Thank you very much for your purchasing our Digital Indicator CSD−891B.
This manual explains installation procedures and connecting method and also operating
method for the Digital Indicator CSD−891B. Make use of it properly after reading through
the manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the
manual at hand after reading it over.
I
Marks and arrangements used in this manual
The following marks are attached to the explanation on the matters that indicate “Don’t do
this.”, “Take care.” and “For reference”.
Be sure to read these items where these marks are attached.
Warning● Warning may cause injury or accident that may harm to the operator.
Don’t do these things described here.
● Caution during operation and working.
Be sure to read the item to prevent malfunction.
Mark during operation.
● Press the switch.
II
For safe operation
Be sure to read this instruction manual before use.
1. Installation place
● Use the instrument where the temperature/humidity specifies with
the range as follows:
Environmental temperature :−10 ℃ to 50 ℃
Environmental humidity:Less than 85 %R.H. (Non condensing)
(1) Location where installation is not allowed.
Warning● Don’t locate the instrument on the places as follows :
It may cause an unexpected faulty in the instrument.
D Do not locate the instrument in direct and/or high temperature area.
D Do not use the instrument in a high humid area.
D Do not install the instrument where there are vibrations and shocks.
D Do not use the instrument where there is excess of dusts and fine particles.
D Do not use the instrument where there are corrosive gas and salt and like
that.
D Do not install the instrument where there is rapid change of temperature
and humidity.
D Do not install the instrument near the devices that are magnetized or
generate an electromagnetic field.
D Do not install the instrument where the instrument may be affected by
radioactivity or radial rays.
D Avoid the location where chemical reaction may take place such as in a
laboratory, or like that.
III
(2) Installation
● When installing the instrument, install as referring to the following
figures and secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental
spaces are as follows:
Outline dimensions
208
200
Front
67
43
4
Side
140.7
4
185
IV
Unit:mm
2. Power supply
Warning● Be sure to check that the power supply is off in connecting each cable.
If the work is done while the power is on, there may have the case that
electric shock to the operator or even may have damage to the
instrument.
● Before supplyingthepower,checkthattheindicationofpowersupply
voltage/specifications for
supply should be the same.
If they are not equal, contact us.
If you use the instrument without checking them, it may cause a
damage in the instrument or electric shock to the operator.
the instrument and the power going to
● Earth wire should be grounded securely.
When earth wire is not connected, it may cause a malfunction of the
instrument or electric shock to the operator.
3. Application note
Warning● Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration. If calibration
will not be made, the correct measuring results may not be obtained
nor which may cause malfunction in the instrument and there may
exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the
similar case when the results are not correct, so make calibration,
again.
Warning● In case of using the instrument, check that the connections are
executed properly. If not connected properly, the correct measuring
result will not be obtained, nor it may cause malfunctions of the
instrument, damage to the peripheral equipments or even more
serious accidents.
V
Warning● When change of setting is made carelessly on the instrument during
measurement, correct measured results may not be obtained and it
may cause malfunction in the instrument and even have the
possibility of damage in peripheral instruments.
Warning● Do not shock the instrument such as throwing something on it.
If neglected, it may cause destruction of the parts and damage to the
electrical circuits.
Warning● Do not push the panel sheet on the instrument with the excessive
strong force nor push it with sharp edge object such as a driver.
If neglected, it may cause a damage to the panel switch and even have
the possibility of damage to resist to environments or operational
performances.
Warning● Don’t remove the cover of the case of the instrument, nor peel off the
panel sheet nor take the instrument into pieces.
If neglected, it may cause a damage to the case and the panel sheet
and even have the possibility of damage to resist to environments or
operational performances.
● At the time of shipment from the factory, the instrument has been
plated with a clear sheet on the panel sheet for protective purpose.
In case of application, use the instrument after removing the
clearsheet first.
VI
History of revision
Date Instruction manual No. Details of revised point
Oct. 2001 DRW. NO.EN294−1143
Jun. 2002 DRW. NO.EN294−1143-A
Sept. 2004 DRW. NO.EN294−1143-B
Apr. 2005 DRW. NO.EN294−1143-C
Aug. 2005 DRW. NO.EN294−1143-D
Version ROM Ver. 1.000 or later
Due to ECN No.FN02−02066
− Change −
Wiring instruction seal affixation is changed right, and the
outline drawing is also changed.
− Additional −
8−2., 13−3. Add F−87
Due to ECN No.FN04−02111
− Change −
7−14. “The backup time is over ten years.” has
changed to “The backup time is about ten years.”
Due to ECN No.FN05−02035
− Addition −
At the warning column in the wiring section, the clause of “
As there is a case which standard wiring color is different,
please confirm the inspection data sheet of the load cell
being used.” is added.
Due to ECN No.FN05−02085
− Correction −
11−8. General specifications
Outline dimensions from “208 mm x 67 mm x 143 mm”
to “208 mm x 67 mm x 140.7 mm”
Due to ECN No.FN10−02013
Jan. 2010 DRW.No.EN294-1143-E
Feb. 2010 DRW.No.EN294-1143-F
May. 2010 DRW.No.EN294-1143-G
Oct. 2010 DRW.No.EN294-1143-H
ROM Ver. 1.800 or later
− Addition −
8−2., 13−3. Add F−84
Due to ECN No.FN10−02013A
− Addition −
8−2. F−84 “Restriction and warning” is added.
Due to ECN No.FN10−02058B
− Addition −
7−3−1. Analog filter
7−7. Detection of stability
7−7−1. Range to detect stability
7−7−2. Time to detect stability
Due to ECN No.FN10−02026B
− Change −
Front cover logo is changed.
Due to ECN No.FN10−02140
− Change −
MInebea logo is changed.
VII
Date Instruction manual No. Details of revised point
Common for
external control
input and open
collector output
output
14LOCK
● The COM.(Terminal No.15) is common for the external control input
(Terminal No.10〜14) and the open collector output(Terminal No.16
〜22).
● Don’t connect with N.C. terminals(Terminal No.23 and 25).
6
4−2. Note on connection
Warning● In case of connection with the instrument, keep strictly to the
following items. If neglected, it may cause an unexpected failure or a
damage to the instrument.
D Be sure to set the power supply to OFF, when the connection will be made.
D Since the terminal boards at front of the instrument is made of resin, take care not to drop
it down or not to apply strong impact.
D Recommended torque to tighten the terminal screws for terminal board should be as
follows.
Torque to tighten the terminal screws
Terminals0.6 N・m
D The suitable crimp type terminal lugs for the terminal board are as follows:
Width of crimp type terminal lugsSuitable crimp type terminal lugs
Terminals6.2 mm or less1.25−3orY−type 1.25−3.5
D Connecting cable with the instrument should be away from the noise source such as power
supply line and/or I/O line for control and so on as far as possible.
D Conduit wiring should be the type of exclusive one, and avoid using with another line
together.
D All of the connections should be executed securely by referring to the Instruction manual
for the instrument.
7
4−3. Connection
4−3−1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so the
connection with another type of strain gage applied transducers shall be proceeded in the same
way.
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.4 and
“Blue” with Terminal No.2 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
※3 When the length of CAB−501 is more than 100 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the remote sensing function not worked
fully.
8
1
Connection with 1 piece of load cell and CSD−891B
Shield
3m
Green ※1
Red
Blue ※1
Attached
short bar
White
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.4 and
“Blue” with Terminal No.2 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
※3 When the length of CAB−501 is more than 100 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the remote sensing function not worked
fully.
9
2
Connection with 1 piece of load cell and a junction box for extension use(B−304) and
CSD−891B
i) When CAB−502(4−cores cable)is used.
Shield
Green ※1
Red
Blue ※1
Attached
short bar
White
3m
Junction box
B−304
CAB−502
※2 (The length of CAB−502 is within 30 m totally.)
Internal wiring diagram of B−304
to load cell
REDWHTBLU
GRN YEL(Shield)
Terminal pitch 9.5 mm
REDWHTBLU
GRN
Suitable crimp type terminal lugs
:1.25−4or2−4
YEL(Shield)
to CSD−891B
10
ii) When CAB−501(6−cores cable) is used
Junction box
3m
B−304
CAB−501
Shield
Green ※1
Red
Blue ※1
Orange
Black
White
※2 (The length of CAB−501 is within 100 m totally.)
Internal wiring diagram of B−304
To Load cell
BLU
Terminal Pitch 9.5 mm
ORN
REDWHT
BLU
RED
BLK
WHT
GRN
GRN
YEL(Shield)
Suitable crimp−type
terminal lugs:
1.25−4or2−4
YEL(Shield)
To CSD−891B
11
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.4 and
“Blue” with Terminal No.2 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
※3 When the length of CAB−501 is more than 100 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the remote sensing function not worked
fully.
3
Connection with 2 to 4 points of load cells and Summing type junction box(B−307) and
CSD−891B
i) When CAB−502(4−cores cable) is used.
Junction box
B−307
CAB−502
Shield
Green ※1
Red
Blue ※1
Attached
short bar
White
※2 (The length of CAB−502 is within 30 m totally.)
12
Internal wiring diagram of B−307
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
1W
1R
1G
1B
1Y
2W
2R
2G
2B
2Y
3W
3R
3G
3B
3Y
4W
4R
4G
4B
4Y
ii) When CAB−501(6−cores cable) is used.
RD
OR
BK
WH
GN
BL
YE
RED
WHT
GRN
BLU
YEL(Shield)
to CSD−891B
Shield
Green ※1
Red
Junction box
Blue ※1
Orange
Black
White
B−307
CAB−501
※2 (The length of CAB−501 is within 100 m totally.)
13
Internal wiring diagram of B−307
WHT
RED
GRN
BLU
YEL(SHIELD)
WHT
RED
GRN
BLU
YEL(SHIELD)
WHT
RED
GRN
BLU
YEL(SHIELD)
WHT
RED
GRN
BLU
YEL(SHIELD)
1W
1R
1G
1B
1Y
2W
2R
2G
2B
2Y
3W
3R
3G
3B
3Y
4W
4R
4G
4B
4Y
RD
OR
BK
WH
GN
BL
YE
RED
ORG
BLK
WHT
GRN
BLU
YEL(SHIELD)
To CSD−891B
14
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.4 and
“Blue” with Terminal No.2 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
※3 When the length of CAB−501 is more than 100 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the remote sensing function not worked
fully.
4
Connection with 2 to 4 points of load cells and Summing type junction box (SB−310) and
CSD−891B.
i) When CAB−502(4−cores cable) is used.
Junction box
SB−310
CAB−502
Shield
Green ※1
Red
Blue ※1
Attached
short bar
White
※2 (The length of CAB−502 is within 30 m totally.)
15
Internal wiring diagram of SB−310
Output port
to CSD−891B
RD
Short bar
Input port No.1
YERDWHBLGR
WH
GR
Input port
No.4
BL
YE
Suitable crimp type
A
terminal lugs:
B
・Y type・Round type
ii) When CAB−501(6−cores cable) is used.
Junction box
SB−310
or
YERDWHBLGR
YERDWHBLGR
φB
φA
Shield
Green ※1
Red
Blue ※1
Orange
Black
White
Input port No.2
YERDWHBLGR
Input port No.3
A:3.2 mm or less
B:7 mm or less
CAB−501
※2 (The length of CAB−501 is within 100 m totally.)
16
Loading...
+ 200 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.