Milwaukee MW600 User Manual

Milwaukee
PORTABLE
DISSOLVED OXYGEN METER
MODEL: MW600
Smart DO Meter
USER MANUAL
PROBE PREPARATION:
The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery into the battery clip connector while observing polarity.
D.O. probes are shipped dry. Assemble the probe and prepare it for use, connect it to the meter and proceed as follows.
1. Remove the red and black plastic cap. This cap is used for shipping purposes only and can be thrown away.
2. Wet the sensor by soaking the bottom 2½ cm of the probe in electrolyte (MA9071) for 5 minutes.
3. Rinse the membrane (supplied with the meter) with electrolyte while shaking it gently. Refill with clean electrolyte.
4. Gently tap the sides of the membrane with a pencil or a rod to disengage air bubbles. To avoid damaging the membrane, do not tap the membrane directly on the bottom.
5. Install O-Ring properly inside the membrane cap.
6. With the sensor facing down, screw the cap clockwise. Some electrolyte will overflow.
7. Examine membrane to verify air is not trapped between the membrane and electrode tip.
When probe is not in use and during polarization, place the protective cap supplied over the electrode tip.
1
PROBE POLARIZATION:
• A dissolved oxygen probe must be polarized to function
properly.
• To polarize the probe, the fully assambled probe must be
connected to the meter and the meter must be on.
• During polarization (and during measurement), approximately
800 mV is applied to the cathode and anode inside the membrane and a chemical reaction occurs. During the polarization period excess oxygen in the electrolyte is consumed. During this phase, probe movement that “moves” the electrolyte will yield jumpy measurements. When a probe is totally polarized moving the probe will not effect the measurement.
• When the meter is turned off, the probe will revert to it’s
prepolarized state. Before using again, the probe will have to be repolarized.
CALIBRATION PROCEDURE:
The calibration is very simple and fast.
• Make sure the probe is ready for measurements
(see Probe Preparation), i.e. the membrane is filled with electrolyte and the probe is connected to the meter.
• Switch the meter on by pressing the ON/OFF
key.
• For an accurate calibration, it is recommended
to wait at least 15 minutes to ensure polarization of the probe.
• Remove the protective cap from the D.O. probe.
2
Zero Calibration:
• Dip the probe into MA9070 zero oxygen
solution and stir gently for 2-3 minutes.
• Wait for the probe to reach it’s lowest
stable reading.
• Adjust the zero D.O. calibration
trimmer (located on the front panel) until the display reads "0.0".
Slope Calibration:
It is suggested to perform the slope calibration in water saturated air.
• Rinse the probe with a large amount of clean water to
remove any residual zero oxygen solution.
• Dry the probe tip and allow a few minutes for the D.O. probe
to stabilize while suspending over a container of water in the air.
• Press and hold the CAL key.
• Adjust the slope trimmer on the
front panel of the meter to read "100%" on the LCD (while still holding the CAL button).
• Release the CAL key and the
LCD will display a value in ppm of oxygen.
The zero calibration of the MW600 is very stable, therefore this procedure needs only to be performed whenever the probe is replaced. However, if most of the measurements are closer to zero, more frequent zero calibration is advised.
Slope calibration can be easily performed on a weekly basis.
3
Loading...
+ 8 hidden pages