USER MANUAL
MI 406
Free Chlorine Meter
Dear Custom er ,
Thank you for choosing a Martini product.
This manual will provide you with the
nec essar y inf ormat ion for the c orrect use of
the instrument.
Please read it carefully before using the
meter. This ins trument is in comp liance with
directives.
SPECIFICATIONS:
Range 0.00 to 5.00 mg/L Cl2
Resolution 0.01 mg/L (0.00-3.50 mg/L);
0.10 mg/L (above 3.50 mg/L)
Precision ±0.04 mg/L @ 1.50 mg/L
Light Source Tungsten lamp
Light Detector Silicon Photocell and
525 nm narrow band interference filter
Method Adaption of the USEPA method 330.5 and Standard Method 4500-Cl G.
Environment 0 to 50°C (32 to 122°F);
100% RH max.
Battery Typ e 1 x 9 volt
Auto-Shut off After 10' of non-use
Dimensions 192 x 104 x 52 mm
(7. 5 x 4.1 x 2” )
Weight 380 g
MEASUREMENT PROCEDURE:
1• Turn the meter on by pressing
ON/OFF.
2• When the LCD displays “- - -”, the
meter is ready.
3• Fill the cuvet with 10
mL of sample, up to the
mark, and replace the
cap.
4• Place the cuvet into the
holder and ensure that
the notch on the cap is
positioned securely into
the groove.
5• Press ZERO and “SIP”
will blink on the display.
6• Aft er a few secon ds the
display will show “-0.0”. The meter is now
zeroed and ready for
measurement.
7• Remove the cuvet. Open it.
8• Add 3 drops of DPD 1 indicator and
3 drops of D PD 2 b uff er to an empt y
cuvet. Immediately add 10 mL of
unrea cted sam ple. Replace the cap
and shake gently.
9• Replace the cuvet into the holder
and ensure that the notch on the cap
is positioned securely into the
groove.
10• Press READ and “SIP”
will blink during
measurement.
11• The instrument directly
displays the
concentration in mg/L of free
chlorine.
Interferences
• Positive error:
- Bromine
- Chlorine Dioxide
- Iodine
- Oxidized Manganese and
Chromium
- Ozone
• Alkalinity above 250 mg/L CaCO3 or
acidity above 150 mg/L CaCO3 will
not reliably d evelop th e full amo unt
of color or it may rapidly f ade.
To resolve this, neutralize the
sample with dil uted HCl or NaOH.
• In case of water with hardness
greater than 500 mg/L CaCO3,
shak e the sam ple for approximately
2’ after adding the powder reagent.