Translation of the original instrument manual9499 050 54103Edition 11.1.010/23/2017
Release 4.50
Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany
Phone: +49.40.67960.303 Fax: +49.40.67960.383
Foreword
Must be followed!
Any information in this document is subject to change without notice and does not represent a commitment on
the part of Minebea Intec unless legally prescribed. This product should be operated/installed only by trained
and qualied personnel. In correspondence concerning this product, the type, name, and release number/serial
number as well as all license numbers relating to the product have to be cited.
Note
The product is partially protected by copyright. It may not be changed or copied, and it may not be used without
purchasing or written permission from the copyright owner (Minebea Intec). Its use constitutes acceptance by
you of the above-mentioned provisions.
1.1Read the manual.........................................................................................................................................................9
1.2This is what instructions look like.............................................................................................................................9
1.3This is what lists look like........................................................................................................................................... 9
1.4This is what menus and soft keys look like.............................................................................................................9
1.5This is what safety instructions look like................................................................................................................. 9
2.4.2Opening the device.........................................................................................................................................12
2.4.3Supply voltage connection............................................................................................................................12
2.6Failure and excessive stress.....................................................................................................................................13
2.8Repairs and maintenance........................................................................................................................................ 14
2.8.1General information ...................................................................................................................................... 14
2.8.2Electrostatically sensitive parts ................................................................................................................... 14
3.2Overview of the device .............................................................................................................................................15
3.4Display and control panel.........................................................................................................................................17
3.4.1General information .......................................................................................................................................17
3.5Overview of connections.........................................................................................................................................28
4.3.1Connecting the screens................................................................................................................................. 32
4.3.2Connecting the equipotential bonding conductor................................................................................... 32
4.4Hardware construction ............................................................................................................................................32
4.5Connecting analog load cells and platforms........................................................................................................45
4.5.1General information ......................................................................................................................................45
4.5.2Connecting a load cell with a 4-wire cable.................................................................................................45
4.5.3Connecting a load cell with a 6-wire cable.................................................................................................47
4.5.4Connecting between 2 and8 load cells (650 Ω) using a 6-wire connection cable ..............................48
4.5.5Connecting load cells of type series PR 6221 ............................................................................................48
4.5.6Testing the measuring circuit ......................................................................................................................49
4.5.7External supply to load cells.........................................................................................................................49
4.5.8Connecting to input/output interface PR 1626/60 ..................................................................................50
4.5.9Connecting an analog weighing platform (CAP… series).........................................................................52
4.6.1General information ......................................................................................................................................54
4.6.31× RS-232 interface and 1× RS-485 interface ............................................................................................59
4.6.4Analog inputs and outputs........................................................................................................................... 72
4.6.9Opto-decoupled inputs and outputs..........................................................................................................86
4.6.10Status LEDs on eldbus card........................................................................................................................93
4.6.16EtherNet IP interface ....................................................................................................................................113
5.1.1General information .....................................................................................................................................116
6.1.1General information ..................................................................................................................................... 117
6.2Application menu [Start] ........................................................................................................................................ 117
7.2Switching on the device..........................................................................................................................................121
7.2.1Setting the date and time........................................................................................................................... 122
7.3Switching o the device ........................................................................................................................................ 123
7.7Finding and connecting a device with a notebook/PC .....................................................................................144
7.8Finding and connecting the device automatically in the network.................................................................. 145
7.9Searching the device in the network with "IndicatorBrowser"........................................................................ 145
7.10Resetting the network address ............................................................................................................................146
7.11Operation using VNC ............................................................................................................................................. 147
7.12Operation via a web browser................................................................................................................................148
7.13.2Date & time ...................................................................................................................................................149
7.14.1General information .................................................................................................................................... 158
7.14.3Selecting the calibration mode..................................................................................................................160
7.14.4Setting maximum load................................................................................................................................ 163
7.14.5Determining the scale interval................................................................................................................... 163
7.14.6Determining the dead load ........................................................................................................................ 164
7.14.7Calibrating with weight............................................................................................................................... 165
7.14.8Calibrating with mV/V................................................................................................................................. 166
7.14.9Calibrating with load cell data (smart calibration).................................................................................. 168
7.14.10Subsequent dead load correction............................................................................................................. 169
7.14.12Calculating the test value ........................................................................................................................... 170
7.14.13Saving the calibration................................................................................................................................... 171
7.14.14Cancelling a calibration............................................................................................................................... 172
7.15.1General information .................................................................................................................................... 179
7.15.2Parameters for serial interface .................................................................................................................. 179
7.15.3Parameters for the xBPI-weighing function............................................................................................ 180
7.15.4Setting up an xBPI platform........................................................................................................................181
7.15.6Setting the xBPI dead load .........................................................................................................................187
7.15.7xBPI calibration with user specied weight.............................................................................................188
7.15.8xBPI calibration with automatic weight detection ................................................................................. 190
7.15.9xBPI calibration with default weight......................................................................................................... 190
7.15.10xBPI calibration with built-in weight .........................................................................................................191
7.16Calibrating digital load cells of type "Pendeo" .................................................................................................. 193
7.16.1General information ....................................................................................................................................193
7.16.2Selecting and conguring RS-485 interface ...........................................................................................194
7.16.3Selecting the load cell type ........................................................................................................................ 195
7.17.1Selecting and conguring serial interfaces .............................................................................................207
7.17.2Date and time ............................................................................................................................................... 215
7.17.8Conguring digital inputs.......................................................................................................................... 230
7.17.9Conguring digital outputs........................................................................................................................233
7.17.10Conguring analog output.........................................................................................................................238
7.18.1Displaying the version................................................................................................................................ 240
7.18.2Displaying the status ...................................................................................................................................241
7.18.4Displaying the alibi memory ......................................................................................................................243
8.2Conguration using the front-panel keys ..........................................................................................................247
8.2.1Conguring production mode using the front-panel keys...................................................................247
8.2.2Conguring interaction mode using the front-panel keys .................................................................. 248
8.2.3Conguring digital inputs and outputs using the front-panel keys................................................... 248
8.2.4Conguring printout using the front-panel keys .................................................................................. 250
8.3Starting the application .........................................................................................................................................253
8.4Conguration via a notebook/PC ........................................................................................................................253
8.4.1Conguring production mode ...................................................................................................................253
8.4.2Conguring digital inputs and outputs....................................................................................................257
8.4.3Conguring material ...................................................................................................................................259
9.1.2Front-panel key test ....................................................................................................................................265
9.1.4Inputs and outputs ..................................................................................................................................... 266
9.2Functions via the WEB site .................................................................................................................................... 271
9.2.1General information .................................................................................................................................... 271
9.2.2Displaying weighing points in a table.......................................................................................................272
9.3Resetting the device to the factory settings using the front-panel keys ......................................................279
9.4Resetting the device to the factory settings ..................................................................................................... 280
9.5Updating new software with FlashIt.................................................................................................................... 281
9.5.1Updating in network with a DHCP service ...............................................................................................281
9.5.2Updating via a point-to-point connection with DHCP service .............................................................283
9.5.3Updating with a xed IP address ............................................................................................................. 285
10.2SPM data in PR 1612 ModBus mode................................................................................................................... 289
12.2.3Reading and writing data .......................................................................................................................... 294
12.2.4Description of the I/O area (read/write window)....................................................................................295
12.2.5Special hints for DeviceNet and EtherNet-IP .........................................................................................300
12.3.5Setup of the eldbus interface ................................................................................................................. 308
13.2Elementary data types........................................................................................................................................... 314
14.1.1Battery for date/time .................................................................................................................................. 321
14.1.2Changing the battery for date/time ......................................................................................................... 321
16.2Error messages for xBPI scales.............................................................................................................................325
16.3Error messages for Pendeo load cells .................................................................................................................325
16.4Error messages during calibration.......................................................................................................................326
16.6.1Weighing point error ...................................................................................................................................330
16.6.2Error in the "EasyFill" application.............................................................................................................. 331
17.1Note on using "free software" ..............................................................................................................................332
17.2Decoding the serial number..................................................................................................................................332
17.3General data ............................................................................................................................................................332
17.3.1Backup battery for time/date ....................................................................................................................332
17.3.2Supply voltage connection 230 V AC........................................................................................................332
17.3.3Supply voltage connection 24 V DC..........................................................................................................332
17.4Eect of ambient conditions ................................................................................................................................333
17.5.3Accuracy and stability .................................................................................................................................334
17.7Documentation on the CD included ....................................................................................................................335
This device has been built and tested in compliance with the safety regulations for
measuring and control equipment for protection class I (protective grounding
conductor) according to IEC 1010/EN 61010 or VDE 0411.
The device was in perfect condition with regard to safety features when it left the
factory.
To maintain this condition and to ensure safe operation, the user must follow the
instructions and observe the warnings in this manual.
2.2Intended use
The device is intended for use as an indicator for weighing functions.
Product operation, commissioning and maintenance must be performed by trained and
qualied personnel who are aware of and able to deal with the related hazards and take
suitable measures for self-protection.
The device reects the state of the art.
The manufacturer does not accept any liability for damage caused by third-party system
components or due to incorrect use of the product. The use of this product signies
recognition of the stipulations listed above.
2.3Incoming goods inspection
The shipment must be checked for completeness. A visual inspection must be performed
to determine if the shipment has been damaged. If there are grounds for a complaint, this
must be brought to the attention of the delivery company immediately. A Minebea Intec
sales or service point must be informed.
2.4Before operational startup
NOTICE
Perform visual inspection.
Before operational startup as well as after storage or transport, inspect the device
visually for signs of mechanical damage.
2.4.1Installation
The device is suitable for installation in control panels or cabinets.
The device has to be installed in an EMC-compliant manner, see Chapter 4.3.
EN-11Minebea Intec
X3 Process Indicator PR 54102 Safety instructions
ComponentProtection classInstallation
Housing frontIP65
Back of the housingIP30
To ensure proper cooling of the device, make sure air circulation around the device is not
blocked. Avoid exposing the instrument to excessive heat, e.g., from direct sunlight, and
vibrations. The ambient conditions specied in Chapter 17.4.1 must be taken into account
at all times.
With outdoor mounting, make sure that adequate weather protection is provided (for
temperatures, see Chapter 17.4.1).
2.4.2Opening the device
WARNING
Working on the device while it is switched on may have life-threatening
consequences.
When removing covers or parts using tools, live parts or terminals may be exposed.
Please note that capacitors in the device may still be charged even after disconnecting
the device from all voltage sources.
Disconnecting the device from the power supply.
This device contains electrostatically sensitive components. For this reason, an
equipotential bonding conductor must be connected when working on the open device
(antistatic protection).
Suitable for any operating environment.
2.4.3Supply voltage connection
The device has no power switch. It is ready for operation immediately after connecting to
a power supply.
2.4.3.1Version 230 V AC
Safe interruption of both supply voltage conductors must be
provided for, either by disconnecting the power connector or
using a separate switch.
The device is equipped with a wide range power supply and
covers AC systems with a frequency of 50/60 Hz and a voltage range of 100 to 240 V AC +10/-15% automatically (without
manual selection).
The power supply is protected against short circuits and
overloads, and disconnects automatically in case of a fault.
If the electrical protection has triggered:
-Disconnect the device from all power sources and wait at
least 1 minute.
-Determine and eliminate the cause of the error.
-Reconnect the device to the supply voltage.
Minebea IntecEN-12
2 Safety instructionsX3 Process Indicator PR 5410
2.4.3.2Version 24 V DC
The PR 5410/01 version is designed for 24 V direct current.
The supply is established via two screw terminals
(- 24 V +). The device is protected against incorrect polarity.
The device is protected in the - line via a fuse on the back
of the device (primary side).
2.4.4Protective conductor connection
2.4.4.1Version 230 V AC
The device must be connected to a protective ground via a protective grounding
conductor (PA) in the power connector.
The power cable contains a protective grounding conductor which must not be
interrupted inside or outside the device.
The protective grounding conductor is connected to the housing rear panel inside the
device.
2.4.4.2Version 24 V DC
The device must be connected to the protective grounding conductor. The connection
can be established via the housing rear panel.
2.5RF interference suppression
The device is designed for operation in the industrial sector and can cause
RF interferences in a residential environment, see Chapter 17.4.3. In this case, the
operator can be required to carry out the appropriate measures.
2.6Failure and excessive stress
If there is any reason to assume that safe operation of the device is no longer ensured,
shut it down and make sure it cannot be used.
Safe operation is no longer ensured if any of the following is true:
-The device is physically damaged.
-The device does not function.
-The device has been subjected to stresses beyond the tolerance limits (e.g., during
storage or transport).
EN-13Minebea Intec
X3 Process Indicator PR 54102 Safety instructions
2.7Important note
Make sure that the construction of the device is not altered to the detriment of safety. In
particular, leakage paths, air gaps (of live parts) and insulating layers must not be
reduced.
Minebea Intec cannot be held responsible for personal injury or property damage caused
by a device repaired incorrectly by an operator or installer.
2.8Repairs and maintenance
2.8.1General information
Repairs are subject to inspection and must be carried out at Minebea Intec.
In case of defect or malfunction, please contact your local Minebea Intec dealer or service
center for repair.
When returning the device for repair, please include a precise and complete description of
the problem.
Maintenance work may only be carried out by a trained technician with expert knowledge
of the hazards involved and the required precautions.
2.8.2Electrostatically sensitive parts
This device contains electro-statically sensitive components. Therefore, potential
equalization must be provided when working on the device (antistatic protection).
2.8.3Replacing fuses
2.8.3.1Replacing fuses in the Ex area
WARNING
Possible explosion due to improper replacement!
PR 5410 used in the Ex area: The replacing of fuses are not allowed!
2.8.3.2Replacing fuses in a safe area
WARNING
Damage due to overheating.
The use of repaired fuses and bypassing the fuse holder is prohibited.
Only the fuses specied in Chapter 17.3.2 and 17.3.3 are permitted.
Minebea IntecEN-14
3 Device descriptionX3 Process Indicator PR 5410
3Device description
3.1General information
The device is equipped with a 6-position, 7-segment display and additional status
indicators. Local operation can be carried out using 6 dual-function keys.
The device contains two applications:
-Standard
-EasyFill
Most functions are supported by both applications.
Some functions depend on the application.
3.2Overview of the device
-Accuracy 10,000 e (class III) for the weighing electronics
-High-speed conversion with measurement times from 10 msec
-Weight display with status via transective, 6-digit, 7-segment display
-Alibi memory
-6 function keys for operation on the front panel
-Enclosed front: IP65, back: IP30
-LAN adapter with 10/100 Mbit/s for data transfer, calibration, parameterization
-Integrated RS-232 interface for, e.g., printer or remote display
-Can be expanded using the following plug-in cards (2 slots):
-Serial input and output board 2× PR 5510/02 or 2× PR 5510/04
1× PR 5510/02 and/or 1× PR 5510/04
-Analog input and output board PR 5510/07
-BCD output board PR 5510/08 or PR 5510/09
-Opto-decoupled input and output board PR 5510/12
-Fieldbus cards PR 1721/3x
-Electrically isolated interfaces (except RS-232)
-Wide range power supply for 100 to 240 V AC, protection class I (protective
grounding conductor)
-Version PR 5410/01 for 24 V direct current
-Plug-in connections inside the device for load cells, inputs/outputs, LAN adapter
-Installation as control panel or control cabinet device
-Calibration using the front-panel keys or PC tool (browser/VNC); calibration of
Pendeo® load cells using the PC tool only
-Calibration using weights according to the mV/V method or directly using load cell
data (smart calibration)
-Software conguration of the interface cards, e.g., for remote display or printer
EN-15Minebea Intec
X3 Process Indicator PR 54103 Device description
-The "EasyFill" application allows for quick and reliable lling and emptying of vessels
(for functional description, see Chapter 6.1).
-Analog test for the weighing electronics
-Overwrite protection using CAL switch on the main board
3.2.1Communication protocols
For the internal RS-232 or external RS-232 + RS-485 (accessories):
-Remote display protocol
-Printer
-ModBus-RTU (slave)
-SMA protocol
-xBPI protocol
-EW-Com protocol
Fieldbus slave (accessories):
-PR 1721/31 ProBus-DP
-PR 1721/32 InterBus-S
-PR 1721/34 DeviceNet
-PR 1721/35 CC-Link
-PR 1721/36 ProNet I/O
-PR 1721/37 EtherNet IP
For the internal LAN:
-ModBus-TCP
-Ethernet TCP/IP
-OPC
3.3Housing
3.3.1General information
The device has an aluminum housing with IP65 front. It can also be used as a control
cabinet version.
The keypad and the display form one unit with the front. A rectangular cut-out is required
for the installation. Cable connections are made at the back of the housing.
The ambient conditions specied for the device must be observed (see Chapter 17.4.1).
Openings for max. 2 plug-in cards are sealed by dummy plates.
Minebea IntecEN-16
3 Device descriptionX3 Process Indicator PR 5410
3.3.2Dimensions
Front ViewSide View
all dimensions in mm
all dimensions in mm
Back View
all dimensions in mm
3.3.3Control panel cut-out
The control panel cut-out must be made before installing the device.
all dimensions in mm
3.4Display and control panel
3.4.1General information
The PR 5410 process indicator can only be operated using the front-panel keys or by
Notebook/PC.
EN-17Minebea Intec
X3 Process Indicator PR 54103 Device description
-Operation using the front-panel keys (see Chapter 7.5.2)
-VNC viewer (see Chapters 3.4.4.4 and 7.11) or
-WEB browser (see Chapter 7.12)
3.4.2Overview
3.4.3Display
3.4.3.1User interface
The display of the user interface shows weight values of up to 7‑digits with decimal point
and plus or minus sign.
Available weight units are t, kg, g or lb.
lb unit is not permitted as legal-for-trade in the EU and EEA.
Above the weight display of the user interface, the currently displayed weight is shown as
a bar graph in relation to the maximum load (Max). When Max reaches 100%, the bar
graph is located on the right.
Minebea IntecEN-18
3 Device descriptionX3 Process Indicator PR 5410
Weight type/plus or minus signDescription
Gross weight
Gross weight in NTEP or NSC mode
Net weight (Net = gross - tare)
Tare weight
Preset tare
The display shows the test value without weight
unit.
No display-Test value
-Gross, not tared
Positive value
Negative value
Standstill/ZeroDescription
Weight value standstill
The gross weight value is within ±¼ d of zero.
Batching mode; ashes when "On hold"; ashes rapidly to indicate error.
EN-19Minebea Intec
X3 Process Indicator PR 54103 Device description
Symbols/mass unitDescription
Value not permissible in legal metrology
(e.g.,‑10-fold resolution, deactivated load cell).
R1Range 1
R2Range 2
R3Range 3
WP AWeighing point A
MaxMaximum load (weighing range)
MinMinimum weight
t, kg, g, lbThese weight units are available.
3.4.3.2Device display
The display of the device permits indication of 6-digit weight values (digit height 18 mm)
with decimal point.
Weight type/plus or minus signDescription
Gross weight
Gross weight in NTEP or NSC mode
Net weight (Net = gross - tare)
Tare weight
Minebea IntecEN-20
3 Device descriptionX3 Process Indicator PR 5410
Weight type/plus or minus signDescription
Fixed tare
No display-Test value
-Gross, not tared
The display shows the test value without weight
unit.
Positive value
Negative value
Standstill/zero/dosingDescription
Weight value standstill
The gross weight value is within ±¼ d of zero.
Filling mode; ashes when batching is stopped; rapidashing indicates error.
Symbols/mass unitDescription
•Range 1
•
•
•
•
•
t, kg, g, lbThese weight units are available.
3.4.4Operating elements
3.4.4.1User interface
The following tables show the basic meanings of symbols on the operator interface.
Indicator keys
Display gross weightSets gross weight to zero, provided that
Range 2
Range 3
-weight value is stable;
-weight is within zero setting range.
This function depends on the conguration.
Display tare weight.Starts a printout.
EN-21Minebea Intec
X3 Process Indicator PR 54103 Device description
Indicator keys
Taring
The current gross weight is stored in
the tare memory, provided that
-weight value is stable;
-the instrument is not in error sta-
tus.
This function depends on the conguration.
Navigation/menu keys
Scroll up in the menu.Conrm input/selection.
Scroll down in the menu.-Backspace
-Pressing the delete key deletes indivi-
dual characters (within an entry).
-Cursor to the left
-Selection
-Cursor to the right
-Selection
Function keys/softkeys
Access the Setup menu.
Information on version number, tted
hardware, 10-fold resolution
-Cancel entry/selection (after a conr-
mation prompt) without saving the change.
-Exit parameters/menu window.
Depending on the settings under
-[Weighingpoint]- [Calib]- [Param]- [Test
mode] the following is displayed by calling
the test with the key later on:
-with "Absolute" the maximum load
-with "Relative" the deviation from test
value.
Softkeys 1…5
Select appropriate menu function, see also
Chapter 3.4.4.2.
No function
Minebea IntecEN-22
3 Device descriptionX3 Process Indicator PR 5410
Alphanumeric keypad
Toggle key
Toggle by pressing:
-between alpha and numerical mode
-during conguration between weight units
Pressing once displays the corresponding rst character, e.g.,"A",
at the cursor position. After pressing twice, "B" is displayed at the
cursor position and after pressing three times, "C" is displayed.
Press thecursor key to nish entering a character or wait
approx. 2 seconds.
If only numeric values are required for input, letters are not enabled.
Input eld
Press thecursor key within an entry to return to the previous
character.
Press thecursor key within an entry to select the next character.
Within an input, pressing thedelete key deletes the character
to the left of the cursor.
In principle
If alphanumeric characters are already present in the input eld of
the selected line, they will be completely overwritten after immediate entry.
If alphanumeric characters are already present in the input eld of
the selected line, you can press thecursor key to select the
characters to be overwritten and overwrite them.
EN-23Minebea Intec
X3 Process Indicator PR 54103 Device description
3.4.4.2Operation using softkeys
The functions of the vesoftkeys below the graphic display are indicated in the
bottommost text line of the display. Softkey functions shown in gray cannot be selected
at the active menu level or with the current access privileges.
In the descriptions of operating sequences which entail the use of softkeys, the softkey
function to be selected is shown in square brackets; the softkey symbol is not displayed;
example: [Calib].
3.4.4.3Navigation key operation
Menu
The cursor keys, theandkeys are used to navigate through the menus.
Parameters
Use thecursor keys to select the individual parameters.
Use thekey to conrm the selection.
The required values | texts are entered via the alphanumeric keys.
Thekey is used to check the ☑ box.
If the list of parameters is long, a vertical bar graph on the left (black and gray) shows
which part of the list is displayed.
An arrow in front of a menu item indicates that there are menu sublevels.
Possible settings and an available selection list is indicated by double arrows.
The parameter is selected using thekey.
3.4.4.4Operation using VNC
User interface, see Chapters 3.4.2, 3.4.3.1 and 3.4.4.1.
Minebea IntecEN-24
3 Device descriptionX3 Process Indicator PR 5410
3.4.4.5Operation using the front-panel keys
The following table shows the basic meanings of the symbols on the front-panel keys.
Indicator keys
Taring
The current gross weight is stored in
the tare memory, provided that
-weight value is stable;
-the instrument is not in error sta-
tus.
The key is locked during the lling
process.
Toggling the display
Net - gross - tare
The key is locked during the lling
process.
Sets gross weight to zero, provided
that
-weight value is stable;
-weight is within zero setting ran-
ge.
Starts a printout.
Depending on the settings under
-[Weighingpoint]- [Calib]- [Param]- [Test
mode] the following is displayed by calling
the test with the key later on:
-with "Absolute" the maximum load
-with "Relative" the deviation from test
value.
The key is locked during the lling
process.
Navigation/menu keys
Scroll up in the menu.Access the Setup menu.
Scroll down in the menu.OK-Conrm input/selection.
-Start/restart the ongoing lling pro-
cess.
EN-25Minebea Intec
X3 Process Indicator PR 54103 Device description
Navigation/menu keys
-Cursor to the left
-Selection
-Cursor to the right
-Selection
3.4.4.5.1Selecting parameters
The selection and modication of parameters are described in the following.
1.Press.
2.Press/OK.
The "Cd" calibration menu ashes.
Exit-Cancel entry/selection (after a conr-
mation prompt) without saving the change.
-Exit parameters/menu window.
-Stop/abort the ongoing lling process.
SEtuPappears on the display.
Cd 000appears on the display.
3.Pressto access the next parameter group.
CP ashes on the display.
4.Pressto access the parameter number.
CP 010appears on the display. The right digit ashes.
5.Press/to select the digits.
6.Press/to increase/decrease the number.
7.Press/OK to select the menu item.
The menu item appears on the display.
8.Press/OK to switch to the parameter selection.
9.Press/to select the parameter values.
10. Press/OK to select the parameter values.
11. Press "Exit" to exit a menu.
If a parameter was changed, SAVE
appears on the display.
12. Press/OK.
YESappears on the display.
Minebea IntecEN-26
3 Device descriptionX3 Process Indicator PR 5410
13. Pressif necessary, to select "no."
14. Press/OK to save the change.
EN-27Minebea Intec
X3 Process Indicator PR 54103 Device description
3.5Overview of connections
Minebea IntecEN-28
Loading...
+ 310 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.