2
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
You will place the tubing connector onto the end of the infusion set to the lled reservoir.
5
41
6 7
2
8
3
With your thumb still on the plunger,
ip over so vial is on top. Release
thumb and pull plunger down to ll
with insulin.
Tap the reservoir to move
air bubbles to top of
reservoir. Push plunger
up to move air into vial.
If needed, pull plunger
back down to amount
of insulin needed for
2-3 days.
Gently push the tubing
connector onto reservoir.
Turn clockwise until locked.
You will hear a click.
To avoid getting insulin on the top of the
reservoir, turn vial over so it is upright. Hold
transfer guard and turn reservoir counterclockwise and remove from transfer guard.
If you see air bubbles, tap
reservoir to move them to
top. Push plunger just a bit to
move them into tubing.
Remove infusion set from package. Remove
the paper that holds the tubing together.
Twist plunger counterclockwise to loosen
and remove.
Select Load Reservoir
from the menu.
Press
to open the menu. If the pump
is locked, you will need to unlock the
pump after pressing
.
Select Next.
THE BACKLIGHT MAY
HAVE TURNED OFF
Press any button to turn
the screen back on.
WARNING If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly ll
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycemia or hypoglycemia.
Insulin
vial
Plunger
Quick-serter
®
Quick-set
®
Tubing connector
Release
button
Tubing
slot
Green side
button
Green side
button
Needle guard
Hub handleBlue hub
Connector
Plunger
CONNECT RESERVOIR TO INFUSION SET
Continued on next page