Milwaukee Instruments MA916B-3, MA917B-1, MA916B-1, MA918B-1 User Manual

pH electrodes are constructed from a special composition glass which senses the hydrogen ion concentration. This
glass is typically composed of alkali metal ions. The alkali
metal ions of the glass and the hydrogen ions in solution undergo an ion exchange reaction, generating a potential
difference. In a combination pH electrode, the most widely used variety, there are actually two electrodes in one body. One portion is called the measuring electrode, the other the reference electrode. The potential generated at the junction site of the measuring portion is due to the free hydrogen
ions present in solution.
The potential of the reference portion is produced by the internal element in contact with the reference fill solution. This potential is always constant. In summary, the measur­ing electrode delivers a varying voltage and the reference electrode delivers a constant voltage to the meter. The volt­age signal produced by the pH electrode is a very small, high impedance signal. The input impedance requires that it be interfaced only with equipment with high impedance cir­cuits.
pH Electrode
basics
pH Electrodes
6
www.milwaukeeinst.com
Milwaukee has a wide assortment of pH and ORP elec­trodes to meet all your specific requirements. Finding the right electrode for a specific application is a very important task and in order to solve this selection prob­lem it is important to consider the following:
• Glass body electrode versus Epoxy (plastic)
body electrode: Glass body electrodes stand
higher temperatures (typically 100°C against 80°C for plastic) and are more resistant to corrosive chemicals and solvents. They are easier to clean and are available in different shapes depending on the application. On the other hand plastic body electrodes are more rugged and the glass bulb is better protected.
• Gel filled electrodes versus refillable electrodes: refillable electrodes last longer since electrolyte can be changed for repeated usage. The response is faster due to a greater outflow of elec­trolyte into the sample and therefore less likely to clog. Gel filled electrodes require less mainte­nance and resist to higher pressure.
• Double reference junction versus Single junc­tion reference: Double junction reference elec-
trodes have a longer live and protects the sample measured from silver contamination from the elec­trolyte. The Silver wire is more protected and therefore gets less contaminated. The single junc­tion electrodes normally costs less and are ideal for general purpose applications
• Conic shaped versus Sphere shaped: The conic-shaped electrode is easier to clean and to maintain (ideal for applications such as dairy). Has a more rugged tip and therefore ideal for penetra­tion. The sphere-shaped has a faster response time due to the larger surface area on the bulb.
Model
MA919B/1 MA924B/1
Measuring Range 0 to 13 pH ±2000 mV Temperature Range -5 to 80 °C -5 to 80 °C Shaft material glass glass
Reference Electrolyte KCL 3.5M KCL 3.5M
Reference Type double Ag/AgCl double Ag/AgCl Reference Junction open open Shape of membrane spheric Platinum ring Max. Pressure 0,1 bar 0,1 bar Connector type BNC BNC Cable length coaxial 1 meter coaxial 1 meter Shaft length 120 mm 120 mm Diameter 8 mm 8 mm Application food laboratory food laboratory
8 mm
M
A
9
2
4
B
/1 =1 m
M
A
9
1
9
B
/1 =1 m
8 mm
120 mm
120 mm
The pH electrode, due to the nature of its construction, needs to be kept moist at all times. In order to operate prop­erly, glass needs to be hydrated. Hydration is required for the ion exchange process to occur. If an electrode should become dry, it is best to place it in some tap water for a half hour to condition the glass.
pH electrodes are like batteries; they run down with time and use. As an electrode ages, its glass changes resistance. This resistance change alters the electrode potential. For this reason, electrodes need to be calibrated on a regular basis. Calibration in pH buffer solution corrects for this change. Calibration of any pH equipment should always begin with buffer 7.0 as this is the "zero point." The pH scale has an equivalent mV scale. The mV scale ranges from +420 to -420 mV. At a pH of 7.0 the mV value is 0. Each pH change corresponds to a change of approx. ±60 mV. As pH values become more acidic the mV values become greater. pH electrodes have junctions which allow the internal elec­trolyte solution of the measuring electrode to leak out into the solution being measured.
pH Electrode
basics
pH Electrodes
7
www.milwaukeeinst.com
Model MA916B/1 - MA916B/3 MA917B/1 MA918B/1
Measuring Range 0 to 13 pH 0 to 14 pH 0 to 12 pH Temperature Range -5 to 100°C (23 to 212°F) 0 to 100°C (32 to 212°F) -5 to 100°C (23 to 212°F) Shaft Material glass glass glass Reference Electrolyte KCl 3.5M + AgCl KCl 3.5M KCl 3.5M + AgCl Reference Junction ceramic, single ceramic, single ceramic, triple Reference Type single, Ag/AgCl double, Ag/AgCl single, Ag/AgCl Shape of membrane spheric spheric conic Max pressure 0.1 bar 0.1 bar 0.1 bar Connector Type BNC BNC BNC Cable length coaxial, 1 or 3 m coaxial, 1 m coaxial, 1 m Shaft length 120 mm 120 mm 120 mm Diameter 12 mm 12 mm 12 mm Application laboratory applications laboratory applications laboratory applications
Glass Conic Tip Sensor
Glass Spheric Sensor
Epoxy Electrode
M
A
9
1
8
B
/1 =1 m
M
A
9
1
7
B
/1 =1 m
M
A
9
1
6
B
/1 =1 m /3 =3 m
12 mm
120 mm
12 mm
120 mm
12 mm
120 mm
MA916B/1 (will be replaced by SE100)
Loading...
+ 4 hidden pages