (FNo.) or exposure value (EV) during spot
measurement.
circle
FNo./EV display
indicator
Preparations
Battery
WARNING
Do not throw batteries into fire. Do not recharge (nonrechargeable batteries), short circuit, heat or disassemble
batteries. Otherwise, there is a risk of causing fire or injury due to
an explosion or fluid leakage.
CAUTION
Do not use any batteries other than those designated for use with
the instrument. When fitting batteries, make sure to align them
according to the polarity shown on the instrument (plus "+" and
minus "-"). Otherwise there is a risk that the batteries may leak or
become damaged, leading to fire, injury or pollution of the
surrounding environment.
1. Preparing
The instrument uses a single
alkaline dry cell (LR-6/1.5 V).
2. Inserting
15. Measuring index circle
The circle’s internal area indicates the measuring area for spot
measurement.
E8
Remove the battery
1
chamber cover by sliding it
lightly in the direction of the
arrow.
Insert the battery with the
2
plus (+) and minus (-) ends
oriented according to the
diagram in the battery
chamber.
★ The meter will not work if the
battery is inserted in the
wrong direction.
E9
Preparations
Preparations
Replace the battery
3
chamber cover.
3. Checking
The instrument automatically
checks the battery when power is
on.
After a new battery is installed, the
display appears as shown below
after the power is turned on.
If you turn on the meter when the
battery power is running low,
"b.c." will appear on the display
for approx. 0.5 sec. before the
normal display appears.
If you turn the power on when
there is not enough battery power
to take measurements, or if the
battery runs low during
measurement, a blinking "b.o."
will display for approx. 1 min. and
then the display will switch off. If
this happens, replace the battery
with a new one.
★ When disposing of used
batteries, observe local waste
disposal regulations.
● The Flash Meter VI can be
operated continuously for about
30 hours with a fresh alkaline
dry cell.
● The Flash Meter VI has a
power-saving function that
automatically turns off the
power after about 10 minutes of
inactivity.
To restart measuring operation
after the display has been
turned off, press the POWER
button. (Pressing the POWER
button restores the meter to the
condition it was in before the
power was turned off. The
settings for film speed, shutter
speed, measuring mode and
display unit, as well as the
existing measurement data and
memory data, are all retained in
memory.)
E10
E11
Preparations
Preparations
Setting film speed
Specify a film speed with the Up/down dial while holding down the ISO
button.
● Turning the control upward
increases the film speed in
increments of 1/3-stop. The
maximum film speed is ISO
8000.
● Turning the control downward
lowers the film speed in
decrements of 1/3-stop. The
minimum film speed is ISO 3.
● Be sure to set film speed to the
correct setting, since all
measurement results are based
on the set value.
● If you change the film speed
after you take a measurement,
the reading will be recalculated
and displayed accordingly.
Setting instant film speed for test
shooting
If the film speed setting used for the final shooting is different from the
instant film speed setting used for test shooting, the instant film speed can
be registered in the meter in advance. Once an instant film speed is
registered for test shooting, the meter converts the measurement result
into a value based on this setting when the instant film ISO button is
pressed after measurement.
Specify an instant film speed with the up/down dial while holding down the
instant film ISO button.
● Turning the control upward
increases the film speed in
increments of 1/3-stop. The
maximum film speed is ISO
8000.
● Turning the control downward
lowers the film speed in
decrements of 1/3-stop. The
minimum film speed is ISO 3.
● If you change the instant film
speed after you take a
measurement, the reading will
be recalculated and displayed
accordingly.
E12
E13
Preparations
Preparations
Selecting a measuring method suitable
for the light-receiving method
Select a measuring method, either incident light measurement or spot
measurement, appropriate to the shooting conditions and nature of the
photograph. The Flash Meter VI can measure exposure in either way. For
incident light measurement, select either the Spherical Diffuser or optional
Flat Diffuser.
1. Incident light measurement
When performing incident light measurements, use the Spherical Diffuser
for three-dimensional subjects such as portraits, and architectural or
landscape photographs. Use the Flat Diffuser when you photograph flat
surfaces such as documents or paintings, or when you want to measure
lighting ratio (See page 58.).
Attaching the Spherical DiffuserRemoving the Spherical Diffuser
Align the index mark (white circle)
of the Spherical Diffuser with the
index of the receptor head. Secure
the diffuser by turning it in the
direction indicated by the arrow
until it stops.
Rotate the diffuser anticlockwise
until it stops, and pull the diffuser
to detach it.
WARNING
CAUTION
Never attempt to look directly at the sun through
the viewfinder of the meter. Doing so will
damage your eyesight.
Do not walk around while looking into the
viewfinder. Doing so may result in a fall or other
accident.
2. Spot measurement
To measure a specific area of a photographic image, select the spot
measurement method (with a light-receiving angle of 1 degree).
To take a spot measurement,
• position the meter near the camera,
• look into the viewfinder at the front of the
meter (data panel side),
• locate the measuring index (circle) at the
center of the viewfinder within the desired
measuring point of the subject, and
• press the spot measuring button.
The allowable measuring distance from the
subject is 1.3 m to infinity (∞).
Measuring index circle
To take an incident light measurement, position the meter near the
subject and aim the Spherical Diffuser directly at the camera.
● The receptor can rotate through a range of 270 degrees, so that you
can use the meter in an almost any photographic configuration.
E14
To stabilize your shooting posture and avoid shaking the meter, turn the
receptor head toward the subject and hold the meter by placing your
hand over it as shown above.
Dioptric adjustment
While looking into the viewfinder for a spot measurement,
adjust the dioptric by turning the dioptric adjustment dial
until the measuring index circle can be clearly seen.
E15
Preparations
Preparations
Difference between incident-light
and Spot (reflected-light) readings
Exposure can be measured in two basic ways. One way is to measure the
light incident on the subject, i.e. the brightness of light illuminating the
subject (illuminance) (see Fig. 1); the other is to measure the light Spot
reflected by the subject, i.e. the intensity of the light reflected from the
subject in the direction of the camera (luminance) (see Fig. 2).
Before selecting the most suitable measuring method, you need to fully
understand the different sources of light you are working with, as well as
the influence of the positions and direction of receptors during
measurement.
Incident-light readings
In general photography, light from the illuminating light source reflects off
the subject and passes through the lens to form an image on the film, and
to expose the film.
To accurately calculate exposure in incident-light readings, you need to
know how much of the illuminating light is actually reflected from the object
to the camera. To do this, you need to know how light or how dark the
subject is, i.e. the reflectance of the subject.
Since a typical value of reflectance for many scenes is 18%*, this value is
used to calculate the light intensity reflected from the subject towards the
camera. The exposure reading (f-number and shutter speed) are then
calculated to reproduce the metered area as a midtone with 18%
reflectance.
E16
Thus, incident-light readings are based on this standard value of 18%
reflectance. This means that areas of subjects having a reflectance higher
than 18% will turn out brighter (e.g. white), while areas of reflectance lower
than 18% will turn out darker (e.g. black). This will produce a clear contrast
in the picture of the subject. From this, we can see that this measuring
method provides for natural tonal range over the entire composition.
* The value of “18%” has been determined to be a typical reflectance value for many
different subjects.
Fig. 3
(a)
Three-
dimensional
subjects
(b)
Flat subjects
To make effective incident-light exposure readings, you must use the
Spherical Diffuser and Flat Diffuser creatively.
When photographing three-dimensional objects such as people, the
highlights and shadow areas of a composition depend on the direction of
the main illuminating light source. Exposure is also influenced by any light
reflected towards the camera from the sides or rear of the subject (Fig. 3
(a)). In these situations, the Spherical Diffuser captures the illuminating
light coming from different directions at the position of the subject, so that
the exposure reading takes into account the contribution of this light on
illuminating the subject. On the other hand, with flat subjects such as
pictures and documents, light from the sides or rear of the subject
generally make little or no contribution to illuminating the subject (Fig. 3(b)).
So, for these situations, accurate exposure readings are made using a Flat
Diffuser to capture only the illuminating light from the front of the subject.
E17
Preparations
Spot (reflected-light) readings
Spot (reflected-light) exposure readings directly measure the amount of
light (luminance) reflected from the subject to the camera. Unlike the case
of incident-light readings, this method does not rely on the assumption of a
standard subject reflectance of 18%. Based on the measured amount of
light falling on the subject, the meter calculates the appropriate exposure
value for reproducing the subject on film at a suitable medium density
(midtone). This means that in Spot (reflected-light) readings, all subjects,
regardless of their reflectance, i.e. regardless of whether they are bright or
dark (white or black), will be reproduced at the same tonal density
(midtone). For this reason, when making Spot (reflected-light) exposure
readings, it is important to decide which area of the subject to measure,
since the reflectance will generally vary quite widely over the composition
under different conditions.
Basic Operation
Here we explain the basics of using the MINOLTA FLASH METER VI
to take exposure readings.
Select a measuring method
● Flash light refers to artificial momentary lighting from light sources such
as electronic flashes, strobe flashes, and speed lights.
● Ambient light refers to continuous lighting from sources such as natural
light (sunlight) and electric lights (including fluorescent lights).
● In either case, both incident-light exposure readings and spot-light
exposure readings can be made.
Are you using a still camera?
Are you using a cine camera?
There are various advanced Spot (reflected-light) readings, such as the
highlight standard exposure method, where an exposure reading is taken
of a bright (white) part of the composition; the shadow standard exposure
method, where a dark (black) part of the composition is measured; and a
method for determining exposure by evaluating the contrast of the
subject and then forecasting how it will come out on film. To make full
use of Spot (reflected-light) readings, refer to specialist books and photo
magazines. You will find that selective metering can give you very
precise control over exposure.
E18
Alt mode
Ambient light
AMBI
mode
(p. 20)
With a still
camera
Select TIME (P.72)
Type of light source to
measure
Are you using a sync cord?
With a sync
cord
CORD
mode
(p. 27)
Flash light
(mixed
light)
Without a
sync
cord
NON CORD
mode
(p. 32)
Alt mode
Select CINE (P.72)
With a cine
camera
AMBI
mode
(p. 24)
E19
Basic Operation
Measuring ambient light
1. With a still camera
Insert a battery (p. 9)
↓
Set film speed (p. 12)
button
Spot measuring
Basic Operation
Incident light
measuring button
Prepare the meter to start
1
taking readings.
E20
Press the MODE button to
2
switch the mode display to
AMBI.
● Changing the measuring mode
retains the memory data but
clears previous measurement
data.
Specify the desired shutter
3
speed with the up/down
dial.
● Shutter speed can be set within
the range of 30 min. to 1/16000
sec.
● Turning the up/down dial
upward increases the shutter
speed. Turning it downward
lowers the shutter speed.
● The shutter speed can also be
changed after meter readings.
Press the measuring
4
button to take readings.
● The meter takes measurements
continuously as you hold down the
incident light measuring button. The
digital display on the data panel
displays the measurement data. At
the same time, the measurement
data are also displayed on the dot
indicator of the analog scale L.
When the measuring button is
released, the meter stops taking
measurements and displays only
the latest measurement result.
● The meter takes measurements
continuously as you hold down the
spot measuring button. The digital
display in the viewfinder displays
the measurement data. At the same
time, the measurement data are
also displayed on the dot indicator
of the analog scale R. When the
measuring button is released, the
meter stops taking measurements.
The latest measurement result
appears on the digital display of the
external data panel and on the dot
indicator of the analog scale R
display.
● Pressing the CLR button
clears the measurement data.
E21
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.