Note : Please read this Instruction Manual carefully before use.
Be sure to follow the items that require attention described in the manual.
Keep the manual at hand so that you can pick it up and read it as soon as
necessity requires.
EN294−1127−K
Forwards
Thank you very much for your purchasing Minebea’s Digital Indicator CSD−701B.
This manual explains installation procedures and connecting method and also operating
method for the Digital Indicator CSD−701B. Make use of it properly after reading through
the manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the
manual at hand after reading it over.
This manual is intended for the technical experts to read.
●The contents of the manual may subject to change for improvement without notice.
I
Marks and arrangements used in this manual
The following marks are attached to the explanation on the matters that indicate “Don’t do
this.”, “Take care.” and “For reference”.
Be sure to read these items where these marks are attached.
Warning● Warning may cause injury or accident that may harm to the operator.
Don’t do these things described here.
● Caution during operation and working.
Be sure to read the item to prevent malfunction.
Mark during operation.
● Press the switch.
II
For safe operation
Be sure to read this instruction manual before use.
1. Installation place
● Use the instrument where the temperature/humidity specifies with
the range as follows:
Environmental temperature :−10 ℃ to 50 ℃
Environmental humidity:Less than 85 %R.H. (Non condensing)
(1) Location where installation is not allowed.
Warning● Don’t locate the instrument on the places as follows :
It may cause an unexpected faulty in the instrument.
・ Don’t locate the instrument in direct and/or high temperature area.
・ Don’t use the instrument in a high humid area.
・ Don’t install the instrument where there are vibrations and shocks.
・ Don’t use the instrument where there is excess of dusts and fine particles.
・ Don’t use the instrument where there are corrosive gas and salt and like
that.
・ Don’t install the instrument where there is rapid change of temperature
and humidity.
・ Don’t install the instrument near the devices that are magnetized or
generate an electromagnetic field.
・ Don’t install the instrument where the instrument may be affected by
radioactivity or radial rays.
・ Avoid the location where chemical reaction may take place such as in a
laboratory, or like that.
III
(2) Installation
● When installing the instrument, install as referring to the following
figures and secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental
spaces are as follows:
Outline dimensions
Front
Rear
Side
Upper
Panel cut size
Unit:mm
IV
2. Power supply
Warning● Be sure to check that the power supply is off in connecting each cable.
If the work is done while the power is on, there may have the case that
electric shock to the operator or even may have damage to the
instrument.
● Before supplying the power, check that the indication of power supply
voltage/specifications for the instrument and the power going to
supply should be the same.
If they are not equal, contact with Minebea.
If you use the instrument without checking them, it may cause a
damage in the instrument or electric shock to the operator.
● Earth wire should be grounded securely.
When earth wire is not connected, it may cause a malfunction of the
instrument or electric shock to the operator.
3. Application note
Warning● Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration. If calibration
will not be made, the correct measuring results may not be obtained
nor which may cause malfunction in the instrument and there may
exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the
similar case when the results are not correct, so make calibration,
again.
Warning● In case of using the instrument, check that the connections are
executed properly. If not connected properly, the correct measuring
result will not be obtained, nor it may cause malfunctions of the
instrument, damage to the peripheral equipments or even more
serious accidents.
V
Warning● When change of setting is made carelessly on the instrument during
measurement, correct measured results may not be obtained and it
may cause malfunction in the instrument and even have the
possibility of damage in peripheral instruments.
Warning● Do not shock the instrument such as throwing something on it.
If neglected, it may cause destruction of the parts and damage to the
electrical circuits.
Warning● Do not push the panel sheet on the instrument with the excessive
strong force nor push it with sharp edge object such as a driver.
If neglected, it may cause a damage to the panel switch and even have
the possibility of damage to resist to environments or operational
performances.
Warning● Don’t remove the cover of the case of the instrument, nor peel off the
panel sheet nor take the instrument into pieces.
If neglected, it may cause a damage to the case and the panel sheet
and even have the possibility of damage to resist to environments or
operational performances.
● At the time of shipment from the factory, the instrument has been
plated with a clear sheet on the panel sheet for protective purpose.
In case of application, use the instrument after removing the
clearsheet first.
VI
History of revision
DateInstruction Manual No.Details of revised point
Dec. 2000
Mar. 2001
Mar. 2001
DRW.NO.EN294−1127First Version ROM Ver. 1.000 or later
ECN No.FN01−02042
− Correction −
P.49 7−1−1. Input signal for external control
“During the holding operation, it is reset condition by short
DRW.NO.EN294−1127−A
DRW.NO.EN294−1127−B
circuit.”→“During the peak hold operation, reset condition
is made by short circuit.”
P.91 9−2−6 Output condition at Positive logic
Output data “Yes”→Condition of transistor “OFF”
Output data “No”→Condition of transistor “ON”
ECN No.FN01−02041ROM Ver.1.300 or later
− Correction −
5−2−1. 6th step.
Add the clause “
5−2−3. 12th step.
Change the display of “SP−L” and “SP−H”
11−1. Specifications for analogInput filter
“2 Hz”→ “1 Hz(At the “0” setting of the digital filter and
stabilized filter)”
− Change −
7−13. “Key lock”→ “Lock of key function”,
According to this change, following sentence is corrected.
8−2. F−06 “Setting key lock”
→ “Setting the lock of key function”
According to this change, following setence is also
corrected.
It can be increased 〜”
Jun. 2001
Jul. 2001
Sep. 2001
May. 2003
Aug. 2003
DRW.NO.EN294−1127−C
DRW.NO.EN294−1127−D
DRW.NO.EN294−1127−E
DRW.NO.EN294−1127−F
DRW.NO.EN294−1127−G
ECN No.FN01−02085
− Correction ‐
Change the outline dimensions (upper part)
− Additions ‐
9−6−2., 9−7−2, 11−11−6. add the “Power consumption”.
−Correction−
9−2−4.
(1)Equivalent circuit of input section “COM.1”→ “COM.”
(2)Equivalent circuit of output section “▼”→ “▽”
13−1. Replacement of fuse140..................................................
13−2. Character’s pattern for display142........................................
13−3. Setting table for functions143............................................
1. General
The instrument is a digital indicator for the application of strain gage applied transducer.
1−1. Features
Main features for CSD−701B are as follows :
(1) Compact size and light weight
48 mm×96 mm×120 mm Approx.0.3 kg(Without any options)
(2) The A/Z function and the peak holding function can be used by the selection.
1
2. Name and function of each point
2−1. Front panel
①
②
④
⑤
⑥
1
Load display section
The load data is shown in the Measurement mode, and status or set value is shown in various
kinds of Calibration mode and Setting mode.
2
Judgement display
Compared results by comparator function can be displayed.
3
Status display
◎Lights up when selecting A/Z or A/Z OFF function in the Function mode.
In the same way, light off when selecting Peak hold function in the Function
mode.
HOLDLights up when between the HOLD and COM.1 at the external control input is
shorted.
A/ZLights up at the time of executing Tare weight cancellation(A/Z ON).
Lights off with the tare weight cancellation clear.
(Effective when selecting the A/Z or A/Z OFF function in the function mode.)
CHECKLights up when the CHECK is ON.
PEAKLights up when selecting the Peak mode.
③
⑩
⑨
⑧
⑦
4
key
Used when shifting to the Function mode.
Also, by pressing this key and the
turned on and off
5
key
Used when calling the S1 set value changeover mode, or carrying digit at the time of various
kinds of settings.
key together, it is used when the CHECK value is
2
6
key
Used when calling the S2 set value changeover mode, or for the increment of values at the time
of various kinds of settings.
Also, by pressing this key and the
zero adjustment).
7
key
Used for making the changeover mode of Peak/Track, or executing the Tare weight
cancellation (A/Z ON).
(Changeover of Peak hold function and A/Z function depends on the setting of Function mode.)
8
key
Used for the reset of Peak value, or for Tare weight cancellation clear(A/Z OFF).
(Changeover of Peak hold function and A/Z function depends on the setting of Function mode.)
9
key
Used for registering set values at the time of various kinds of settings.
Also, by pressing this key and the
turned on and off.
Also, by pressing this key and
adjustment).
key together, it will execute the set of zero(One−touch
key together, it is used when the CHECK value is
key together, it will execute the set of zero(One−touch zero
10
Position of pasting the Unit seal
As necessity requires, paste the Unit seal attached.
2−2. Rear panel
②
1
Terminal board
Connects with external control input, contact output, various kinds of strain gage applied
transducer such as load cell, analog output, AC power supply, and a grounding wire.
2
Installing section for options
Whichever one can be installed from the optional BCD−OUT, RS−232C, RS−422 or serial
interface.
①
3
3. Installation procedures
3−1. Installation place
● Use the instrument where the temperature/humidity specifies within
the range as follows:
Environmental temperature:−10 ℃ to 50 ℃
Environmental humidity:85 %RH or less(Non condensing.)
3−2. Location where installation is not allowed.
Warning● Don’t locate the instrument on the places such as follows:
It may cause an unexpected faulty in the instrument.
D Do n ot expose the instrument in direct sunlight and/or high temperature area.
D Do n ot use the instrument in a high humid area.
D Do n ot install the instrument where there is high mechanical vibrations and shock.
D Do n ot use the instrument where there are excess of dusts and fine particles.
D Do n ot install the instrument where there include any corrosive gas or any salty
atmosphere.
D Do not install the instrument where there is rapid change of temperature and humidity.
D Do n ot install the instrument near the devices that are magnetized or generate an
electromagnetic field.
D Do n ot install the instrument where there may suffer radioactivity or radioactive rays.
D Avoid the location where chemical reaction may take place such as in a laboratory, or like
that.
4
3−3. Installation
● When installing the instrument, install as the following figures and
secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental spaces are as
follows:
Outline dimensions
Front
Side
Rear
Panel cut size
Upper
Unit:mm
5
3−4. Applicable environment
Warning: The instrument may subject to use in a highly humid area or in full of powder dust.
In such a case, use the instrument by inserting the panel mount gasket attached between the con-
trol panel (cabinet) and the main body
.
By inserting the panel mount gasket, the front panel section becomes IP65 (Inter-
national Protection Code) or equivalent in dust−proof and water−proof construc-
tion.
* Care should be taken when handling the panel mount gasket, since there are up
and down directions.
6
4. Connecting method
External
control
applied
p
Anal
input
4−1. Layout of the terminal boards
There is one terminal board containing 21 point of terminals in the panel.
Layout of terminal boards are shown in the following figure.:
Termin
al No.
1A11
2BStrain gage
3C
4D14S1
5E15S2
6A−OUT +
7A−OUT −
8COM.1
9ZERO
10HOLD
DescriptionsApplications
transducer
og output
External control
Termin
al No.
12
13COM.2
16F. GFrame ground
17SOURCEAC power supply
18N.C.
19SOURCEAC power supply
20N.C.
21Ground
DescriptionsApplications
PEAK/TRACK
or A/Z
RESET or A/Z
OFF
External control
input
Contact output
● The COM.1(Terminal No.8) and COM.2(Terminal No.13) are isolated.
7
4−2. Note on connection
Warning● In case of connection with the instrument, keep strictly to the
following items. If neglected, it may cause an unexpected failure
or a damage to the instrument.
D Be sure to set the power supply to OFF, when the connection will be made.
D Since the terminal boards at rear side of the instrument is made of resin, take care not to
drop it down or not to apply strong impact.
D Recommended torque to tighten the terminal screws for terminal board should be as
follows:
Torque to tighten
the terminal screws
Terminal board0.6 N・m
D The suitable crimp type terminal lugs for the terminal board are as follows:
Width of crimp
type terminal lugs
Terminal board6.2 mm or less1.25−3 or Y type 1.25−3.5
D Connecting cable with the instrument should be away from the noise source such as power
supply line and/or I/O line for control and so on as far as possible.
D Conduit wiring should be the type of exclusive one, and avoid using with another line
together.
D All of the connections should be executed securely by referring to the Instruction manual
for the instrument.
Suitable crimp type
terminal lugs
8
4−3. Connection
4−3−1. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure
transducer and so on. Here, we will describe the example of connections with load cell, so the
connection with other type of strain gage applied transducers shall be proceeded in the same
way.
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
1
Connection with 1 piece of load cell and CSD−701B
3m
※1※
1
RED
WHT
BLU
GRN
Shield
9
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
2
Connection with 1 piece of load cell and Junction box for extension use(B−304) and
CSD−701B.
Junction box
3m
※2 (Total length of CAB−502 is within 30 m)
Internal wiring diagram of B−304
to load cell
Terminal pitch 9.5 mm
B−304
CAB−502
REDWHTBLU
REDWHTBLU
※1※
RED
BLU
GRN YEL(Shield)
Suitable crimp type terminal lugs
:1.25−4or2−4
YEL(Shield)
GRN
1
WHT
GRN
Shield
to CSD−701B
10
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
3
Connection with 2 to 4 pieces of load cells, Summing type junction box(B−307) and
CSD−701B.
Junction box
B−307
CAB−502
※2 (Total length of CAB−502 is within 30 m)
Internal wiring diagram of B−307
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
WHT
RED
GRN
BLU
YEL(Shield)
1W
1R
1G
1B
1Y
2W
2R
2G
2B
2Y
3W
3R
3G
3B
3Y
4W
4R
4G
4B
4Y
※1※
RED
BLU
1
WHT
GRN
Shield
RED
RD
OR
BK
WHT
WH
GN
BL
YE
GRN
BLU
YEL(Shield)
to CSD−701B
11
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
※2 When the length of CAB−502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the
resistance of cable makes the input voltage of the instrument
decreased.
4
Connection with 2 to 4 pieces of load cells, Summing type junction box(SB−310) and
CSD−701B.
SB−310
CAB−502
※2 (Total length of CAB−502 is within 30 m)
Internal wiring diagram of SB−310
Output port
to CSD−701B
RD
WH
GR
Input port
No.4
BL
YE
※1※
RED
BLU
Short bar
YERDWHBLGR
YERDWHBLGR
YERDWHBLGR
YERDWHBLGR
1
WHT
GRN
Shield
Input port No.1
Input port No.2
Input port No.3
Suitable crimp type
terminal lugs:
A
B
or
φB
φA
A:3.2 mm or less
B:7 mm or less
・Y type・Round type
12
4−3−2. Connection with external control inputs
Connections with external control input “ZERO”,“HOLD”, “PEAK/TRACK” and “RESET”
should be made according to the below figures by using a contact point or an open collector
between the each terminal and terminal No. 8 at “COM.1”
Refer to the paragraph 7−1 for the function of each input.
or
Shield
Warning● Connections with external control outputs should be made securely
according to the figures. If neglected, it may cause an unexpected
failure and/or malfunction to the instrument.
● For the connections with external control inputs, be sure to apply
shielded cable, and the shielded cable should be connected with GND
terminal of the instrument.(Terminals No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
13
4−3−3. Connection with contact outputs
Connections with contact outputs “S1” and “S2” and external load should be made by using
each terminal and terminal No. 13 at “COM.2”.
At the same time, take care that the load should not exceed the rated load of contact output.
Rated load of contact outputAC125 V0.1 A (Resistance load)
DC30 V0.5 A (Resistance load)
load
Surge preventive element
〜
or
Shield
or
Warning● Connections with contact outputs should be made securely according
to the figures and also within the rated capacity of the instrument. If
neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
● For the protection from the contact of the instrument, connect the
surge preventive element that satisfies the characteristics of external
load to connect. If neglected, it may cause unexpected failure and/or
malfunction due to the effects from damage/melt down of the contact
and so on.
● For the connections with contact outputs, be sure to apply shielded
cable, and the shielded cable should be connected with GND terminal
of the instrument.(Terminal No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
14
4−3−4. Connection with the power supply and the earth
Connections with the power supply and the earth should be made as the following figure.
Grounding should be the D class with single earth.
Power supply voltageAC100 V to AC240 V
(Allowable variable range:AC85 V to AC264 V)
Frequency for power supply50/60 Hz
Power consumptionApprox.11 VA at maximum. (at AC100 V)
D class with single earth
〜
AC100 V to AC240 V
(Allowable variable range
:AC85 V to AC264 V)
Warning● Connections with the power supply and the earth should be made
securely according to the figures and also within the rated capacity of
the instrument. If neglected, it may cause an unexpected cause of
failure.
● Grounding should be the D class with single earth.
If neglected, it may cause an unexpected malfunction due to the
effects of noise from other equipments.
15
4−3−5. Connection with analog outputs
(1) In case of standard instrument (voltage output)
The instrument prepares output voltage for analog outputs as a standard.
Connections with voltage output should be made as the following figure.
Voltage outputDC0 V to 5 V
Over−rangeAt “−OL” displayApprox.−1.25 V
At “OL” displayApprox.6.25 V
Load resistance5 kΩ or more
Shield
−
V
+
Warning● Connections with voltage outputs should be made securely according
to the figures and also within specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
● For the connections with voltage outputs, be sure to apply shielded
cable, and the shielded cable should be connected with the F.G.
terminal of the instrument(Terminal No.16).
If not connected, it may cause malfunction due to the effects from
external noises and so on.
16
(2) In case of option (current output)
Connections with current output should be made as the following figure.
Current outputDC4 mA to 20 mA
Over−rangeAt “−OL” displayApprox.0 mA
At “OL” displayApprox.24 mA
Load resistance260 Ω or less
Shield
−
A
+
Warning● Connections with current outputs should be made securely according
to the figures and also within the specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
● For the connections with current outputs, be sure to apply shielded
cable, and the shielded cable should be connected with the F.G.
terminal of the instrument.(Terminal No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
17
Loading...
+ 127 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.