Milton Roy LMI P1 Instruction Manual

Page 1
LMI ELECTRONIC DOSING PUMPS
Code : P1
Basic manual Instruction supplement Spare parts list for drive assembly Declaration « CE » of conformity List of « Technical assistance » and « spare parts » departments
07/98
Page 2
INSTRUCTION MANUAL
FOR INSTALLATION, OPERATING, AND MAINTENANCE.
LMI ELECTRONIC DOSING PUMPS
Diaphragm liquid end
Basic manual
This manual should be made available to the person responsible for installation,
operating and maintenance.
Date : 02/2012 Ref : 160.0101
.001 Rev. D
Replaces Rev. 0 09/98
Page 3
Page 4
LMI, Doseur à membrane, L001, Rev.B, 1098
CONTENTS
I – DESCRIPTION
I – 1. Unpacking I – 2. Description I – 3. Accessories I – 4. Precautions
II – INSTALLATION
II – 1. Hydraulic installation II – 2. Electrical installation
III – START UP
III – 1. Priming III – 2. Output adjustment controls III – 3. Calibration III – 4. Volumetric calibration in external mode III – 5. Pressure control
IV – MAINTENANCE
IV – 1. Spare parts replacements IV – 2. Checking pump for proper zeroing (stroke knob) IV – 3. Troubleshooting IV – 4. EPU Resistance chart
Page 5
0.0 Precautions
The pump described here may be used for the dosing of liquid products in connection with the applications set out in this manual. The following precautions should be taken when working with LMI metering pumps. Please read this section carefully prior to installation.
Storage precautions
- Storage period of less than six months:
The product should preferably be stored in its original packing and be protected from bad weather.
- Storage period of more than six months:
Store the pump in its original packing. It should be stored covered by a heat-sealing plastic cover and desiccant bags should be used (quantity dependent on volume and planned storage period). Store away from the influence of bad weather.
Protective Clothing
ALWAYS wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump. Additional precautions should be taken depending on the solution being pumped. Refer to MSDS precautions from your solution supplier.
Water Pre-Prime
All LMI pumps are pre-primed with water when shipped from the factory. If your solution is not compatible with water, disassemble the Pump Head Assembly. Thoroughly dry the pump head, valves, seal rings, balls and Liquifram (diaphragm). Reassemble head assembly tightening screws in a crisscross pattern. Refill the pump head with the solution to be pumped before priming the pump. (This will aid in priming.)
Solution Compatibility
Determine if the materials of construction included in the liquid handling portion of your pump are adequate for the solution (chemical) to be pumped. Always refer to the solution supplier and the LMI Chemical Resistance Chart for compatibility of your specific LMI metering pump. Contact your local MRE distributor for further information.
Tubing Connections
Inlet and outlet tubing or pipe sizes must not be reduced. Outlet tubing size must not be increased. Make certain that any tubing is SECURELY ATTACHED to fittings prior to start-up (see Section ., Tubing Connections). ALWAYS use LMI supplied tubing with your pump, as the tubing is specifically designed for use with the pump fittings. It is recommended that all tubing be shielded and secure to prevent possible injury in case of rupture or accidental damage. If tubing is exposed to sunlight, black UV resistant tubing should be installed. Check tubing frequently for cracks and replace as necessary.
Vinyl Tubing
Your carton may contain a roll of clear vinyl tubing; this is only for connection to the return line of the FastPrimeHead and must not be used as discharge tubing.
Connectors
All connectors shall be tightened by hand. An additional turn of 1/8 to 1/4 might be necessary to ensure tightness. Any excessive tightening or using a toolmight damage the connecting device and the seals or the liquid end.
All LMI pumps have straight screw machine threads on the head and fittings and are sealed by the O-rings. DO NOT use Teflon® tape or pipe dope to seal threads. Teflon® Tape may only be used on the 1/2" NPT thread side of the Injection Check Valve, the stainless steel liquid end connections, or if piping is directly connected to the pipe threads of the suction or discharge fittings.
Installation
Always comply with local regulations and requirements. Always adhere to your local plumbing codes and requirements. Be sure installation does not constitute a cross connection. Check local plumbing codes for guidelines. Milton Roy Europe is not responsible for improper installations.
Back Pressure/Anti-syphon valve
If you are pumping downhill or into low or no pressure system, a back pressure/antisyphon device such as LMI’s Four­Function Valve should be installed to prevent over pumping or siphoning. Contact your distributor for further information..
Electrical Connections
WARNING: To reduce the risk of electrical shock, the metering pump must be plugged into a properly grounded grounding-type receptacle with ratings conforming to the data on the pump nameplate.. Do not use adapters! All wiring
must conform to local electrical codes. If the supply cord is damaged, it must be replaced by the manufacturer, stocking distributor, or authorized repair centre in order to avoid a hazard..
Environment
The ambient temperature must be less than 50°C. Avoid direct exposure to sunlight. The pump must be accessible and situated in a ventilated room.
Page 6
Noise
The noise level of your equipment is less than 65 dB (A).
Maintenance
CAUTION - ELECTRICAL RISK: Before performing any maintenance work, deactivate the pump and make sure that all electrical connections are disconnected. CAUTION: : DISCHARGE LINE DEPRESSURIZATION If you are pumping downhill or into low or no pressure system, a back pressure / antisyphon device such as LMI’s Four-Function Valve should be installed to prevent over pumping or siphoning. Contact your LMI distributor for further information. CAUTION: CHEMICAL AND INFLAMMABLE PRODUCTS: Take all the precautions necessary when performing work on equipment that is dosing chemical and/or inflammable products. Contact your local Milton Roy Europe distributor for more information.
Earth leakage protection
CAUTION: It is essential to install an earth leakage circuit breaker calibrated on the basis of the electrical values noted on the equipment plate.
Cleaning
Make sure that your equipment is clean at all times and that no external items interfere with its functioning.
Page 7
I - 1
LMI, Doseur à membrane, L001, Rev.B, 1098
PART I – DESCRIPTION
This manual addresses the installation, maintenance and troubleshooting procedures for electronic dosing pumps. An instruction supplement and a liquid end sheet complete this manual.
Please read them carefully before installing your pump.
I – 1. UNPACKING
The packaging must be carefully examined on receipt in order to ensure that the contents have not sustained any obvious damage.
The carton contains the items necessary to a proper installation.
I - 2 - DESCRIPTION
Your pump consists of two parts :
1. The drive assembly
2. The liquid handling assembly
The pump code which appears on the packaging and on the identification plate consists of two codes according to drive assembly type and liquid handling assembly type.
Dosing pump Function valve * Ceramic weight*
Tubing Injection check valve* Foot valve*
* depending on model
DRIVE
A753
LIQUID HANDLING ASSY
352SM
Page 8
I - 2
LMI, Doseur à membrane, L001, Rev.B, 1098
Output adjustment controls
According to pump type, output adjustment controls are available :
Stroke adjustment Each pump is equipped with a stroke adjustment. Stroke control provides adjustment of percent of maximum diaphragm travel.
Note : For B & C Series, stroke adjustment only when pump is running.
Speed adjustment A1, A7, B1, B7, C1, C7, J5, P1 Series pumps are equipped with a speed adjustment. Speed control provides adjustment of the percent of maximum strokes per minute.
Pressure control adjustment B7 and C7 Series Pumps are equipped with a pressure control adjustment. Pressure control provides the adjustment of the pump’s pressure capability and power consumption, reducing heat, pipe shock and pulsation.
For more details, refer to Instruction supplement
I - 3. ACCESSORIES
4-Function valve
This accessory is supplied as standard equipment or as option. The functions are :
1. Anti-syphon (automatic)
Prevents syphoning when pumping downhill or into a vacuum.
2. Back pressure (automatic)
Supply approximately 1,7 bar back pressure to prevent over pumping when little or no system back pressure is present.
3. Priming (manual)
Prime your pump while it is connected to a pressurized line.
4. Line depressurization (manual)
By pulling both knobs (yellow and black) the discharge line will drain back to your supply tank.
3-Function valve
This accessory is supplied as standard equipment or as option (according to Series 300 models). The functions are :
1. Priming (manual)
Prime your pump while it is connected to a pressurized line.
2. Pressure relief (automatic)
Provides protection against excessive system pressure.
3. Line depressurization (manual)
By opening the relief valve, the discharge line will depressurize without having loosen or disconnect discharge tubing.
Bleed/4-Function valve
This accessory is supplied as standard equipment or as option (according to Series 300 models). The functions are :
1. Anti-syphon (automatic) Prevents syphoning when pumping downhill or into a vacuum.
2. Back pressure (automatic) Supply approximately 1,7 bar back pressure to prevent over pumping when little or no system back pressure is present.
3. Continuous bleed (automatic) This valve is designed to permit an adjustable, continuous bleed or degasing or entrapped vapors. It is ideal for pumping small volumes of solutions having low vapor pressures.
4. Line depressurization (manual) By opening bleed port, the discharge lline will depressurize, without having to loosen tubing or fittings.
Foot valve
(Accessory available according to the model) The foot valve makes easier priming and keep the
pump primed in suction lift applications. The strainer prevents sediment to pass through and
valves to clog.
Injection check valveCanne d’injection
The injection check valve prevents backflow from a treated line.
I – 4. PRECAUTIONS
Always wear protective clothing, face shield, safety glasses and gloves when working on or near your metering pump.
All LMI pumps are shipped from factory with water in pump head to make easier priming.
If the pump head is empty, refill it with water or solution compatible with pumped solution before starting up the pump (see note on the Liquid End Sheet).
Solution compatibility
Verify that the liquid handling components are compatible with the pumped solution (refer to Liquid End Sheet).
Make certain that all tubing is securely attached to fittings prior to start-up.
Tubing supplied allows to match with pump maximum specifications.
All fittings should be hand tightened.Overtightening or use of a pipe wrench can cause damage to the fittings, causing the pump to lose prime or not function.
All LMI pumps have straight ¾’’- 16N or 7/8 ’’- 14N or 1"- 12 N machine threads on the head and fittings.
The seal rings prevent leaks. Do not use Teflon tape between fittings and pump head.
Page 9
II - 1
LMI, Doseur à membrane, L001, Rev.B, 1098
PART II – INSTALLATION
II – 1. HYDRAULIC INSTALLATION
The pump can be mounted in one of two ways :
Flooded suction (ideal installation).
Suction lift – when suction lift is less than 1,5 m for
solutions having a specific gravity of water. Your LMI metering pump must be mounted so that the suction and discharge valves are vertical.
Flooded suction
The pump is mounted at the base of the storage tank. This installation is the most trouble-free, and is
recommended for very low outputs, solutions that gasify, and high viscosity solutions. (Install shut off valve on the tank).
1 Solution tank 9 Vent 4 Pump 10 Tee 8 Injection check valve
Fig. 2.1a : Flooded suction
Suction lift
Wall bracket mount. The pump may be mounted directly above the solution tank. A pump mounted in this manner allows for easy changing of solution tank.
1 Solution tank 4 Pump 6 Pressure line 8 Injection check valve
10 Tee
A 50 mm space for
sediment accumulation
Fig. 2.1b : Suction lift
Page 10
II - 2
LMI, Doseur à membrane, L001, Rev.B, 1098
Tank mount. The pump may be mounted on a molded tank. Tanks from 60 to 1000 liters are available for this utilisation.
1 Solution tank 2 Foot valve (equipped
with strainer) 4 Pump 6 Flow 8 Injection check valve
10 Te 11 Ceramic weight
A 50 mm space for
sediment accumulation
Shelf mount (customer supplied). The pump may be mounted on a shelf. An LMI mounting kit is available for securing the pump to a shelf. (Series A , B, C).
1 Solution tank 2 Foot valve (equipped
with strainer) 4 Pump 6 Flow 8 Injection check valve
10 Te 11 Ceramic weight
A 50 mm space for
sediment accumulation
Fig. 2.1c : Tank mount
Fig. 2.1d : Shelf mount
Page 11
II - 10
LMI, Doseur à membrane, L002, Rev.B, 1098
Pump location
Locate pump in an area convenient to solution tank and electrical supply. The pump should not be subjected to ambient temperatures above 50°C. Avoid exposing to direct sunlight the pump.
Tubing connections
For installation and trouble-free running, use
only recommended tubing size (refer to spare parts list).
Do not use clear vinyl tubing on the discharge
side of the pump. The pressure created by the pump can rupture the vinyl tubing.
Before installation, all tubing must be cut with
a clean square end.
Remove the plugs which protect the head
connections and valves.
Do not use pliers or pipe wrench on coupling
nuts or fittings.
Refer also to the Liquid End Sheet.
Foot valve/Ceramic weight
The foot valve must be sit in a vertical position at the bottom. Position approximately 50 mm off the bottom if the tank or drum contains sediment. The ceramic weight positions the foot valve in a vertical position. Refer to the Liquid End Sheet for details of connections
1
Solution tank
4
Pump
2
Foot valve
11
Ceramic weight
Fig. 2.2a : Foot valve/Ceramic weight
Injection check valveCanne d’injection
Connect the injection check valve to your discharge (outlet) line (refer to the Liquid End Sheet). For in-line injection, use fitting or pipe tee with reducing bushing to ½’’ (12x17). Use Teflon tape or pipe dope to seal the pipe threads only.
When installing the injection check valve, be sure to position it so that the valve enters the botton of your pipe in a vertical position (see fig. 2.2b). Variations left and right within 30° are acceptable. For low pressure (under 1,7 bar) or open tank use, the injection nozzle prevents siphoning but not flow increase. In this case,we recommend the use of a 4-Function valve. Refer to the Liquid End Sheet for details of connections.
Page 12
II - 4
LMI, Doseur à membrane, L001, Rev.B, 1098
8 Injection check valve B Use Teflon tape 10 Tee C Do not use Teflon tape 12 Reducing bushing ½ ‘’ D
± 30° variation acceptable
13 Pipe cross section
Fig. 2.2b : Injection check valve
II – 2. ELECTRICAL INSTALLATION
Check the specifications of the pump and compare them with the voltage available on your installation before making connections.
WARNING : To reduce the risk of electrical shock, the metering
pump must be plugged into a fused grounded outlet with ratings conforming to the pump specification chart (see Technical characteristics : refer to peak power to determine the appropriate fuse size).
DO NOT USE ADAPTERS. All wiring must be conform to local electrical codes.
OVER-VOLTAGE LMI pumps include a Varistor vonnected on the supply
voltage terminals. This device is an over-voltage protection. It will burn if
the supply voltage exceeds 150 V for 115 V rated pumps or 275 V for 230 V rated pumps to protect the pulser.
As soon as it happends, it means the pump got a too high voltage peak. After supply voltage control, replace the Varistor and the pump will run again.
Page 13
III - 1
LMI, Doseur à membrane, L001, Rev.B, 1098
PART III – START UP
Pumps are shipped from the factory with water in the pump head to aid in priming. After a long storage, refill the pump head with water or solution compatible with pumped solution (see note on the Liquid End Sheet).
III – 1. PRIMING
After connecting the pump, plug in or switch the pump on. While the pump is running, set the speed knob at 80% and the stroke knob at 100%. If the pump is equipped with a pressure control knob (B7, C7), set the switch on « INTERNAL » and turn the pressure control knob fully clockwise.
III – 2. OUTPUT ADJUSTMENT CONTROLS
Once the pump has been primed, an appropriate output adjustment must be made.
The identification plate indicates the output at max. stroke and max. speed (100%) and max. pressure. The graduations of stroke knob and speed knob correspond to percentage of this max. output.
Note : P0 pumps have no stroke adjustment. Example : Use max. output = 2,3 L/H. If the pump is set at 40%
speed and 60% stroke lenght, the appropriate pump out put is 2,3 L/H x 0.40 x 0.60 = 0,55 L/H.
Note : Remember that 6x8 tubing is 0,22 ml/cm which is 0,8 % wrong for 1 l measuring pot.
III – 3. CALIBRATION
Once installation is complete and the approximate output has been determined, the pump should be calibrated to adjust speed and stroke for your actual desired output.
1. If equipped (B7 – C7 series), make certain Pressure control knob is turned fully clockwise.
2. Be sure the pump is in normal service (i.e. including factors such as injection pressure, fluid viscosity, suction lift …).
3. Place the foot valve in a graduated container.
4. Start the pump at maximum frequency and let it run to remove air trapped in suction line (bleed valve open when using a 4FV). Close bleed after priming.
5. Turn the pump off. Refill graduated container to a level strarting point.
6. Using a stopwatch or timer, turn the pump on for a measured amount of time (50 pump strokes minimum). The longer the time period, the more confident you can be of the results.
7. The volume and the time unit are known. Calculate the output. If the output is too low or too great, adjust speed and or stroke, estimating required correction.
8. Adjust stroke length knob to desired value refering to calibration.
2 Foot valve 10 Tee 4 Pump 14 Graduate container
(1 l min.) 6 Pressure line E Suction lift 8 Inj. check valve
Fig. 3.2a : Calibration
Page 14
III - 2
LMI, Doseur à membrane, L001, Rev.B, 1098
III – 4. VOLUMETRIC CALIBRATION IN EXTERNAL MODE
1. Since pump output is governed by an external device such as Flowmeter-pulser, Liquitron Current-to­frequency converter or 4-20 mA DC signal from an instrument with an LMI Analog-to-digital converter, only the ouptut per stroke may be calibrated.
2. With pump primed and discharge tubin connected to the injection point as it would be in normal service, place foot valve in a graduated container.
3. Swtch pump to Internal mode with speed knob set at 100 until air is exhausted from suction line and pump head.
4. Adjust control pressure (B7 – C7 series).
5. Turn the pump off. Refill graduated container to a level strarting point.
6. Switch pump on and count the number of strokes. For a good accuracy, count at least 100 strokes. Switch pump off.
7. Note the volume pumped during the calibration period and the number of strokes. Divide into the volume the number of strokes to determine the volume of solution pumped per stroke.
8. Adjust stroke length knob to desired value refering to calibration.
III – 5. PRESSURE CONTROL
Adjust pressure control (B7 - C7 series) : while unit is running, turn pressure control knob fully clockwise then turn slowly counter-clockwise until unit just begins to stall. This is the optimmum pressure control setting for your installation. Adjust pressure control to prolong pump life.
Increase setting if back pressure is increased.
F : Excess pressure
Fig. 3.3a : Control pressure
Note : On series A9, B9, C9, this setting is in the main menu.
Page 15
IV - 1
IV – MAINTENANCE
IV – 1. SPARE PARTS REPLACEMENTS
WARNING : Always wera protective clothing, face
shield, safety glasses and gloves when performing any maintenance or replacement on your pump. LMI metering pumps are designed for trouble-free operation, yet routine maintenance of elastomeric parts is essential for optimmum performance. Refer to the spare parts list. Replacements frequency will depend on your particular application. We recommend replacing these parts at least one a year.
Diaphragm replacement
1. Depressurize the pump (see Liquid End Sheet).
2. Place the foot valve into a container of water or
other neutralizing solution. Turn the pump on to flush the head assembly. Lift the foot valve out of the solution and continue to pump air into the pump head until the pump head is purged of water or neutralizing solution. If the liquid cannot be pumped due to diaphragm rupture, using protective gloves, carefully disconnect the suction and discharge tubing. Remove the screws to the head and immerse the head in water or other neutralizing solution.
3. While running, set the stroke knob to zero and turn
the pump off.
4. Dismantle the liquid end. Unscrew the diaphragm.
Clean the diaphragm recess and the disk. Check the size of replacement diaphragm, the code (0,5 - 0,9 ­1,8 - 3,0- 6,0) is indicated on the diaphragm and the disk or spacer (see fig. 4.1a).
5. Reinstall the disk (A, J, P series).
6. Start the pump and turn the stroke knob to the
setting indicated on stroke setting chart which matches the pump model number located on the pump dataplate. Stop the pump.
7. Grasp the outer edge of the diaphragm and adjust
by screwing it in or out so that the center of the diaphragm is flush with the outside of the spacer edge (see fig. 4.1b). Once the diaphragm is properly positionned, remount the pump head to the spacer using the screws. Tighten in a crisscross pattern. For the diaphragm size 3.0 (show on the diaphragm) put the glue (loctite 221) on the fixing screws. Screw the liquid end’s screws at 0.6N.m then tighten them at 3.5N.m Note : Tighten the liquid end’s screws diametrically opposite
AFTER ONE WEEK OF OPERATION, RECHECK THE SCREWS AND TIGHTEN IF NECESSARY.
16 Diaphragm 18 Code 17 Spacer
Fig. 4.1a : Diaphragm identification
15 Straight edge G Gap 16 Diaphragm H No gap 17 Spacer
Fig. 4.1b : Diaphragm fitting
Pump series
Stroke knob setting
AA - A – B – J – P (tous modèles) Cx0 – Cx1 – Cx2 Ex0 – Ex1 – Ex2
90 %
Cx3 – Cx4 – Cx5 Ex3 – Ex4
70 %
Page 16
IV - 2
LMI, Doseur à membrane, L001, Rev.B, 1098
Ball, seal ring and injection check valve spring replacement or set of cartridges replacement (depending on model)
Refer to the spare parts list for the proper sparte parts kit number.
1. Depressurize the pump (see Liquid End Sheet).
2. Flush and purg the pump head before dismantling (see above).
3.During dismantling, note the orientation of each item.
4. Dismantling the injection check valve. Unscrew the valve body and replace the spring, the ball and the seal ring. Refer to the spare parts list drawing.
5. Replace the seal rings and balls in the pump head and the foot valve. Order of assembly changes depending on valve location.
6. In case of cartridge liquid end, see spare parts list drawing to identify the orientation of each item.
IV – 2. CHECKING PUMP FOR PROPER ZEROING (STROKE KNOB)
1. With pump running, turn stroke knob counter­clockwise toward zero or end of black or red band.
2. Listen to the clicking as the pump is running. The pump shoukd operate quietly at the zero position (no clicking). If the pump continues to click at zero or stops clicking before zero is reached, the pump zero must be reset.
Push on knob
1. Remove stroke knob from the pump by grasping the knob firmly and pulling it toward you.
2. Pry off the yellow cap.
3. Place the knob on a flat surface.
4. Using needle nose pliers, squeeze the inner section together while lifting the outer section up.
5. Push the inner section back onto the « D » shaped stroke shaft.
6. With the pump running, zero the pump by turning the inner sectionof the knob counter-clockwise until the pump stops clicking.
7. Position the outer section of the knob so that the pointer aligns with zero on the nameplate or end of the black or red band.
8. Push down on the outer section (a snap sound indicates parts are locked together).
9. Replace the yellow cap over the outer section of the knob, aligning the tabs on the cap with the slots inside the knob.
20 Shaft 22 Knob base 21 Knob pointer 24 Knob pointer
Fig. 4.2a : Push on knob
Page 17
IV - 3
LMI, Doseur à membrane, L001, Rev.B, 1098
IV – 3. TROUBLESHOOTING
Pump will not prime
Pump not turned on or plugged in. Turn on pump/plug in pump.
Output dials not set properly. Always prime pump with speed at 80 % and stroke at 100 %.
Foot valve not in vertical position on bottom of tank. See Chapter II – 1. Section Foot valve/ceramic weight.
Pump suction lift too high. Maximum suction lift is 1,5 m. Pump with High viscosity liquid handling assemblies require flooded suction.
Suction tubing is curved or coiled in tank. Suction tubing must be vertical. Use ceramic weight supplied with pump. (see Chapter II – 1. Section Foot valve/ceramic weight).
Fittings are over tightened. Do not overtighten fittings. This causes seal rings to distort and not seat propertly which causes pump to leak back or lose prime.
Air trap in suction valve tubing. Suction tubing should be as vertical as possible. AVOID FALSE FLOODED SUCTION (see Chapter II – 1. Section Flooded suction).
Too much pressure at discharge (pump without 4­function valve). Shut off valves in pressurized line. Disconnect tubing at injection check valve (see Chapter III – 1. Priming). When pump is primed, reconnect discharge tubing.
Pump loses prime
Solution container ran dry. Refill container with solution and reprime (see Part III).
Foot valve not in vertical position on bottom of tank. See Chapter II – 1. Section Foot valve/ceramic weight.
Pump suction lift too high. Maximum suction lift is 1,5 m. Pump with High viscosity liquid handling assemblies require flooded suction.
Suction tubing is curved or coiled in tank. Suction tubing must be vertical. Use ceramic weight supplied with pump. (see Chapter II – 1. Section Foot valve/ceramic weight).
Fittings are over tightened. Do not overtighten fittings. This causes seal rings to distort and not seat propertly which causes pump to leak back or lose prime.
Air trap in suction valve tubing. Suction tubing should be as vertical as possible. AVOID FALSE FLOODED SUCTION (see Chapter II – 1. Section Flooded suction).
Air leak on suction side. Check for pinholes, cracks. Replace if necessary.
Leakage at tubing
Worn tubing ends. Cut tubing about 25 mm (1’’) off tubing and then replace as before.
Loose or crack fitting. Replace fitting if cracked. Carefully hand tighten fittings. Do not use pipe wrench. Once fitting comes into contact with seal ring, tighten an additional 1/8 or ¼ turn.
Worn seal rings. Replace balls and seal rings (see part IV).
Solution attacking Liquid handling assembly material. Consult Technical Assistance Department for alternate materials compatible with the pumped solution.
Low output or failure to pump against pressure
Pump’s maximum pressure rating is exceeded by injection pressure. Injection pressure cannot exceed pump’s maximum pressure. See pump data plate.
Worn seal rings. Replace them : see Chapter IV –1.
Ruptured diaphragm. Replace diaphragm : see Chapter IV –1. Section Diaphragm replacement.
Incorrect stroke length. Check zero on pump/Re-zero pump (See Chapter IV – 2.).
Tubing run on discharge may be too long. Longer tubing runs may create frictional losses sufficient to reduce pump’s pressure rating.
Clogged foot valve strainer. Remove foot valve strainer when pumping slurries or when solution particles cause strainer to clog.
Failure to run
Pump not turned on or plugged in. Turn on or plug in pump. Check also programming (A9, B9, C9 Series).
EPU failure. Disassemble pump and measure the resistance of the EPU across the EPU wires. Resistance reading should be in accordance to the table (see Chapter IV – 4.). Also check EPU leads to ground.
Pulser failure. The pulser should be replaced if EPU checks out OK.
Page 18
IV - 4
LMI, Doseur à membrane, L001, Rev.B, 1098
Excessive pump output
Syphoning. Pumping downhill without a 4-FV. Move injection point to a pressurized location or install an 4-FV.
Little or no pressure at injection point. If pressure at injection point is less than 1,5 bar (25 psi), an 4-FV should be installed.
Excessive strokes per minute. Replace pulser or control panel.
IV – 4. EPU RESISTANCE CHART
Pump series Voltage Coil resistance (ohms)
20° C (68°F)
Ax4, Ax5, Ax6 115 V 76 – 87 Px4, Px5, Px6, Px8 H9
230 V 307 – 353
Ax7, Ax8 115 V 152 – 176 Px2, Px3 230 V 583 – 671
Bx1, Bx2, Bx3, Bx4 115 V 43 –49
230 V 167 – 193
Cx0, Cx1, Cx2, Cx3, Cx4, Cx5 115 V 22.8 – 26,2 E70, E71, E72, E73, E74 230 V 91 – 105
J54D, J55D, J56D 12VDC 1.1 – 1.3
Page 19
Instruction Supplement
LMI PUMP SERIES P1
Date : 10.98
Replaces Rev. 0 , 07.98
REF. : 161.5304.001 Rev. A
1 Foot valve 9 Injection check valve 2 Ceramic weight 10 Discharge tubing 3 Suction tubing 11 Spacer E.P.U. 4 Suction fitting 12 Metering pump housing 5 Return line (pressure relief) 13 Speed knob 6 Pump head 14 Stroke knob 7 4-Function valve (option) 15 Power cord 8 Coupling nut
Metering pump component diagram
Page 20
LMI PUMP SERIES P1
2/3
Control panel detail
Wiring diagram
1 Identification plate 2 Stroke knob 3 Speed knob 4 Stroke light
1 Potentiometer 2 Stroke led 3 Pulser assembly 4 Power cord 5 Mov/cap assembly 6 EPU assembly
23 P12x, P13x : cut 1 resistor
Other P1 : leave both resistors BLU Blue BRN Brown
GRN/YEL Green/yellow
YEL Yellow
Page 21
LMI PUMP SERIES P1
3/3
TECHNICAL CHARACTERISTICS
P12x P13x P14x P15x P16x P18x
Output
- max. (l/h)
- min (ml/h)
0.75
2.25
1.6
4.8
2.2
4.4
3.8
7.6
7.6
15.2
12
24 Max. pressure (bar) 10.3 7.6 17.3 7.6 3.5 1.5 Stroke capacity (ml)
- min.
- max.
0.07
0.22
0.13
0.44
0.07
0.37
0.13
0.63
0.25
1.27
0.40
2.00
Stroke frequency (cps/mn)
- min.
- max.
1
60
1
60
1
100
1
100
1
100
1
100 Diaphragm size (in2) 0.5 0.9 0.5 0.9 1.8 1.8 Min. recommanded stroke lenght (%) 30 30 20 20 20 20 Peak power (W) 75 75 150 150 150 150 Average power (Wh) 11 11 22 22 22 22
Note : The « x » which appears in the code designates both voltage and power cord/plug type.
Page 22
Pipe Thread Connection 1/4" or 1/2"
Typical injection check valve installation
Liquid End Sheet
1 ”pipe tee
Reduction*
1 ”to ½ ”NPT
injection check valve
CORRECT
Use Teflon tape on
the thread only
CORRECT
Do NOT use
Teflon tape on
machined thread
injection check
valve
Flow
Flow
40°Variation acceptable
* customer supplied
NPT pipe connection
DO NOT USE CLEAR VINYL TUBING ON THE DISCHARGE SIDE OF THE PUMP.
(Pressure can rupture tubing.)
DO NOT USE PLIERS OR PIPE WRENCH ON COUPLING NUTS OR FITTINGS.
Tubing Connections
Your pump will be supplied with one of the following tubing connections. Before installation, all tubing must be cutted with a clean square end. Valve and head connections from the factory may be capped. Remove and discard these caps before connecting the tubing.
Apply Teflon© tape.
(Customer supplied pipe
connections)
Injection Check Valve and Discharge Tubing Installation
The Injection Check Valve prevents backflow from a treated line. Connect the Injection Check Valve to your “DISCHARGE” (outlet) line. An NPTF fitting or pipe tee with a reducing bushing to 1/2" NPTF will accept the injectioncheck valve. Use Teflon® tape or pipe dope to seal the pipe threads only.
When installing the Injection Check Valve, be sure to position it so that the valve enters the bottom of your pipe in a vertical position. Variations left and right within 40° are acceptable (see illustration below).
After cutting an appropriate length of discharge tubing, connect tubing to the injection check valve then back discharge
Make sure it does not crimp or come into contact with hot or sharp surface (see Tubing Connections).
When pumping solutions, make certain that all tubing is securely attached to the fittings. It is recommended that tubing or pipe lines be shielded to prevent possible injury in case of rupture or accidental damage. Always wear protective clothing and face shield when working on or near your metering pump.
Flexible pipe connection with connection Nozzle
A : Nut C : Connection Nozzle B : End fitting D : Connection Encase properly pipe into connection nozzle as shown in the drawing
Flexible connection without connection Nozzle
Mounting Steps : Manual mounting
1
At the startup of the pump check that the assembly does not leak
2
: Fit the part 3 on the part 4 in mechanical stop
3
Pull on the part 3 for checking the assembly
Page 23
Use Ceramic Weight
pump pump
Foot Valve Must Reamin Vertical
CORRECTINCORRECT
2.0in (50mm) for Sediment Accumulation
solution tank
Foot Valve Tilted Sideways
Foot Valve Tubing Connection
WILL NOT PRIME
Foot Valve/Suction Tubing Installation
The Foot Valve acts as a check valve to keep the pump primed in suction lift applications.
The foot valve is designed to be submersed in the solution tank or drum and must sit in a vertical position at the bottom. Position approximately 2 inches (50mm) off the bottom if the tank or drum contains sediment. The ceramic weight, when installed, positions the foot valve in a vertical position.
NOTE: Pump models equipped with high-viscosity liquid end s are not equipped with foot valves. Flooded suction is recommended A 1/2" NPT connector is included for flooded suction installations.
1. Cut a piece of suction tubing to a length so that the foot valve hangs just above the bottom of the solution container. Maximum recommended vertical suction lift is 5 feet (1.5 m).
2. Attach the foot valve to one end of the suction tubing.
3. Slide the ceramic weight over the tubing end until it contacts the top of the foot valve coupling nut.
4. Place foot valve and tubing into the solution tank. Check that the foot valve is vertical and approximately 2 inches (50 mm) from the bottom of the tank or drum(see ullustration). Connect the Connect the other end of the tubing to the suction side (bottom side) of the pump head (see Tubing Connections).
Multi-Function Valve Installation
To install the multi-function valve, remove the yellow screw cap on the top of the pump head and screw in the valve so that it contacts the seal ring. An additional 1/8 – 1/4 turn may be necessary to prevent leakage.
1/4" O.D. tubing connects to the side of the valve and acts as a return line to the solution tank. To ensure priming, this tubing must NOT be submerged in the solution.
This return line tubing must be secured to ensure pumped solution will safely return to supply tank.
Depressurizing Discharge Line
(pumps equipped with multi-function
valves)
It is possible to depressurize the discharge line and pump head without removal of tubing or loosening of fittings.
1. Be sure injection check valve is properly installed and is operating. If a gate valve or globe valve has been installed downstream of injection check valve, it should be closed. Be certain relief tubing from the multi-function valve is connected and run to solution reservoir.
2. Turn Pressure Relief knob 1/4 turn.
3. Open the anti-syphon valve, if equipped
4. The discharge line is now depressurized.
5. If injection check valve is of higher elevation than
pump head, disconnecting tubing at the injection check valve end will allow air to enter and cause solution to drain back to tank.
Discharge
To Pump Head
¼ ” Tubing
Coupling nut
To solution Tank
Bottom Tubing
in Fitting
Push and hold while
tightening connector.
1 Inch
(25 mm)
Multi-Function Valve Tubing Connection
Page 24
1 Flapper valve 2 Injection check valve body 3 Injection check valve spring 4 Check valve ball 5 Seal ring 6 Cartridge valve 7 Cartridge valve o-ring 8 Cartridge valve washer 9 Valve seat 10 Clamp ring 11 Ferrule 12 Clamp sleeve 13 Tubing adapter 14 Coupling nut 15 Discharge tubing 16 Valve housing 17 Multi-function valve 18 High-viscosity spring 19 Liquifram 20 Pump head 21 Pump head screw 22 Suction tubing 23 Foot valve seat 24 Foot valve screen 25 High-viscosity valve seat 26 H.V. tubing clamp 27 H.V. suction tubing 28 H.V. Tubing x 1/2 NPT connector 29 Injection check valve assembly 30 Discharge valve assembly 31 Suction valve assembly 32 Pump head assembly 33 Foot valve assembly 34 Injection Seat PTFE 35 Ceramic Weight 36 Return Line 37 Cap ASM (Black Knob) 38 Cap ASM (Yellow Knob) 39 Multi-Function Valve Body 40 Nut Multi-Function Valve 41 Screw Multi-Function Valve 42 Return Line Coupling Nut 43 Adjustment Screw B/4-FV 44 Cap B/4-FV 45 Plug B/4-FV 46 Gasket B/4-FV 47 Small O-Ring B/4-FV 48 Large O-Ring B/4-FV
Liquid End Parts List
NOTE: This illustration is a visual
representation of all LE components. Liquid ends will not include all parts shown.
40
42
40
36
38
39
48
47
46
45
44
41
43
41
42
36
41
39
37
17
17
17
37
39
38
40
42
36
3FV
B/4FV
4FV
Page 25
Bleed 4-FV
Start-Up/Priming for Pump Supplied with Multi-Function Valve
Read this entire section completely before proceeding.
When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now start priming the pump.
1. Plug in or switch the pump on.
2. While the pump is running, set the speed knob at
80 %and the stroke knob at 100%.
If the pump is equipped with pressure control, turn fully clockwise.
3. 1/4 turn open the relief side (black knob) of the multi-function valve.
3A. (Bleed 4FV only) With screwdriver rotate bleed
adjustement screw counter-clockwise 2 full turns. When solution begins to flow through translucent bleed return tubing, the pump is primed. Stop pump.
4. The suction tubing should begin to fill with solution from the tank.
5. A small amount of solution will begin to discharge out the return line of the multi-function valve. Once this happens,1/4 turn or release the knob and SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, or disconnect the power cord.)
6. The pump is now primed.
6A. (Bleed 4FV only)
a. Start pump and let pump inject solution into
the discharge line.
b. Close the bleed adjustment screw by rotating
it clock-wise with a screwdriver.
c. Now adjust the pump stroke length and/or
speed (frequency) to a range approximately 25% higher than you would normally want for the process.
d. Slowly rotate bleed adjustment screw counter-
clockwise until just a small amount of solution begins to trickle down inside the bleed return tubing. A small amount of solution pumped back to the tank with each stroke of the pump will allow gas and air to escape without air or gas locking in the pump head.
NOTE: If the pump does not self-prime, remove the multi-function valve on the discharge side of the pump head. Remove the check valve and pour water or solution into the port until the head is filled. Replace valve, then follow start-up/priming steps.
Start-Up/Priming without Multi­Function Valve
Read this entire section completely before proceeding.
When all precautionary steps have been taken, the pump is mounted, and the tubing is securely attached, you may now prime the pump.
1. Plug in or switch on the pump.
2. While the pump is running, set the speed knob at
80% and the stroke knob at 100%.
If the pump is equipped with pressure control, turn fully clockwise.
3. The suction tubing should begin to fill with solution from the tank.
4. Once the solution begins to exit the pump head on the dis-charge side, SHUT THE PUMP OFF. (If pump is not equipped with an on/off switch, disconnect the power cord).
5. The pump is now primed.
NOTE: If the pump does not self-prime, remove the
fitting on the discharge side of the pump head. Remove the ball and pour water or solution into the port until the head is filled. Replace valve, then follow start-up/ priming steps.
Rotate clockwise
to close Bleed
Adjustment Screw
Anti-Syphon
Knob
Negative pressure at discharge closes diaphragm and prevents syphoning.
Bleed
(Relief) Port
Space open
to drain
Pump discharge stroke lifts diaphragm to allow for fluid flow.
Page 26
F
DECLARATION "CE" DE CONFORMITE
Nous, déclarons que le matériel désigné ci-après a été conçu et fabriqué suivant les directives et spécifications suivantes : Directive basse tension 2006/95/EEC
Directive CEM 2004/108/EEC
GB
"EC" DECLARATION OF CONFORMITY
We, certify that the equipment designated below has been designed and manufactured in accordance with the specifications of the following :
Low voltage Directive 2006/95/EEC
EMC Directive 2004/108/EEC
D
EG-KONFORMITÄTSERKLÄRUNG
Wir, verklären, daß die nachgestehend bezeichneten Gerätschaften im Einklang mit folgenden Richtlinien und Spezifikationen geplant und hergestellt wurden :
EG-Richtlinie 2006/95 für Niederspannung
EG-Richtlinie 2004/108 zur elektromagnetischen Verträglichkeit
NL
EG FABRIKANTENCONFORMVERKLARING
De ondergetekenden, verklaren geheel onder eigen verantwoor delijkheid dat het produkt waarop deze verklaring betrekking heeft in overeenstemming is met:
Richtlijn 2006/95/EEG inzake elektrisch materiaal bestemd voor gebruik binnen b epaalde spanningsgrenzen;
Richtlijn 2004/108/EEG inzake elektromagnetische compatibiliteit.
I
DICHIARAZIONE DI CONFORMITA' "CE"
La società, Dichiara che il materiale sotto specificato è stato progettato e realizzato in conformità con l e seguenti direttive e specifiche:
Direttiva bassa tensione 2006/95/EEC
Direttiva CEM 2004/108/EEC
E
DECLARACION "CE" DE CONFORMIDAD
Nosotros, declaramos que el material detallado a continuación está diseñado y fabricado de acuerdo a la siguientes directivas y normas :
Directiva basa tensión 2006/95/EEC
Directiva EMC 2004/108/EEC
TYPE / TYP / TIPO LMI : AA7.. / AA9.. / B1.. / B7.. / B9.. / C1.. / C7.. / C9.. / H9.. / J5.. / P0.. / P1.. / P5../ E7
Alain LAMPERIER Le Directeur Production Production Manager
AQFORM 289
10/2010
Page 1/1
Page 27
PS-GARANTIE /1
1
GUARANTEE
The vendor guarantees his products according to the Milton Roy Europe . general conditions of sale.
The guarantee for components and sub-assemblies not fabricated by the vendor is limited tothat given by the supplier.
The vendor’s guarantee only covers the replacement or the repair, at his cost and in his factory, of all parts acknowledged by his technical services as being defective due to an error in conception, of material or of execution.
It is the purchasers responsability to prove the said defects. The guarantee does not cover the replacement of wear parts mentioned in part V - Preventive Maintenance.
The vendor reserves the right to modify all or part of his products in order to satisfy the guarantee. The guarantee does not cover charges arising from dismantling, assembly, transport and movements.
The replacement of one or several parts, for whatever reason, does not prolong the period of guarantee.
The guarantee is not applicable notably in the following cases : installation not in accordance with standard current practice.
deterioration or accident resulting from negligence. lack of surveillance or maintenance. modifications to conditions of use. chemical corrosive or erosive attack. The proposed materials of construction are recommendations subject in all cases to verification and acceptance by the client. The recommendations, based on the experience of the vendor and the best available information, do not guarantee against wear or chemical action.
The guarantee ceases : if the storage of the material, outwith the vendors factory, does not conform to his
recommendations or to current standard practices. in case of work or dismantling of the material by someone who does not respect written recommendations of the instruction manual (when replacing wear parts). if parts from another origin are substituted for the original parts supplied by the manufacturer.
The purchaser cannot call on guarantee claims to justify differing payments.
INDUSTRIAL OWNERSHIP
This manual can only be used by the purchaser or the user. It cannot be distributed, published, reproduced (partially or totally) or generally communicated to third parties without the advance, formal written authorisation of the vendor. Any breach of these rules may result in legel action being taken.
Page 28
FRANCE
Tel.: +33.(0)2.32.68.30.00 / Fax: +33.(0)2.32.68.30.93 E-mail: contact@miltonroy-europe.com Assistance technique / Technical assistance: ?el.: +33.(0.2.32.68.30.02 - Fax: +33.(0)2.32.68.3 0.96 Pièces de rechange / Spare parts: ?el.: +33 (0)2.32.68.30.01 - Fax: +33.(0)2.32.68.30.92 E-mail: quote.sp@miltonroy-europe.com 10 Grande Rue 27360 Pont-Saint-Pierre - FRANCE
www.miltonroy
-europe.com
ESPANA / PORTUGAL
Sistemas de Dosificación y Mezcla de Fluidos, S.L.
Tel.: +34 91 4741914 - Fax : +34. 91 3459707 - E-mail: info@sdm-sistemas.com C/ Embajadores, 100 28012 Madrid - ESPANA
www.sdm
-sistemas.com
ITALIA / MALTA
Milton Roy Italia Srl
Tel.: +39.(0)39.65.78.505 - Fax: +39 (0)39 60.56.906 E-mail: sergio.testa@miltonroy-europe.com Centro Direzionale Colleoni Via Paracelso, 16 Palazzo Andromeda Ingresso 2 20041 Agrate Brianza (Ml) - ITALIA
www.miltonroy
-europe.com
UNITED-KINGDOM / NORTHERN EUROPE
Milton Roy UK, Ltd
TECHNICAL ASSISTANCE AND SPARE PARTS : Tel.: +44 (0)191.5375.256 - Fax: +44.(0)191.5490.911 E-mail: victoria.richardson@miltonroy-europe.com North Hylton Road Sunderland, SR5 3JD - UNITED-KINGDOM
www.miltonroy
-europe.com
NORTH AND SOUTH AMERICA
LMI Milton Roy Milton Roy Company
Tel.: 215.293.0401 - Fax: 215.293.0445 Tel.: 215.441.0800 - Fax: 215.441.8620 E-mail: info@lmipumps.com E-mail: info@miltonroy.com
201 Ivyland Road 201 Ivyland Road Ivyland, PA 18974 - USA Ivyland, PA 18974 - USA
www.lmipumps.com
www.miltonroy.com
FOR OTHER COUNTRIES:
Visit our web site at www.miltonroy-europe.com for a complete list of our distributors network around the world. For further details you can contact our headquarter located in France at contact@miltonroy-europe.com.
Loading...