Minebea OPT-563B Quick Instruction Manual

1
DIGITAL TRANSMITTER OPT-563B
Quick Instruction Manual
Minebea Co., Ltd.
EN294-1723
2
I
Introduction
Thank you for purchasing the Minebea OPT-563B Digital Transmitter for flange type torque transducers using optical transmission.
This instruction manual is a [quick] instruction manual primarily intended to describe the procedure for calibration by inputting numerical torque values.
For detailed instructions on how to use the OPT-563B, refer to Instruction Manual EN294-1596*.
Note that handling or operating the device incorrectly may result in malfunctions. Read the instruction manual EN294-1596* thoroughly before use for safety and optimal results.
The guaranteed accuracy when calibrating by inputting numerical torque values will be approximately 1/500 (±0.2% RO).
Note also that symmetry correction cannot be adjusted using numerical input.
Keep this instruction manual in a location where it is readily accessible to end users.
Pictograms and conventions used in this manual
This manual uses the following pictograms to indicate actions to avoid at all times, aspects requiring caution, and other noteworthy matters.
Be sure to read the descriptions provided alongside these pictograms.
WARNING
This indicates circumstances in which incorrect handling may result in death or serious injury to users. Avoid the actions described here at all times.
CAUTION
This indicates circumstances in which incorrect handling may result in injury to users or damage to property.
This indicates operating or procedural precautions or restrictions. Always read the details included here to avoid malfunctioning.
II
Revision History
Date Manual No. Revision reason (details)
2015/12 DRW. No.EN294-1723 1st issue
III
Contents
INTRODUCTION ...................................................................................................................................................... I
PICTOGRAMS AND CONVENTIONS USED IN THIS MANUAL ........................................................................... I
REVISION HISTORY ............................................................................................................................................... II
CONTENTS ............................................................................................................................................................ III
1. P A RT NAMES AND FUNCTIONS ................................................................................................................... 5
1-1. Front Panel .......................................................................................................................................................................................... 5
1-2. Rear Panel ............................................................................................................................................................................................ 7
2. OPERATING INSTRUCTIONS ........................................................................................................................ 8
2-1. Key ................................................................................................................................................................................................. 8
2-1-1. When Operated in Measurement Mode ........................................................................................... 8
2-2. Key ................................................................................................................................................................................................. 8
2-2-1. When Operated in Measurement Mode ........................................................................................... 8
2-2-2. When Operated in Other Modes ....................................................................................................... 8
2-3. Key ................................................................................................................................................................................................. 8
2-3-1. When Operated in Measurement Mode ........................................................................................... 8
2-3-2. When Operated in Other Modes ....................................................................................................... 8
2-4. Key .................................................................................................................................................................................................. 8
2-4-1. When Operated in Measurement Mode ........................................................................................... 8
2-4-2. When Operated in Other Modes ....................................................................................................... 9
2-5. Key .................................................................................................................................................................................................. 9
2-5-1. When Operated in Measurement Mode ........................................................................................... 9
2-5-2. When Operated in Other Modes ....................................................................................................... 9
2-6. Key .................................................................................................................................................................................................. 9
2-6-1. When Operated in Measurement Mode ........................................................................................... 9
2-6-2. When Operated in Other Modes ....................................................................................................... 9
2-7. Key .................................................................................................................................................................................................. 9
3. CALIBRATION ............................................................................................................................................... 10
3-1. Settings Required for Calibration ............................................................................................................................................. 10
IV
3-2. Settings to be Altered As Necessary After Calibration ................................................................................................. 10
3-3. Calibration Procedure ................................................................................................................................................................... 11
3-3-1. Calibration in Measurement Mode .................................................................................................. 11
3-3-2. Selecting Calibration Mode ............................................................................................................ 12
3-3-3. Minimum Scale Division Setting ..................................................................................................... 12
3-3-4. Torque Setting ................................................................................................................................ 13
3-3-5. Actual Torque Setting ..................................................................................................................... 14
3-3-6. Zero Calibration .............................................................................................................................. 15
3-3-7. Span Calibration ............................................................................................................................. 16
3-3-8. Calibration End ............................................................................................................................... 17
3-4. Zero/Span Fine Adjustment Function ................................................................................................................................... 18
3-4-1. Selecting Zero Point Fine-adjustment Mode .................................................................................. 18
3-4-2. Selecting Span Point Fine-adjustment Mode ................................................................................. 19
4.
FUNCTION MODE ......................................................................................................................................... 20
4-1. Selecting Function Mode ............................................................................................................................................................. 20
4-2. Function List .................................................................................................................................................................................... 21
5
1. Part Names and Functions
1-1. Front Panel
(1) Torque units display
Displays the measurement units set.
(2) Torque display
Displays the torque reading and [OL] (for over limit).
(3) Status display
Indicates the OPT-563B status.
H : Illuminates when the torque meter light level is normal.
M : Illuminates when the torque meter light level is low.
Note that [M] does not illuminate if used in conjunction with the DBX-001 power supply box.
A/Z : Illuminates when the A/Z function operates.
CHECK : Illuminates when the key is pressed to turn CHECK on.
LOCK : Illuminates when external control input LOCK to COM1 is short-circuited.
ERROR : Illuminates when an error signal is output.
(4) Rotation speed units display
Displays the rotation speed units.
(4) Rotation speed units display
(2) Torque display
(1) Torque units display
(3) Status display
(6) ~ (12) Key switches
(13) Torque analog output
trimmer section
(14) Rotation speed analog
output trimmer section
(5) Rotation speed display
6
(5) Rotation speed display
Displays the rotation speed and [OS] (for overspeed).
(6) key
Turns on the A/Z function. Also moves the setting digit to the left when setting numerical values.
(7) key
Turns off the A/Z function. Also moves the setting digit to the right when setting numerical values.
(8) key
Increments the digit selected when setting numerical values.
(9) key
Decrements the digit selected when setting numerical values.
(10) key
Used to select function mode. Also used to return to measurement mode without entering the value when setting. Depress for at least 2 seconds to select function mode.
(11) key
Used to turn the CHECK value on or off. Depress for at least 2 seconds to turn on the CHECK value.
(12) key
Used for registering settings.
(13) Torque analog output trimmer section
Torque analog output (VOUT and IOUT) zero point and span point trimmer adjustment (fine adjustment) section
(14) Rotation speed analog output trimmer section
Rotation speed analog output (ROUT) zero point and span point adjustment (fine adjustment) section
7
1-2. Rear Panel
(1) Power supply terminal board
Connected to the AC power supply and grounding wire.
(2) Protective ground terminal
Connect this to ground to eliminate noise effects such as static electricity. Do not connect any wires other than the grounding wire.
This connector is internally linked to F.G. on the power supply terminal board.
(3) External control input/output connector
Used to connect the external control device.
(4) Torque transducer/rotation speed detector connector
Connect the signal cable to the torque transducer/rotation speed detector (MP-9820).
(5) Analog output connector
Connect the signal cable to the analog input device.
(6) Optional part mounting section
One optional part (RS-232C, RS-422/485, PROFIBUS, or CANopen) can be mounted.
(1) Power supply terminal board
(2) Protective ground terminal
(6) Optional part mounting section
(3) External control input/output connector
(4) Torque transducer/rotation speed detector connector
(5) Analog output connector
8
2. Operating Instructions
This section describes how to operate the equipment using the keys on the front panel.
CAUTION
Stop measuring before operating the keys. There is a risk of malfunctioning if the keys are operated while measurement is in progress.
Key operations in measurement mode are enabled by depressing for approximately 1 second.
2-1. Key
2-1-1. When Operated in Measurement Mode
This selects function mode and is displayed on the setting display.
Function setting or other modes can be selected in this mode.
2-2. Key
2-2-1. When Operated in Measurement Mode
Pressing this button activates the auto-zero function and zeros the torque reading if the value displayed on the torque display is 10% or less of the maximum value. [A/Z] is also displayed on the status display.
2-2-2. When Operated in Other Modes
Pressing the key when a setting is displayed causes the flashing digit to move to the left in the sequence 10
1
, 102, 103, 104. (Note that the range of movement will vary depending on the number of
digits and sign displayed.)
2-3. Key
2-3-1. When Operated in Measurement Mode
This cancels the auto-zero function, and [A/Z] disappears on the status display.
2-3-2. When Operated in Other Modes
Pressing the key when a setting is displayed causes the flashing digit to move to the right in the sequence 10
4
, 103, 102, 101. (Note that the range of movement will vary depending on the number of
digits and sign displayed.)
2-4. Key
2-4-1. When Operated in Measurement Mode
This has no effect.
9
2-4-2. When Operated in Other Modes
Setting increment
Pressing the key when a setting is displayed increments the value in sequence from 0 to 9, and then back to 0.
Fine adjustment increment
Pressing the key for fine adjustment of the zero point, span point, asymmetry correction, linearization correction, or analog output increments the corresponding value.
2-5. Key
2-5-1. When Operated in Measurement Mode
This has no effect.
2-5-2. When Operated in Other Modes
Setting decrement
Pressing the key when a setting is displayed decrements the value in sequence from 0 to 1, and then back to 0.
Fine adjustment decrement
Pressing the key for fine adjustment of the zero point, span point, asymmetry correction, linearization correction, or analog output decrements the corresponding value.
2-6. Key
2-6-1. When Operated in Measurement Mode
[CHECK] illuminates on the status display, and the value (CHECK value) set in function mode F-20 is output from the torque analog output (V-OUT/I-OUT). For further details, refer to Section 9-2.
2-6-2. When Operated in Other Modes
This has no effect.
2-7. Key
Pressing the key registers the setting altered within the system.
10
3. Calibration
Calibration refers to the process of adjusting the display reading to match the torque acting on the torque transducer in order to ensure that the electrical signal from the torque transducer is displayed as an accurate torque figure.
For example, this adjustment ensures that an accurate reading of 1000.0 N·m is given when a torque of 1,000 N·m acts on the torque transducer.
The equipment must always be calibrated before using for the first time or after replacing a torque transducer.
When calibrating the equipment before using for the first time, calibrate by entering the numerical values, referring to the calibration certificate sheet provided when the torque transducer was purchased.
When calibrating the equipment after replacing a torque transducer, calibrate by entering the numerical values, referring to the calibration certificate sheet provided with the torque transducer purchased.
3-1. Settings Required for Calibration
(1) Minimum scale division
The minimum measurement interval. The available settings are [1], [2], [5], and [10]. The value set for [SPAN value/MIN. SCALE DIVISION] will form the display resolution.
(2) SPAN value
The maximum torque value that can be measured by the torque transducer.
(3) ZERO calibration
The setting to be calibrated to ensure that the equipment torque reading is zero when no torque is applied to the torque transducer (initial torque state). This can be calibrated either using the torque value (initial torque state) or by entering the torque transducer output value.
(4) SPAN calibration
The setting to be calibrated to ensure that the variations in the electrical signal from the torque transducer are displayed on the equipment correctly as the torque when a torque is applied to the torque transducer. This can be calibrated either using the torque value (span torque state) or by entering the torque transducer output value.
(5) Calibration data selection
The equipment can save up to four sets of calibration data. Select the location for saving the calibration data by selecting Calibration Data 1 to 4 before calibrating using F-59. If the equipment is jointly calibrated at Minebea, the calibration data will be stored to [Calibration Data 1].
3-2. Settings to be Altered As Necessary After Calibration
For details of the setting instructions, refer to Instruction Manual EN294-1596*.
(1) Symmetry correction
This function corrects torsional span errors in the clockwise and counter-clockwise directions.
(2) Linearization correction
This function minimizes measuring errors by correcting up to five points, excluding zero and span.
(3) Change polarity
This function inverts the polarity of the torque transducer output.
(4) Decimal place
Sets the decimal place for the torque reading on the equipment.
(5) Units
Adds units to the torque reading on the equipment.
Recalib rate as necessary if the usage environment changes.The maximum readout resolution for valid performance is 10,000. In span calibration, use a value of at least 2/3 of the span to minimize calibration errors.
11
3-3. Calibration Procedure
3-3-1. Calibration in Measurement Mode
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Connect the torque transducer to the equipment.
Turn on and wait for approximately 10 minutes for the torque transducer and equipment to stabilize.
Select calibration mode.
Select the minimum torque units.
Select the minimum torque units.
Set the maximum torque value that can be measured by the torque transducer.
Calibrate the zero point with the torque transducer at the initial torque state or by setting the zero point torque for the torque transducer.
Calibrate the span point with the span torque acting on the torque transducer or by setting the span point torque for the torque transducer.
Calibration is complete. Perform steps 10 onward as described below where necessary.
Steps 4 to 9 must be carried out in sequence. The calibration results will not be updated if the proced ure is not
completed up to step 9.
The A/Z function is canceled once calibration is completed (up to step 9). The A/Z function is canceled if step 10, 11, or 12 is performed. For details of steps 11 to 15, refer to Instruction Manual EN294-1596*.
Section 3-3-3
Minimum scale division setting
Section 3-3-4
Torque setting
Step 11
Ste
p
12 Step 13
Ste
p
14
Section 3-3-5
Actual torque setting
Section 3-3-7
Span calibration
Section 3-3-6
Zero calibration
Step 15
After turning on powe
r
Calibration mode selection
End of calibration
Decimal place
setting
Units setting
Symmetry correction
function
Linearization
correction
function
Polarity change
function
Step 10
Section 3-4
Zero/span fine
adjustment
function
Torque transducer connection
12
3-3-2. Selecting Calibration Mode
Calibration mode can be selected from measurement mode as follows.
appears in the upper part when the key is depressed for approximately 2 seconds in measurement mode.
The upper display reads when the key is pressed.
The upper display reads and
flashes on the lower display, indicating
calibration mode has been selected when the key is pressed. (If the equipment has been previously calibrated, the minimum scale division registered at that time will be displayed.)
3-3-3. Minimum Scale Division Setting
Set the minimum scale division for torque.
Use the or key to set the required minimum scale division on the lower display. Select from [1], [2], [5], or [10].
* Enter value (1) on the calibration certificate sheet. (No decimal point) For example, if the minimum digit is 100.00 N·m, set to [d-01].
: Selects the minimum scale division.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting. (If the equipment has been previously calibrated, the torque value registered at that time will be displayed.)
The maximum readout resolution for valid performance is 10,000.
Measurement mode
To minimum scale division setting
Registered minimum scale division
Registered minimum scale division
Minimum scale division used
To torque setting
Registered torque value
13
3-3-4. Torque Setting
Set the maximum torque displayed.
Use the keys to change the torque on the lower display to the required value.
* Enter value (2) on the calibration certificate sheet. (No decimal point) For example, if the maximum indication is 100.00 N·m, set to [10000].
: Alters the value of the digit to be
altered.
: Selects the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting.
New torque value
To actual torque setting
Registered torque value
New torque value
14
3-3-5. Actual Torque Setting
Set actual torque acting on the torque transducer (or torque transducer output). When calibrating by entering numerical values, set the same value as set in "3-3-4 Torque Setting".
* Enter value (3) on the calibration certificate sheet. (No decimal point)
For example, if the maximum indication is 100.00 N·m, set to [10000].
Use the keys to change the torque on the lower display to the required value.
: Alters the value of the digit to be altered.
: Selects the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting.
Actual torque value or torque meter output
To zero calibration
New torque value
15
3-3-6. Zero Calibration
Register the input at the zero point.
To calibrate by entering the torque value for the torque transducer, press the key and then press
the key.
Calibrate zero by entering the zero point torque (frequency) for the torque transducer.
Calibration by entering the torque value for the torque
transducer
Press the key to display in the
lower display, and then press the key to display the value.
The value displayed will be the output frequency for the torque transducer at the initial torque state previously registered.
Set the torque value (frequency) corresponding to the zero point in 1 Hz intervals.
* Enter the torque calibration value zero frequency included on the calibration certificate sheet for the torque transducer.
* Enter value (4) on the calibration certificate sheet. (No decimal point)
For example, if the torque calibration value (zero) is
9.957 kHz, set to [9957].
: Alters the value of the digit to be altered.
: Selects the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting.
Zero calibration error display
: The display flashes for approximately 2 seconds when the torque transducer frequency output is less than 9,500 Hz or if the value entered is less than 9,500 Hz.
: The display flashes for approximately 2 seconds when the torque transducer frequency output exceeds 10,500 Hz or if the value entered exceeds 10,500 Hz.
To span calibration
Hz
Registered torque transducer output value
Hz
Set torque transducer output value
16
3-3-7. Span Calibration
Register the input at the span point.
To calibrate by entering the torque value for the torque transducer, press the key and then press
the key.
Calibrate span by entering the span point torque (frequency) for the torque transducer.
Calibration by entering the torque value for the torque
transducer
Press the key to display in the
lower display, and then press the key to display
. The value displayed will be the output frequency for the torque transducer at the span point previously registered. Set the torque value (frequency) corresponding to the span point in 1 Hz intervals.
* Enter the torque calibration value span frequency included on the calibration certificate sheet for the torque transducer.
* Enter value (5) on the calibration certificate sheet. (No decimal point)
For example, if the torque calibration value (span) is
14.913 kHz, set to [14913].
: Alters the value of the digit to be altered.
: Selects the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting.
Span calibration error display
: The display flashes for approximately 2 seconds when [span point torque transducer output frequency or span point torque transducer value entered] - [zero point torque transducer output frequency or zero point torque transducer value entered] 0 or when the torque transducer output frequency is less than 14,500 Hz or if the value entered is less than 14,500 Hz.
: The display flashes for approximately 2 seconds when the torque transducer output frequency exceeds 15,500 Hz or if the value entered exceeds 15,500 Hz.
In span calibration, use a torque value of at least 2/3 of the display torque to minimize calibration errors.
Hz
Registered torque transducer output value
Hz
Set torque transducer output value
17
3-3-8. Calibration End
Displays after span calibration ends.
Press the key to exit calibration mode.
The system switches to measurement mode and the data set is registered.
Th e calib ration values (zero point and span) will not be entered unless the procedure is performe d in sequence
up to this step.
Calibration can be performed only in the counter-clockwise torsional direction. The A/Z function is canceled once calibration is completed.
Measurement mode
18
3-4. Zero/Span Fine Adjustment Function
This function fine-adjusts the zero point and span point if there are discrepancies between the actual torque and the reading displayed.
3-4-1. Selecting Zero Point Fine-adjustment Mode
Zero point fine-adjustment mode can be selected from measurement mode as follows.
appears in the upper part when the key is
depressed for approximately 2 seconds in measurement mode.
The upper display reads when the key is pressed twice.
The upper display reads and the reading on the lower display flashes, indicating zero point fine-adjustment mode
has been selected when the key is pressed. The lower display displays the torque value currently entered. Set the torque transducer to the initial torque state, and adjust the display to [ZERO].
: Increments the torque.
Hold down the key to increment the torque continuously.
: Decrements the torque.
Hold down the key to decrement the torque continuously.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next
step.
Press the key after setting.
is displayed after zero point fine-adjustment.
Press the key to exit zero point fine-adjustment mode. The system switches to measurement mode and the data set is registered.
Measurement mode
Current torque value
(twice)
Measurement mode
19
3-4-2. Selecting Span Point Fine-adjustment Mode
Span point fine-adjustment mode can be selected from measurement mode as follows.
.
appears in the upper part when the key is depressed for approximately 2 seconds in measurement mode.
The upper display reads when the key is pressed three times.
The upper display reads and the reading on the lower display flashes, indicating span point fine-adjustment
mode has been selected when the key is pressed. The lower display displays the torque value currently entered. Apply an actual torque to the torque transducer, and adjust the display to [actual torque value].
: Increments the torque. Hold down the key to increment the torque continuously.
: Decrements the torque. Hold down the key to decrement the torque continuously.
: Cancels the setting and returns to measurement
mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting.
is displayed after span point fine-adjustment.
Press the key to exit span point fine-adjustment mode. The system switches to measurement mode and the data set is registered.
The change polarity setting will be temporarily canceled for zero/span fine-adjustment. (Restored after
adjustment)
The A/Z function is canceled once the zero/span fine-adjustment function ends.
Measurement mode
Current torque value
(3 times)
Measurement mode
20
4. Function Mode
4-1. Selecting Function Mode
Function mode can be selected from measurement mode as follows.
appears in the upper part when the
key is depressed for approximately 2 seconds in
measurement mode.
The upper display reads when the
key is pressed.
Use the keys to select the required function number on the upper display.
: Alters the value of the digit to be altered.
: Selects the digit to be altered.
: Cancels the setting and returns to measurement mode.
: Registers the value displayed and proceeds to the next step.
Press the key after setting. The lower display changes to display the function details.
Use the keys to select the required function number on the lower display.
The upper display reads when the
key is pressed, indicating that function registration
is complete.
Press the key to exit function mode and return to measurement mode.
Measurement mode
Value set
Value registered
To measurement mode
Value set
21
4-2. Function List
Function No. Item
Setting
value
Details
F-01 Decimal place
0 No decimal point
1 1000.0
2 100.00
3 10.000
4 1.0000
F-02 Torque display units
0 N·m
1 kN·m
2 V
F-04 Display cycles
0 4 cycles/s
1 20 cycles/s
F-05 Torque analog filter
0 1 Hz
1 10 Hz
2 30 Hz
3 50 Hz
4 100 Hz
5 300 Hz
6 500 Hz
7 1 kHz
F-06 Torque digital filter
00~88
No. of moving-average samples = 2
m+2n
m: 101 setting, n: 100 setting
00
F-07 Rotation speed analog filter
0 1 Hz
1 10 Hz
F-08 Rotation speed digital filter
00~88
No. of moving-average samples = 2
m+2n
m: 101 setting, n: 100 setting
00
F-09
Torque stabilization filter No. of averaging
0 No. of moving-average samples: 1
1 No. of moving-average samples: 2
2 No. of moving-average samples: 4
3 No. of moving-average samples: 8
4 No. of moving-average samples: 10
5 No. of moving-average samples: 12
6 No. of moving-average samples: 14
7 No. of moving-average samples: 16
8 No. of moving-average samples: 32
F-10
Torque stabilization filter Time width setting
000~999
Unit: 10 ms 000: Torque stabilization filter off
000
F-11
Torque stabilization filter Data width setting
00~99
Unit: DIGIT 00: Torque stabilization filter off
20
22
Function No. Item
Setting
value
Details
F-15
Rotation speed stabilization filter No. of averaging
0 No. of moving-average samples: 1
1 No. of moving-average samples: 2
2 No. of moving-average samples: 4
3 No. of moving-average samples: 8
4 No. of moving-average samples: 10
5 No. of moving-average samples: 12
6 No. of moving-average samples: 14
7 No. of moving-average samples: 16
8 No. of moving-average samples: 32
F-16
Rotation speed stabilization filter Time width setting
000~999
Unit: 10 ms 000: Rotation speed stabilization filter off
000
F-17
Rotation speed stabilization filter Data width setting
00~99
Unit: DIGIT 00: Rotation speed stabilization filter off
20
F-18 Key lock
0000
~1111
0000
0: Off 1: On
10
0
digit:
10
1
digit:
10
2
digit:
10
3
digit:
F-19 Polarity change
0
Torque in counter-clockwise direction is positive.
1 Torque in clockwise direction is positive.
F-20 CHECK value
0~20
Refer to Section 9-2-3.
16
F-21
Torque analog output Max. display value
1~99 999
Unit: DIGIT
10 000
F-22
Rotation speed analog output Max. display value
1~27 500
Unit: DIGIT
25 000
F-23 Rotation direction polarity
0 +
1 -
F-24
Rotation detection input signal format
0 Off
1 Measured using 2 rotation detectors
2
Measured using 1 rotation detector. Rotation direction detected using external input signal
3
Measured using 1 rotation detector. Rotation direction cannot be detected.
23
Function No. Item
Setting
value
Details
F-25 No. of rotation detection gears
0 120
1 240
2 360
F-26
Selecting A/Z data save destination
0 RAM
1 EEPROM
F-30 RS-232C operating mode
0 Command mode
1 Stream mode
F-31 RS-232C stream output target
0 Torque display data
1 Input torque A/D data
2 Rotation speed display data
3 Frequency data
F-32
RS-232C/RS-422/RS-485 Baud rate
0 1,200 bps
1 2,400 bps
2 4,800 bps
3 9,600 bps
4 19,200 bps
5 38,400 bps
6 57,600 bps
7 115,200 bps
F-33
RS-232C/RS-422/RS-485 Data bit length
0 7 bit
1 8 bit
F-34
RS-232C/RS-422/RS-485 Parity bit
0 None
1 Odd
2 Even
F-35
RS-232C/RS-422/RS-485 Stop bit
0 1 bit
1 2 bit
F-36
RS-232C/RS-422/RS-485 Terminator
0 CR
1
CRLF
F-37
RS-232C/RS-422/RS-485 Decimal point
0 No
1 Yes
F-38 RS-422/485 address
00~31
00
F-39 RS-422/485 selection
0 RS-422
1 RS-485
F-40 RS-485 data delay time
000~999
Unit: 10 ms Can be set in range 0~9.99 s.
001
24
Function No. Item
Setting
value
Details
F-41
PROFIBUS Station No.
000~125
000
F-42
CANopen Node ID
001~127
001
F-43
CANopen Baud rate
0 10 kbps
1 20 kbps
2 50 kbps
3 100 kbps
4 125 kbps
5 250 kbps
6 500 kbps
7 800 kbps
8 1 Mbps
F-45
CANopen PDO output frequency
0~100
0: Off Unit: 1 ms
100
F-50 Maintenance 1 00000 0~99 999 (Do not use)
F-51 Maintenance 2 00000 0~99 999 (Do not use)
F-55 Calibration prohibited 0000
0: Calibration permitted 1: Calibration prohibited 10
0
digit: Calibration data 1
10
1
digit: Calibration data 2
10
2
digit: Calibration data 3
10
3
digit: Calibration data 4
F-56 Symmetry correction clear 0
Clear data corrected using symmetry correction function.
F-57
Clearing counter-clockwise linearization correction
0
Clear data corrected using linearization correction function. (Counter-clockwise direction)
F-58
Clearing clockwise linearization correction
0
Clear data corrected using linearization correction function. (Clockwise direction)
F-59 Calibration data selection
0 Calibration data 1
1 Calibration data 2
2 Calibration data 3
3 Calibration data 4
F-60 Calibration 1 Increment 1 Datum value (initial value)
F-61
Calibration 1 Maximum display value
10000 Datum value (initial value)
F-62
Calibration 1 Actual torque value
10000 Datum value (initial value)
F-63 Calibration 1 ZERO A/D 0x1FFFF Datum value (initial value)
F-64 Calibration 1 +SPAN A/D 0x3AAAA Datum value (initial value)
F-65 Calibration 1 -SPAN A/D 0x5555 Datum value (initial value)
F-66
Calibration 1 ZERO frequency conversion value
10000 Datum value (initial value)
F-67
Calibration 1 +SPAN frequency conversion value
15000 Datum value (initial value)
F-68
Calibration 1 -SPAN frequency conversion value
5000 Datum value (initial value)
25
Function No. Item
Setting
value
Details
F-70 Calibration 2 Increment 1 Datum value (initial value)
F-71
Calibration 2 Maximum display value
10000 Datum value (initial value)
F-72
Calibration 2 Actual torque value
10000 Datum value (initial value)
F-73 Calibration 2 ZERO A/D 0x1FFFF Datum value (initial value)
F-74 Calibration 2 +SPAN A/D 0x3AAAA Datum value (initial value)
F-75 Calibration 2 -SPAN A/D 0x5555 Datum value (initial value)
F-76
Calibration 2 ZERO frequency conversion value
10000 Datum value (initial value)
F-77
Calibration 2 +SPAN frequency conversion value
15000 Datum value (initial value)
F-78
Calibration 2 -SPAN frequency conversion value
5000 Datum value (initial value)
F-80 Calibration 3 Increment 1 Datum value (initial value)
F-81
Calibration 3 Maximum display value
10000 Datum value (initial value)
F-82
Calibration 3 Actual torque value
10000 Datum value (initial value)
F-83 Calibration 3 ZERO A/D 0x1FFFF Datum value (initial value)
F-84 Calibration 3 +SPAN A/D 0x3AAAA Datum value (initial value)
F-85 Calibration 3 -SPAN A/D 0x5555 Datum value (initial value)
F-86
Calibration 3 ZERO frequency conversion value
10000 Datum value (initial value)
F-87
Calibration 3 +SPAN frequency conversion value
15000 Datum value (initial value)
F-88
Calibration 3 -SPAN frequency conversion value
5000 Datum value (initial value)
F-90 Calibration 4 Increment 1 Datum value (initial value)
F-91
Calibration 4 Maximum display value
10000 Datum value (initial value)
F-92
Calibration 4 Actual torque value
10000 Datum value (initial value)
F-93 Calibration 4 ZERO A/D 0x1FFFF Datum value (initial value)
F-94 Calibration 4 +SPAN A/D 0x3AAAA Datum value (initial value)
F-95 Calibration 4 -SPAN A/D 0x5555 Datum value (initial value)
F-96
Calibration 4 ZERO frequency conversion value
10000 Datum value (initial value)
F-97
Calibration 4 +SPAN frequency conversion value
15000 Datum value (initial value)
F-98
Calibration 4 -SPAN frequency conversion value
5000 Datum value (initial value)
F-99 Memory clear -
Returns function settings to default settings.
26
SAMPLE
(1) (2)(3)
(4) (5)
.
D The contents of this manual may subject to change without notice.
HEAD QUARTER : MINEBEA CO., LTD.
410673 Miyota, Miyotamachi, Kitasakugun, Naganoken 3890293, Japan
0267322200 .0267311350
Measuring Components Business Unit
FUJISAWA PLANT 111, Katase, Fujisawashi Kanagawaken, 2518531 Japan
0466227152 .0466221701
KARUIZAWA PLANT 410673 Miyota, Miyotamachi, Kitasakugun, Naganoken 3890293, Japan
0267311309 .0267311350
HOMEPAGE ADDRESS http://www.minebeamcd.com
Loading...