Milwaukee Instruments SMS510 User Manual

SMS510
RANGE ±1000 mV RESOLUTION 1 mV ACCURACY (@25°C) ±5 mV SETPOINT 0 to 600 mV ALARM active when the measure is
lower than setpoint
OUTPUT POWER SOCKET active when the measure is
lower than setpoint
ORP ELECTRODE MA921B/2 (included) ENVIRONMENT 0 to 50°C, 95% RH max. POWER SUPPLY 12 VDC (included) POWER DRIVERS 115VAC, 2A, 60Hz or 230VAC, 1A, 50Hz DIMENSIONS 160 x 80 x 40 mm WEIGHT 220 g (meter only)
Milwaukee
OPTIONAL ACCESSORIES:
MA921B/2 ORP Electrode with BNC connector and 2m cable M10000B Electrode rinse solution, 20 mL sachet (25 pcs) MA9015 Electrode storage solution, 220 mL bottle MA950 Portable meter wall fixing kit
WARRANTY:
This instrument is warranted from all defects in materials and manufacturing for a period of two years from the date of purchase. The electrode is warranted for a period of 6 months. If during this period, the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either dealer or our office and the repair will be effected free of charge. Note: We reserve the right to modify the design, construction and appearance
of our products without advance notice.
ISMIL510 12/01
Milwaukee
MODEL: SMS510
Smart Monitoring System
USER MANUAL
PORTABLE
ORP METER
OPERATION:
SETPOINT:
• Connect the supplied 12 VDC power adapter to the meter and to the mains.
• Connect the ORP electrode to the BNC socket on the bottom of the meter.
• Always remove the electrode protective cap before taking any measurement. If the elec­trode has been left dry, soak the tip (bottom 2.5 cm) in rinse solution (M10000B) for a few minutes to reactivate it.
• Turn the instrument on by pressing the ON/OFF key.
• Allow the reading to stabilize and the meter will start con­tinuous monitoring.
• The meter is factory calibrated.
• A blinking alarm will indicate when the measured ORP value is lower than the setpoint.
NOTE:
• The power contact is not protected by fuse inside the meter. It is recommended to protect it outside, against failure.
• WARNING: when the alarm led is blinking, the mains voltage is present in the output power socket.
• The setpoint can be selected by adjusting the central front knob to the desired value.
• The selectable range for
SMS510 is from 0 to 600 mV.
INSTALLATION PROCEDURE: