Warning may cause injury or accident that may harm to the operator。
Don’t do these things described here.
It is a description when the occourence only of the assumption of danger by which the user owes injury when handling is mistaken, and the
material damage is assumed。
It is attention and a limitation in the operation and work
Be sure to read the items to prevent the malfunction.
Forward
Thank you very much for purchasing Minebea’s Transmitter, model CSA-522B. This
manual explains installation procedures and connecting method and also operating
method for the transmitter. Make use of it properly after reading through the manual
carefully. This manual is intended for the technical experts to read.
Marks and arrangements used in this manual
The following marks are put to the explanation on the matters that indicate “Don’t do
this.”, “Take care” and “For reference”.
Be sure to read these items where these marks are attached.
Warning
Caution
II
The temperature and humidity must use it in the place within the following ranges.
Do not set up this instrument in the following places. It might cause an unexpected faulty in the instrument.
For safe operation
Be sure to read this instruction manual before use.
1. Installation place
Caution
● Environmental temperature : -10 ℃ to 50 ℃
● Environmental humidity : 85 %RH or less (Non condensing)
Warning
① Location where installation is not allowed.
● Don’t locate the instrument in direct sunshine and/or high temperature area.
● Don’t use the instrument in the high humid area.
● Don’t install the instrument in the place with the vibrations and shock.
● Don’t use the instrument where there is excess of dusts and fine particles.
● Don’t use the instrument where there are corrosive gas and salt and like that.
● Don’t install the instrument where there is rapid charge of temperature and humidity.
● Don’t install the instrument near the devices that are magnetized or generate an
electromagnetic field.
● Don’t install the instrument where the instrument may be affected by radioactivity or radial
rays.
● Avoid the location where chemical reaction may take place such as in a laboratory, or like that.
III
Install based on the following sizes to set up this instrument, and secure the space for surroundings of this container.
Unit:mm
Front
Side
Rear
Panel cut size
or
2-φ4
(20)
② When you set up this instrument
Warning
Each dimensions of the instrument and required dimensions for the environmental spaces are as
follows:
IV
Install each cable while turned off the power supply. Might it get an electric shock, and this instrument be damaged when working with the
power supply enters.
Confirm the indication of the power supply voltage of the instrument and the specifications are corresponding to the supplied power supply
before turning on the power supply. Consult with MINEBEA when not matching. There is danger of causing damage and the electric shock of
this instrument when using it like the uncertainty.
Groundline should be connected securely. When groundline is not connected, it may cause a malfunction of the instrument or el ectric shock to
the operator
Before this instrument is newly used, or when the strain gage applied transuducer is exchanged with a new one, please execute the calibration.
If not calibrating, it may cause not obtaining the correct measurement result, or a malfunction in the instrument and there i s a possibility of the
damage of peripherals. Moreover, there is a similar possibility that the result is incorrect even if the calibration has been already made. Please
take the calibration again.
Please confirm connecting wires is correctly executed when you use this instrument. If it is not connected correctly, it may cause the
malfunctions on the instrument, and the damage to the peripheral equipments or even worse serious accidents.
CautionIt causes not obtaining a correct measurement result, and the malfunction when the setting change is carelessly executed while
measuring it with this instrument and there is a possibility of the damage of peripherals.。
Please do not give the impact such as throwing the thing at this instrument.
There is a possibility of causing the malfunction on this instrument, or causing the damage of the electric circuit.
2. Power supply
Warning
Warning
Caution
3. Application note
Caution
Caution
Caution
Caution
V
Please observe the following conditions strictly when this instrument suits the above−mentioned standard.
If neglected, there is a possibility of not suiting the above−mentioned standard.
4. CE conformity standard
This instrument has suited the following standard.
EN61326−1:2006
“Electrical equipment for measurement, control, and laboratory use − EMC requirements”“Immunity test requirements for equipment intended for use in industrial locations”
EN61010−1:2001
“Safety requirements for electrical equipment for measurement, control and laboratory use”
The operating condition to conform this standard is as follows :
Caution
4-1. Place of installation
Please set up this instrument in the shielded case or control panel where EMC measures are
given.
4-2. Wiring
① Shield processing
Please make sure to shield all the signal cable by using the shielded cable or using the conduit
piping including the storage case and control panel.
Please make sure to shield the power cable by using the conduit piping including the storage case
and control panel.
② Grounding
Please make sure to apply the grounding through the case and control panel where EMC
measures are given with protective earth terminal.
VI
Date
Manual No.
Revision reason (Contents)
2007/06
DRW. NO.EN294-1348
First version
2009/06
DRW. NO.EN294-1348-A
Due to ECN.FN09-02035
Changed CE conformity standard EN61326-1:2006
-Addition-
“Immunity test requirements for equipment intended for use in
industrial locations”
2016/10
DRW. NO.EN294-1348-B
Due to ECN.FN10-02121A
・Change length of terminals (15) to (20)
・Minebea logo is changed
Due to ECN.FN16-02057
・「 MINEBEA Co., Ltd. Measuring Components Business Unit」
on the cover is deleted
2016/10
DRW. NO.EN294-1348-C
Due to ECN.FN10-02121B
・Delete the extra page number .「next page of cover and table
of contents page number Ⅱ」
・Repair the position of word .「Warning」、「!」
・Delete the extra white page .「next page of history of
revision」
・Add the white page .「next page of table of contents」
2016/10
DRW. NO.EN294-1348-D
Due to ECN.FN10-02121C
・Delete the extra white pages .
「next page of History of revision」「next page of Index」
History of revision
Index
FORWARD ........................................................................................................................................................ I
MARKS AND ARRANGEMENTS USED IN THIS MANUAL ........................................................................ I
FOR SAFE OPERATION ................................................................................................................................ II
1. INSTRATION PLACE ............................................................................................................................................................................................... II
2. POWER SUPPLY ................................................................................................................................................................................................... IV
3. APPLICATION NOTE ............................................................................................................................................................................................ IV
4. CECONFORMITY STANDARD ........................................................................................................................................................................... Ⅴ
HISTORY OF REVISION ..............................................................................................................................VI
1. EACH FUNCTION AND NAME .............................................................................................................1
2-1.ALLOCATION OF THE TERMINALS ................................................................................................................................................................ 3
2-2.NOTES ON CONNECTIONS.............................................................................................................................................................................. 4
2-3.CONNECTION WITH STRAIN GAGE APPLIED TRANSDUCER ...................................................................................................................... 4
2-4.CONNECTION WITH POWER SUPPLY AND GROUND ................................................................................................................................... 6
2-5.CONNECTION WITH ANALOG OUTPUT.......................................................................................................................................................... 7
2-6.CONNECTION WITH THE AUTO ZERO BY THE EXTERNAL CONTROL INPUT ........................................................................................... 8
4. FUNCTION AND OPERATION ............................................................................................................ 13
4-1.SETTING OF ZERO ADJUSTMENT ............................................................................................................................................................. 13
4-2.SETTING OF GAIN ADJUSTMENT ............................................................................................................................................................... 13
4-3.SETTNG OF CALIB VALUE.......................................................................................................................................................................... 13
4-4.SETTING OF FREQUENCY RESPONSE ......................................................................................................................................................... 13
4-5.SETTING OF BRIDGE POWER SUPPLY VOLTAGE ...................................................................................................................................... 13
5. HOW TO CHANGE THE FUSE ............................................................................................................ 14
6-1.AUTO ZERO (CSA522B-P99) .................................................................................................................................................................. 15
6-2.POWER SUPPLY VOLTAGE AC110V(CSA522B-P61) ...................................................................................................................... 16
6-3.POWER SUPPLY VOLTAGE AC200V(CSA522B-P63) ...................................................................................................................... 16
II
6-4.POWER SUPPLY VOLTAGE AC220V(CSA522B-P64)...................................................................................................................... 16
8-5.STANDARD SPECIFICATION AT THE SHIPMENNT ..................................................................................................................................... 24
It lights by turning on the power supply (POWER switch ON).
② ZERO Fine control trimmer
It is a trimmer for the fine control of zero.
③ ZERO Coarse control switch
It is a switch for the coarse control of zero. Some ±0.3mV/V input is shifted.
④ GAIN Fine control trimmer
It is a trimmer for the fine control of amplification degree.
⑤ GAIN Coarse control switch
It is a trimmer for the coarse control of amplification degree. It is 1000 times in L, 2000 times in M,
and 3000 times in H.
⑥ CALIB switch
It is a switch that turns the CAILB value on and off.
⑦ FILTER switch
It is a switch that changes the frequency response.
It is 25 kHz in OFF, and 1 Hz or 30 Hz in ON. (Change with the DIP switch on the circuit board.)
⑧ GV switch
The power supply voltage of the bridge to the strain gage applied transducer is selected.
⑨ MONITOR terminal
It is a terminal to monitor the output voltage value.
⑩ AUTO ZERO LED (Options)
LED lights while auto zero executions
⑪ AUTO ZERO ON switch (Options)
The analog output when pushing is adjusted to 0 volt.
Please release the auto zero when you calibrate.
⑫ AUTO ZERO OFF switch (Options)
Auto zero is released.
2
① Termnal board
③ I-OUT SPAN fine control trimmer
② I-OUT ZERO fine control trimmer
④ protective
earth terminal
1-2. Rear pannel
① Terminal board
The vaious strain gage applied transducers like load cell, etc., voltage output, current output,
external control input (Option), grounding wire and AC power supply are connected.
② I-OUT ZERO fine control trimmer
It is the fine control trimmer of the current output (DC4 mA).
③ I-OUT SPAN fine control trimmer
It is the fine control trimmer of the current output (DC20 mA).
④ Protective earth terminal
Connect the grounding line to suit the CE Conformity Standard.
Do not connect excluding the grounding wire.
Loading...
+ 26 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.