Milwaukee Instruments MW11 User Manual

WARRANTY:
Milwaukee
This instrument is warranted from all defects in materials and manufacturing for a period of two years from the date of purchase. If during this period, the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either dealer or our office and the repair will be effected free of charge. Note: We reserve the right to modify the design, construction and appearance
of our products without advance notice.
SPECIFICATIONS:
MW 11
RANGE 0.00 to 3.50 ppm RESOLUTION 0.01 ppm ACCURACY ±0.03 ppm ±3% of reading @ 25 °C TYPICAL EMC DEV. ±0.01 ppm LIGHT SOURCE Light Emitting Diode @ 525 nm LIGHT DETECTOR Silicon Photocell METHOD Adaptation of USEPA method 330.5. The reaction
between total chlorine and the DPD reagent
causes a pink tint in the sample.
ENVIRONMENT 0 to 50°C (32 to 122 °F)
max. 95% RH non-condensing
BATTERY TYPE 1 x 1.5V AAA AUTO-SHUT OFF After 2 minutes of non-use DIMENSIONS 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5”) WEIGHT 64 g (2.25 oz.)
OPTIONAL ACCESSORIES:
Reagent Sets
2720216 Total Chlorine powder reagent, 25 pcs.
Other Accessories
3000100 Glass cuvettes with caps (2 pcs) 3000200 Tissue fow wiping cuvettes (2 pcs) 3000300 1.5V AAA batteries (4 pcs)
Milwaukee
USER MANUALT
TOTAL CHLORINE
MODEL: MW11
ISTMW11 08/10
PRELIMINARY EXAMINATION:
Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occured during shipment, please notify your Dealer. Each MW 11 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Six powder reagents for Total Chlorine
• 1 x 1.5V AAA Battery
• Instruction Manual
FUNCTIONAL DESCRIPTION:
1. Dust cover
2. Cuvette with cap
3. Cuvette holder
4. Liquid Crystal Display
5. Button
OPERATION:
• Turn the meter on by pressing the button. After all the segments are dis­played, “C.1”, “Add” appears with “Press” blinking, the meter is ready.
• Fill the cuvette with 10 mL of unre­acted sample and replace the cap. Place the cuvette into the meter and close the meter’s cap.
• Press the button. When the display shows “Add”, “C.2” with “Press” blink­ing the meter is zeroed.
10 mL
• Remove the cuvette, open it and add the content of one packet of 2720216 reagent. Replace the cap and shake gently for 20 seconds. Replace the cuvette into the meter.
• Wait for 2 minutes and 30 seconds and then press the button or press and hold the button until the timer is displayed on the LCD.
• The instrument directly displays the concentration of total chlorine in ppm. The meter automatically turns off after 2 minutes.
BATTERY REPLACEMENT:
To save the battery, the instrument shuts down after 2 minutes of non-use. One fresh battery lasts for a minimum of 5000 measurements, depending on the light level. When the battery capacity is under 10 % “bAt” appears on the LCD at start-up. If the battery is empty and accurate measurements can’t be taken any more, the instrument shows “bAd” then ”bAt” each for 1 second and turns off. To restart the instrument, the battery must be replaced with a fresh one. To replace the instrument’s battery, follow the steps:
• Turn the instrument off by holding the button until the meter
shuts off.
• Turn the instrument
upside down and remove the battery cover with a screwdriver.
• Remove the battery from
its location and replace it with a fresh one.
• Insert the battery cover
and replace the screw with a screwdriver.