Much of the material found in this report is taken from the manufacturer's
Instruction Manual and Service Manual. In many cases the material is quoted
7/21/2004 Page 1 of 22 Spectronic 501 & 601.EDP.doc
.
Page 2
Milton Roy
Spectronic 501/601
Spectrophotometer
Table of Contents
HOME TECHNICIAN'S NOTES CONTENTS
1. Equipment Needed
2. Parts and Supplies Needed
3. Introduction to equipment
4. Special Precautions
5. Initial Preparation
6. QA - Operational Verification
7. PMQA Procedures
8. Check List / Log Forms
9. Specifications
10. Technician's Notes
11. Factory Modifications:
12. Troubleshooting
13. Block Diagram
14. Manufacturer & Vendor Information
15. Appendix
7/21/2004 Page 2 of 22 Spectronic 501 & 601.EDP.doc
Page 3
Milton Roy
Spectronic 501/601
Spectrophotometer
Test Equipment Required
HOME TECHNICIAN'S NOTESCONTENTS
Generic Test Equipment Equipment Specific Test Equipment
Leakage Current Meter / Safety Analyzer 1 Spectrophotometric Function Test Set
Milton Roy Standards #333150 or
(Harleco #64666 or equivalent)
Digital Volt Meter
Hand Tools
Leakage Current Meter / Safety Analyzer
Digital Volt Meter
Hand Tools
Eye Protection (UV light and shattering
glass)
Protective gloves (UV light and lint-less)
1 Holmium test cuvette
335101-10001 SPECTRONIC 501/601
1
Operator's Manual
335101-10020 SPECTRONIC 501/601 Service
1
Manual
1 Stray Radiant Energy (SRE) filter
7/21/2004 Page 3 of 22 Spectronic 501 & 601.EDP.doc
Page 4
Milton Roy
Spectronic 501/601
Spectrophotometer
Parts and Supplies Needed
HOME TECHNICIAN'S NOTESCONTENTS
SUPPLIES PARTS
Paper Towel 1
General purpose cleaner 1 343424: Deuterium lamp
Canned Air (Dry) 1 332879: 2.0 ampere Slo-Blo fuse
Lint-less wipes (Kim-Wipes)
Disinfectant (Clean Conscience)
335423 Tungsten-halogen lamp: 6V - 10 Watt
(Phillips type 6605)
7/21/2004 Page 4 of 22 Spectronic 501 & 601.EDP.doc
Page 5
Milton Roy
Spectronic 501/601
Spectrophotometer
Introduction to Equipment
HOME TECHNICIAN'S NOTES CONTENTS
The Milton Roy 501/601 Spectrophotometer is a simple instrument designed for quick and easy
absorbency or transmittance measurements at single wavelengths. The 501 has a range of 325 to
999nm, while the 601 includes the UV range, (195-999nm), with the addition of a Deuterium lamp.
These units come with a variety of options such as a heated or cooled sample holder, a recorder
output, and an RS232 output. The units are designed for single wavelength measurements but a
scanning program could be written via the RS232 communications port. All necessary information is
given in the operator’s manual in section 9. Though the instrument could not be very accurate between
wavelengths far apart, it could be effective for short scans.
Operation is very basic. Enter the desired wavelength, zero the blank and read the sample. Results can
be printed out to a printer but it needs to accept serial data and at the set parameters of the unit – see
sections 7 and 8 of the operator’s manual.
Be aware of the precautions listed at the beginning of this procedure and servicing this unit can be very
simple.
7/21/2004 Page 5 of 22 Spectronic 501 & 601.EDP.doc
Page 6
Milton Roy
Spectronic 501/601
Spectrophotometer
Special Precautions
HOME TECHNICIAN'S NOTES CONTENTS
1. The Tungsten-Halogen lamp operates under high pressure. To avoid accidents, protect the lamp
from fingerprints, abrasions, and scratches when handling. Wear eye protection when adjusting the
lamp interchange mirror, and never operate the lamp with the shield removed.
2. Allow enough time for the shield and lamp to cool. If the spectrophotometer has been operating
even for a short time, the shield and lamp will be very hot. Temperatures are high enough to cause
contact burns, so allow at least 5 minutes before removing the shield and lamp.
3. Use a clean lint-free cloth or wear lint-free gloves to handle the new lamp: fingerprints on
the lamp detract from its peak performance and shorten the life of the lamp.
SHATTERING: The tungsten lamp operates under high pressure. To avoid accidents,
protect the lamp from fingerprints, abrasions, and scratches when handling. Wear eye
protection when adjusting the lamp interchange mirror, and never operate the lamp with
the shield removed.
4. The deuterium lamp operates at an extremely high voltage, and has dangerous voltages on the
terminals even when the lamp is turned off. Be sure to turn off the power and disconnect the power
cord before you attempt to replace the lamp.
ULTRAVIOLET: The short wavelength UV radiation emitted by the deuterium lamp can be
5.
harmful to the eyes and skin. Wear eye protection and avoid exposing skin whenever the lamp
cover is removed and the lamp is on (for example, when aligning the lamp mirror).
6. User Settings: If the unit is running, save that configuration in the unit’s non-volatile memory by
pressing the following keystrokes: 1,4 <SECOND> <DIAG>. Any changes you make will then be
replaced by these saved default values when the unit is powered on next unless you re-enter
different values via the keystrokes listed.
7/21/2004 Page 6 of 22 Spectronic 501 & 601.EDP.doc
Page 7
Milton Roy
Spectronic 501/601
Spectrophotometer
Initial Preparation
HOME TECHNICIAN'S NOTESCONTENTS
Test Parameter Storage:
The SPECTRONIC 501/601 can store the existing set of test parameters as the "default" values in nonvolatile memory. Once stored, these values will be present the next time you turn your
spectrophotometer on. Any changes made to these parameters during subsequent use of the
spectrophotometer will be temporary and replaced with the stored “default” values when power is
cycled.
The parameters that are stored are:
_______ Current wavelength (wavelength shown at startup).
_______ LMP SAV – Lamp saver status. <SECOND> <LMP SAV>
_______ LMP CH – Wavelength defining the change from visible to UV. <SECOND> <LMP CH>
_______ Data mode <%T, A or C >: (shown at startup - %T, A or C)
_______ Factor – +
_______ SIG AV – Noise reduction factor. <SECOND> <SIG AV>
_______ HI LIM – The highest acceptable final result <SECOND> <HI LIM>
_______ LO LIM – The lowest acceptable final result <SECOND> <LO LIM>
1. Historical: optional but beneficial in some cases
1.1. CHECK THE CURRENT EQUIPMENT LOG SHEET FOR:
1.1.1. General trends
1.1.2. Previous Problems and work performed on previous PMS.
1.2. RUN AN EQUIPMENT HISTORY FOR THIS SPECIFIC PIECE OF EQUIPMENT
1.2.1. Check for reoccurring problems.
1.2.2. Check for excessive repairs
1.2.3. Check for unresolved or intermittent problems.
1.3. RUN EQUIPMENT ERROR RECORDS (IF APPLICABLE).
1.4. CHECK ANY SERVICE REPORTS THAT THE CUSTOMER MIGHT HAVE FROM THE MANUFACTURER
OR OTHER SERVICE VENDORS.
2. PRELIMINARY: This does not apply to two-wire and double-insulated equipment.
2.1. LEAKAGE CURRENT:
2.1.1. Chassis to Earth Ground (power-cord ground wire intact).
2.1.2. Chassis to Earth Ground (Open ground wire):
2.1.3. Ground Wire Resistance:
2.2. CLEAN EXTERIOR AND INTERIOR ACCESSORIES
2.3. VISUALLY INSPECT THE UNIT
2.3.1. all cords and cables
2.3.2. verify the viability of the strain relief,
2.3.3. Knobs and buttons condition.
3. Post notification of intent to Pm the unit. Give an appropriate length of time prior to servicing, if
applicable.
4. Contact customer to notify him of your arrival and post in the instrument's usage log, if
applicable.
5. Store changed settings Enter the following keystrokes: 1,4 <SECOND FUNCTION> <DIAG>. The
new settings will now be the default settings.
6. Clean exterior and interior accessories
6.1. Do Not Sterilize. May be cleaned by wiping with a dry cloth.
7/21/2004 Page 9 of 22 Spectronic 501 & 601.EDP.doc
Page 10
Preventative Maintenance
(Continued)
HOME TECHNICIAN'S NOTESCONTENTS
7. LEAKAGE CURRENT:
7.1. Chassis to Earth Ground (power-cord ground wire intact).
7.2. Chassis to Earth Ground (Open ground wire):
7.3. Ground Wire Resistance:
8. Turn on and verify that it goes through the Self-Test
9. NOTE: For effective operation, the spectrophotometer must warm up for one-half hour after
being turned on. Allow an additional 15 minutes for stabilization when you turn a lamp on or off.
10. LAMP AND MIRROR ALIGNMENT: Proper alignment of the lamps and mirror is essential for
optimal performance of the spectrophotometer. (Far right Letter on display = lamp in use: T for
Tungsten, D = Deuterium and B = both.
11. The alignment itself consists of two parts:
11.1. Adjusting the position of the light beam hitting and passing through the filters.
11.2. Minimizing the absorbency reading.
12. The position of the mirror determines the amount of light hitting and passing through the slit. The
amount of light in turn directly affects your results and the overall performance of the
spectrophotometer. These steps help ensure peak performance of you instrument.
12.1. If the instrument is off, turn it on and wait for it to initialize. If it is on, turn it off and
then on again so that it initializes correctly.
12.2. Spectronic 601 only--Ensure that the wavelength is set to a value greater than 400 nm
and press <SECOND FUNCTION> <DEUT> to turn off the deuterium lamp. The lamp
indicator on the display should be a "T".
12.3. Remove the maintenance/lamp access door by loosening the lamp/maintenance
compartment access screw three to five complete turns and lifting up the door at the
back edge.
12.4. Examine the illuminated filter on the filter wheel to determine the position of the light.
Light should be centered on the filter and on the slit, as shown in Figure 6-3a.
7/21/2004 Page 10 of 22 Spectronic 501 & 601.EDP.doc
Page 11
Preventative Maintenance
(Continued)
HOME TECHNICIAN'S NOTES CONTENTS
12.5. If the light is not passing through the center of the filter as shown in Figure 6-3b, adjust
the Tungsten-Halogen lamp adjustment screw (22) and/or the Tungsten-Halogen lamp
vertical adjustment screw (30), as indicated in Figure.6.4, until the light is centered on
the filter.
13.
14. Figure 6.4
14.1. Press 6, 3, 0, <SECOND FUNCTION>, <GO TO λ > to set the wavelength to 630 nm.
14.2. Use the %T/A/C key to put the instrument into absorbency mode (A).
14.3. Press <1> <%T/A/C > and wait for the instrument to display an absorbency value. (It
should be near 1.0).
14.4. Use a flathead screwdriver with an insulated handle to turn the Tungsten-Halogen lamp
adjustment screw (22) until the instrument displays a minimum absorbency reading. If the instrument displays a negative absorbency reading, stop and repeat step 14.3.
14.5. Use a phillips screwdriver to turn the Tungsten-Halogen lamp vertical adjustment screw
(30) until the instrument displays a minimum absorbency reading. If the instrument
displays a negative absorbency reading, stop and repeat step 14.3 and 14.4
14.6. Turn on Deuterium Lamp: <SECOND FUNCTION>, <DEUT>
14.7. Spectronic 601 only--Press 2,4,0, <SECOND FUNCTION>, <GO TO λ > to set the
wavelength to 240 nm.
14.8. Spectronic 601 only--Use the %T/A/C key to put the instrument into absorbency
mode.
14.9. Spectronic 601 only--Press <1> <%T/A/C > and wait for the instrument to display an
absorbency value. (It should be near 1.0).
14.10. Spectronic 601 only--Use a flathead screwdriver with an insulated handle to turn the
deuterium lamp adjustment screw (21) until the instrument displays a minimum absorbency reading. If the instrument displays a negative absorbency reading, stop and
repeat step 14.9 and 14.10
7/21/2004 Page 11 of 22 Spectronic 501 & 601.EDP.doc
Page 12
Preventative Maintenance
(Continued)
HOME TECHNICIAN'S NOTES CONTENTS
14.11. When adjustments are complete, turn the instrument off and then on again to re-
initialize it.
14.12. Reposition the maintenance/lamp access door and secure it by tightening the
lamp/maintenance compartment access screw.
15. REVISION (DIAG 2): Press 2 <SECOND> <DIAG> and record the software revision.
16. DISPLAY TEST (DIAG 3):
16.1. Checks the spectrophotometer’s display by testing all segments. First, sixteen '*'s will
appear, then sixteen "0"s. The final display scrolls from left to right, presenting the
characters A through Z, and 0 through 9. To stop the test, press any key after the *’s
and 0’s have been displayed.
17.1. This routine turns on the Deuterium lamp, rotates the mirror, and sets the appropriate
gain. It also moves the monochromator from 659 to 653nm and displays the measured
peak wavelength. Press <CLEAR> to exit. Result should be 656 ± 2nm. Record value on
log sheet.
17.2. If out of specification perform the reprogramming of the EEPROM per EEPROM ERROR /
MONO ERROR CORRECTION PROCEDURES of August 13, 1997 - see at end of
bioengineering’s copy of the operation manual.
18.1. This routine is similar to diagnostic 6, except that monochromator scans from 489 to
483nm. Press <CLEAR> to exit. Result should be 486 nm (± 2nm). Record value on log
sheet.
19. PEAK NOISE MEASUREMENT (DIAG 8):
19.1. Measures noise for approximately 15 seconds and displays peak-to-peak noise in
absorbency. The value is computed on the displayed data, and applies the <SIG AV>
value that is in effect before selecting this diagnostic. The result should be <0.0015 A>
19.1.1. Press: 0, <SECOND> <SIG AV> to set for no filtering.
19.1.2. Press: 8, <SECOND> <DIAG> to run test.
19.1.3. Wait for the "Final PPK" value. Record value on log sheet.
19.1.4. Press <CLEAR> to exit.
19.1.5. Repeat steps 19.1.1 – 19.1.5 for SIG AV filtering values of 1 and 2.
20. STRAY RADIANT ENERGY (SRE) MEASUREMENT (DIAG 10): Optional
20.1. Checks SRE by displaying transmittance to within 0.01% resolution. To use this
diagnostic, move the spectrophotometer to the appropriate wavelength, set 0A/100%T,
and insert the appropriate SRE filter (or Blocking beam assy. if no SRE filter - see NOTE
below). Then, enter 10. <SPECIAL FUNCTION> <DIAG>. The SRE measurement
will be displayed in the form "SRE=XX.XX T", and will be held for 0.5 seconds (for a
longer display, depress and hold down any key as long as you wish to see the display).
NOTE: If SRE filters are not available - use a blocking beam assy. (the holmium filter
turned sidewise works well). Set wavelength at 340nm, run and record reading. Reset
wavelength to 220nm and run and record again.
Remove the SRE or Blocking-Beam Filter.
7/21/2004 Page 12 of 22 Spectronic 501 & 601.EDP.doc
Page 13
Preventative Maintenance
(Continued)
HOME TECHNICIAN'S NOTES CONTENTS
21. LAMP ALIGNMENT (DIAG 13): Inappropriately named, as it actually is a gain measurement.
21.1. Prompts for lamp choice, (601 only), then moves to the appropriate wavelength.
Converts display to a linear-responding bar graph. The left-hand digit indicates the
relative gain achieved in hexadecimal format with 0 the lowest and F the highest.
21.1.1. Press: 1,3 <SECOND> <DIAG>.
21.1.2. Select lamp by pressing <TUNG> or <DEUT>.
21.1.3. Record gain on log sheet for corresponding lamp.
21.1.4. Press <CLEAR> to go back to the lamp select prompt or press <CLEAR> again to
exit.
22. KEYBOARD TEST (DIAG 15):
22.1. Simply verifies that each key is working. Press the key and observe the display; if
keystroke does not produce the character listed in the chart below, contact you
authorized Milton Roy service representative. (To exit from this test turn the
spectrophotometer off and then on again.)
KEY: DISPLAYED AS:
<CLEAR > C
<SEND> S
<SECOND FUNCTION> N
<DOWN ARROW> D
<UP ARROW> U
<%T/A/C> T
<+/-> +
<.> .
<0-9> NUMBER PRESSED
23. PRINTER TEST (DIAG 16): Optional
23.1. Checks printer operation by transmitting ASCII codes for SPECTRONIC 501/601's
character set. Operates continuously until you press <CLEAR> to exit.
24. Photometric Accuracy: Use Neutral-Density Filters at 546.1 nm: run 10 times each.
24.1. 0.3 ABS Filter: record both average and Std Deviation
24.2. 0.5 ABS Filter: record both average and Std Deviation
24.3. 1.0 ABS Filter: record both average and Std Deviation
24.4. Holmium Oxide Filter:
24.4.1. At 546.1 nm: record both average and Std Deviation
24.4.2. At 360.8 nm: record both average and Std Deviation
7/21/2004 Page 13 of 22 Spectronic 501 & 601.EDP.doc
Page 14
Preventative Maintenance
(Continued)
HOME TECHNICIAN'S NOTES CONTENTS
25. PHOTOMETRIC LINEARITY TEST
25.1. Photometric Linearity: all three of each test tube set is within their respective tolerances.
(post readings below).
25.2. Readings of Test Kit:
Wavelength Relative
Concentration
420 nm 1.00 0.824 - 0.602 15.0 - 25.0
0.75 0.602 - 0.456 25.0 - 35.0
0.50 0.409 - o.310 39.0 - 49.0
0.25 0.215 - 0.149 61.0 - 71.0
520 nm 1.00 0.824 - 0.602 15.0 - 25.0
0.75 0.602 - 0.456 25.0 - 35.0
0.50 0.409 - o.310 39.0 - 49.0
Actual
Absorbance
Absorbance
tolerances
Actual
%T
%T
tolerances
0.25 0.215 - 0.149 61.0 - 71.0
670 nm 1.00 0.824 - 0.602 15.0 - 25.0
0.75 0.602 - 0.456 25.0 - 35.0
0.50 0.409 - o.310 39.0 - 49.0
0.25 0.215 - 0.149 61.0 - 71.0
26.
27. DRIFT TEST:
27.1. Empty the sample compartment
27.2. Enter: 5,0, 0, <SECOND FUNCTION>, <GO TO ^>
27.3. Make sure it is in ABS mode: <SECOND FUNCTION>, %T/A.C>
27.4. Record readings: Initial reading and reading after 1 hour.
7/21/2004 Page 14 of 22 Spectronic 501 & 601.EDP.doc
20. PEAK FIND TEST: (Diag test # 6 and # 7)
A. 656 nm (± 2.0 nm): <6>, <SECOND FUNCTION>, <DIAG>
B. 486 nm (± 2.0 nm): <7>, <SECOND FUNCTION>, <DIAG>
21. PEAK NOISE test: (Diag. Test # 8)
A. SA = 0 (<0.0015A):
<8>, <SECOND FUNCTION>, <DIAG>
<0>, <SECOND FUNCTION>, <SIG AV>
B. SA = 1 (<0.0015A): <1>,, <SECOND FUNCTION>, <SIG AV>
<8>, <SECOND FUNCTION>, <DIAG>
601
L-2910
Service
Manual
Section
C. SA = 2 (<0.0015A):
<8>, <SECOND FUNCTION>, <DIAG>
<2>,, <SECOND FUNCTION>, <SIG AV>
22. STRAY RADIANT ENERGY (SRE): (Diag test # 10)
A. 340 nm:
•
• put into %T mode. . . <%T/A/C> until %T
• empty sample chamber,
• <SECOND FUNCTION>, <%T/A/C (100%T/ZEROA)>,
• insert blocking filter (Holminum - sideway),
• 1,0, <SECOND FUNCTION>, <DIAG>
3,4,0, <SECOND FUNCTION>, <YES (Go to λ)>
B. 220 nm:
•
• put into %T mode. . . <%T/A/C> until %T
• empty sample chamber,
• <SECOND FUNCTION>, <%T/A/C (100%T/ZEROA)>,
• insert blocking filter (Holminum - sideway),
• 1,0, <SECOND FUNCTION>, <DIAG>
2,2,0, <SECOND FUNCTION>, <YES (Go to λ)>
23. LAMP GAIN: (Diag. # 13)
A. VIS (0 - F) (630nm): 1,3, <SECOND FUNCTION>,
<DIAG> Select <9 - TUNG> for lamp.
B. UV (0 - F) (240nm): <Clear>, select <6 - DEUT> for lamp
7/20/2004Page 2 of 3Spectronics601_ChkLst.xls
Page 17
Milton Roy - SPECTRONICS 601 - Spectrophotometer
A
PMQA Check-List
Dept:Procedure:
SSET #:
Location:Mfg:m/n:s/n
See the Corresponding PMQA Procedure for details
24. PHOTOMETRIC ACCURACY:
5,4,6, .,1,
a. 0.3 abs Neutral Density Filter @ 546.1 nm:
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Insert 0.3 abs N.D. Filter and Read value:
• Remove N.D. Filter. Average/ Std. Deviation:
b. 0.5 abs Neutral Density Filter @ 546.1 nm:
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Insert 0.5 abs N.D. Filter and Read value:
• Remove N.D. Filter. Average/ Std. Deviation:
c. 1.0 abs Neutral Density Filter @ 546.1 nm:
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Insert 1.0 abs N.D. Filter and Read value:
• Remove N.D. Filter. Average/ Std. Deviation:
d. Holmium Oxide Filter @ 546.1 nm:
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Insert Holmium Filter and Read value:
• Remove Filter. Average/ Std. Deviation:
e. Holmium Oxide Filter @ 360.8 nm:
• 3, 6, 0, ., 8, <SECOND FUNCTION>, <GO TO λ>
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Insert Holmium Filter and Read value:
• Remove Filter. Average/ Std. Deviation:
25. DRIFT TEST: abs, 500 nm
• 5, 0, 0, <SECOND FUNCTION>, <GO TO λ>
• Empty sample compartment
• <SECOND FUNCTION>, <%T/A/C>
• Read value: initially and after 1 hour
• Remove Filter. Initial reading & 1-hr reading:
<SECOND FUNCTION>, <YES - GO TO λ>
Milton Roy
Dates performed:
Technician's Initials:
601
L-2910
Service
Manual
Section
REMARKS:
7/20/2004Page 3 of 3Spectronics601_ChkLst.xls
Page 18
Milton Roy
Spectronic 501/601
Spectrophotometer
Specifications
HOME TECHNICIAN'S NOTESCONTENTS
Light sources Tungsten-Halogen (325 to 999nm)
Deuterium (195 to 380 nm) (SP 601 only)
Monochromator Modified Czerny-Turner with holographic grating (1200 lines per
millimeter).
Detector Single silicon photodiode
Wavelength range 501: 325-999nm, 601: 195-999 nm
Wavelength accuracy
Wavelength precision Better than 0.5 nm
Spectral slit width <= 5 nm
Stray radiant energy <= 0.05% T at 220nm and 340 nm
Photometric accuracy
Photometric stability
Noise Less than 0.004 A near 2 A nm, Less than 0.0015 A near 0 A
Photometric Ranges Absorbency: -0.1 to 3.0 A, Transmittance: 0.0 to 125% T, concentration:
Sample compartment
dimensions
Power requirements
Overall dimensions 19.7 cm H x 44.5 cm W x 49 cm D; (7.75" H x 17.5" W x 14.3"D).
Weight 15 kg (35 lbs.)
Environment Operating 50-105 F; 10-40.5 C. relative humidity 10-60%
Storage -40 to 105 F; -40 to 40.5 C; relative humidity not to exceed 60% (Allow
± 1.0 nm
± 0.005 A near 1.0 A
± 0.003 A per hour near 0 A
-9999 to + 9999
10.2 cm (4 in) High x 10.6 cm (4.2 in) Wide x 17.1 cm (6.5 in) Deep
100/115 and 220/240 V. ± 10%; 50/60 Hz, ± 2 Hz; 2.0 A
unit to adjust to room temperature for 24 hours when taking it out of
storage.)
7/21/2004 Page 16 of 22 Spectronic 501 & 601.EDP.doc
Page 19
Milton Roy
Spectronic 501/601
Spectrophotometer
Technician’s Notes
HOME TECHNICIAN'S NOTES CONTENTS
1 Thermo Spectronics
2 Most parts and manuals are no longer available (1/1/02).
has bought out Milton Roy.
7/21/2004 Page 17 of 22 Spectronic 501 & 601.EDP.doc
Page 20
Milton Roy
Spectronic 501/601
Spectrophotometer
Factory Modifications
HOME TECHNICIAN'S NOTESCONTENTS
EEPROM ERROR / MONO ERROR CORRECTIONS PROCEDURES
The following procedure should allow for correction of the EEPROM error that is occurring with the
501/601 Spectrophotometer.
This problem is usually caused by power line anomalies (surges, spikes, or low voltage conditions). A
line conditioner may be required if this is a recurring problem.
If the power line is okay, then a problem may exist on the CPU board in the unit. Replacement of the
board or the EEPROM may be required. An internal power supply problem could also be a cause.
See FAX from Milton Roy at the end of bioengineering’s operator manual.
7/21/2004 Page 18 of 22 Spectronic 501 & 601.EDP.doc
Page 21
Milton Roy
Spectronic 501/601
Spectrophotometer
Troubleshooting
HOME TECHNICIAN'S NOTES CONTENTS
See Section 6.4 MAINTENANCE & TROUBLESHOOTING page 11 in bioengineering’s copy of Spectronic
601 Operation Manual.
See Section 6.4.2 ERROR MESSAGES page 12 in bioengineering’s copy of Spectronic 501/601 operation
manual.
See the August 13, 1997 FAX from Spectronic Instruments, Inc: EEPROM ERROR/ MONO ERROR
CORRECTIONS PROCEDURES at the back of bioengineering’s Spectronic 601 operation manual.
7/21/2004 Page 19 of 22 Spectronic 501 & 601.EDP.doc
Page 22
Milton Roy
Spectronic 501/601
Spectrophotometer
Block Diagram
HOME TECHNICIAN'S NOTES CONTENTS
7/21/2004 Page 20 of 22 Spectronic 501 & 601.EDP.doc
Page 23
Milton Roy
Spectronic 501/601
Spectrophotometer
Manufacturer & Vender Information
HOME TECHNICIAN'S NOTES CONTENTS
7/21/2004 Page 21 of 22 Spectronic 501 & 601.EDP.doc
Page 24
Milton Roy
Spectronic 501/601
Spectrophotometer
Appendix
HOME TECHNICIAN'S NOTES CONTENTS
7/21/2004 Page 22 of 22 Spectronic 501 & 601.EDP.doc
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.