The Milltro nics CompuFlo III is a
microprocessor based integrator designed
specifically for b u lk dry solids in-line
weighing applications.
The CompuFlo III integrates a signal
proportional to material flowrate (such as
from a Milltronics flowmeter) over time to
provide material totalization.
The Com puFlo III provides:
» Rate and Total displays
» an analog mA output proportional to rate
» a relay contact f or remote total izer or PLC operation
» 2 alarm/ control relays based on rate and/or On-line Calibration
» 2 communication ports for computer , PLC, or serial printer c onnection
The CompuFlo I II has three modes of operation;
program, calib rate and run.
MULTI SPAN NO: 1R2:RATE HI R1:SPEED LO
1263957
TOTAL
T
150%100%50%0%-50%
The operator programs the CompuFlo III via the
keypad to provide displays and outputs in the
desired units of measure. The Comp uFlo III
displays the name, range of acceptable values,
(and additional information/instruction) of the
parameter (programming feature) accessed.
In the calibrate mode, the system Zero (no flow) and Span (100% flowrate) calibration is established. A test
reference (test weight) is used to simulate the force of material impacting the flowmeter sensing plate during
the Span calibration. A mater ial test is normally performed to verify the calibration accuracy. If necessary, the
material test results may be used to correct any deviation in calibration accuracy.
Calibration accuracy may be checked at any time. The results of th is check are automatically compared to the
system calibration and the deviation is displayed. The operator may choose to continue operation with the
previous calibration or the results of the calibration check.
All programming and calibration information is stored in non-volatile memor y. If desired, the security
parameter may be set to ensur e the CompuFlo III programming and calibration is not inadvertent ly altered in
the run mode.
The run mode is the normal operating mode, providing dynamic measurement displays, analog mA output,
and a variety of other features.
PL-3681 – 1
ABOUT THIS MANUAL
GENERAL
This instruction manual conta ins informat ion essential to successful CompuFlo III operation.
SPECIFICATIONS, includes CompuFlo III environmental, phys ical, and operational characteristics.
INSTALLATION, indicates the typical mounting and interconnection procedure.
START UP, prepares the Compu Flo III for programming.
PROGRAM MING, details activities required to ta ilor t he CompuFlo III to the par ticular application.
CALIBRATION, includes the CompuFlo III input signal to output feature corr elation procedure.
PARAMETERS, describes the function, factory setting, and range of possible settings,
of each parameter.
PROGRAM MING CH A RT, provides a space to record programmed parameter values for futur e reference.
OPERATION, describes initial and normal CompuFlo III operation.
COMMUNICATION, provides CompuFlo III serial communication information.
APPENDICES, contains Quick Start , Alphabetical Parameter L ist, Maintenance and Spare Parts,
Error Messages and Trouble Shooting information.
FIGURES, includes a complete set of installation and interconnection drawings.
CONVENTIONS
Throughout this instruction manual the following symbols and presentation practices have been utilized:
[1234567ABCDEFG]alpha numeric characters enclosed within these brackets repr esents information
displayed by the CompuFlo III display.
[T/h]units of measure were arbitrarily chosen from the imperial set. The actual units
displayed or other wise applicable will be dependent upon o perator programming.
indicates the following CompuFlo III keypad button is to be pressed.
"RUN" or alpha numeric characters enclosed within quotation marks or this symbol
ENTER
represent a key on the CompuFlo III keypad.
Italics
are used to identify the defined word(s) or phrase within a definition statement.
Informat ion of a hig her order of importance, is indented from
both left and right margins and in bold print.
PL-3681 – 2
■
SPECIFICATIONS
COMPUFLO III
Power:» 100/115/200/230 V AC ±10%, jumper
selectable, 50/60 Hz, 65 VA
Fuse:» FU 1» 3/ 4 Amp MDL SLO-BLO or equivalent
» FU 2» 1/ 4 Amp MDL SLO-BLO or equivalent
Operating» -20 to 50°C (-5 to 122°F)
Temperature:
Accuracy:» 0.1% of span, typical
Resolution:» 0.02% of span, typical
Inputs:» Load cell:» -20 to +45 mV DC per load cell
» 2 load cells max.
» LVDT:» 0 to 0.75 Vrms, 3 kHz
» Multispan:» 4 contact inputs for up to 8 spans
» Keypad:» 20 key sealed membrane with tactile feedback
» On-Line:» 4-20 mA proportional to weight
(Calibration)
» Auto Zero» contact input for enable
» Printer:» contact input for print request
Belden 8404 or equivalent
Barrier Strips:» For hazardous location flowmeters
Enclosure Heater:» Recommended for operating temperatur es below -10 ° C
(14°F) or excessive humidity
Relay:» Food Grade remote totalizer relay
CVCC:» For bipolar current loop/RS-232C conversion
BIC-II:» For 6 port bipolar current loop/RS-232C or RS-422 conversion.
PL-3682 – 2
■
INSTALLATION
CompuFlo III
The CompuFlo III should be mounted in an area that is within the specified temper ature range, and that is
suitable for NEMA 4 style enclosures. The front cover should be accessible for keypad operation, display
viewing, and have sufficient room to swing open.
It is advisable to keep the CompuFlo III away from high voltage or current runs, contact ors, S CR control
drives and RF interference.
For proper totalizer relay operation, the CompuFlo III must be mounted within 30° from vertical.
Do not mount the CompuFlo III in direct sunl ight without the use of a sun shield.
INTERCONNECTION
Remov e the enclosure bottom plate prior t o drilling or punching cable/conduit entrance(s ), to ensure metal
filings do not come in contact with the CompuFlo III electronics.
Load cell or LVDT, rate output, On-Line Calibration input, communication, and low voltage contact wiring may
be run in a common conduit separated from power and high voltage contact wiring. Refer to Figures.
The power ground (TB1-50) is internally connected to SHLD connections (TB1-13,26, TB2-5), and to the
mother board bottom mounting lugs. COM connections (TB1-15,29,32,36,37, a nd TB2-8) are connected to
power ground via jumper J2 (below TB1, between the mounting lugs). If the COM connections must be
isolated from power ground, remove J2.
Refer to FIGURES
All wiring must be perf ormed in accordance with all governing regulations.
HAZARDOUS LOCATIONS
The CompuFlo III cannot be mounted in an area classed as hazardous. Where the associated flowmeter is
mounted in a hazardous area, barrier strips must be used to isola te the CompuFlo III from the hazardous area
for intrinsic safety. Encapsulated LVDT transducing elements are also available. Contact Milltronics or your
local representative for hazardous application information.
PL-3683 – 1
■
START UP
HARDWARE SET UP
UNIT I.D.
The CompuFlo III identification code factory setting is 00,
as set by SW1 contacts 1 - 4.
When more than one CompuFlo III is connected to a
host computer or PLC, the identification code of each
CompuFlo III must be set to a differ ent value.
UNITSW 1
I. D.1 2 3 4
00O O O O
01C O O O
02O C O O
03C C O O
04O O C O
05C O C O
06O C C O
07C C C O
SW1
UNIT I. D. #
1248
METRIC
0000
IMPERIAL
UNITSW 1
I. D.1 2 3 4
08O O O C
09C O O C
10O C O C
11C C O C
12O O C C
13C O C C
14O C C C
15C C C C
LOCKOUT
MAIN
TOTAL
RESET
TEST
NORMAL
O = switch contact open C = switch contact closed
UNITS OF MEASURE
Two sets of Units are available; Imperial and Metric. Leave SW1 contact 5 (SW1-5) in the open position
for Imperial, or closed for Metric units. R efer to PARAMETERS / P0 - RATE UNITS for the Rate Unit
options available.
SW1-5 must be set to the desired position b ef ore i nitial pro grammi ng is performed.
To alter the Units of Measure of a programmed CompuFlo III, set SW1-5 as desired,
then perform a Reset. Refer to Parameters / P99 Reset.
LOCKOUT
Leave SW1 contact 6 (SW1-6) in the open position for normal applications. Close SW1-6 for Lockout Mode
operation, after all programming and calibration is complete. Refer to Calibration / Calibration Alternatives.
TOTALIZER RESET
Leave SW1 contact 7 (SW1-7) in the open position for normal operation. When this contact is closed, the
Main Total Display cannot be reset.
TEST MODE
Leave SW1 contact 8 (SW1-8) in the open position for normal operation. The closed position is reserved for
Milltronics service personnel.
PL-3684 – 1
ALTERNATE
TB2 BAUD
000
4
X
4
2
PRINTER
REMOTE
TERMINAL
SERIAL
TEST /
ADJUST
SW2
3
0
1
X
ENGLISH
SPARE
PORT 2 BAUD RATE
The CompuFlo III TB2 com munication port baud rate is independent of the PORT 1 baud rate, and is set by
SW2 contacts 1 and 2. Set the baud as required by the device connected. The factory setting is 300 baud.
BaudSW2
Rat e1 2
300O O
O = switch contact open
1200C O
2400O C
C = switch contact closed
9600C C
NORMAL
LANGUAGE
Leave SW2 contact 3 (SW2-3) in the open position for English, or closed for alternate language, CompuFlo III
displays.
REMOTE DISPLAY
Leave SW2 contact 6 (SW2-6) open for normal operation. If a remote mounted terminal is used, SW2-6 of the
remote terminal (only) must be closed.
PORT 2 COMMUNICATION MODE
Leave SW2 contact 7 (SW2-7) open for normal operation, or closed if a serial printer is connected to PORT 2.
TEST/ADJUST MODE
Leave SW2 contact 8 (SW2-8) in the ope n position for normal operation. In th e closed position, a test menu is
displayed.
When the keypad test is selected, a pictorial representat ion of the keypad is displayed. As each keypad
button is depressed, the display indicates successful operation.
When the display contrast adjustment is selected, pressing "4" increases, and pressing "8" decreases the
display contrast.
PL-3684 – 2
PRE START UP CHECKS
After the Compu Flo III has been properly installed and interconnected, the following checks should be
performed prior to connecting power:
»all cable connections are correct and secure.
»the proper voltage has been selected (J1 on the mother board).
»SW1 and SW2 have been set as required.
»the POWER switch is in the OFF position.
Connect power to TB1 -51,52 and ground to TB1-50 of the CompuFlo III. Ensure all connections are secure.
Do not operate the CompuFlo III with the
ground (earth) wire disconnected.
INITIAL START UP
Se t the Co mpuFl o III POWER switch to the ON position. The CompuFlo III will enter the following
initialization routine:
COMPUSCALE/FLO III(REV DATE)(EPROM 2 REV)
TEST :(MESSAGES)
PL-3684 – 3
When the display initializa tion routine is complete, the display revert s to the last operati ng mode accessed. If
the CompuFlo III is not programmed, the paramet er entry display is accessed.
The CompuFlo III display is divided into the following fields:
MULTISPAN
FEATURE
ECAL
VALUE
RELAY
CALIB
CYCLE
UNITS
AUX 1
AUX 2
-50%
150%100%RAT E0%
RATE BAR
GRAPH
FIELD DESCRIPTION
MULTISPANcurrent span calibration selected. If Multispan 0 is selected, the field is blank.
RELAYcurrent relay status (run mode). If re lays are not used, or in non-alarm, the field is blank.
FEATUREpara meter number (parameter entry), calibration type (calibrate mod e), Rate or Total (run mod e)
ECALindicates ECAL (simulated rate signal) is on. If ECAL is off, the field is blank.
CYCLEfull cycles remaining before calibration completion (calibrate mode).
CALIBcalibrate mode indicator. When displayed, tota lizers are on HOLD (unless otherwise specified).
VALUEthe current va lue associated with the featur e selected.
UNITSthe units of measure associated with the value displayed
AUX 1parameter name (parameter entry), auxiliary message (cali b rate mode), not used (run mode).
AUX 2parameter value range (paramet er entry) , addit ional information or error message (any mode).
BAR GRA PHrepresents current rate, displayed as percent of span (run mode).
PL-3684 – 4
■
PROGRAMMING
GENERAL
Information specific to a particular application is entered into the Com puFlo III memory, via the keypad, prior
to the calibration procedure.
KEYPAD
Button Operation Function
–
0
9
parameter entrynumeric entries
4
8
RUN
PAR
ZERO
SPAN
ALT
DISP
RESET
TOTAL
parameter entryincrease: analog mA output (P17 and P18 only), ECAL value (P22 only)
parameter entrydecrease: analog mA output (P17 and P18 only), ECAL value (P22 only)
parameter entrydecimal point
parameter entrynegative entry or display prev ious parameter
any modeenter the run mode, return to previously selected display
any modeaccess parameter for entry or viewing
any modeaccess zero calibration, view current zero count
any modeaccess span calibration, view current span count
any modeenter the run mode, select rate or total display
any modereset displayed totalizer value to zero (refer to Start Up\Hardware
parameter entrydisplay parameter value, or store altered parameter value,
or display next parameter
calibrate modeactivate zero or span calibration, or accept calibration results
any modewith total displayed, switch to alternate total (ma in or sub)
PL-3685 – 1
PARAMETER ENTRY
Application information must be entered into the appropriate CompuFlo III memory locations . These locations
are identified by a Paramet er Number (P#). The information entered is referred to as a Parameter Value
(Value ). To enter informat ion, the operator accesses the desired parameter number, v iews the parameter
value displayed, and then accepts or alters the current parameter value.
Common Parameter Entry Display Fields
FEATURE
VALUE
UNITS
AUX 1
AUX 2
-50%
150%100%RATE0%
RATE BAR
GRAPH
DIRECT PARAMETER ACCESS
To directly access any parameter, followed by the parameter number.
PAR
Only valid parameter numbers will be accepted. Refer to PARAMETERS for the list of valid parameter
numbers.
eg.[PARAMETER]is displayed in the F EAT URE field, if PARAMETER 0 is desired
[PARAMETER 0]is displayed in the F EAT UR E field,
PAR
0
[RATE UNITS]is displayed in the AUX 1 field.
To view the current paramet er value and auxiliary messages for the displayed parameter,
ENTER
[PARAMETER 0]is displayed in the FEAT UR E field
[ 1]is displayed in the VAL UE field (unless previously altered)
[RATE UNITS ]is displayed in the AUX 1 field
[1:T/h;2: LT/h;3: lb/ h;4: lb/m in]is displayed in the AUX 2 field (with SW1-5 open)
AUTOMATIC PARAMETER ACCESS
Upon accessing any parameter value, pressing "ENTER" will autom atically access the next parameter in the
programming sequence. Pressing "-" will automat ically access the previous parameter.
eg.with the VA LUE field of PARA METER 0 displayed...
ENTER
[PARAMETER 1]is displayed in the FEAT UR E field
[ 0.0000]is displayed in the VAL UE field (unless previously altered)
[RATE]is displayed in the AUX 1 field
[ENTER VALUE]is displayed in the AUX 2 field
all fields associated with PARAMETER 0 are displayed
PL-3685 – 2
ALTERING PARAMETER VALUES
Once the desired parameter number is accessed, the parameter value may be accept ed or altered as
required.
eg.if the VALUE f ield of PARAMETER 0 is 1 (T/h), however 2 (LT/h) is desired
2
ENTER
[ 2]is displayed in the VALUE fi eld
the new value (2) is now stored in memory
LAST PARAMETER RECALL
From the Calibrate or Run mode, the last parameter accessed in the Program mode, may be recalled.
the last parameter accessed is displayed.
ENTER
PAR
PARAMETER RESET
To reset all parameters to factory settings, refer to PARAMETE RS/P99 M aster Reset.
PROGRAMMING SEQUENCE
Typically, the CompuFlo III is programmed as follows:
a) Referring to PARAMETERS, copy the selected parameter v alues to the spaces provided on
the PROGRAMMING CHART fold out.
b) Perform a MASTER RESET (P99).
c) Access each parameter ( in sequence from P 0) and enter the parameter values
from the PROG RAMM ING CHA RT.
d) Perform a Zero and Span calibration (refer to CALIBRATION)
e) Perform Span Adjust (P8) or Factoring (P9) procedures, as req uired.
Remember to copy any parameter value alterations onto
the PROGRAMMING CHART for future refer ence.
QUICK START PROCEDURE
The Quick Start Procedure describes the minimum star t up, programming, and calibration requirements of the
CompuFlo III, to obta in successful RUN mode operation for most applications.
If desired, refer to APPENDICES /Quick Start. Parameters which are not included in the procedure may be
program med after the RU N mode is entered.
■
PL-3685 – 3
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.