Milwaukee ORP 57 User Manual

ACCESSORIES
MI 57P Replaceable electrode for ORP 57 MA9015 Electrode storage solution, 230 ml MA9016 Electrode cleaning solution, 230 ml MA9020 ORP test solution (200/275 mV), 230
ml bottle M10000B Electrode rinse solution, 20 ml sachet, 25 pcs.
SPECIFICATIONS RANGE ±1000 mV (ORP)
-5.0 to 60.0°C / 23.0 to 140.0°F RESOLUTION 1 mV
0.1°C / 0.1°F ACCURACY (@25 °C) ±2 mV ±0.5°C / ±1°F TYPICAL EMC DEVIATION ±2 mV ±0.3°C / ±0.6°F
CALIBRATION Fa ct ory c alibrated PROBE Replaceable MI 57P ENVIRONMENT -5 to 50°C; 100% RH
max.
BATTERY TYPE 4 x 1.5V; IEC LR44, A76 BATTERY LIFE approx. 300 hours of use AUTO-OFF after 8 minutes of non-use DIMENSIONS 200 x dia 38 mm WEIGHT 100 g
USER MANUAL
ORP 57
Pocket-size
ORP/°C/°F Meter
www.milwaukeeinst.com
Measuring unit on main LCD Instability indicator Main LCD
Low battery indicator Temperature rating Secondary LCD
OPERATIONAL GUIDE
• Remov e th e el ect r od e and tur n th e meter on b y pressing the ON/OFF button. All the used segments on th e LCD w ill b e visib le for 1 sec ond or as long as the but ton is pressed.
• Immerse the electrode in the solution to be tested. Stir gently and wait for stable reading, i.e. the inst ability ind icator (h ourglass) on the LCD tur ns off.
• The ORP (mV) value will be displayed on the main LCD, whil e t h e tem p eratur e is show n on th e secondary LCD.
To freeze the display, while in measurement mode, press and hold the SET/HOLD button. The “HO LD” m ess age appear s on the s ec ondar y display and the reading will be frozen on the LCD. Press an y but t on t o r etur n to normal m ode.
To turn the meter off, press the ON/OFF button. The “OFF” message will appear on the secondary display. Release the button. Note: T he meter i s fact ory c alibrat ed. C ontact the nearest Service Center for periodic recalibration.
°C/°F SELECTION
• Press and hold the ON/OFF button until the secondary LCD shows “TEMP” and the current temperature unit (e.g. TEMP °C).
• Use SET/HOLD to select the unit (°C or °F); then pr ess again O N/ O F F t o r et u r n t o t h e n o r m al mode.
ELECTRODE REP LACEMENT
• Remove the protective cap and unscrew the plastic ri ng on th e top of th e electr od e.
• P u ll ou t t h e MI57P elec tr ode and r ep lac e it with a new on e.
• Make sure the gaskets are in place before screwing back the ring.
BATTERY REPLACEMENT
When the batteries become weak, the battery symbol on the LCD will light up to advise that only a f ew hour s of working time is remaining. The meter is also provided with BEPS (Battery Error Prevention System), which avoids any erroneous readings due to low battery level by automatically switching the meter off. It is recommended to replace the batteries immediately. To replace the batteries unscrew the battery compartment cap and replace all four 1.5V batteries while paying attention to their polarity. Make s ure t h e g as k et is in place befor e s c rewing back the cap. Batteries should only be replaced in a non­hazardous area using the battery type specified in this instruction manual.
WARRANTY
This instrument is warranted against defects in materials and manufacturing for a period of two years from the date of purchase. Probe is warranted for 6 months. If during this period the repair or replacement of parts is required, where the damage is not due to negligence or erroneous operation by the user, please return the parts to either distributor or our office and the repair will be done free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. Milwaukee/Martini instruments reserves the right to make improvements in design, construction and appearance of its products without advance notice.