Milwaukee MA886 Instruction Manual

www.milwaukeeinst.com
INSTRUCTION MANUAL
Milwaukee Refractometer
MA886
Refractometer for SodiumRefractometer for Sodium
Refractometer for Sodium
Refractometer for SodiumRefractometer for Sodium Chloride MeasurementsChloride Measurements
Chloride Measurements
Chloride MeasurementsChloride Measurements
www.milwaukeeinst.com
1
Instruction Manual MA886 Refractometer for Sodium Chloride Measurements
TABLE OF CONTENTS
FUNCTIONAL DESCRIPTION ............................................................................. 3
GENERAL DESCRIPTION .................................................................................... 5
SPECIFICATIONS .............................................................................................. 6
PRINCIPLE OF OPERATION................................................................................ 6
MEASUREMENT GUIDELINES .............................................................................. 8
CALIBRATION PROCEDURE ............................................................................... 8
MEASUREMENT PROCEDURE ........................................................................... 10
CHANGING MEASUREMENT UNIT ................................................................... 11
CHANGING TEMPERATURE UNIT ..................................................................... 12
MAKING A STANDARD SODIUM CHLORIDE SOLUTION .................................... 12
ERROR MESSAGES ........................................................................................... 13
BATTERY REPLACEMENT................................................................................... 14
WARRANTY ..................................................................................................... 15
Remove the instrument from the packing materials and examine carefully to ensure no damage has occurred during shipping. If any damage has occurred, notify your Dealer.
Each instrument is supplied with:
• 9 V battery
• Instruction manual
Note: Save all packing material until you are sure that the instrument
functions correctly. A defective instrument must be returned in its original packing.
22
2
22
www.milwaukeeinst.com
sale@ponpe.com
www.ponpe.com
FUNCTIONAL DESCRIPTION
DISPLAYDISPLAY
DISPLAY
DISPLAYDISPLAY
A. BATTERY STATUS ICON (BLINKS WHEN LOW BATTERY CONDITION DETECTED) B. MEASUREMENT IN PROGRESS TAG C. SETUP: FACTORY CALIBRATION TAG D. CAL: CALIBRATION TAG E. AUTOMATIC TEMPERATURE COMPENSATION
(BLINKS WHEN TEMPERATURE EXCEEDS 10-40 °C / 50-104 °F RANGE) F. PRIMARY DISPLAY (DISPLAYS MEASUREMENT AND ERROR MESSAGES) G. TEMPERATURE UNITS H. SECONDARY DISPLAY (DISPLAYS TEMPERATURE MEASUREMENTS; WHEN BLINKING,
TEMPERATURE HAS EXCEEDED OPERATION RANGE: 0-80 °C / 32-176 °F) I. RANGE INDICATOR
3
Instruction Manual MA886 Refractometer for Sodium Chloride Measurements
FRONT PANELFRONT PANEL
FRONT PANEL
FRONT PANELFRONT PANEL
A. LIQUID CRYSTAL DISPLAY (LCD) B. READ KEY (USER MEASUREMENT) C. ZERO KEY (USER CALIBRATION) D. RANGE KEY (USER MEASUREMENT UNIT) E. ON/OFF F. STAINLESS STEEL SAMPLE WELL AND PRISM G. SECONDARY DISPLAY H. PRIMARY DISPLAY
BOTTOMBOTTOM
BOTTOM
BOTTOMBOTTOM
I. BATTERY COVER J. BATTERY COMPARTMENT
44
4
44
www.milwaukeeinst.com
sale@ponpe.com
www.ponpe.com
GENERAL DESCRIPTION
SIGNIFICANCE OF USE
Thank you for choosing Milwaukee. This instruction manual will provide you the necessary information for correct use of the meter.
The MA886 is an optical instrument that employs the measurement of the refractive index to determine sodium chloride concentration in aqueous solutions used in food preparation. It is not intended for sea water salinity measurements. The measurement of refractive index is simple and quick and provides the user an accepted method for NaCl analysis. Samples are measured after a simple user calibration with deionized or distilled water. Within seconds the instrument measures the refractive index of the solution. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for measurements where you need them.
The instrument utilizes internationally recognized references for unit conversion and temperature compensation. It can display the measurement of NaCl concentration 4 different ways: g/100 g, g/100 mL, Specific Gravity, and °Baumé.
Temperature (in °C or °F) is displayed simultaneously with the measurement (on 3 of the ranges) on the large dual level display along with icons for Low Power and other helpful message codes.
Key features include:
• Waterproof models offers IP65 waterproof protection
• Automatic Temperature Compensation (ATC)
• Battery operation with Low Power indicator
• Automatically turns off after 3 minutes of non-use.