Minebea CSD-701B Instruction Manual

MINEBEA CO., LTD.
INSTRUCTION MANUAL
DIGITAL INDICATOR CSD−701B
Note : Please read this Instruction Manual carefully before use.
Be sure to follow the items that require attention described in the manual. Keep the manual at hand so that you can pick it up and read it as soon as necessity requires.
EN2941127K
Forwards
Thank you very much for your purchasing Minebea’s Digital Indicator CSD701B.
This manual explains installation procedures and connecting method and also operating
method for the Digital Indicator CSD701B. Make use of it properly after reading through
the manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the manual at hand after reading it over.
This manual is intended for the technical experts to read.
The contents of the manual may subject to change for improvement without notice.
I
Marks and arrangements used in this manual
The following marks are attached to the explanation on the matters that indicate “Don’t do
this.”, “Take care.” and “For reference”.
Be sure to read these items where these marks are attached.
Warning Warning may cause injury or accident that may harm to the operator.
Don’t do these things described here.
● Caution during operation and working.
Be sure to read the item to prevent malfunction.
Mark during operation.
Press the switch.
II
For safe operation
Be sure to read this instruction manual before use.
1. Installation place
● Use the instrument where the temperature/humidity specifies with
the range as follows
Environmental temperature :−10 to 50 Environmental humidity Less than 85 %R.H. (Non condensing)
(1) Location where installation is not allowed.
Warning Don’t locate the instrument on the places as follows :
It may cause an unexpected faulty in the instrument.
Don’t locate the instrument in direct and/or high temperature area.Don’t use the instrument in a high humid area.
Don’t install the instrument where there are vibrations and shocks.Don’t use the instrument where there is excess of dusts and fine particles.Don’t use the instrument where there are corrosive gas and salt and like
that.
Don’t install the instrument where there is rapid change of temperature
and humidity.
Don’t install the instrument near the devices that are magnetized or
generate an electromagnetic field.
Don’t install the instrument where the instrument may be affected by
radioactivity or radial rays.
Avoid the location where chemical reaction may take place such as in a
laboratory, or like that.
III
(2) Installation
● When installing the instrument, install as referring to the following
figures and secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental
spaces are as follows:
Outline dimensions
Front
Rear
Side
Upper
Panel cut size
Unitmm
IV
2. Power supply
Warning Be sure to check that the power supply is off in connecting each cable.
If the work is done while the power is on, there may have the case that
electric shock to the operator or even may have damage to the instrument.
● Before supplying the power, check that the indication of power supply
voltage/specifications for the instrument and the power going to
supply should be the same.
If they are not equal, contact with Minebea. If you use the instrument without checking them, it may cause a
damage in the instrument or electric shock to the operator.
● Earth wire should be grounded securely.
When earth wire is not connected, it may cause a malfunction of the
instrument or electric shock to the operator.
3. Application note
Warning Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration. If calibration
will not be made, the correct measuring results may not be obtained
nor which may cause malfunction in the instrument and there may
exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the similar case when the results are not correct, so make calibration,
again.
Warning In case of using the instrument, check that the connections are
executed properly. If not connected properly, the correct measuring
result will not be obtained, nor it may cause malfunctions of the
instrument, damage to the peripheral equipments or even more
serious accidents.
V
Warning When change of setting is made carelessly on the instrument during
measurement, correct measured results may not be obtained and it
may cause malfunction in the instrument and even have the
possibility of damage in peripheral instruments.
Warning Do not shock the instrument such as throwing something on it.
If neglected, it may cause destruction of the parts and damage to the
electrical circuits.
Warning Do not push the panel sheet on the instrument with the excessive
strong force nor push it with sharp edge object such as a driver.
If neglected, it may cause a damage to the panel switch and even have
the possibility of damage to resist to environments or operational
performances.
Warning Don’t remove the cover of the case of the instrument, nor peel off the
panel sheet nor take the instrument into pieces. If neglected, it may cause a damage to the case and the panel sheet
and even have the possibility of damage to resist to environments or
operational performances.
● At the time of shipment from the factory, the instrument has been
plated with a clear sheet on the panel sheet for protective purpose.
In case of application, use the instrument after removing the
clearsheet first.
VI
History of revision
Date Instruction Manual No. Details of revised point
Dec. 2000
Mar. 2001
Mar. 2001
DRW.NO.EN2941127 First Version ROM Ver. 1.000 or later
ECN No.FN0102042
Correction
P.49 711. Input signal for external control “During the holding operation, it is reset condition by short
DRW.NO.EN2941127A
DRW.NO.EN2941127B
circuit.”“During the peak hold operation, reset condition
is made by short circuit.” P.91 926 Output condition at Positive logic
Output data “Yes”Condition of transistor “OFF” Output data “No”Condition of transistor “ON”
ECN No.FN0102041 ROM Ver.1.300 or later
Correction
521. 6th step.
Add the clause “ 523. 12th step. Change the display of “SPL” and “SPH” 111. Specifications for analog Input filter
“2 Hz”“1 Hz(At the “0” setting of the digital filter and
stabilized filter)”
Change 713. “Key lock”“Lock of key function”,
According to this change, following sentence is corrected. 82. F06 “Setting key lock”
“Setting the lock of key function”
According to this change, following setence is also corrected.
It can be increased
Jun. 2001
Jul. 2001
Sep. 2001
May. 2003
Aug. 2003
DRW.NO.EN2941127C
DRW.NO.EN2941127D
DRW.NO.EN2941127E
DRW.NO.EN2941127F
DRW.NO.EN2941127G
ECN No.FN0102085
Correction
Change the outline dimensions (upper part)
Additions
962., 972, 11116. add the “Power consumption”.
Correction
924.
(1)Equivalent circuit of input section “COM.1”“COM.” (2)Equivalent circuit of output section “▼”→ “▽”
Due to ECN No.FN0102134
Change
Operating temperature range
“0 ℃ to 50 ℃”→ “−10 ℃ to 50 ℃”
Due to ECN No.FN0302040
Correction
435.(2) “4mA to 20mA” to “DC4mA to 20mA”
724., (1), 7154., 62., 911., 921., 922.
931., 117., “Display interlock”“PEAK/Net weight”
Due to ECN No.FN0302118
Correction 112. Specifications for digital section
Display range “99 999”“9 999”
VII
Date
Apr. 2005
Nov. 2007
Instruction Manual No. Details of revised point
Due to ECN No.FN0502035
Addition
DRW.NO.EN2941127H
DRW.NO.EN2941127I
At the warning column in the wiring section, the clause of “ As there is a case which standard wiring color is different,
please confirm the inspection data sheet of the load cell being used.” is added.
Due to ECN No.FN0702126
Correction 923. “DC37PN made by JAE”
to “Equivalent to DC37PNR made by JAE”
934. “DE9SN (JAE)”
to “Equivalent to DE9SNR made by JAE”
May, 2010
Oct. 2010
DRW.NO.EN2941127J
DRW.NO.EN2941127K
Due to ECN No.FN1001019A
Addition 1110.Accessories panel mounting gasket is added.
Due to ECN No.FN1002140
Change Minebea logo is changed.
VIII
INDEX
Forwards .......................................................
Marks and arrangements used in this manual ....................................
For safe operation ............................................................
1. Installation place .........................................................
2. Power supply ............................................................
3. Application note .........................................................
History of revision ............................................................
1. General 1.................................................................
11. Features 1............................................................
2. Name and function of each point 2..........................................
21. Front panel 2.........................................................
22. Rear panel 3..........................................................
3. Installation procedures 4...................................................
31. Installation place 4....................................................
32. Location where installation is not allowed. 4.............................
33. Installation 5.........................................................
34. Applicable environment 6..............................................
4. Connecting method 7......................................................
41. Layout of the terminal boards 7.........................................
42. Note on connection 8..................................................
43. Connection 9..........................................................
431. Connection with strain gage applied transducers 9..................
432. Connection with external control inputs 13..........................
433. Connection with contact outputs 14.................................
434. Connection with the power supply and the earth 15...................
435. Connection with analog outputs 16.................................
5. Calibration procedures 18...................................................
51. Preparations 18........................................................
52. Calibration procedures 18...............................................
521. Calibration method to register the output of strain gage applied transducer
at the time of maximum display after setting the load to zero 19........
522. Calibration procedures to register the output of strain gage applied
transducer at the time of zero and the maximum display 24............
523. Calibration method to register by reading output value of
strain gage applied transducer in the conditions of
zero/actual load application individually 29..........................
524. Zero fine adjustment 35............................................
525. Span fine adjustment 37...........................................
526. Calibration procedure to apply registration again for zero point only 39
53. Selection of calibration methods on each condition 41......................
531. In case of executing the calibration on the instrument newly. 41.......
532. When the calibration is executed again. 45...........................
54. Setting the prohibition against calibration 45..............................
6. Operation procedure 46.....................................................
61.
key 46............................................................
611. Operations in Measurement mode 46...............................
62.
key 46............................................................
621. When operating in the measurement mode 46........................
622. Operation is made in another mode 47..............................
63.
key 47............................................................
631. When operated in the Measurement mode. 47........................
632. When operated in another modes 48................................
64.
key 48............................................................
641. When operated in the Measurement mode. 48........................
642. When operated in another modes 48................................
65.
key 49............................................................
651. When operated in the Measurement mode 49........................
652. When operated in another modes. 49................................
66.
key 49............................................................
7. Function and operation 50...................................................
71. External control input signal and contact output signal 50..................
711. Input signal for external control 50.................................
712. Contact output signal 51...........................................
713. Equivalent circuit 51..............................................
72. Comparator 52.........................................................
721. ON/OFF of comparator S1 and S2 52................................
722. Change of set value 52.............................................
723. Operation on comparator S1 and S2 54..............................
724. Comparative target for comparator S1 and S2 55.....................
725. Hysteresis on comparator 56.......................................
73. How to use the filter 58.................................................
731. Digital filter 58...................................................
74. Zero tracking 58........................................................
741. What is zero tracking? 58..........................................
742. Setting related with zero tracking. 59...............................
743. Cancellation for compensation by zero tracking 59...................
75. Stabilized filter 60
751. What is the Stabilized filter? 60.....................................
752. Setting related with the Stabilized filter. 60..........................
76. Change of Peak function and A/Z function 61..............................
77. How to use the peak hold. 62.............................................
78. Various functions concerning display 63..................................
781. Selection of target of display 63.....................................
782. Selection of position of decimal point display 63......................
783. Load display range 63.............................................
79. Selection the target for HOLD 63.........................................
710. Change of bridge power supply voltage 64.................................
711. Tare weight cancellation (A/Z) 64........................................
712. Zero set 64.............................................................
713. Lock of key function 65.................................................
714. CHECK value 65.......................................................
715. How to use the analog output 66.........................................
7151. Scaling of analog output 66........................................
7152. Fine adjustment 1 on analog output 67..............................
7153. Fine adjustment 2 on analog output 70..............................
......................................................
7154. Selection of the target of analog output 72...........................
716. Memory location for setting data and so on 72.............................
717. Prohibition of calibration 72.............................................
718. Check mode 73.........................................................
7181. Operating procedure for the check mode 73..........................
719. Monitor mode 78.......................................................
8. Function mode 80..........................................................
81. Setting method for function mode 80.....................................
82. Function of Function data 82............................................
9. Options 88.................................................................
91. Current output 88......................................................
911. Related functions 88..............................................
912. Specifications for the current output 88.............................
92. BCD output 89.........................................................
921. Related function 89...............................................
922. Specifications for BCD output 89...................................
923. Pin configurations for the BCD output connector 89..................
924. Equivalent circuit for input/output 90...............................
925. Timing chart 91..................................................
92
6. Output condition 92...............................................
927. Selection of output logic for P.C.(Print command),
and selection of its width 93........................................
93. RS232C interface 94..................................................
931. Related function 94...............................................
932. Specifications for interface 94......................................
933. Procedures of data transfer 95......................................
934. Pin configurations for connector pin 96.............................
935. Data format 97...................................................
936. Communication error process 100...................................
94. RS422 interface 101....................................................
941. Related functions 101..............................................
942. Specifications on interface 102......................................
943. Procedure of data transmission 102..................................
944. Pin layout and wiring of Connector 103..............................
945. Data format 104...................................................
946. Process of communication error 106.................................
95. Serial interface 107......................................................
9
51. Related function 107...............................................
952. Specifications for Interface 107......................................
953. Connecting method and internal equivalent circuit 108................
954. Data format 109...................................................
955. Explanation of format data 110......................................
956. Explanation of output type 110......................................
96. Power supply voltage DC12 V(CSD701BP66) 111........................
961. Layout of the terminal boards 111...................................
962. Connection with the power supply and the earth 112...................
97. Power supply voltage DC24 V(CSD701BP67) 113........................
971. Layout of the terminal boards 113...................................
972. Connection with the power supply and the earth 114...................
10. Trouble shooting 115........................................................
101. Execute trouble shooting 116.............................................
102. Optional check 125......................................................
103. Error display 131........................................................
11. Specifications 132...........................................................
111. Specifications for analog section 132.......................................
112. Specifications for digital section 132.......................................
113. Front panel sheet key function 133........................................
114. External control function 133.............................................
115. Comparator function 134.................................................
116. Contact output signal 134................................................
117. Various kinds of functions 134............................................
118. General specifications 134................................................
119. Standard specifications at the shipment 135................................
1110. Accessories 135..........................................................
1111. Options 135.............................................................
11111. Current output 135................................................
11112. BCD output 135...................................................
11113. RS232C interface 135.............................................
11114. RS422 interface 137..............................................
11115. Serial interface 137................................................
1111
1112. Outline dimensions 138..................................................
6. Power supply voltage 137...........................................
12. Warranty 139................................................................
121. Warranty 139...........................................................
122. Repair 139..............................................................
13. Appendix 140................................................................
131. Replacement of fuse 140..................................................
132. Character’s pattern for display 142........................................
133. Setting table for functions 143............................................
1. General
The instrument is a digital indicator for the application of strain gage applied transducer.
11. Features
Main features for CSD701B are as follows :
(1) Compact size and light weight
48 mm×96 mm×120 mm Approx.0.3 kg (Without any options)
(2) The A/Z function and the peak holding function can be used by the selection.
1
2. Name and function of each point
21. Front panel
④ ⑤
1
Load display section
The load data is shown in the Measurement mode, and status or set value is shown in various kinds of Calibration mode and Setting mode.
2
Judgement display
Compared results by comparator function can be displayed.
3
Status display
Lights up when selecting A/Z or A/Z OFF function in the Function mode.
In the same way, light off when selecting Peak hold function in the Function mode.
HOLD Lights up when between the HOLD and COM.1 at the external control input is
shorted.
A/Z Lights up at the time of executing Tare weight cancellation(A/Z ON).
Lights off with the tare weight cancellation clear.
(Effective when selecting the A/Z or A/Z OFF function in the function mode.)
CHECK Lights up when the CHECK is ON.
PEAK Lights up when selecting the Peak mode.
⑩ ⑨ ⑧
4
key
Used when shifting to the Function mode.
Also, by pressing this key and the
turned on and off
5
key
Used when calling the S1 set value changeover mode, or carrying digit at the time of various
kinds of settings.
key together, it is used when the CHECK value is
2
6
key
Used when calling the S2 set value changeover mode, or for the increment of values at the time
of various kinds of settings.
Also, by pressing this key and the
zero adjustment).
7
key
Used for making the changeover mode of Peak/Track, or executing the Tare weight
cancellation (A/Z ON).
(Changeover of Peak hold function and A/Z function depends on the setting of Function mode.)
8
key
Used for the reset of Peak value, or for Tare weight cancellation clear(A/Z OFF).
(Changeover of Peak hold function and A/Z function depends on the setting of Function mode.)
9
key
Used for registering set values at the time of various kinds of settings.
Also, by pressing this key and the
turned on and off.
Also, by pressing this key and
adjustment).
key together, it will execute the set of zero(Onetouch
key together, it is used when the CHECK value is
key together, it will execute the set of zero(Onetouch zero
10
Position of pasting the Unit seal
As necessity requires, paste the Unit seal attached.
22. Rear panel
1
Terminal board
Connects with external control input, contact output, various kinds of strain gage applied
transducer such as load cell, analog output, AC power supply, and a grounding wire.
2
Installing section for options
Whichever one can be installed from the optional BCDOUT, RS232C, RS422 or serial
interface.
3
3. Installation procedures
31. Installation place
● Use the instrument where the temperature/humidity specifies within
the range as follows:
Environmental temperature :−10 ℃ to 50 ℃ Environmental humidity :85 %RH or less(Non condensing.)
32. Location where installation is not allowed.
Warning Don’t locate the instrument on the places such as follows:
It may cause an unexpected faulty in the instrument.
D Do n ot expose the instrument in direct sunlight and/or high temperature area.
D Do n ot use the instrument in a high humid area.
D Do n ot install the instrument where there is high mechanical vibrations and shock.
D Do n ot use the instrument where there are excess of dusts and fine particles.
D Do n ot install the instrument where there include any corrosive gas or any salty
atmosphere.
D Do not install the instrument where there is rapid change of temperature and humidity.
D Do n ot install the instrument near the devices that are magnetized or generate an
electromagnetic field.
D Do n ot install the instrument where there may suffer radioactivity or radioactive rays.
D Avoid the location where chemical reaction may take place such as in a laboratory, or like
that.
4
33. Installation
● When installing the instrument, install as the following figures and
secure the space around the instrument.
Each dimensions of the instrument and required dimensions for the environmental spaces are as follows:
Outline dimensions
Front
Side
Rear
Panel cut size
Upper
Unitmm
5
34. Applicable environment
Warning: The instrument may subject to use in a highly humid area or in full of powder dust.
In such a case, use the instrument by inserting the panel mount gasket attached between the con-
trol panel (cabinet) and the main body
.
By inserting the panel mount gasket, the front panel section becomes IP65 (Inter-
national Protection Code) or equivalent in dustproof and waterproof construc-
tion.
* Care should be taken when handling the panel mount gasket, since there are up
and down directions.
6
4. Connecting method
External
control
applied
p
Anal
input
41. Layout of the terminal boards
There is one terminal board containing 21 point of terminals in the panel.
Layout of terminal boards are shown in the following figure.
Termin
al No.
1 A 11
2 B Strain gage
3 C
4 D 14 S1
5 E 15 S2
6 AOUT 7 AOUT
8 COM.1
9 ZERO
10 HOLD
Descriptions Applications
transducer
og output
External control
Termin
al No.
12
13 COM.2
16 F. G Frame ground
17 SOURCE AC power supply
18 N.C.
19 SOURCE AC power supply
20 N.C.
21 Ground
Descriptions Applications
PEAK/TRACK or A/Z
RESET or A/Z OFF
External control input
Contact output
● The COM.1(Terminal No.8) and COM.2(Terminal No.13) are isolated.
7
42. Note on connection
Warning In case of connection with the instrument, keep strictly to the
following items. If neglected, it may cause an unexpected failure
or a damage to the instrument.
D Be sure to set the power supply to OFF, when the connection will be made.
D Since the terminal boards at rear side of the instrument is made of resin, take care not to
drop it down or not to apply strong impact.
D Recommended torque to tighten the terminal screws for terminal board should be as
follows
Torque to tighten
the terminal screws
Terminal board 0.6 Nm
D The suitable crimp type terminal lugs for the terminal board are as follows:
Width of crimp
type terminal lugs
Terminal board 6.2 mm or less 1.253 or Y type 1.253.5
D Connecting cable with the instrument should be away from the noise source such as power
supply line and/or I/O line for control and so on as far as possible.
D Conduit wiring should be the type of exclusive one, and avoid using with another line
together.
D All of the connections should be executed securely by referring to the Instruction manual
for the instrument.
Suitable crimp type
terminal lugs
8
43. Connection
431. Connection with strain gage applied transducers
The instrument can connect with strain gage applied transducers, such as load cell, pressure transducer and so on. Here, we will describe the example of connections with load cell, so the
connection with other type of strain gage applied transducers shall be proceeded in the same
way.
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the resistance of cable makes the input voltage of the instrument decreased.
1
Connection with 1 piece of load cell and CSD701B
3m
※1※
RED
WHT
BLU
GRN
Shield
9
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+”
direction is required, connect “Green” with Terminal No.2 and
“Blue” with Terminal No.4 individually. As there is a case which
standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the resistance of cable makes the input voltage of the instrument decreased.
2
Connection with 1 piece of load cell and Junction box for extension use(B304) and
CSD701B.
Junction box
3m
※2 (Total length of CAB502 is within 30 m)
Internal wiring diagram of B304
to load cell
Terminal pitch 9.5 mm
B304
CAB−502
RED WHTBLU
RED WHTBLU
※1※
RED
BLU
GRN YEL(Shield)
Suitable crimp type terminal lugs
1.254or2−4
YEL(Shield)
GRN
WHT
GRN
Shield
to CSD701B
10
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+” direction is required, connect “Green” with Terminal No.2 and “Blue” with Terminal No.4 individually. As there is a case which standard wiring color is different, please confirm the inspection data
sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the resistance of cable makes the input voltage of the instrument decreased.
3
Connection with 2 to 4 pieces of load cells, Summing type junction box(B307) and
CSD701B.
Junction box
B307
CAB−502
※2 (Total length of CAB502 is within 30 m)
Internal wiring diagram of B307
WHT RED GRN
BLU
YEL(Shield)
WHT RED GRN
BLU
YEL(Shield)
WHT RED GRN
BLU
YEL(Shield)
WHT RED GRN
BLU
YEL(Shield)
1W 1R
1G 1B
1Y 2W 2R
2G 2B
2Y 3W
3R 3G 3B
3Y 4W 4R
4G 4B
4Y
※1※
RED
BLU
WHT
GRN
Shield
RED
RD OR BK
WHT
WH GN BL YE
GRN BLU
YEL(Shield)
to CSD701B
11
※1 When tension is applied with the application of tension type or
universal(compression/tension) type of load cell, and display of “+” direction is required, connect “Green” with Terminal No.2 and “Blue” with Terminal No.4 individually. As there is a case which standard wiring color is different, please confirm the inspection data sheet of the load cell being used.
2 When the length of CAB502 is more than 30 m totally, there may
have the case that the accuracy is out of warranty because the resistance of cable makes the input voltage of the instrument decreased.
4
Connection with 2 to 4 pieces of load cells, Summing type junction box(SB310) and
CSD701B.
SB310
CAB−502
※2 (Total length of CAB502 is within 30 m)
Internal wiring diagram of SB310
Output port
to CSD701B
RD
WH
GR
Input port No.4
BL
YE
※1※
RED
BLU
Short bar
YERD WH BLGR
YERD WH BLGR
YERD WH BLGR
YERD WH BLGR
WHT
GRN
Shield
Input port No.1
Input port No.2
Input port No.3
Suitable crimp type
terminal lugs
A
B
or
φB
φA
A3.2 mm or less B7 mm or less
Y type Round type
12
432. Connection with external control inputs
Connections with external control input “ZERO”,“HOLD”, “PEAK/TRACK” and “RESET”
should be made according to the below figures by using a contact point or an open collector
between the each terminal and terminal No. 8 at “COM.1”
Refer to the paragraph 71 for the function of each input.
or
Shield
Warning Connections with external control outputs should be made securely
according to the figures. If neglected, it may cause an unexpected failure and/or malfunction to the instrument.
● For the connections with external control inputs, be sure to apply
shielded cable, and the shielded cable should be connected with GND
terminal of the instrument.(Terminals No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
13
433. Connection with contact outputs
Connections with contact outputs “S1” and “S2” and external load should be made by using
each terminal and terminal No. 13 at “COM.2”.
At the same time, take care that the load should not exceed the rated load of contact output.
Rated load of contact output AC125 V 0.1 A (Resistance load)
DC30 V 0.5 A (Resistance load)
load
Surge preventive element
or
Shield
or
Warning Connections with contact outputs should be made securely according
to the figures and also within the rated capacity of the instrument. If
neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
For the protection from the contact of the instrument, connect the
surge preventive element that satisfies the characteristics of external
load to connect. If neglected, it may cause unexpected failure and/or malfunction due to the effects from damage/melt down of the contact
and so on.
● For the connections with contact outputs, be sure to apply shielded
cable, and the shielded cable should be connected with GND terminal
of the instrument.(Terminal No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
14
434. Connection with the power supply and the earth
Connections with the power supply and the earth should be made as the following figure. Grounding should be the D class with single earth.
Power supply voltage AC100 V to AC240 V
(Allowable variable rangeAC85 V to AC264 V)
Frequency for power supply 50/60 Hz
Power consumption Approx.11 VA at maximum. (at AC100 V)
D class with single earth
AC100 V to AC240 V (Allowable variable range
AC85 V to AC264 V)
Warning Connections with the power supply and the earth should be made
securely according to the figures and also within the rated capacity of
the instrument. If neglected, it may cause an unexpected cause of
failure.
● Grounding should be the D class with single earth.
If neglected, it may cause an unexpected malfunction due to the effects of noise from other equipments.
15
435. Connection with analog outputs
(1) In case of standard instrument (voltage output)
The instrument prepares output voltage for analog outputs as a standard. Connections with voltage output should be made as the following figure. Voltage output DC0 V to 5 V
Overrange At “OL” display Approx.1.25 V
At “OL” display Approx.6.25 V
Load resistance 5 kΩ or more
Shield
V
Warning Connections with voltage outputs should be made securely according
to the figures and also within specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to
the instrument.
● For the connections with voltage outputs, be sure to apply shielded
cable, and the shielded cable should be connected with the F.G. terminal of the instrument(Terminal No.16).
If not connected, it may cause malfunction due to the effects from
external noises and so on.
16
(2) In case of option (current output)
Connections with current output should be made as the following figure. Current output DC4 mA to 20 mA
Overrange At “OL” display Approx.0 mA
At “OL” display Approx.24 mA
Load resistance 260 Ω or less
Shield
A
Warning Connections with current outputs should be made securely according
to the figures and also within the specified load resistance.
If neglected, it may cause an unexpected failure and/or malfunction to the instrument.
● For the connections with current outputs, be sure to apply shielded
cable, and the shielded cable should be connected with the F.G.
terminal of the instrument.(Terminal No.16)
If not connected, it may cause malfunction due to the effects from
external noises and so on.
17
Loading...
+ 127 hidden pages