MiniMed 670g Quick Reference Manual

Page 1
1
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
START HERE
1
5
2
1 2 3 4
6
3 4
Wash your hands. Press
. Select Options.
Select Reservoir & Tubing. Select New Reservoir.
Select Rewind.
Remove the infusion set you have been using by loosening the adhesive and pulling away from body.
Remove the used reservoir from the pump.
Remove from package. Make sure insulin vial is at room temperature to reduce the risk of air bubbles.
Follow the next steps to ll reservoir with insulin and connect to the infusion set tubing.
Pull plunger down to the amount that you
plan to ll with insulin.
Wipe vial with alcohol. Place vial on table. Firmly press the blue transfer guard onto vial.
Push and hold plunger down.
Plunger rod
Plunger
Transfer guard
Reservoir
Transfer guard
Insulin vial
FILL RESERVOIR & CONNECT TO THE INFUSION SET TUBING
Continued on next page
Page 2
2
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
You will place the tubing connector onto the end of the infusion set to the lled reservoir.
5
41
6 7
2
8
3
With your thumb still on the plunger,
ip over so vial is on top. Release thumb and pull plunger down to ll
with insulin.
Tap the reservoir to move air bubbles to top of reservoir. Push plunger up to move air into vial.
If needed, pull plunger back down to amount of insulin needed for 2-3 days.
Gently push the tubing connector onto reservoir. Turn clockwise until locked. You will hear a click.
To avoid getting insulin on the top of the reservoir, turn vial over so it is upright. Hold transfer guard and turn reservoir counter­clockwise and remove from transfer guard.
If you see air bubbles, tap reservoir to move them to top. Push plunger just a bit to move them into tubing.
Remove infusion set from package. Remove the paper that holds the tubing together.
Twist plunger counter­clockwise to loosen and remove.
Select Load Reservoir from the menu.
to open the menu. If the pump is locked, you will need to unlock the pump after pressing
.
Select Next.
THE BACKLIGHT MAY
HAVE TURNED OFF
Press any button to turn
the screen back on.
WARNING If insulin or any liquid gets inside the tubing connector, it can temporarily block the vents that allow the pump to properly ll
the infusion set. This may result in the delivery of too little or too much insulin, which could cause hyperglycemia or hypoglycemia.
Insulin vial
Plunger
Quick-serter
®
Quick-set
®
Tubing connector
Release button
Tubing slot
Green side button
Green side button
Needle guard
Hub handleBlue hub
Connector
Plunger
CONNECT RESERVOIR TO INFUSION SET
Continued on next page
Page 3
3
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
1 2
3 4
1 2 3
Place reservoir into pump. Select Next.
Select Load and keep holding
.
When you see this screen, select Next.
After you see drops, press
and select Next.
Select Fill and keep holding
until you see
drops at the end of tubing, then let go.
Turn clockwise until you feel reservoir lock into place.
Place blue hub into serter, placing the handle in the tubing slot.
Holding the serter with one hand, gently press infusion set to secure. Be careful not to push all the way into serter. Do not hold or press on the blue button.
Drops at end of tubing
21
Now place the lled reservoir into the reservoir compartment of the pump.
Follow these steps to load the reservoir and ll the tubing.
Next, follow the steps to insert the infusion set into your body.
PLACE RESERVOIR INTO PUMP
LOAD RESERVOIR AND FILL TUBING
INSERT INFUSION SET
Be careful not push or hold release button
Blue hub
Tubing slot
Continued on next page
Page 4
4
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
3
7
11
4
8
12
5
9 10
6
1
2
Holding needle guard,
pull o the paper that
covers the adhesive.
Hold serter against cleaned site.
Hold infusion set. Pull blue handle straight out to remove needle.
Pull green handle down until you hear it click.
Press the two green side buttons at the same time.
Fold blue handle until locked.
Choose an insertion site from the shaded areas shown here. Wipe with alcohol or other antiseptic.
Press down the release button to unlock.
Turn to loosen needle guard and pull.
Pull serter away from body. Press adhesive against skin.
Release button
Release button
Green side buttons
Needle guard
Needle guard
Adhesive
Continued on next page
Page 5
5
MINIMED® 670G SYSTEM | CHANGING THE QUICK-SET INFUSION SET QUICK REFERENCE GUIDE
1
3
2
4
Select Fill.
Select Fill Now.
Select Fill amount and enter :
 0.300 if using 6mm cannula  0.500 if using 9mm cannula
Then press
.
The Home screen displays
the insulin as it lls the cannula.
You will now ll the cannula, the little tube under your skin, with insulin.
Your infusion set change is now complete!
FILL CANNULA
NOTE: Your pump will remember the Fill amount that you
used last. Always verify that the Fill amount is correct.
 If it is correct, press
to Fill Now and press
.
 If it is incorrect, press
. Change to correct amount and
press
. Select Fill Now.
NOTE: Select Stop Filling if you need to stop, for example, if you notice the
Total amount is incorrect. This should
rarely happen if you have veried the
Fill amount on the previous screen.
950M16510-011 20170401 ©2017 Medtronic MiniMed, Inc. All rights reserved.
Loading...