Milwaukee pH 58 User Manual

USER MANUAL
pH 58
Pocket-size
pH/OR P /° C/°F Meter
Automatic Temperature Measuring unit
Compansation Indicat or on main LCD Instability Indicator Main LCD
Calibration Indicator Temperature Low Battery Indicator reading Secondary LCD
OPERATI ONAL GUIDE
• Rem ov e th e el ec tr ode c a p an d im m ers e th e meter in MA9015 st orage soluti on for 2 hours to activ ate th e elec t r ode.
• Turn the meter on by pressing th e ON/OFF button. All t he used segm ents on t he LCD wi ll be visible for 1 s econd or as long as the button is pressed.
• Immerse th e electrode i n the solut ion to be tested. Stir g ently and w ait for the readi ng to stabilize, i.e. the instability indicator (hourglass) on t he LCD turns off.
• Select the pH or ORP (mV) mode by pressing the SE T/HOLD button.
• The displayed pH value is automatically com p ens at e d for t em p er at ur e. T h e OR P (m V ) value will be displayed on the main LCD, while the temperature will always be displayed on the sec ondary L CD.
To free ze the di splay, whil e in measu rement mode, press an d hold the SET/HOLD button. The “HOLD” message appears on the secondary display and the reading will be frozen on the LC D. Press any butt on to return to the norm al me asurement mode.
To turn the meter off, press the ON/OFF button. The “OF F” message will appear on the secondary display. Releas e t he button.
Note:
• Before t aking any m easurem ent, make s ure the meter is calibrated (t he CA L tag is on).
• Af ter us e alw ays t urn th e met er OFF , rin se the electrode with water to minimize contamination and store it with a few drops of storage (MA9015) or pH7 (M10007) solution in the protective c ap.
NEVER USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSE.
CALIBRATION PROCEDURE The ORP (mV) range is factory calibrated.
Contact the nearest Service Center for recalibration, if necessary. For better acc urac y, f requ ent cali brati on of th e tester is recommended. Calibration is also necessary after electrode replacement, after testing aggressive chemicals and where extrem e accu r acy is r equi red.
• Tu rn the meter on and select the pH mode with the SET/ HOLD button.
• Press and hol d th e O N/OFF /CAL but t on until the “OFF” m essage on the sec ondary LCD is substitut ed by “CAL”. Release t he button.
• The instrum ent enters th e calibration m ode by displaying “pH 7.01 USE” (or “pH 6.86 USE” if the NIST buf fer set was s elected).
• For a single-point cali bration, imm erse the electrode in any buffer, i .e. pH 4.01, 7.01 (or
6.86), 10.01 ( or 9. 18).
• The meter activates the automatic buffer recognition. If no valid bu ffer is detected, the
meter keeps the USE indic ation active for 12 seconds, and then replaces it with WRNG, indicating th at the sample being measu red is not a valid buffer. Ot herwise, if a valid buffer is detected, then its value is shown on the primary display, and REC appears on the secondary LCD.
• If the pH 7.01 (or p H 6.86) w as used, press the S ET but ton to exit th e Calib ration mod e and the “OK 1” message will appear on the display. The first calibration point is stored and the meter returns to normal measurement mode.
For better accuracy, it is always recommended to perform a 2-point calibration.
• For a two-point calibration, immerse the electrode in pH 7.01 (or pH 6.86) buffer solution.
• After the first point has been accepted, t he meter will t hen ask for the second bu ffer and the messag e “pH 4. 01 USE” will appear.
• Rinse the electrode and imm erse it in the second solution (pH 4.01, 10.01 or 9.18).
• If a vali d buff er va lue i s det ected , th e REC message is displayed and the meter completes the calibration proc edure. The LCD shows the accepted value with the “OK 2” message and the instrument returns to the norm al m easur ement mode. O ther wise, if no valid buffer is d etected, the meter displays the WR NG m essage.
Note: When the calibration procedure is completed, the CAL t ag i s t urned on.
To quit the proced ure and return t o the last cali brati on data, after enter ing the cali brati on mode press th e ON/ OF F but ton. The secondary LCD displays “ESC” for 1 second and then the meter returns to the normal mea surement mode.
To reset to the default values and clear a previous calibration, press the SET/HOLD button aft er ent ering th e c alibration mode and bef ore t he first poi nt is accepted. The secondary LCD displays “CLR” for 1 second, the meter resets to the default calibration and the CA L tag on th e LCD turns off.
SETUP Setup mode allows the selection of temperature (°C or °F) and pH buffer set for calibration. T o enter the S etup mode, press the ON/OFF button until “CAL” on the secondary LCD is replaced by “TEMP” and t he current temperature unit (e.g. TEMP °C). Then:
for °C/°F selection: use the SET/HOLD button; then press the ON/OFF button once to enter the buffer set selection or twice to ret urn to the norm al me asurement mode.
to change the calibration buffer set: after setting the temperature unit, press ON/OFF once and select the buffer set (“pH 7.01 BUFF” or “pH 6.86 BUFF” for NIST) by using the SET/HOLD button. Press ON/OFF to ret urn to the normal measurement mod e.
ELECTRODE RE PLACE MENT
• Remove th e protect ive c ap and unscr ew the plastic ring on t he top of the electr ode.
• Pu ll out the MI58P el ectrode and replac e it with a n ew one.
• Make sur e the gaskets ar e in place bef ore screwi ng back the ring.
BATTERY REPLACEMENT Wh en th e b att eri es bec om e w eak, th e b att ery sym bol on t he L CD w il l l igh t up to ad vi se t hat only a few hours of w orking tim e is rem aining. The meter i s als o provided w ith BEPS (Battery Error Prevention System), which avoids any erroneous r eadi ngs due to low battery l evel by automatically sw itching the meter o ff. It is recommended to replace the batteries immediately. To r eplac e the b atter ies u nscr ew the b atte ry compartment cap and replace all four 1.5V batteries while paying attention to their polarity. Make sure the gasket is in place before screwing back the cap. Batteries should only be replaced in a nonhazardous area using the battery type specified i n t his instructi on manual.
ACCESSORIES MI 58P Replaceable electrode for pH 58
M10004B pH 4.01 buffer, 20 ml sachet, 25 pcs. M10007B pH 7.01 buffer, 20 ml sachet, 25 pcs. M10010B pH 10.01 buffer, 20 ml sachet, 25
pcs.
MA9004 pH 4.01 buffer, 230 ml bottle MA9006 pH 6.86 buffer, 230 ml bottle MA9007 pH 7.01 buffer, 230 ml bottle MA9009 pH 9.18 buffer, 230 ml bottle MA9010 pH 10.01 buffer, 230 ml bottle MA9015 Electrode storage solution, 230 ml MA9016 Electrode cleaning solution, 230 ml MA9020 ORP test solution (200/275 mV), 230
ml bottle M10000B Electrode rinse solution, 20 ml sachet, 25 pcs.
WARRANTY
This instrument is warranted against defects in materials and manufacturing for a period of two years from t he d at e of purchase. Probe is warranted for 6 months. If during this period the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, pl ease return the parts to either distributor or our office and the repair will be done free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. Milwaukee/Martini instruments reserves the right to make improvements in design, construction and appearance of its products without advance notice.
SPECIFICATIONS RANGE -2.00 to 16.00 pH
±1000 mV (ORP)
-5.0 to 60.0°C / 23.0 to 140.0°F
RESOLUTION
0.01 pH / 1 mV
0.1°C / 0.1°F
ACCURACY (@25°C) ±0.05 pH / ±2 mV
±0.5°C / ±1°F
TYPICAL E M C DEVIATI O N ±0.02 pH
±2 mV ±0.3°C / ±0.6°F
TEMPERATURE COMPENSATION
Automat ic f or pH, from -5 to 60°C CALIBRAT ION, pH Au tomatic, 1 or 2 poin ts with 2 sets of memorized buffers(pH 4.01,
7.01, 10.01 or 4.01, 6.86, 9.18)
CALIBRATION, ORP Factory calibrated PROBE Replaceable MI 58P ENVIRONMENT -5 to 50° C; 100% RH max. BATTERY TYPE 4 x 1.5V; IEC LR44, A76 BATTERY LIFE approx. 250 h ours of use AUTO-OFF after 8 minut es of non-use DIMENSIONS 200 x dia 38 mm WEIGHT 100 g