Milwaukee Mi453 Instruction Manual

www.milwaukeetesters.com
1
www.milwaukeetesters.com
Milwaukee WMilwaukee W
Milwaukee WMilwaukee W
Milwaukee W
ine Line L
ine Line L
ine L
ab Photometer
Mi453
Reducing SugarsReducing Sugars
Reducing SugarsReducing Sugars
Reducing Sugars
INSTRUCTION MANUAL
Instruction Manual Mi453 REDUCING SUGARS Photometer for wine analysis
22
22
2
FUNCTIONAL DESCRIPTION ............................................................................... 3
GENERAL DESCRIPTION ..................................................................................... 4
SPECIFICATIONS ............................................................................................... 6
GUIDE TO DISPLAY CODES ................................................................................. 7
GENERAL TIPS FOR AN ACCURATE MEASUREMENT ................................................. 8
MEASURENT PROCEDURE ................................................................................... 9
ERROR MESSAGES ........................................................................................... 13
BATTERY REPLACEMENT .................................................................................... 14
ACCESSORIES ................................................................................................. 15
TABLE OF CONTENTS
www.milwaukeetesters.com
3
FUNCTIONAL DESCRIPTION
FRONT PANEL
A. LID B. CUVET HOLDER C. LIQUID CRYSTAL DISPLAY (LCD) D. ON/OFF KEY, TO TURN THE METER ON AND
OFF E. ZERO KEY, TO START THE ZERO MEASUREMENT F. READ KEY, TO START THE SAMPLE MEASUREMENT G. TIMER KEY, TO ACTIVATE THE COUNTDOWN
MODE TIMER
DISPLAY
A. BATTERY STATUS ICON B. HOURGLASS ICON C. LAMP STATUS INDICATOR D. MEASURE STATUS E. MEASUREMENT UNIT F. MAIN DISPLAY G. PARAMETER NUMBER INDICATOR H. TIMER MODE INDICATOR I. SECONDARY DISPLAY
Instruction Manual Mi453 REDUCING SUGARS Photometer for wine analysis
44
44
4
GENERAL DESCRIPTION
Thank you for choosing Milwaukee. This instruction manual will provide you the necessary information for correct use of the meter.
Mi453 is an auto-diagnostic portable microprocessor meter. It has an advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated.
The auto-diagnostic feature of this meter ensures always optimal measurement conditions to perform most precise readings. The light level is automatically adjusted each time a zero-measurement is made, and the temperature of the lamp is controlled to avoid overheating.
SIGNIFICANCE OF USE
The determination of concentration of reducing sugars (RS) is one of the most important parameters that need to be measured during the wine making process. Following the increase of RS during maturation of grapes can help to decide when to start harvest. Having the highest possible sugar content is important because this is the main parameter that defines the commercial value of grapes. During the alcoholic fermentation instead, the decrease of sugars can be followed to decide when fermentation is completed, or allows making corrective actions if the content of RS is too low to obtain the desired alcohol degree or sweetness.
The predominant RS in grape products are glucose and fructose (hexoses). After reaction with excess alkaline cupric tartrate (Fehling reagents), the RS content can be determined colorimetricly. The Fehling method is not an exact determination but an index of the reducing sugar concentration, because the reaction depends upon the amount and kind of RS present. When the reducing sugar content is known at the beginning of fermentation, the potential alcohol degree can be estimated multiplying the sugar concentration (in g/L) by 0.06. Phenols interfere in the Fehling reaction and therefore red wine must be decolorized prior to analysis. Wine also contains non-fermentable reducing sugars like pentose which will also be analysed by this method.
www.milwaukeetesters.com
5
This meter is supplied with:
• Four sample cuvets and caps
• Reagents for 20 tests (Mi553-0, Mi553A-0, Mi553B-0)
• One 200 PL automatic pipette with Instruction Sheet
• One 1000 PL automatic pipette with Instruction Sheet
• Two plastic tips for 200 PL automatic pipette
• Two plastic tips for 1000 PL automatic pipette
• One spoon
• One funnel
• Filter paper (25 pcs)
• Four 1.5V AA batteries
• Tissue for wiping cuvets
• Instruction Manual
• Instrument Quality Certificate
Typical content of reducing sugars in must and wine
Must sweet must 20-25 % 200-250 g/L
normal 10-20 % 100-200 g/L
in fermentation 4-12.5 % 40-125 g/L
Wine Sweet 2.5-12.5 % 25-125 g/L
Semi sweet 0.8-2.5 % 8-25 g/L
Almost dry 0.2-0.8 % 2-8 g/L
Dry 0-0.2 % 0-2 g/L